K170846 · Jiangsu Caina Technology Co., Ltd. · FMI · Dec 13, 2017 · General Hospital
Device Facts
Record ID
K170846
Device Name
Disposable Insulin Pen Needle
Applicant
Jiangsu Caina Technology Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Dec 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Disposable Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
Device Story
Disposable, single-use sterile needle for subcutaneous insulin delivery. Device consists of a 304 stainless steel cannula, polypropylene hub, polyethylene needle cap, and paper seal. Screws onto standard insulin pen injectors. Used by patients or clinicians in clinical or home settings. Manual operation; no electronic or software components. Provides a fluid path for insulin from the pen reservoir to the subcutaneous tissue. Benefits include reliable insulin administration via standard pen injection systems.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including ISO 7864 (sterile needles), ISO 9626 (tubing), ISO 11608-2 (injection systems), and ISO 10993 series (biocompatibility: cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis). Packaging integrity verified via ASTM F88, F1140, F2096, and F1929.
Technological Characteristics
Materials: 304 stainless steel cannula, polypropylene hub/protector, polyethylene cap, polydimethylsiloxane lubricant. Dimensions: 29G-32G (various lengths). Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 7864, ISO 9626, ISO 11608-2, ISO 10993, ASTM F88, F1140, F2096, F1929, USP <85>.
Indications for Use
Indicated for subcutaneous insulin injection using pen injector devices. Not intended for neonates, newborn infants, or children.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 13, 2017
Jiangsu Caina Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai 200120. China
Re: K170846
Trade/Device Name: Disposable Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: November 9, 2017 Received: November 13, 2017
Dear Ms. Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tina Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170846
Device Name
Disposable Insulin Pen Needle
Indications for Use (Describe)
The Disposable Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K170846
- 1. Date of Preparation: 12/12/2017
- 2. Sponsor Identification
Jiangsu Caina Technology Co., Ltd. No.23, Huanxi Road, Zhutang Town, Jiangyin, Jiangsu, 214425, China
Establishment Registration Number: 3005670221
Contact Person: Jun Lu Position: General Manager Tel: +86-510-86205183 Fax: +86-510-86215183 Email: jun.lu@cainamedical.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)
## Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
- Identification of Subject device 4.
Trade Name: Disposable Insulin Pen Needle Common Name: Insulin Pen Needle
### Regulatory Information
Classification Name: Hypodermic single lumen needle; Classification: II Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital
#### Indications for Use:
The Disposable Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
#### Device Description
The subject device, Disposable Insulin Pen Needle, is a single-use device, which is designed for used with a pen injector for the subcutaneous injection of insulin. It consists of needle tube, inner sheath, outer sheath, hub and sealed paper. The hub can be connected screwed onto the insulin pen. The subject device is not intended for neonates, newborn infants or children. The subject device is used in conjunction with following pen injector during clinical use.
| K Number | Product Name | Manufacturer |
|----------|--------------|-----------------------|
| K123766 | NovoPen | Novo Nordisk Inc |
| K142518 | HumaPen | Eli Lilly and Company |
- 5. Identification of Predicate Device
510(k) Number: K152824 Product Name: MEDT Pen Needle Manufacturer: Shandong Caremed Medical Products Co., Ltd.
#### 6. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
- > ISO 7864: 1993 Sterile hypodermic needles for single use
{5}------------------------------------------------
- > ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices
- > ISO 11608-2:2012 Needle-based injection systems for medical use- Requirements and test methods-Part 2: Needles
- > ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals
- > ASTM F88/F88M-09 Standard test method for seal strength of flexible barrier materials
- > ASTM F1140/F1140M-13 Standard test methods for internal pressurization failure resistance of unrestrained packages
- > USP 39-NF 34 <85> Bacterial Endotoxins Test
- > ASTM F2096-11 Standard Test Method For Detecting Gross Leak in Medical Packaging by Internal Pressurization (Bubble Test)
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
The biocompatibility testing, including cytotoxicity, irritation, sensitization, systemic toxicity and hemolysis were referenced to demonstrate that the patient-contact materials of subject device meets the requirements of following ISO 10993 series standards.
- > ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity;
- > ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization;
- > ISO 10993-11:2006, Biological evaluation of medical devices Part 11: Tests for systemic toxicity;
- > ASTM F756-13, Standard Practice for Assessment of Hemolytic Properties of Materials.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
- 8. Substantially Equivalent (SE) Comparison
| | Item | Subject device | Predicate Device |
|---|---------------------|----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| | | | K152824 |
| | Product Code | FMI | FMI |
| | Regulation Number | 880.5570 | 880.5570 |
| | Indications for Use | The Disposable Insulin Pen Needle is intended<br>for use with pen injector devices for the<br>subcutaneous injection of insulin. | The MEDT Pen Needle is intended for use with pen<br>injector devices for the subcutaneous injection of<br>insulin. |
| C | Cannula | 304 Stainless Steel | 304 Stainless Steel |
#### Table 1 Comparison of Technology Characteristics
{6}------------------------------------------------
| Hub | Polypropylene | Polypropylene |
|--------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Lubricant | Polydimethylsiloxane | Polydimethylsiloxane |
| Needle Cap | Polyethylene | Polyethylene |
| Needle Hub<br>Protector | Polypropylene | Polypropylene |
| Sealed<br>Paper | Paper | Paper |
| Operation Mode | Manual | Manual |
| Needle Gauge | 29G, 30G, 31G, 32G | 29G, 30G, 31G, 32G |
| Needle Dimension<br>(mm) | 0.23×4<br>0.25×4, 0.25×5, 0.25×6, 0.25×8,<br>0.30×8, 0.30×10<br>0.33×12 | 0.23×4, 0.23×5, 0.23×6,<br>0.25×4, 0.25×5, 0.25×6, 0.25×8,<br>0.30×5, 0.30×6, 0.30×8,<br>0.33×10, 0.33×12 |
| Performance | Comply with ISO 7864, ISO 9626, and ISO<br>11608-2 | Comply with ISO 7864, ISO 9626, and ISO<br>11608-2 |
| Sterile | EO sterilized, SAL: 10-6 | EO sterilized, SAL: 10-6 |
| Single Use | Single use | Single use |
| Labeling | Conform with 21 CFR 801 | Conform with 21 CFR 801 |
| Biocompatibility | Conform with ISO 10993 standards | Conform with ISO 10993 standards |
The subject device has eight needle sizes; the predicate device heedles sizes. There are seven needle sizes of subject device are identical to those of the predicate device, only one needle size (0.30×10) of the subject device is different from the predicate device; this size (0.30×10) has same needle gauge as the predicate, but has a longer length than the predicate. The needle size (0.30×10) meets the performance requirements of ISO 7864. ISO 9626 and ISO 11608 standards, and the needle length (10mm) is one common insulin needle length for clinical use, therefore we think the difference on needle length will not raise new problem for the subject device.
All the subject devices are complied with the same performance standards. Through performance bench testing, the subject device has demonstrated that it is substantially equivalent to the predicate.
#### Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.