K170856 · Contec Medical Systems Co.,Ltd · IYO · Nov 17, 2017 · Radiology
Device Facts
Record ID
K170856
Device Name
CMS600P2 B-Ultrasound Diagnostic System
Applicant
Contec Medical Systems Co.,Ltd
Product Code
IYO · Radiology
Decision Date
Nov 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
CMS600P2 B-Ultrasound Diagnostic System in a general-purpose, digital ultrasound diagnostic system for abdomen,gynecology, obstetric, urology,and small-parts application. The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, transvaginal, peripheral vascular,and musculo-skeletal (both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate. The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional. This device is not intended for home use.
Device Story
CMS600P2 is a digital, general-purpose ultrasound diagnostic system; utilizes convex and linear transducers (2.0-10.0 MHz) to acquire ultrasound signals; employs digital beam forming (DBF), real-time dynamic aperture (RDA), and dynamic receive focusing (DRF) to process signals; produces real-time B-mode, M-mode, and combined B/M images on an integrated LCD display. Used in clinical settings by physicians or qualified healthcare professionals; not for home use. Provides measurement, calculation, and reporting features for clinical diagnosis. Benefits include high-resolution, stable imaging for various anatomical assessments; supports image storage, cine loops, and management functions to assist clinical decision-making.
Clinical Evidence
No clinical study included. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (IEC 61157), and performance testing (IEC 60601-2-37). Biocompatibility testing performed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Digital ultrasound system; uses convex and linear transducers. Materials: ABS probe cover, silicone elastomer acoustic lens. Technologies: DBF, RDA, DRA, DRF, DFS. Connectivity: Standalone system with LCD display, keypad, and video memory. Power: DC15V/Adapter 100-240V. Software: Embedded operating system. Sterilization: Non-sterile.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in adults, pregnant women, pediatric, and neonate patients. Applications include abdomen, gynecology, obstetrics, urology, small parts, fetal organ, neonatal cephalic, transvaginal, peripheral vascular, and musculoskeletal (conventional and superficial) studies.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
EMP-2100 Full Digital Ultrasound Diagnostic Device (K081873)
Submission Summary (Full Text)
{0}------------------------------------------------
November 17, 2017
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Contec Medical System Co., Ltd. % Mr. Ray Wang General Manager Beijing Believe Technology Service Co., Ltd. 5-1206. Build 332. DaFangJu. No.25 BanBiDian Road LiYuan Town, TongZhou District, Beijing, 101121 CHINA
Re: K170856
Trade/Device Name: CMS600P2 B-Ultrasound Diagnostic System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: October 18, 2017 Received: October 23, 2017
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Mr. Ray Wang
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K170856
#### Device Name
CMS600P2 B-Ultrasound Diagnostic System
CMS600P2 B-Ultrasound Diagnostic System in a general-purpose, digital ultrasound diagnostic system for abdomen,gynecology, obstetric, urology,and small-parts application.
The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, transvaginal, peripheral vascular,and musculo-skeletal (both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate.
The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional.
This device is not intended for home use.
Please refer to the acoustic output declaration for each transducer as following pages.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
EF
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
System: CMS600P2 B-Ultrasound Diagnostic System
Transducer:
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|----------------------|----------------------------------|-------------------------|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Dopple<br>r | Combine<br>d<br>(Specify)<br>B/M | Other*<br>(Specify<br>) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | N | N | | | | N | |
| | Abdominal | N | N | | | | N | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| Fetal<br>Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | | | | N | |
| | Small Organ (Specify) | N | N | | | | N | Note 2,3 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | N | N | | | | N | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal(Conve<br>ntional) | N | N | | | | N | |
| | Musculo-skeletal<br>(Superficial) | N | N | | | | N | |
| | Intravascular | | | | | | | |
| | Other (Specify) | N | N | | | | N | Note 1 |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | | | | N | |
| | Other (Specify) | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combine<br>d<br>(Specify)<br>B/M | Other*<br>(Specify<br>) |
| Ophthalm<br>ic | Ophthalmic | | | | | | | |
| | Fetal | N | N | | | | N | |
| | Abdominal | N | N | | | | N | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| Fetal<br>Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | | | | N | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal(Conve<br>ntional) | N | N | | | | N | |
| | Musculo-skeletal<br>(Superficial) | N | N | | | | N | |
| | Intravascular | | | | | | | |
| | Other (Specify) | N | N | | | | N | Note 1 |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combine<br>d<br>(Specify) | Other*<br>(Specify) |
| Ophthalm<br>ic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | N | N | | | | N | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | Other*<br>(Specify) |
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Dopple<br>r | Combine<br>d<br>(Specify) | |
| Ophthalm<br>ic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| Fetal<br>Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | N | N | | | | N | Note 2,3 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | | | | N | |
| | Other (Specify) | | | | | | | |
Additional comment:Combined mode : B/M
Note 1:urology: prostate and vesica urinaria; Note 2: Thyroid Gland; Note 3: Hip Joint
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{4}------------------------------------------------
### System: CMS600P2 B-Ultrasound Diagnostic System
Transducer:
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Additional comment:Combined mode : B/M
Note 1:urology: prostate and vesica urinaria
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging,
Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{5}------------------------------------------------
#### System: CMS600P2 B-Ultrasound Diagnostic System E6.5-80R13-A16A Transducer:
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Additional comment:Combined mode : B/M
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{6}------------------------------------------------
### System: CMS600P2 B-Ultrasound Diagnostic System Transducer:
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Additional comment:Combined mode : B/M Note 2: Thyroid Gland
Note 3: Hip Joint
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{7}------------------------------------------------
# Tab #4 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K170856
- 1. Date of Preparation: 11/18/2017
- 2. Sponsor Identification
Contec Medical System Co., Ltd. No.112 Qinhuang West Street, Economic & Technical Development Zone, Qinhuangdao, Hebei, 066004, China. Establishment Registration Number: 3006979678
Contact Person: Xueyong Li Position: Quality Manager Tel: 86-355-8015490 Fax: 86-355-8015490 Email:_lxy1011@163.com
- 3. Designated Submission Correspondent
Mr. Ray Wang
Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd., LiYuan Town, TongZhou District, Beijing, 101121, China
Tel: +86-18910677558 Fax: +86-10-52214696 Email: ray.wang@believe-med.com
{8}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: CMS600P2 B-Ultrasound Diagnostic System Common Name: Diagnostic Ultrasound System with Accessories Model(s): CMS600P2
Regulatory Information
Classification Name: System, Imaging, Pulsed Echo, Ultrasonic ; Transducer, Ultrasonic, Diagnostic
Classification:II Product Code:IYO & ITX Regulation Number:21 CFR 892.1560 & 21 CFR 892.1570 Review Panel:Radiology:
Intended Use Statement:
CMS600P2 B-Ultrasound Diagnostic System in a general-purpose, digital ultrasound diagnostic system for abdomen,gynecology, obstetric, urology,and small-parts application.
The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, transvaginal, peripheral vascular,and musculo-skeletal (both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate.
The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional.
This device is not intended for home use.
Please refer to the acoustic output declaration for each transducer as following pages.
#### Device Description
● The apparatus is high resolution ultrasound diagnostic system, which adopts digital beam forming(DBF), real-time dynamic aperture(RDA), real-time dynamic beam apodization (DRA), real-time dynamic receive focusing (DRF), digital-control dynamic frequency scanning (DFS), frame correlation etc. technologies. The image is clear, stable, and high resolution.
● It adopts embedded operating system, which greatly improve the product security and compatibility, and enhance the data processing ability.
● The apparatus is taken more conveniently for its humane operation interface and management system, flexible expansibility and compatibility, pop-up menu and keyboard design.
● Users can chose different kinds of display modes,such as:B, 2B, 4B, BM, M. Image real-time, freeze, memory, calling up and large capability cine loop can be realized; the device possesses multilevel scanning depth, dynamic range, sound output power and focus number, focus space, focus position adjusting etc. functions.
● With controllable frame correlation, gamma correction, image reversing, histogram, depth advance, zoom magnification, no distortion real-time part magnification etc. image processing functions. It is convenient for choice of optimal diagnostic image with image pigeonhole, browse, management functions. Measure, calculation, and report features are provided to facilitate image measurement.
{9}------------------------------------------------
- The device supports linear probe and convex probe. The frequency of the device is 2.0
MHz-10.0 MHz. It has extensive application that is suitable for clinic diagnosis, such as abdomen, obstetrics, gynecology, small organs, cardiology and urology etc. .
- The device is consist of three parts: mainframe, transducer(probe), adapter.
- The transducers provided are listed below in table 7-1, the specification of the system and transducers please refer to Tab 9 section 7 Technical Specifications. The picture and design drawing of the subject device and its transducer is presented in Tab 9 section 2 Device Illustration.
| Transducer Model | Type | Frequency | Application |
|------------------|--------------------|-----------|------------------------------|
| C3.5-80R60-A16A | convex probe | 3.5 MHz | Abdomen, obstetrics, urology |
| L7.5-80L40-A16A | linear probe | 7.5 MHz | small organs |
| E6.5-80R13-A16A | endo-vaginal probe | 6.5 MHz | Gynecology and transvaginal |
Image /page/9/Figure/6 description: The image shows the title of a table. The title is "Table 7-1 Transducer List". The title is written in a clear, sans-serif font and is centered on the page.
Image /page/9/Figure/7 description: The image shows the text "Fig 7-1 Working Frame of CMS600P2". The text appears to be a figure label or title, possibly from a technical document or manual. The text indicates that the figure is related to the working frame of a device or system named CMS600P2. The figure is labeled as "Fig 7-1", suggesting it is the first figure in chapter 7.
Image /page/9/Figure/8 description: The image shows a block diagram of a system. The system includes components such as a probe, beamforming, A/D converter, TGC control, multi-route pulse control, FPGA, image processing, arithmetic software package, ARM, LCD display, image and video memory, and keypad input. The components are connected by arrows, indicating the flow of data or control signals within the system.
- ર . Identification of Predicate Device(s)
Predicate Device :
510(k) Number: K081873 Product Name: Full Digital Ultrasound Diagnostic Device Model Name: EMP-2100 Manufacturer: Shenzhen Emperor Electronic Technology Co., Ltd.
{10}------------------------------------------------
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- a. IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance
- b. IEC 60601-1-2:2007, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Tests
- IEC 60601-2-37:2007 Medical electrical equipment Part 2-37: Particular requirements for the C. basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- IEC61157:1992 Requirements for the Declaration of the Acoustic Output of Medical Ultrasound d. Diagnostic Apparatus
- ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity. e.
- f. ISO 10993-10:2010 Standard, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity
- Clinical Test Conclusion 7.
No clinical study is included in this submission.
{11}------------------------------------------------
# 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device(s) | Predicate Device(s) | Remark |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Device name | CMS600P2 B-Ultrasound Diagnostic System | EMP-2100 Full Digital<br>Ultrasound Diagnostic Device | / |
| Classification Name | System, Imaging, Pulsed echo,<br>Ultrasonic<br>Transducer, Ultrasonic, Diagnostic | System, Imaging, Pulsed echo,<br>Ultrasonic<br>Transducer, Ultrasonic, Diagnostic | SE |
| Product Code | IYO<br>ITX | IYO<br>ITX | SE |
| Regulation Number | 892.1560<br>892.1570 | 892.1560<br>892.1570 | SE |
| Comparison<br>Statement | The proposed device has same classification information as the predicate device. | | |
| Intended Use | CMS600P2 B-Ultrasound Diagnostic System in a<br>general-purpose, digital ultrasound diagnostic system for<br>abdomen,gynecology, obstetric, urology,small-parts, and<br>cardiology application.<br>The system is intended to use for the following type of studies:<br>fetal organ, abdominal, pediatric, small organs, neonatal<br>cephalic, cardiac, transvaginal, peripheral vascular,and<br>musculo-skeletal (both conventional and superficial). The<br>device is intended to adult, pregnant woman, pediatric and<br>neonate.<br>The system is a prescription device intended to be used by or on | EMP-2100 Full Digital Ultrasound Diagnostic Device in a<br>general-purpose, digital ultrasound diagnostic system for<br>abdomen, gynecology,obstetric, urology,small-parts,and<br>cardiology application.<br>The system is intended to use for the following type of<br>studies:fetal organ, abdominal, pediatric, small organs,<br>Neonatal cephalic, cardiac, transvaginal, peripheral<br>vascular,and musculo-skeletal (both conventional and<br>superficial). The device is intended to adult, pregnant<br>woman, pediatric and neonate.<br>The system is a prescription device intended to be used by or | SE |
# Table 7-1 Comparison of Technology Characteristics
{12}------------------------------------------------
| the order of a physician or similar qualified health care | | | on the order of a physician or similar qualified health care | | | | | | |
|-----------------------------------------------------------|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|-------------------------------------|---------------------------------------|-----------------------------------------------------------------------------------------------------|--------------------------|------------------|-----------|
| professional. | | | professional. | | | | | | |
| This device is not intended for home use. | | | This device is not intended for home use. | | | | | | |
| Comparison | The proposed device has similar intended use as the predicate device. | | | | | | | | |
| | Main Unit Technical Specifications | | | | | | | | |
| Probe mode | C3.5-80R60<br>-A16A<br>Convex probe | L7.5-80L40-<br>A16A<br>linear | E6.5-80R13-<br>A16A<br>endo-vaginal | C128-50<br>Convex | L096-42C<br>Linear | C080-13G<br>Micro Convex | C096-20A<br>Micro Convex | / | |
| | Scan angle/width | 110° | 6.3mm | 114.6° | 128° | 40mm | 120° | 88° | Analyse1 |
| | Maximum number of<br>Active elements for a<br>single pulse | 5 | 5 | 5 | 32 | 32 | 32 | 32 | Analyse 1 |
| Frequency | 3.5MHz | 7.5MHz | 6.5MHz | 3.5MHz | 6.5MHz | 6.5MHz | 3.5MHz | Analysel | |
| Number of Element | 80 | 80 | 80 | 128 | 96 | 80 | 96 | Analyse1 | |
| Size of Element(mm) | 0.48x7.5 | 0.5x6.3 | 0.33x7.3 | 0.41x13.4 | 0.34 x8.2 | 0.26 x8.2 | 0.24 x11.4 | Analyse 1 | |
| Spacing of<br>element(mm) | 0.48 | 0.5 | 0.33 | 0.49 | 0.42 | 0.34 | 0.32 | Analysel | |
| Array<br>dimensions(mm) | 38.4x7.5 | 40x6.3 | 26.4x7.3 | 62.8x13.4 | 40.3 x8.2 | 27.2 x8.2 | 30.7 x11.4 | Analyse 1 | |
| Detect Depth(mm) | ≥160 | ≥50 | ≥40 | ≥200 | ≥90 | ≥90 | ≥90 | Analysel | |
| Resolution<br>(mm) | Lateral | ≤3(Depth≤80)<br>≤4(80≤130) | ≤2(Depth≤<br>40) | ≤2(Depth<br>≤30) | ≤3(depth≤8<br>0)≤4(80 <<br>depth≤130) | ≤2(depth<br>≤60) | ≤2(depth<br>≤60) | ≤2(depth<br>≤40) | Analyse 1 |
| | Axial | ≤2(Depth≤80)<br>≤3(80 ≤1(Depth<br>≤50) ≤1(Depth<br>≤40) ≤1(80<<br>depth≤130) ≤0.5<br>(depth≤80) ≤0.5<br>(depth ≤60) ≤0.5<br>(depth≤40) Analyse 1 | | | | | | | |
| | | ≤130) | ≤5 | ≤3 | ≤4 | ≤3 | ≤3 | ≤3 | Analyse1 |
| Blind Zone(mm) | | | ≤5 | ≤3 | ≤4 | ≤3 | ≤3 | ≤3 | Analyse1 |
| Focus Position | | 4 focus, adjustable position | | 4 focus, position is not adjustable | | 4-steps dynamic focusing | | | Analyse1 |
| Monitor Size | | 10.1 inch | | | | 10 inch | | | SE |
| Display Mode | | B, 2B, BM, M, 4B | | | | B, B+B, B+M, M, 4B | | | SE |
| Image gray scale | | 256 level | | | | 256 | | | SE |
| Image Storage | | 2048 frame | | | | 16 | | | Analyse1 |
| Cine Loop | | 600 frame | | | | 256 frame | | | Analyse1 |
| Image Flip | | Up/down, left/right, black/white | | | | Up/down, left/right | | | Analyse1 |
| Image Process | | Gamma (gray scale) correction, tissue harmonic imaging,<br>histogram, local magnification | | | | Dynamic range, edge enhancement, frame correlation, smooth,<br>IP, gray map,r-correction, rejection | | | Analyse1 |
| Measurement | | Distance, circumference, area, volume, heart, pregnant age,<br>fetal weight, expected date | | | | Distance, circumference, area,volume, histogram, profile, time<br>velocity heart rate | | | Analyse1 |
| Notation | | Date, time, name, No., sex, age, doctor, hospital name, probe<br>frequency, etc. | | | | Date, Time, Name, Sex, Age, Doctor, Hospital name, probe<br>frequency, etc. | | | SE |
| Net weight | | 2.3 kg (include probe) | | | | 10kg | | | SE |
| Power supply | | DC15V; Adapter:100 V~240 V, 50 Hz/60 Hz | | | | AC 100-240V, 47-60Hlz | | | Analyse2 |
| Dimensions(mm) | | 292 mm(L) × 232 mm(W) × 45 mm(H) | | | | 365 (height) X292 (width) X380 (depth) | | | SE |
| Configuration | | mainframe, transducer(probe) | | | | mainframe, transducer(probe) | | | SE |
| Acoustic | Ispta. 3 | | | | | | | | |
| Output<br>Isppa. 3 | | Meet the requirements of Track 3 | | | | Meet the requirements of Track 1 | | | Analyse1 |
| Parameter | MI | | | | | | | | |
| Skin<br>Contacted | | Probe Cover (ABS) | | | | Probe Cover (ABS) | | | SE |
| Material | | Acoustic Lens (Silicone elastomer) | | | | Acoustic Lens (Silicone elastomer) | | | SE |
{13}------------------------------------------------
# 510(k) Summary
{14}------------------------------------------------
| Operation<br>Environment | Temperature: +10°C~ +40°C<br>Relative humidity: 30%~ 75%<br>Atmosphere pressure: 700 hPa ~ 1060 hPa | Temperature 5°C -40°C ,<br>relative humidity up to 90%RH<br>air pressure 70-106Kpa | SE |
|--------------------------|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|----------|
| Storage environment | Temperature: -10°C~+55°C,<br>Relative Humidity: ≤93%, no condensation.<br>Atmosphere pressure: 700 hPa ~ 1060 hPa, | Temperature -5-40°C ,<br>relative humidity up to 90%RH (no water I drop)<br>air pressure 70-106Kpa | SE |
| Sterile | No | No | SE |
| Single Use | No | No | SE |
| Comparison<br>Statement: | The proposed device has the similar main unit specifications with the predicate device. | | |
| Applied Standards: | | | |
| Biocompatibility | ISO10993-5&ISO10993-10 | ISO10993-5&ISO10993-10 | SE |
| Electrical Safety | IEC60601-1 | IEC60601-1 | SE |
| EMC | IEC60601-1-2 | IEC60601-1-2 | SE |
| Performance | IEC 60601-2-37;IEC61157 | UD2 | Analyse1 |
| Comparison<br>Statement | The proposed probe has similar applied Standards with the predicate device. | | |
{15}------------------------------------------------
9. Substantially Equivalent (SE) Conclusion
SE Analysis :
The subject device has same classification information, same intended use, same indication for use, similar product design, similar specification, same safety elements, similar applied Standards as predicate device.
The differences are included as followings:
Analyse 1: The device and predicate device have difference in performance specification, such as scan/angle/width, maximum number of elements for a single pulse, frequency, number of element, size of element, spacing of element, array dimensions, detect depth and resolution. But the propose device have tested for measurement accuracy by accuracy testing and software validation, so these difference can prove the effectiveness of propose device.
The proposed device has tested according with the safety standard IEC 60601-1, IEC60601-1-2 for the safety. The proposed device has tested according with the accuracy report, IEC60601-2-37 report, IEC61157 Acoustic Output report for the effectiveness. The proposed device has tested according with the ISO 10993 series standard for the Biocompatibility .
Therefore, the differences above between the proposed device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety. No new technology is applied in the subject device. Most of the main aspects on effectiveness and safety between the proposed device and predicate device are same. The differences are slight so that no substantial influence on the effectiveness and safety.
Analyse 2 : Although the Power supply specifications of CMS600P2 is different from the predicate device,but both the predicate device and the proposed device has passed the IEC60601-1 safety test, we believe these differences will not affect the effectiveness and safety compared with the predicate device.
Conclusion: The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.