K170426 · Andersen Sterilizers, Inc. · JOJ · Oct 30, 2017 · General Hospital
Device Facts
Record ID
K170426
Device Name
AN85/AN86 EO Indicators
Applicant
Andersen Sterilizers, Inc.
Product Code
JOJ · General Hospital
Decision Date
Oct 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
AN85/AN86 EO Indicators are single-use Type 1 process indicators used to distinguish between processed and unprocessed packaged medical devices, through a visible color change from yellow-green to blue. They are intended for the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc. Critical process parameters for the cycle are summarized in Table 5-1.
Device Story
Adhesive-backed Type 1 process indicators; used to distinguish processed from unprocessed medical devices following ethylene oxide (EO) sterilization. Device contains pH-sensitive ink printed in dots (AN85) or stripes (AN86) laminated between plastic layers. Upon exposure to EO, chemical reaction alters pH, causing color shift from yellow-green to blue. Used in clinical or healthcare settings; operated by personnel managing sterilization cycles. Provides visual confirmation of EO exposure to healthcare providers; assists in inventory management of sterile goods. Benefits patient by ensuring medical devices have undergone intended sterilization process.
Clinical Evidence
Bench testing only. Performance characterized in a Chemical Indicator Evaluator Resistometer (CIER) and Anprolene AN75 sterilizer. Testing confirmed functionality (ISO 11140-1 Type 1 compliance), endpoint color stability (stable 12 months), and 2-year shelf life. Biocompatibility confirmed via non-toxic ingredients.
Technological Characteristics
Adhesive-backed Type 1 process indicator; plastic film construction; proprietary pH-sensitive indicator ink. Conforms to AAMI/ANSI/ISO 11140-1. Sensing principle: chemical reaction with ethylene oxide causing pH-dependent color change (yellow-green to blue).
Indications for Use
Indicated for use as a single-use Type 1 process indicator to distinguish between processed and unprocessed packaged medical devices during a 12-hour ethylene oxide sterilization cycle at 20-29°C in the Anprolene AN75 Sterilizer.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the words "U.S. FOOD & DRUG ADMINISTRATION" on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the text.
October 30, 2017
Andersen Sterilizers, Inc. William Andersen President 3154 Caroline Drive Haw River, North Carolina 27258
Re: K170426
Trade/Device Name: AN85/AN86 EO Indicators Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: September 28, 2017 Received: October 6, 2017
Dear William Andersen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K170426
Device Name AN85/AN86 EO Indicators
Indications for Use (Describe)
AN85/AN86 EO Indicators are single-use Type 1 process indicators used to distinguish between processed packaged medical devices, through a visible color change from yellow-green to blue. They are intended for the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc. Critical process parameters for the cycle are summarized in Table 1.
Table 1. Critical sterilization cycle parameters in the Anprolene AN75 Ethylene Oxide Gas Sterilizer
| Ethylene Oxide | Temperature | Relative Humidity | Ethylene Oxide Exposure Time | Total Cycle Time |
|----------------|-------------|-------------------|------------------------------|------------------|
| 17.6 g ± 5% | 20-29°C | 35-90% | 12 hours | 14 hours |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|× Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### K170426
# 510(k) Summary
| 5.1 Applicant's Name and Address | | | |
|----------------------------------|--|--|--|
| | | | |
Andersen Sterilizers, Inc. Establishment Registration Number 3004634710 3154 Caroline Drive Haw River, NC 27258
### 5.2 Contact Person
William K. Andersen, BE, MD, FAAOS President Phone: 336-376-8622 Fax: 336-376-5428
### 5.3 Date of Preparation
September 28, 2017
### 5.4 Device
| Proprietary Name | AN85/AN86 EO Indicators |
|------------------|----------------------------------------------------|
| Common Name | Indicator, Physical/Chemical Sterilization Process |
| Classification | Class II (21 CFR 880.2800) Chemical Indicator |
| Product Code | JOJ |
#### 5.5 Predicate Device
Device Name 510(k) number Manufacturer
AN85/AN86 EO Indicators K150644 Andersen Sterilizers, Inc.
The predicate AN85 and AN86 EO Indicators (K150644) are approved for use with the EOGas 4 Ethylene Oxide Gas Sterilization system manufactured by Andersen Sterilizers, Inc.
In this submission, the sterilization temperature and time claims of the predicate devices were modified in order to indicate the AN85 and AN86 EO Indicators for use in the Anprolene AN75 Ethylene Oxide Gas Sterilization System. No modifications were made to the manufacturing method, technology, or intended use.
# 5.6 Device Description
AN85/AN86 EO Indicators are adhesive-backed Type 1 process indicators for ethylene oxide sterilization that conform to AAMI/ANSI/ISO 11140-1.
AN85/AN86 EO Indicators contain a pH indicator. The pH indicator ink is printed in dots on the AN85 and in stripes on the AN86. The indicator ink is laminated between two layers of plastic material and changes color from yellow-green to blue by chemical reactions when exposed to ethylene oxide.
{5}------------------------------------------------
# 5.7 Indications for Use
AN85/AN86 EO Indicators are single-use Type 1 process indicators used to distinguish between processed and unprocessed packaged medical devices, through a visible color change from yellow-green to blue. They are intended for the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc. Critical process parameters for the cycle are summarized in Table 5-1.
Table 5-1. Critical sterilization cycle parameters in the Anprolene AN75 Ethylene Oxide Gas Sterilizer
| Ethylene Oxide | Temperature | Relative Humidity | Ethylene Oxide Exposure Time | Total Cycle Time |
|----------------|-------------|-------------------|------------------------------|------------------|
| 17.6 g ± 5% | 20-29°C | 35-90% | 12 hours | 14 hours |
# 5.8 Technological Characteristics
AN85/AN86 EO Indicators contain a proprietary pH indicator which changes color by chemical reactions when exposed to ethylene oxide, allowing differentiation of ethylene oxide processed vs. unprocessed devices.
### 5.9 Performance Testing
The performance of AN85/AN86 EO Indicators was characterized in a Chemical Indicator Evaluator Resistometer (CIER) as well as in an Anprolene AN75 Ethylene Oxide Gas Sterilizer using the 12 hour cycle at 20-29°C. Performance testing is summarized in Table 5-2.
Table 5-2. Summary of bench tests performed to demonstrate safety and effectiveness of AN85/AN86 EO Indicators
| Test | Description | Result |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Functionality | 1) Critical parameters include temperature, time, and gas concentration at<br>a relative humidity of 35-90%;<br>2) ISO 11140-1 Type 1 process indicator;<br>3) Indicate EO exposure in the CIER and in the Anprolene AN75<br>sterilizer. | Pass |
| Biocompatibility | Not direct or indirect patient-contacting devices;<br>Non-toxic ingredients;<br>Provides reasonable assurance for safety. | Pass |
| Endpoint Color<br>Stability | Stable for at least 12 months at 20-25°C and 30-50% relative humidity<br>away from direct sunlight;<br>Stability demonstrates reasonable assurance for effectiveness. | Pass |
| Shelf Life | Maintains performance specifications throughout the shelf life of 2 years. | Pass |
{6}------------------------------------------------
The bench studies demonstrate that AN85/AN86 EO Indicators perform as intended to indicate that the devices have been exposed to ethylene oxide, and perform as safely and effectively as the legally marketed predicate device, the AN85/AN86 EO Indicators (K150644).
# 5.10 Device Comparison
AN85/AN86 EO Indicators used in the Anprolene AN75 Sterilizer are substantially equivalent to the AN85/AN86 EO Indicators (K150644) used in the EOGas 4 sterilizer. Both the subject and predicate indicators have the same intended use, design, and technical characteristics. A comparison between the indicators is listed in Table 5-3.
| | Predicate AN85/AN86 EO<br>Indicators (K150644) | Subject AN85/AN86 EO<br>Indicators | Comparison |
|-----------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Intended Use | Process indicator to indicate<br>exposure to EO<br>Adhesive-backed labels | Process indicator to<br>indicate exposure to EO<br>Adhesive-backed labels | Identical |
| Sterilization<br>Method | EOGas 4 sterilizer;<br>100% EO process;<br>3 hr EO exposure at 50±3°C | Anprolene AN75 sterilizer;<br>100% EO process;<br>12 hr EO exposure at 20-<br>29°C | Difference does not<br>alter safety or<br>effectiveness |
| Design | Indicator changes color when<br>exposed to EO | Indicator changes color<br>when exposed to EO | Identical |
| Indicator<br>Agent | pH indicator | pH indicator | Identical |
| Device<br>Materials | Plastic films;<br>Plastic layer above and<br>below the ink prevents direct<br>contact with the ink. | Plastic films;<br>Plastic layer above and<br>below the ink prevents<br>direct contact with the ink. | Identical |
| Endpoint<br>Color<br>Change | Yellow-green to blue color | Yellow-green to blue color | Identical |
| Technology | Chemical reactions with EO<br>change the pH and the color<br>of the indicator ink. | Chemical reactions with<br>EO change the pH and the<br>color of the indicator ink. | Identical |
| Performance | Type 1 process indicator | Type 1 process indicator | Equivalent |
Table 5-3. Comparison between AN85/AN86 EO Indicators and the predicate device
# BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
Based on the intended use, technological characteristics, performance data, and nonclinical tests performed, the subject AN85/AN86 EO Indicators are substantially equivalent to, and are as safe and as effective as, the legally marketed predicate device, the AN85/AN86 EO Indicators used in the EOGas 4 sterilization system, cleared under K150644.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.