K152824 · Shandong Caremed Medical Products Co., Ltd. · FMI · Mar 7, 2016 · General Hospital
Device Facts
Record ID
K152824
Device Name
MEDT PEN NEEDLE
Applicant
Shandong Caremed Medical Products Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Mar 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MEDT Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
Device Story
MEDT Pen Needle is a single-use, sterile, manual hypodermic needle designed for subcutaneous insulin delivery. It consists of a stainless steel needle, polypropylene hub, and polyethylene inner sheath, packaged with an outer sheath and Tyvek paper. The device is intended for use by patients or clinicians with compatible pen injectors (e.g., NovoPen Echo, HumaPen Luxura). Operation involves attaching the needle to the injector, performing the injection, and disposing of the needle after a single use. The device facilitates insulin administration, aiding in glycemic control for patients with diabetes. It is provided sterile via ethylene oxide (EO) sterilization.
Clinical Evidence
Bench testing only. No clinical data. Testing included physical, mechanical, and chemical evaluations per ISO 7864, ISO 9626, and ISO 11608-2. Biocompatibility testing (cytotoxicity, intracutaneous reactivity, skin sensitization, acute systemic toxicity, hemolysis, and pyrogenicity) followed ISO 10993 and ASTM F756-13. Sterility and shelf-life were validated per USP and ISO 10993-7 standards.
Technological Characteristics
Materials: Stainless steel (needle), polypropylene (hub/outer sheath), polyethylene (inner sheath). Energy: Manual. Dimensions: 29G-32G gauges; lengths 4mm-12mm. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 7864, ISO 9626, ISO 11608-2, ISO 10993, ASTM F88/F88M-09, ASTM F756-13.
Indications for Use
Indicated for subcutaneous injection of insulin using pen injector devices in patients requiring insulin therapy.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized human figure in profile, with three faces overlapping to represent the department's focus on health and human well-being. The emblem is simple and modern in design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2016
Shandong Caremed Medical Products Co., Ltd. % Ms. Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CHINA
Re: K152824
Trade/Device Name: MEDT Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: January 28, 2016 Received: February 3, 2016
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K152824
Device Name MEDT Pen Needle
Indications for Use (Describe)
The MEDT Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: K152824
- 1. Date of Preparation: 03/03/2016
- Sponsor Identification 2.
## Shandong Caremed Medical Products Co., Ltd.
(Yantai) Industrial Park, DingTao County (The East of Zhanqian Road), Shandong, 274100, China
Contact Person: Lin Yanyan Position: Management representative Tel: 86-530-7397700 Fax: 86-530-7397711 Email: lyyan@shanghaicarelife.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)
## Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
- 4. Identification of Proposed Device
Trade Name: MEDT Pen Needle Common Name: Insulin Pen Needle
Regulatory Information Classification Name: Needle, Hypodermic, Single Lumen; Classification: II; Product Code: FMI; Regulation Number: 21 CFR 880.5570;
{4}------------------------------------------------
Review Panel: General Hospital;
Intended Use Statement:
The MEDT Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
Device Description
The proposed device, MEDT Pen Needle, is a single-use device, which is designed for used with a pen injector for the subcutaneous injection of insulin. It consist of three components, which are (1) needle, (2) hub and (3) inner sheath; it is sealed in a package consists of (1) a outer sheath and (2) Tyvek Paper.
- ડ. Identification of Predicate Device(s)
510(k) Number: K133059 Product Name: Insulin Pen Needle Manufacturer: Wenzhou Beipu Science & Technology Co., Ltd
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
## Physical, Mechanical and Chemical Tests performed on the proposed device
- > ISO 7864: 1993 Sterile hypodermic needles for single use
- > ISO 9626:1991 AMD 2001 Stainless steel needle tubing for the manufacture of medical devices
- A ISO 11608-2:2012 Needle-based injection systems for medical use- Requirements and test methods-Part 2: Needles
### Test Item
- Materials Surface finish Cleanliness Limits for acidity and alkalinity Size designation Dimensions Stiffness Resistance to breakage Resistance to corrosion
{5}------------------------------------------------
Cleanliness Limits for acidity or alkalinity Limits for extractable metals Size designation Colour coding Needle hub Sheath Needle tube Needle point Performance Materials Dimensions Determination of flow rate through the needle Bond between hub and needle tube Needle points Freedom form defects Lubrication Dislocation of measuring point at patient end Determination of functional compatibility with needle-based injection systems Ease of assembly and disassembly Sterility
Sterile Barrier Packaging Testing performed on the proposed device:
Seal Strength: ASTM F88/F88M-09 Standard test method for seal strength of flexible barrier materials
Sterilization and Shelf Life Testing performed on the proposed device:
| EO residue | ISO 10993-7:2008 Biological evaluation of medical devices- Part<br>7: Ethylene oxide sterilization residuals |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| ECH residue | ISO 10993-7:2008 Biological evaluation of medical devices- Part<br>7: Ethylene oxide sterilization residuals |
| Bacteria Endotoxin Limit | USP 37-NF 32 <85> Bacterial Endotoxins Limits |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package were performed on<br>accelerated aging samples to verify the claimed shelf life of the<br>device |
| | Sterility Test was performed per USP 38-NF 33 <71> Sterility<br>Tests |
Biocompatibility test on the proposed device:
{6}------------------------------------------------
| Cytotoxicity test | ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity |
|--------------------------------|------------------------------------------------------------------------------------------------------------------|
| Intracutaneous Reactivity test | ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization |
| Skin Sensitization Test | ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization |
| Acute Systemic Toxicity Test | ISO 10993-11:2006 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity |
| Hemolysis Test | ASTM F 756-13 Standard practice for assessment of hemolytic properties of materials |
| Pyrogen Test | USP 38-NF 33 <151> Pyrogen test |
- 7. Substantially Equivalent (SE) Comparison
| Item | Proposed Device<br>K152824 | Predicate Device<br>K133059 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | FMI | Same |
| Regulation Number | 880.5570 | Same |
| Intended Use | The MEDT Pen Needle is intended for<br>use with pen injector devices for the<br>subcutaneous injection of insulin. | Same |
| Configuration and<br>material | Needle tube Stainless Steel<br>hub Polypropylene<br>Inner sheath Polyethylene<br>Outer sheath Polypropylene<br>Tyvek paper Paper | Needle tube Stainless Steel<br>hub Polypropylene<br>Tube sheath Polypropylene<br>Hub sheath Polypropylene<br>Sealed paper Paper |
| Operation mode | Manual | Same |
| Needle Gauge | 29G, 30G, 31G, 32G | Same |
| Needle Dimension<br>(mm) | $0.23 \times 4, 0.23 \times 5, 0.23 \times 6,$<br>$0.25 \times 4, 0.25 \times 5, 0.25 \times 6, 0.25 \times 8,$<br>$0.30 \times 5, 0.30 \times 6, 0.30 \times 8,$<br>$0.33 \times 10, 0.33 \times 12$ | $0.23 \times 4,$<br>$0.25 \times 4, 0.25 \times 5, 0.25 \times 6, 0.25 \times 8,$<br>$0.30 \times 8, 0.30 \times 10,$<br>$0.33 \times 12$ |
| Performance | Comply with ISO 7864, ISO 9626, and<br>ISO 11608-2 | Same |
| Compatible pen<br>injectors | NovoPen Echo<br>HumaPen Luxura | Same |
| Sterile | EO sterilized, SAL: 10-6 | Same |
| Single Use | Single use | Same |
| Endotoxin Limit | 20 EU per product | Same |
- Table 1 Comparison of Technology Characteristics
{7}------------------------------------------------
| Labeling | Conform with 21 CFR 801 | Same |
|------------------------------|------------------------------------------|------|
| Biocompatibility | Conform with ISO 10993 standards | Same |
| Cytotoxicity | No cytotoxicity | |
| Intracutaneous<br>Reactivity | No Intracutaneous reactivity | |
| Skin Sensitization | No significant evidence of sensitization | |
| Acute Systemic<br>Toxicity | No systemic toxicity | |
| Hemolysis | No evidence of hemolysis | |
| Pyrogen | No pyrogen | |
## Comparison discussion:
The subject device has the same indications for use and the same principle of operation as the predicate device. Both subject and predicate devices are provided as sterile, nonpyrogenic and individually packaged.
The differences between the subject and the predicate devices include:
- . The material used for Inner sheath of the subject device is Polyethylene. The material used for Tube sheath of the predicate device is Polypropylene.
- . The subject and predicate devices have same needle gauges but different needle lengths. The subject device contains twelve (12) needle sizes. The predicate device contains eight (8) needle sizes.
The performance testing of the subject device met all acceptance criterial. These differences do not have impacts to product technology and performance to the subject device, and do not raise new issues of product performance.
#### 8. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.