K170975 · Televere Systems · MQB · May 1, 2017 · Radiology
Device Facts
Record ID
K170975
Device Name
Televere Digital Imaging System
Applicant
Televere Systems
Product Code
MQB · Radiology
Decision Date
May 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The Televere Digital Imaging System is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used. Allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities. Not for mammography.
Device Story
Televere Digital Imaging System captures digital X-ray images for general radiography. System comprises PerkinElmer XRpad 4343F-N flat panel detector (FPD), PC-based computer, power supply, and TigerView Professional software. FPD uses amorphous silicon sensor array with gadolinium oxysulfide (GadOx) scintillator to convert X-rays to digital signals. Software manages image acquisition, display, editing (resize, contrast, crop), storage, printing, and distribution within PACS environment. Used in clinical settings by trained personnel; integrates with high-frequency X-ray generators via interface power unit or automatic exposure detection (AED). Provides high-quality images for physician review to support clinical decision-making. Benefits include digital workflow efficiency and diagnostic image quality comparable to or better than predicate systems.
Clinical Evidence
Bench testing included electrical safety (IEC 60601-1) and EMC (IEC 60601-1-2) compliance. Performance testing followed FDA guidance for solid-state digital X-ray panels, including MTF and DQE measurements. Clinical images were provided for inspection but were not required to establish substantial equivalence. No clinical trials were conducted.
Technological Characteristics
Amorphous silicon sensor array with GadOx scintillator; 4318 x 4320 pixel matrix; 100 µm pixel pitch; 16-bit A/D conversion. Connectivity via Gigabit Ethernet. Power: 100-240 Vac. Dimensions: 460 x 460 x 15 mm. Software: TigerView Professional (PACS). Standards: IEC 60601-1, IEC 60601-1-2, UL Listed.
Indications for Use
Indicated for digital radiographic imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities in patients requiring general radiographic examinations. Not indicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
ViZion + DR, Digital Flat Panel X-Ray Detector System (K123644)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 1, 2017
Televere Systems % Robert Bakin Regulatory Consultant Technology & Business Law Advisors, LLC 1244 Capuchino Avenue Burlingame, California 94010
Re: K170975
Trade/Device Name: Televere Digital Imaging System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: March 21, 2017 Received: April 3, 2017
Dear Robert Bakin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michal D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170975
Device Name Televere Digital Imaging System
#### Indications for Use (Describe)
The Televere Digital Imaging System is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used. Allows imaging of the skull, chest, shoulders, pelvis, and extremities. Not for mammography.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary 510(k) Number K170975
Image /page/3/Picture/1 description: The image shows the logo for TELEVERE SYSTEMS. The word "TELEVERE" is in orange, and the word "SYSTEMS" is in gray. To the right of the logo are the words "ACQUIRE, STORE, VIEW, SHARE" in gray.
Televere Systems, LLC 1160 Cavalier Road Arnold, MD 21012 Email: rquaal@tigerview.com Establishment Registration Number: 2954358 Phone: 571-215-3507 CONTACT: Robert E. Bakin Email: rbakin@tblawadvisors.com Technology & Business Law Advisors, LLC
1244 Capuchino Avenue, Burlingame, CA 94010 Prepared April 26, 2017
## 1. Identification of the Device:
Proprietary-Trade Name: Televere Digital Imaging System Classification Name: Stationary X-ray System Product Code MQB, Common/Usual Name: Digital X-Ray Receptor Panel Device Class/Regulation Number: Class II per regulation 892.1680
- 2. Equivalent legally marketed device: ViZion + DR, Digital Flat Panel X-Ray Detector System K123644 Classification Name: Stationary X-ray System Product Code MQB, Common/Usual Name: Digital X-Ray Receptor Panel Device Class/Regulation Number: Class II per regulation 892.1680
## 3. Associated legally marketed devices:
The Software, TigerView Professional (PACS Software) K061035. This is the software used with this device.
Classification Name: System, Image Processing, Radiological, Code LLZ Common/Usual Name: Picture Archiving and Communications System Device Class/Regulation Number: Class II per regulation 892.2050. Similar Flat Panel Detector: XRpad 4343F MED Flat Panel Detector, K142698 Product Code MQB Classification Name: Stationary X-ray System Common/Usual Name: Digital X-Ray Receptor Panel Device Class/Regulation Number: Class II per regulation 892.1680
- 4. Indications for Use (intended use) The Televere Digital Imaging System is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used. Allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities. Not for mammography.
{4}------------------------------------------------
- 5. Description of the Device: The Televere Digital Imaging System consists of a combination of digital imaging software, a flat panel display (FPD), a PC-based computer and a power supply. The imaging software has previously been cleared by FDA (i.e., K061035, TigerView Professional). This previously cleared software provides basic image adjustment features: An image management system that allows the physician to acquire, display, edit (e.g., resize, adjust contrast, crop, etc.), review, store, print, and distribute medical images within a Picture Archiving and Communication System (PACS) environment. TigerView Professional runs on standard PCcompatible computers and is compatible with capture devices which attach to the computer using a Network Adaptor, USB port, PCI slot, parallel port, memory card, S-video port on a video capture card, or SCSI card. The FPD component seeking clearance (PerkinElmer Model XRPad 4343F N) has not been previously cleared by FDA. The FPD uses a large-area amorphous silicon sensor array with a gadolinium oxysulfide (GadOx) scintillator for displaying high quality images over a wide range of dose settings and is intended to be integrated into a complete X-ray system. This premarket notification seeks clearance for a combination finished device consisting of the combination of the previously cleared digital imaging software and the FPD. The combination of the digital imaging software and the FPD does not affect the safety or efficacy of either component device alone, or in combination.
Integration-level requirements/restrictions: This device will work with most modern highfrequency X-ray generators currently on the market. Sync with the generator is achieved through the interface power unit, supplied. Alternately the panel has automatic exposure detection (AED). If this mode is used, the precautions listed in the panel user manual must be observed. The user should confirm that the generator/tubestand combination proposed employs only Electronic Product Radiation Control certified components. The user should use only trained service engineers during the integration process and fully test the system prior to use on patients.
- 6. Safety and Effectiveness, comparison to predicate device. The results of clinical image inspection, bench, and test laboratory results indicates that the new device is as safe and effective as the predicate device. Clinical images collected demonstrate equal or better image quality as compared to our predicate.
| | Viztec LLC: ViZion + DR, Digital Flat Panel X-Ray<br>Detector System K123644 | Televere Digital Imaging System K162521<br>Using XRpad 4343F-N panel |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | ViZion + DR is intended for digital image<br>capture use in general radiographic<br>examinations, wherever conventional screen-<br>film systems may be used, excluding<br>fluoroscopy, angiography and mammography.<br>ViZion + allows imaging of the skull, chest,<br>shoulders, spine, abdomen, pelvis, and<br>extremities. | The Televere Digital Imaging System is<br>intended for digital image capture use in<br>general radiographic examinations,<br>wherever conventional screen-film systems<br>may be used. Allows imaging of the skull,<br>chest, shoulders, spine, abdomen, pelvis,<br>and extremities. Not for mammography.<br>(Essentially identical) |
| Configuration | Digital Panel and Software only, no generator<br>or stand provided. | SAME |
| Digital Panel | 3072×3072 pixels with 139µm pitch | 4318 × 4320 100 µm Pixel Pitch. Better<br>Resolution |
| Internal<br>image storage | 200 images | 1 GB DDR3, 4 GB SDHC card, number of<br>images depends on image size. Example for<br>10 mb image there would be room for 500<br>images. |
## 7. Substantial Equivalence Chart
{5}------------------------------------------------
| | Viztec LLC: ViZion + DR, Digital Flat Panel X-Ray<br>Detector System K123644 | Televere Digital Imaging System K162521<br>Using XRpad 4343F-N panel |
|---------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Image<br>acquisition<br>time | ≤ 5 Seconds | SAME |
| DICOM 3 | YES | YES via our software cleared in K061035 |
| A/D<br>Conversion | 12 Bit | 16 Bit |
| Scintillator | Gadolinium Oxysulfide (GadOx) | Gadolinium Oxysulfide (GadOx) |
| MTF | 54% (1 cy/mm), 23% (2 cy/mm) (not measured<br>at 4 cy/mm) for RQA5 | 55% (1 cy/mm), 25% (2 cy/mm) 5% (4<br>cy/mm) for RQA5. Better than predicate |
| DQE | 30% (0 cy/mm), 18% (1 cy/mm), 3% (3 cy/mm)<br>for RQA5. | 40% (0 cy/mm), 30% (1 cy/mm), 10% (3<br>cy/mm) for RQA5. Better than predicate |
| Interface | Gigabit Ethernet Port | SAME |
| Panel<br>Dimensions | 495 (w) x 510 (l) x 38 mm (h) | 460 mm (w) × 460 mm (l) × 15 mm (h) |
| Power Source | External Power Supply 100-240 Vac | SAME |
| Electrical<br>safety and<br>EMC | Electrical Safety per IEC 60601-1 and EMC per<br>IEC 60601-1-2. UL Listed | SAME |
The comparison table reveals there are no new issues of safety or effectiveness raised by the modification to the scintillator of the digital x-ray panel.
- 7. Summary of Bench Testing Conducted: IEC Standards have been employed for: Electrical Safety and Electromagnetic Compatibility. The panel power supply is UL Listed. We compared MTF and DQE measurements for the original and modified scintillator versions of the panel and found them to be similar enough not to make a notable difference in acquired images. As compared to the predicate panel the MTF and DQE performance of the new panel is superior. The digital panel manufacturer conducted performance testing according to the FDA guidance document for solid state digital x-ray panels. Risk Analysis and System operation verification tests were conducted in accordance with FDA guidance documents. Since no software modifications were required, confirmation testing was performed. Labeling was developed to comply with the FDA solid state panel guidance document.
- 8. Summary of Clinical Testing: Sample clinical images from the new device were also provided however they were not necessary to demonstrate substantial equivalence.
- 9. Conclusion: After analyzing bench, clinical image, and external laboratory testing to applicable standards, it is the conclusion of Televere that the Televere Digital Imaging System K162521 is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device. The new and predicate devices provide values for MTF and DQE that are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.