The Menicon Saline Rinse Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The Menicon Saline Rinse Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.
Device Story
Menicon Saline Rinse Solution is a sterile, non-preserved 0.9% NaCl (normal saline) solution provided in single-use 5 mL plastic containers. It is used by contact lens wearers to rinse soft, rigid gas permeable, or hard contact lenses after disinfection and before insertion. It also serves as an insertion solution for scleral lenses, a rinse for lens cases, and for mid-day lens rinsing. By providing a preservative-free environment, it avoids potential irritation from osmolarity agents or preservatives found in multipurpose solutions when held against the cornea. The device is intended for over-the-counter use by patients.
Clinical Evidence
No clinical data. Bench testing only. The solution is a standard 0.9% saline meeting USP requirements; biocompatibility was established via adherence to ISO 10993 and FDA guidance for contact lens care products.
Technological Characteristics
0.9% NaCl (normal saline) solution; non-preserved; sterile; single-use 5 mL plastic resin container with twist-off cap. Biocompatibility evaluated per ISO 10993 and FDA guidance for contact lens care products.
Indications for Use
Indicated for users of soft (hydrophilic), rigid gas permeable, and hard contact lenses, including large diameter (scleral) lenses, requiring a sterile, non-preserved saline rinse for lens insertion, case rinsing, or mid-day rinsing following proper disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Optics Laboratory, Inc. Eye-Cept Sterile Saline Solution (K110221)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 28, 2015
Menicon Co., Ltd. % Ms. Ellen M. Beucler Vice President Foresight Regulatory Strategies, Inc. 187 Ballardvale Street, Suite 180 Wilmington, MA 01887
Re: K151768
Trade/Device Name: Menicon Saline Rinse Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN, MRC Dated: June 30, 2015 Received: June 30, 2015
Dear Ms. Beucler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151768
Device Name Menicon Saline Rinse Solution
#### Indications for Use (Describe)
The Menicon Saline Rinse Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The Menicon Saline Rinse Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.
Type of Use (*Select one or both, as applicable*)
| <input type="checkbox"/> <span>Research Use (Part 21 CFR 601. Subpart D)</span> | <input checked="" type="checkbox"/> <span>Compassionate Use (21 CFR 601 Subpart E)</span> |
|---------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Menicon. The logo consists of a red and green circular graphic to the left of the company name. The company name "Menicon" is written in a stylized, black font.
Page 1 of 3
# K 151768
#### 510(k) Summary
#### MENICON SALINE RINSE SOLUTION
## 1. Applicant Information
Menicon Co., Ltd. 21-19, Aoi 3-Chome, Naka-ku Nagoya 460-0006 Japan
| Contact Person: | Tohru Kawaguchi, Ph.D. |
|-----------------|-------------------------|
| Phone: | +81-52-935-1676 |
| Fax: | +81-52-935-1633 |
| email: | tohru@menicon-net.co.jp |
| Date Prepared: | July 8, 2015 |
#### 2. Device Information
| Classification name: | Contact Lens Care Products |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Device classification: | Class II |
| Regulation number: | 21 CFR 886.5918 (Rigid Gas Permeable Contact Lens Care Products)<br>21 CFR 886.5928 (Soft (hydrophilic) Contact Lens Care Products) |
| Product code: | LPN |
| Proprietary name: | Menicon Saline Rinse Solution |
#### 3. Predicate Devices
Menicon Co., Ltd. claims substantial equivalence to K093367, Cachet Pharmaceuticals Pvt. Ltd. Sterile Saline Solution and K110221, Optics Laboratory, Inc. Eye-Cept Sterile Saline Solution.
#### 4. Description of Device
The Menicon Saline Rinse Solution is a sterile unit dose non-preserved, 0.9% NaCl (normal saline) solution indicated for rinsing soft (hydrophilic), rigid gas permeable (RGP) and hard contact lenses prior to insertion. The rinsing solution removes debris and bacteria following proper disinfection as recommended by the eye care practitioner.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Menicon. The logo consists of a circular graphic on the left and the word "Menicon" on the right. The circular graphic is divided into two halves, with the top half being red and the bottom half being green. A white curved line runs through the center of the circle, separating the two halves. The word "Menicon" is written in a black, sans-serif font.
Page 2 of 3
This sterile, normal saline solution can be used to rinse contact lens cases, rinse lenses as needed throughout the day and to fill the concave posterior surface of scleral lenses prior to insertion to provide a more natural environment than currently approved multipurpose solutions which contain preservatives and osmolarity agents that are not designed to be held against the cornea for extended periods of time.
# 5. Indications for Use
The Menicon Saline Rinse Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The Menicon Saline Rinse Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses. as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.
## 6. Performance Data
#### Non-Clinical Data
Biocompatibility tests were unnecessary for this application. The solution is a 0.9 % saline solution that meets the requirements of USP saline therefore testing was not required. The components are safe for the intended use.
## Clinical Data
Clinical studies involving the saline rinse solution were unnecessary for this application. Lens care solutions used with this Saline Rinse Solution are already cleared for use as cleaning, rinsing, disinfection and storage solutions for contact lenses.
## Conclusion
Based upon the composition of the saline rinse solution and previous test data presented, the Menicon Saline Rinse Solution is as safe, as effective and performs as well as the predicate devices. A comparison of the new device and the predicate devices is presented in Table 1.
## 7. Substantial Equivalence
The claim of substantial equivalence to the previously cleared to K093367, Cachet Pharmaceuticals Pvt. Ltd Sterile Saline Solution and K110221, Optics Laboratory, Inc. Eye-Cept Sterile Saline Solution, is supported by the following Comparison of Characteristics in Table 1.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Menicon. The logo consists of a red and green circular symbol on the left, followed by the word "Menicon" in a stylized, cursive font. The symbol appears to be a stylized eye or lens, with the red portion on top and the green portion on the bottom.
Page 3 of 3
The Menicon Saline Rinse Solution and the predicates are similar in composition and intended use. Therefore, Menicon Saline Rinse Solution is substantially equivalent to the predicate device.
| Table 1<br>Comparison of Characteristics | | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| | Menicon Saline Rinse Solution | Cachet Pharmaceuticals<br>Sterile Saline Solution | Optics Laboratory Inc.<br>Eye-Cept Sterile Saline Solution |
| Device Name | Contact Lens Care Product | Contact Lens Care Product | Contact Lens Care Product |
| Trade Name | Menicon Saline Rinse Solution | Sterile Saline Solution | Eye-Cept Sterile Saline Solution |
| Document Number | K151768 | K093367 | K110221 |
| Classification | Ophthalmic | Ophthalmic | Ophthalmic |
| Product Code | LPN | LPN | LPN |
| Regulation Number | 21 CFR 886.5918<br>21 CFR 886.5928 | 21 CFR 886.5928 | 21 CFR 886.5928 |
| Class | II | II | II |
| Intended Use | For use following proper lens disinfection<br>as recommended by the eye care<br>practitioner. The Menicon Saline Rinse<br>Solution is for rinsing soft (hydrophilic),<br>rigid gas permeable and hard contact<br>lenses prior to lens insertion. This solution<br>may also be used as an insertion solution<br>for large diameter (scleral) contact lenses,<br>as a rinse for contact lens cases, and may<br>be used as needed throughout the day to<br>rinse contact lenses. | For rinsing and wetting of soft<br>(hydrophilic) contact lenses | For rinsing soft contact lenses after<br>cleaning and for wetting soft contact<br>lenses after disinfection before use. |
| Volume | 5 mL | - | 10 mL |
| Preservative Free | yes | yes | yes |
| Container Usage | single use | single use | single use |
| Sterility | Sterile | Sterile | Sterile |
| Materials | Plastic resin container<br>with twist off cap | Plastic resin container<br>with twist off cap | Plastic resin container<br>with twist off cap |
| Biocompatibility | Components used in this lens case have<br>been evaluated in accordance with Part<br>10993 of the ISO standard for Biological<br>Evaluation and the FDA Guidance for<br>Contact Lens Care Products | Components used for this product<br>have been evaluated in accordance<br>with FDA Guidance for Contact Lens<br>Care Products | Components used for this product<br>have been evaluated in accordance<br>with FDA Guidance for Contact Lens<br>Care Products |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.