EMP-2100 FULL DIGITAL ULTRASOUND DIAGNOSTIC DEVICE
Applicant
Shenzhen Emperor Electronic Technology Co. , Ltd.
Product Code
IYO · Radiology
Decision Date
Jul 16, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
EMP-2100 Full Digital Ultrasound Diagnostic System is a general-purpose, digital ultrasound diagnostic system for abdomen, gynecology, obstetric, urology, small-parts, and cardiology application. The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, cardiac, transvaginal, peripheral vascular, and musculo-skeletal(both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate. The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional. This device is not intended for home use.
Device Story
Portable digital B/W ultrasound system; uses array of probes (linear, convex, microconvex) to emit ultrasound waves into body tissue; captures returned echoes; processes signals to display images in B, B/B, 4B, B/M, and M modes. Operated by physicians or qualified healthcare professionals in clinical settings. Provides real-time visualization of internal structures; aids in diagnostic examinations of abdomen, gynecology, obstetrics, urology, small parts, and cardiology. Output displayed on 10" SVGA monitor; assists clinicians in anatomical assessment and diagnostic decision-making. Benefits patients through non-invasive diagnostic imaging.
Clinical Evidence
Bench testing only. Clinical measurement effectiveness test conducted to evaluate measurement accuracy. Electrical safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2) testing performed. Biocompatibility testing (cytotoxicity, sensitization, irritation) conducted per ISO 10993-1:2003.
Technological Characteristics
Portable B/W ultrasound system; 10" SVGA display; 256 gray scales; 2.5-7.5 MHz frequency range. Modes: B, B/B, 4B, B/M, M. Digital scanning converter 576x460x8 bits. Power: 100-240V AC. Weight: 10 kg. Non-sterile, reusable transducers. Connectivity: PAL-D video output.
Indications for Use
Indicated for diagnostic ultrasound imaging in adults, pregnant women, pediatric patients, and neonates for studies including fetal organ, abdominal, pediatric, small organs, neonatal cephalic, cardiac, transvaginal, peripheral vascular, and musculo-skeletal (conventional and superficial). Not for home use.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
DP-6600 Digital Ultrasonic Diagnostic Imaging System (K060949)
Submission Summary (Full Text)
{0}------------------------------------------------
# Ko81873
JUL 1 6 2008
moer of Shenzhen Emperor Electronic Technology Co., Ltd
# Appendix VI 510(k) Summary
As Required by CFR 807.92
1. Submission Date: May 21, 2008
#### 2. Sponsor Information
Shenzhen Emperor Electronic Technology Co., Ltd. 2/F, Building7, Tian'an Nanyou Industrial Area, Nanshan District, Shenzhen, 518054, China Contact Person: Ms. Cheng Guiyuan Tel: +86-755-26073381 Fax: +86-755- 26419886
#### 3. Submission Correspondent:
Ms. Diana Hong Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D, Zhongshan Zhongxin Mansion No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030, China.
#### 4. Subject Device Information
- Trade/Proprietary Name: EMP-2100 Full Digital Ultrasound A Device Diagnostic Device
- Device Classification Name: System, Imaging, Pulsed echo, Ultrasonic ア Transducer, Ultrasonic, Diagnostic
- A Device Common Name: Ultrasonic pulsed echo imaging system
- ア Review Panel: Radiology
- ア Product Code: 90-IYO & 90-ITX
- ア Regulation Number: 892.1560
- > Device Class: II
- A Review Category: Tier II
- A Intended Use:
EMP-2100 Full Digital Ultrasound Diagnostic System is a general-purpose, digital ultrasound diagnostic system for abdomen, gynecology, obstetric, urology, small-parts, and cardiology application.
The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, cardiac, transvaginal,
{1}------------------------------------------------
# Peror Shenzhen Emperor Electronic Technology Co., Ltd
peripheral vascular, and musculo-skeletal(both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate.
The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional. This device is not intended for home use.
#### 5. Predicate Device:
DP-6600 Digital Ultrasonic Diagnostic Imaging System with added transducers K-number: K060949
### Product Code: IYO
#### Intended Use:
The system is a general-purpose, fully digital ultrasound system for abdominal, gynecologic and obstetric, small parts, and cardiac applications.
The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, cardiac, transvaginal, and peripheral vascular and musculo-skeletal (both conventional and superficial).
This device is intended to adult, pregnant woman, pediatric and neonate. The system is a prescription device intended to be used by or on the order of a physician or similarly qualified health care professional. This Device is not intended for home use.
#### Manufactured by:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
#### 6. Device Description
EMP-2100 Full Digital Ultrasound Diagnostic Device is a portable digital ultrasonic diagnostic B/W system applied in ultrasound diagnostic examination of abdomen, gynecology, obstetric, urology, small-parts, and cardiology application.
It is designed to produce ultrasound waves into body tissue and to present the returned echo information on the monitor. The resulting information is displayed in five display modes as listed below. This system is a Track 1 device that employs an array of probes that include linear array, convex linear array and microconvex linear array with a frequency range of approximately 2.5MHz~7.5MHz.
The subject device is not for life-supporting or life-sustaining, not for implant. The device or transducers are not provided sterile and the transducer does not need sterilization and the transducer is reusable but does not need re-sterilization.
{2}------------------------------------------------
Shenzhen Emperor Electronic Technology Co., Ltd
| Transducer Model | Type | Nominal<br>Frequency | Application | Track |
|------------------|--------------|----------------------|-----------------------------------------------------------------------------|-------|
| C080-13G | Micro Convex | 6.5 MHz | gynecology, obstetric, urology and transvaginal | 1 |
| L096-42C | Liner Probe | 6.5 MHz | small-parts, neonatal cephalic, peripheral vascular and<br>musculo-skeletal | 1 |
| C096-20A | Micro Convex | 3.5 MHz | cardiology, pediatric, abdomen, gynecology and obstetric | 1 |
| C128-50 | Convex | 3.5 MHz | abdomen, pediatric, gynecology and obstetric | 1 |
The system consists of a main unit, a display and four transducers.
# 7. Technology Characteristics
Table VI-2 Technology Characteristics
| Display Mode | B,B/B,4B,B/M,M |
|-----------------------------|----------------------------------------------------------------------------------------------|
| Gray Scales | 256 |
| Frame rate | 60 frames/second max.<br>(various according different probes and the frequency) |
| Display | 10"SVGA |
| Video output | PAL-D |
| Digital Scanning Converter | 576 ×460 ×8 bits |
| Zoom | 11 levels, partial zoom |
| Body mark | 38 types indicating probe positions |
| Power Supply | AC 100~240V, 47~60Hz |
| Input Power | 200 VA |
| Continuous Operation Period | >8 hours |
| Dimensions | 365 (height) ×292 (width) ×380 (depth) |
| Weight | Net weight 10 Kg |
| Working Condition | Temperature 5°C -40°C, relative humidity up to 90%RH |
| Storage Condition | Temperature -5-40°C, relative humidity up to 90%RH (no water<br>drop) air pressure 70-106Kpa |
## 8. Effectiveness and Safety Considerations
### Effectiveness:
Clinical Measurement Effectiveness Test was conducted on the subject device to evaluate its measurement accuracy.
{3}------------------------------------------------
Shenzhen Emperor Electronic Technology Co., Ltd
Safety Considerations:
The Electrical Safety Testing following IEC 60601-1 and Electromagnetic Compatibility Testing following IEC 60601-1-2 was conducted on the subject device
Guideline "INFORMATION FOR MANUFACTURERS Per FDA SEEKING MARKETING CLEARANCE OF DIAGNOSTIC ULTRASOUND SYSTEMS AND TRANSDUERS" dated May 1, 1997 and with regard to Table I Initial Evaluation Tests for Consideration and Table 2 Supplementary Evaluation Tests for Consideration in ISO 10993-1:2003(E), Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing, the necessary tests for Biocompatibility Testing includes: Cytotoxicity, Sensitization, Irritation.
#### 9. Substantially Equivalent Analysis
The subject device has same classification information, same indication for use, similar product design, same performance effectiveness, performance safety as the predicate device.
The differences are included as followings:
Scanning angle
Linear probe width
Micro-convex probe scanning angle
Power supply: 110V±10%, 230V±10% vs 100~240VAC±10%
Volume and Weight: 306 mm (W)*426 mm (L)*357 mm (H), 12KG vs 86 mm(W)*385 mm(L)*306 mm(H), 11KG
Image Process: the subject device does not have Line Density as the predicate device Storage: 16 images in main unit vs 100 images in main unit
Operation Environment: Temperature 5°C -40°C relative humidity 30-80%RH vs Temperature 5°C-40°C ,relative humidity 30-85%RH
Storage Environment .: Temperature -5-40°C ,relative humidity 30-80%(no water drop) air pressure 86-106Kpa vs Temperature -22℃ -55℃ . relative humidity 30-85% (no condensation) air pressure 70-106Kpa
Display Mode
The differences above between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
No new unique technology is applied in the subject device. Most of the main aspects on effectiveness and safety between the subject device and predicate device are same. The
{4}------------------------------------------------
PCI CI Shenzhen Emperor Electronic Technology Co., Ltd
differences are slight so that no substantial influence on the effectiveness and safety.
## 10. Substantially Equivalent Determination
.
The subject device, EMP-2100 Full Digital Ultrasound Diagnostic Device, is substantially equivalent to the predicate device, DP-6600 Digital Ultrasonic Diagnostic Imaging System with added transducers
:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 1 6 2008
Shenzhen Emperor Electronic Technology Co., Ltd % Mr. Morten Simon Christensen Staff Engineer & FDA Accredited Person Program Coordinator, Program Reviewer Underwriters Laboratories, Inc. UL Silicon Valley Office 455 E. Trimble Road SAN JOSE CA 95131-1230
Re: K081873
Trade/Device Name: EMP-2100 Full Digital Ultrasound Diagnostic System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO and ITX Dated: June 24, 2008 Received: July 2, 2008
Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the EMP-2100 Full Digital Ultrasound Diagnostic System, as described in your premarket notification:
#### Transducer Model Number
| C128-50 |
|----------|
| C080-13G |
| L096-42C |
| C096-20A |
{6}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Robert A. Phillips, Ph.D. at (240) 276-3666.
Sincerely yours.
Arque M. Whan
A Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
# Indication for Use
510(k) Number:
Device Name: EMP-2100 Full Digital Ultrasound Diagnostic System
### Indications for Use:
EMP-2100 Full Digital Ultrasound Diagnostic System is a general-purpose, digital ultrasound diagnostic system for abdomen, gynecology, obstetric, urology, small-parts, and cardiology application.
The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, cardiac, transvaginal, peripheral vascular, and musculo-skeletal(both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate.
The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional. This device is not intended for home use.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
![]()
(Division Sign-off)
Page _1_ of __ 1
Division of Reproductive. Abdomina Radiological Devices 510(k) Number
{8}------------------------------------------------
#### Fill out one form for each ultrasound system and each transducer.
#### Device Name: EMP-2100 Full Digital Ultrasound Diagnostic System (C128-50 Transducer)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | | | |
|-------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | | | | | | N | | |
| Abdominal | | N | N | | | | | | N | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Aoguih Whag
ductive. Andominal 510(k) Nu
{9}------------------------------------------------
#### Fill out one form for each ultrasound system and each transducer.
Device Name: EMP-2100 Full Digital Ultrasound Diagnostic System (C080-13G Transducer) Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation A B M PWD CWD Color Amplitude Color Combined Other Doppler Doppler Velocity (B/M) (specify) Imaging Ophthalmic Fetal Abdominal Intraoperative (specify) Intraoperative Neurologica Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Transesophageal Transrectal Transvaginal N N N Transurethral N N N Intravascular Peripheral Vascular Laparoscopic Musculo-skeletalConventional Musculo-skeletal Superficial
N= new indication; P= previously cleared by FDA; E= added under Appendix E Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Other (specify)
forpaitm Whay
(Division Sian-Off) 510(k) Number
{10}------------------------------------------------
#### Fill out one form for each ultrasound system and each transducer.
# Device Name: EMP-2100 Full Digital Ultrasound Diagnostic System (L096-42C Transducer)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | | |
|------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurologica | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | N | N | | | | | | N | |
| Neonatal Cephalic | | N | N | | | | | | N | |
| Adult Cephalic | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | | | | | | N | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletalConventional | | N | N | | | | | | N | |
| Musculo-skeletal Superficial | | N | N | | | | | | N | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Comments:
が出来ました。 2007年07月27日 10:00:00 【送料無料】 2007年07月10日 10:00 【送料無料】 2007/08/07 11:00 【送料無料】 2007/08/08 11:00 【送料無料】 2007/08/08 11:00 【
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
のお気になる。
Vornthethun
Prescription Use (Per 21 CFR 801.109)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(Division Sign-Off)/
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K081873
{11}------------------------------------------------
#### Fill out one form for each ultrasound system and each transducer.
### Device Name: EMP-2100 Full Digital Ultrasound Diagnostic System (C096-20A Transducer)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) |
|------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | N | N | N | | | | | | N | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurologica | | | | | | | | | | |
| Pediatric | N | N | N | | | | | | N | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletalConventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| Cardiac | N | N | N | | | | | | N | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
______________________________________________________________________________________________________________________________________________________________________________
Tomik Whing
(Division Sign-Off) Reproductive. Abdominal and
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.