K170995 · Edan Instruments, Inc. · DPS · Jun 2, 2017 · Cardiovascular
Device Facts
Record ID
K170995
Device Name
Electrocardiograph
Applicant
Edan Instruments, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Jun 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SE-18 18-lead electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
Device Story
The SE-18 is an 18-lead electrocardiograph used in hospitals and healthcare facilities by trained professionals. It acquires ECG signals via body surface electrodes; processes signals using digital filters (anti-baseline drift, AC, EMG, low pass) and the Smart ECG Measurement and Interpretation Program (SEMIP); and displays waveforms on a 15-inch screen. The device includes an external sampling box, main control board, thermal printer, and connectivity modules (Ethernet, Wi-Fi, USB). Recorded ECGs can be stored in flash memory or transmitted to a PC. The SEMIP program provides automated measurements and interpretive statements to clinicians on an advisory basis to aid in heart disease diagnosis. The device does not deliver substances to or extract substances from the patient.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing per IEC 60601-2-25.
Indicated for adult and pediatric patients requiring ECG signal acquisition via body surface electrodes for heart disease analysis and diagnosis by healthcare professionals.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Predicate Devices
EDAN Instrument, Inc. Electrocardiograph, models SE-1515 (K152427)
EDAN Instrument, Inc. Electrocardiograph, models SE-12, SE-12 Express, SE-1200, and SE-1200 Express (K160876)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2017
Edan Instruments, Inc. Mr. Doug Worth Sr. Dir. US RA/OA 1200 Crossman Ave, Suite 200 Sunnyvale, California 94086
Re: K170995
Trade/Device Name: SE-18 Electrocardiograph Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: March 30, 2017 Received: April 3, 2017
Dear Mr. Doug Worth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
M.A. Hillemann
for.
for __________________________________________________________________________________________________________________________________________________________________________ Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if know n)
### K170995
Device Name SE-18 Electrocardiograph
### Indications for Use (Describe)
The SE-18 18-lead electrocardiograph is intended to acquire ECG signals fromadult and pediatric patients through body surface ECG electrodes. The electrocardiographis only intended o be used in hospitals or healthcare healthcare professionals. The cardiogramrecorded by the electrocardio graph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
| Type of Use (Select one or both, as applicable) | <div> <span> <span style="padding-right:5px;"> <div style="display:inline-block; vertical-align:middle;"> <img alt="Checked" src="checkbox_checked.png" style="width:15px; height:15px; vertical-align:middle;"/> </div> </span>Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> <div> <span> <span style="padding-right:5px;"> <div style="display:inline-block; vertical-align:middle;"> <img alt="Unchecked" src="checkbox_unchecked.png" style="width:15px; height:15px; vertical-align:middle;"/> </div> </span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
PRAStaff@fda.hhs.gov "An agencymay not conductor sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Paperwork Reduction Act (PRA) Staff
FORM FDA 3881 (8/14)
EF PSC. Publishing Services (301) 443-6740
{3}------------------------------------------------
### 510(K) Summary
## Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc<br>#15 Jinhui Road,Jinsha Community,<br>Kengzi Sub-District, Pingshan District,<br>Shenzhen, 518122 P.R.China.<br>518067 P.R. China<br>Tel: +86(0755) 26858736<br>Fax: +1 (408) 418-4059 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person:<br>Preparing date: | Alice Yang<br>March 30, 2017 |
| 2. Device name and<br>classification: | Device Name: Electrocardiograph<br>Model: SE-18<br>Classification Name/ Product code:<br>870.2340 Electrocardiograph/DPS<br>Regulatory Class: Class II |
| 3. Predicate Device(s): | 1. EDAN Instrument, Inc. Electrocardiograph, models SE-1515,<br>K152427 (Primary)<br>2. EDAN Instrument, Inc. Electrocardiograph, models SE-12, SE-12<br>Express, SE-1200, and SE-1200 Express, K160876 (Reference) |
| 4. Reason for<br>Submission | New model |
| 5. Pre-Submission,<br>IDE | Not applicable, there is no prior submission. |
| 6. Device Description: | The SE-18 acquires and displays an 18 leads waveforms, which can<br>also be printed by an integrated thermal printer with effective<br>recording width 216mm. |
| | Digital filtering techniques similar to those incorporated on the<br>SE-1515 and SE-12 have been used in SE-18: including |
{4}------------------------------------------------
Anti-baseline drift filter, AC filter (50/60Hz), EMG Filter and Low pass Filter, which can help the user to record a higher quality ECG. The EDAN Instruments "Smart ECG Measurement and interpretation program" (SEMIP) is included in this machine. The SEMIP program is completely integrated and provides the clinician with a detailed analysis of the ECG signal to aid in the interpretation of the ECGs. The recorded ECG can be saved in flash memory or sent to a PC by Ethernet or WIFI. During the examination, there are no substances delivered to and/or extracted from the patient. SE-18'sfunction block diagram is shown as Fig 1.It's made up of an external ECG Sampling Box, SE-18 Main Control Board with RAM, FLASH, Ethernet module, WIFI module and USB module, a thermal printer module, a power supply module, a keyboard, and 15' 1024*768 display module. The SE-18 18-lead electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG
7. Intended Use: electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
### 8. Predicate Device Comparison
The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters items, which are shown in the following tables in details.
| Item | Proposed device<br>(SE-18) | Predicate device<br>(SE-1515) | Remark | |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|------|
| 510(k) Number | Current Submission | K152427 | ---- | |
| Indications for Use | | | | |
| Intended use | The SE-18 18-lead<br>electrocardiograph is intended<br>to acquire ECG signals from<br>adult and pediatric patients<br>through body surface ECG<br>electrodes. The<br>electrocardiograph is only<br>intended to be used in<br>hospitals or healthcare<br>facilities by doctors and<br>trained healthcare<br>professionals. The cardiogram<br>recorded by the<br>electrocardiograph can help<br>users to analyze and diagnose<br>heart disease. However, the<br>interpreted ECG with<br>measurements and<br>interpretive statements is<br>offered to clinicians on an<br>advisory basis only. | SE-1515 PC ECG is intended to<br>acquire, process and store ECG<br>signals from adult and pediatric<br>patients undergoing stress<br>exercise test or resting test. The<br>SE-1515 PC ECG is intended to<br>be used only in hospitals and<br>healthcare facilities by doctors<br>and trained healthcare<br>professionals. The cardiogram<br>recorded by the SE-1515 PC<br>ECG can help users to analyze<br>and diagnose heart diseases.<br>However, the ECG with<br>measurements and interpretive<br>statements is offered to clinicians<br>on an advisory basis only. | Differen<br>t | |
| SW Algorithm | SEMIP | SEMIP | Same | |
| Sampling Box | DE18 | DE18/DE15/DX12/DP12 | Differen<br>t | |
| Safety Specifications | | | | |
| Safety Standards | IEC 60601-1:2005/A1:2012<br>EN 60601-1:2006/A1:2013<br>IEC 60601-1-2:2007<br>EN 60601-1-2:2007/AC:2010<br>IEC 60601-2-25:2011 | IEC 60601-1:2005/A1:2012<br>EN 60601-1:2006/A1:2013<br>IEC 60601-1-2:2007<br>EN 60601-1-2:2007/AC:2010<br>IEC 60601-2-25:2011 | Same | |
| Anti-electric-shock<br>type: | Class I with internal power<br>supply | Class II | Differen<br>t | |
| Anti-electric-shock<br>degree: | Type CF with<br>defibrillation-proof | Type CF with defibrillation-proof | Same | |
| Degree of protection<br>against harmful ingress<br>of water: | Ordinary equipment (Sealed<br>equipment without liquid<br>proof) | Ordinary equipment (Sealed<br>equipment without liquid proof) | Same | |
| Disinfection/sterilization method: | Refer to the user manual for details | Refer to the user manual for details | Same | |
| Degree of safety of application in the presence of flammable gas: | Equipment not suitable for use in the presence of flammable gas | Equipment not suitable for use in the presence of flammable gas | Same | |
| Working mode: | Continuous operation | Continuous operation | Same | |
| EMC: | CISPR 11 Group 1, Class A | CISPR 11 Group 1, Class A | Same | |
| Patient Leakage Current: | NC<br>SFC | $<10μA (AC) / <10μA (DC)<50μA (AC) / <50μA (DC)$ | $<10μA (AC) / <10μA (DC)<50μA (AC) / <50μA (DC)$ | Same |
| Patient Auxiliary Current: | NC<br>SFC | $<10μA (AC) / <10μA (DC)<50μA (AC) / <50μA (DC)$ | $<10μA (AC) / <10μA (DC)<50μA (AC) / <50μA (DC)$ | Same |
| Environmental Specifications | | | | |
| Temperature | | | | |
| Operating | +5°C (+41°F) ~ +40°C (+104°F) | +5°C (+41°F) ~ +40°C (+104°F) | Same | |
| Transport/Storage | -20°C (-4°F) ~ +55°C (+131°F) | -20°C (-4°F) ~ +55°C (+131°F) | Same | |
| Humidity | | | | |
| Operating | 15%RH~95%RH<br>Non-Condensing | 25% RH~80% RH<br>Non-Condensing | Different | |
| Transport/Storage | 15%RH~95%RH<br>Non-Condensing | 25% RH~93% RH<br>Non-Condensing | t | |
| Pressure | | | | |
| Operating | 70 kPa ~106 kPa | 86 kPa ~106 kPa | Different | |
| Transport/Storage | 70 kPa ~106 kPa | 70 kPa ~106 kPa | t | |
| Performance Specifications | | | | |
| HR Recognition | | | | |
| HR Range: | 30 bpm ~300 bpm | 30 bpm ~300 bpm | Same | |
| Accuracy: | $\pm$ 1 bpm | $\pm$ 1 bpm | Same | |
| ECG Unit (only DE18 is listed for SE-1515) | | | | |
| Leads: | 18 standard leads | 18 standard leads | Same | |
| Acquisition Mode: | 18 leads acquisition<br>simultaneously | 18 leads acquisition<br>simultaneously | Same | |
| Sampling Frequency | 16kHz | 16kHz | Same | |
| A/D: | 24bits | 24 bits | Same | |
| Resolution: | 0.1575 μV/LSB | 0.1575uV/LSB | Same | |
| Time Constant: | ≥3.2s | ≥3.2s | Same | |
| Frequency Response: | 0.01~300Hz(-3dB) | 0.01~300Hz(-3dB) | Same | |
| Gain: | 2.5, 5, 10, 20, 10/5, AGC (mm/mV) | 2.5, 5, 10, 20, 10/5, AGC (mm/mV) | Same | |
| Input Impedance: | ≥100M Ω (10Hz) | ≥100MΩ (10Hz) | Same | |
| Input Circuit Current: | ≤0.01μA | ≤0.01μA | Same | |
| Input Voltage Range | ≤±5 mVpp | ≤±5 mVpp | Same | |
| Calibration Voltage: | 1mV±2% | 1mV ± 2% | Same | |
| DC Offset Voltage: | ±600mV | ±600mV | Same | |
| Minimum Amplitude: | 20 μVp-p | 20 μVp-p | Same | |
| Noise: | ≤12.5μVp-p | ≤12.5μVp-p | Same | |
| Multichannel crosstalk | ≤0.5mm | ≤0.5mm | Same | |
| Filter | AC | 50Hz/60Hz/Off | 50Hz/60Hz/Off | Same |
| | DFT | 0.01Hz/0.05Hz/0.32Hz/0.67Hz | 0.01Hz/0.05Hz/0.32Hz/0.67Hz | |
| | EMG | 25Hz/35Hz/45Hz/Off | 25Hz/35Hz/45Hz/Off | |
| | LOWPASS Filter | 300Hz/270Hz/150Hz/<br>100Hz/75Hz | 300Hz/270Hz/150Hz/100Hz/75Hz | |
| | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.