Bausch + Lomb Sensitive Eyes Plus Saline Solution is a sterile, preserved solution for rinsing soft (hydrophilic) and gas permeable contact lenses after cleaning and before lens insertion with chemical (not heat) and hydrogen peroxide systems.
Device Story
Bausch + Lomb Sensitive Eyes Plus Saline Solution is a sterile, isotonic, buffered aqueous solution used for rinsing contact lenses. It contains boric acid, sodium borate, potassium chloride, sodium chloride, polyaminopropyl biguanide (0.00003%), and edetate disodium (0.025%). The device is packaged in a multi-use plastic bottle with a tamper-evident seal. It is intended for over-the-counter use by contact lens wearers to rinse soft (hydrophilic) and rigid gas permeable lenses after cleaning and prior to insertion. The solution functions via mechanical rinsing to remove debris. It is compatible with chemical and hydrogen peroxide disinfection systems. The product supports a 90-day discard date after opening.
Clinical Evidence
No clinical data. Safety and effectiveness were established through bench testing, including antimicrobial preservative efficacy testing (ISO 14730), regimen evaluation (FDA 1997 Guidance), and lens compatibility testing (ISO 11981).
Technological Characteristics
Isotonic, buffered aqueous solution. Ingredients: boric acid, sodium borate, potassium chloride, sodium chloride, polyaminopropyl biguanide (0.00003%), edetate disodium (0.025%). Form factor: multi-use plastic bottle. Standards: ISO 14730 (antimicrobial efficacy), ISO 14729 (regimen testing), ISO 11981 (lens compatibility).
Indications for Use
Indicated for rinsing soft (hydrophilic) and gas permeable contact lenses after cleaning and before insertion with chemical or hydrogen peroxide systems. For use by contact lens wearers.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
Bausch + Lomb Sensitive Eyes Plus Saline Solution (P800024/S16)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one in front of the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 11, 2017
Bausch & Lomb Incorporated Ms. Melissa Thomas Senior Manager, Regulatory Affairs 1400 North Goodman Street Rochester, New York 14609
Re: K170483
Trade/Device Name: Bausch + Lomb Sensitive Eyes Plus Saline Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN, MRC Dated: February 14, 2017 Received: February 16, 2017
Dear Ms. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Denise L. Hampton -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K170483
Device Name Bausch + Lomb Sensitive Eyes Plus Saline
#### Indications for Use (Describe)
Bausch + Lomb Sensitive Eyes Plus Salinon is a sterile, preserved solution for rinsing soft (hydrophilic) and gas permeable contact lenses after cleaning and before lens insertion with chemical (not heat) and hydrogen peroxide systems.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY K170483
### Bausch + Lomb Sensitive Eyes Plus Saline Solution
#### 1. Submitter Information
| Primary | Alternate |
|------------------------------------|----------------------------------|
| Melissa Thomas | Glenn Davies, O.D. |
| Senior Manager, Regulatory Affairs | Sr. Director Regulatory Affairs, |
| 1400 North Goodman Street | 1400 North Goodman Street |
| Rochester, NY 14609 | Rochester, NY 14609 |
| Tel. (585) 338-6045 | Tel. (585) 338-8215 |
| Fax (585) 338-0702 | Email: Glenn.Davies@bausch.com |
| Email: Melissa.Thomas@bausch.com | |
Summary Prepared: April 2017
#### 2. Device Name
| Trade Name: | Bausch + Lomb Sensitive Eyes Plus Saline Solution |
|------------------------|-----------------------------------------------------------------------------------------------------|
| Classification: | Accessories, Contact Lens Care Products |
| Device classification: | Class II |
| Regulation Number: | 886.5928 Soft Contact Lens Care Products<br>886.5918 Rigid Gas Permeable Contact Lens Care Products |
| Product Code: | LPN, MRC |
#### 3. Predicate Device
Bausch + Lomb Sensitive Eyes Plus Saline Solution (P800024/S16) Menicon Saline Rinse Solution (K151768)
#### 4. Description of the Device
Bausch + Lomb Sensitive Eyes Plus Saline Solution is a sterile, isotonic, buffered solution that contains boric acid, sodium borate, potassium chloride, sodium chloride, preserved with polyaminopropyl biguanide (0.00003%) and edetate disodium (0.025%). The sterile solution is packaged in a plastic bottle with a tamper evident seal and labeled with a lot number and expiration date.
{4}------------------------------------------------
#### 5. Intended Use
Bausch + Lomb Sensitive Eyes Plus Saline Solution is a sterile, preserved solution for rinsing soft (hydrophilic) and gas permeable contact lenses after cleaning and before lens insertion with chemical (not heat) and hydrogen peroxide systems.
#### 6. Description of Safety and Substantial Equivalence
A series of preclinical testing was performed to demonstrate the safety and effectiveness of Bausch + Lomb Sensitive Eyes Plus Saline Solution as described in Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products, May 1, 1997. A brief summary of the test results is provided below:
#### Microbiology
Studies were performed to establish discard dating of 90 days for Sensitive Eves Plus Saline Solution. The test articles were evaluated initially and again following simulated use for the proposed discard date of 90 days.
The results of these evaluations demonstrate that Sensitive Eyes Plus Saline Solution meets the requirements of ISO 14730, Annex C Procedure II. Ophthalmic optics - Contact lens care products - Antimicrobial preservative efficacy testing and guidance on determining discard for the designation of a 90 day discard date.
Additionally a modified regimen procedure allowing for a rinse with Sensitive Eves Plus Saline Solution In Place of Tap Water was conducted for both Boston Conditioning Solution and Boston Advance Conditioning Solution. The results of both evaluations demonstrate that modified regimen with Sensitive Eyes Plus Saline Solution meets the FDA performance criteria for reqimen evaluation as described in the May 1, 1997 Guidance for Industry, Pre-market Notification (510(k)) Guidance Document for Contact Lens Care Products when used in a 10 second rub, 5 second rinse (per lens side) regimen. The products used in this regimen also meet the performance criteria established in ISO Standard 14729: 2001/Amd. 1:2010: (E) for regimen testing.
#### Biocompatibility
Biocompatibility tests were unnecessary for this application. The Sensitive Eyes Plus Saline Solution was previously cleared for use and expansion of the indication did not prompt the need for new testing.
#### Lens Compatibility
The results of lens compatibility studies per ISO 11981 demonstrate Bausch + Lomb Sensitive Eyes Plus Saline Solution is compatible with rigid gas permeable contact lenses.
#### Clinical Data
Clinical studies involving Sensitive Eyes Plus Saline Solution were unnecessary for this application. The Sensitive Eyes Plus Saline Solution has been commercially available in the market for over 25 years with well demonstrated safety and efficacy.
{5}------------------------------------------------
### 7. Substantial Equivalence
The cumulative results of laboratory testing sponsored by Bausch + Lomb demonstrate that the safety, effectiveness and performance of Bausch + Lomb Sensitive Eyes Plus Saline Solution when used with gas permeable contact lenses are substantially equivalent to the current product which is indicated for use with soft contact lenses.
| Feature | Menicon Saline Rinse<br>Solution (K151768) | Sensitive Eyes Plus<br>Saline Solution<br>(P800024/S16) | Sensitive Eyes Plus<br>Saline Solution<br>(K170483) |
|------------------------------------------------|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Product Code | LPN, MRC | LPN | LPN, MRC |
| Regulation Number | 21 CFR 886.5918<br>21 CFR 886.5928 | 21 CFR 886.5928 | 21 CFR 886.5918<br>21 CFR 886.5928 |
| Indicated for Soft<br>Lenses | Yes | Yes | Yes |
| Indicated for Rigid<br>Gas Permeable<br>Lenses | Yes | No | Yes |
| Rinse Required for<br>Cleaning | Unknown | Thoroughly Rinse<br>5 Seconds Each Side | Thoroughly Rinse<br>5 Seconds Each Side |
| Discard After Opening | Unknown | NA | 90 day Discard Date |
| Primary Container | Plastic resin container<br>with twist off cap | Plastic Bottle<br>Multiple Sizes | Plastic Bottle<br>Multiple Sizes |
| Container Usage | Single Use | Multi Use | Multi Use |
| Preserved | No | Yes | Yes |
| Sterile | Yes | Yes | Yes |
| Technology<br>Features | Isotonic<br>Aqueous Solution | Isotonic<br>Aqueous Solution | Isotonic<br>Aqueous Solution |
| Materials | Plastic Resin Container<br>With Twist Off Cap | Plastic Resin Container<br>With Cap | Plastic Resin Container<br>With Cap |
| Principles of<br>Operation | Mechanical Rinsing | Mechanical Rinsing | Mechanical Rinsing |
| Directions for Use | Unknown | Rinse After Cleaning<br>(5 sec. per side)<br>Rinse After Disinfection<br>(20 seconds)<br>Carton Onsert<br>Bausch.com | Rinse After Cleaning<br>(5 sec. per side)<br>Rinse After Disinfection<br>(20 seconds)<br>Carton Onsert<br>Bausch.com |
### SUBSTANTIAL EQUIVALENCE SUMMARY TABLE
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.