K150644 · Andersen Sterilizers, Inc. · JOJ · Nov 18, 2015 · General Hospital
Device Facts
Record ID
K150644
Device Name
AN85/AN86 EO Indicators
Applicant
Andersen Sterilizers, Inc.
Product Code
JOJ · General Hospital
Decision Date
Nov 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The AN85/AN86 EO Indicators are single-use Class 1 process indicators used to distinguish between processed and unprocessed packaged medical devices, through a visible color change from yellow-green to blue. They are intended for the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc.
Device Story
AN85/AN86 EO Indicators are adhesive-backed, single-use Class 1 chemical process indicators. They consist of a pH-sensitive indicator ink printed in dots (AN85) or stripes (AN86) on a substrate, laminated between two plastic film layers. When exposed to ethylene oxide gas during a 3-hour, 50°C sterilization cycle in the EOGas 4 Sterilizer, the ink undergoes a chemical reaction, changing color from yellow-green to blue. This visual change allows healthcare personnel to differentiate between processed and unprocessed medical device packaging. The device is used in clinical settings where EOGas 4 sterilization is performed. It provides a qualitative visual confirmation of exposure to the sterilization process, aiding in inventory management and ensuring that only processed items are used in clinical procedures.
Clinical Evidence
Bench testing only. Performance characterized in a Chemical Indicator Evaluator Resistometer (CIER) and the EOGas 4 Sterilizer. Testing confirmed functionality (color change) under specified parameters (17.6g EO, 50°C, 35-90% RH), endpoint color stability (12 months at 20-25°C), and a 2-year shelf life. Biocompatibility is not applicable as the device is not patient-contacting.
Technological Characteristics
Class 1 chemical process indicator; adhesive-backed plastic film construction; pH-sensitive indicator ink; conforms to AAMI/ANSI/ISO 11140-1; non-patient contacting; shelf life 2 years; storage 20-30°C, 30-50% RH.
Indications for Use
Indicated for use as a Class 1 process indicator to distinguish between processed and unprocessed packaged medical devices during ethylene oxide sterilization cycles in the EOGas 4 Sterilizer.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community or interconnectedness. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the organization's name and national affiliation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2015
Andersen Sterilizers, Inc. William K. Andersen, BE, MD, FAAOS President 3154 Caroline Drive Haw River, NC 27258
Re: K150644
Trade/Device Name: AN85/AN86 EO Indicators Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization process indicator Regulatory Class: Class II Product Code: JOJ Dated: October 20, 2015 Received: October 22, 2015
Dear Dr. William K. Andersen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin |-Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K150644
Device Name AN85/AN86 EO Indicators
Indications for Use (Describe)
The AN85/AN86 EO Indicators are single-use Class 1 process indicators used to distinguish between processed packaged medical devices, through a visible color change from yellow-green to blue. They are intended for the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc. Critical process parameters for the cycle are summarized in Table 1.
Table 1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer
| Ethylene Oxide | Temperature | Relative Humidity | Ethylene Oxide Exposure Time | Total Cycle Time |
|----------------|-------------|-------------------|------------------------------|------------------|
| 17.6 g ± 5% | 50°C ± 3°C | 35-90% | 3 hours | 3.5 hours |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|× Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
- 5.1 Applicant's Name and Address
Andersen Sterilizers, Inc. 3154 Caroline Drive Haw River, NC 27258
- 5.2 Contact Person
William K. Andersen, BE, MD, FAAOS President Phone: 336-376-8622 Fax: 336-376-5428
- 5.3 Date of Preparation November 17, 2015
- 5.4 Device
| Proprietary Name | AN85/AN86 EO Indicators |
|------------------|----------------------------------------------------|
| Common Name | Indicator, Physical/Chemical Sterilization Process |
| Classification | Class II (21 CFR 880.2800) Chemical Indicator |
| Product Code | JOJ |
- 5.5 Predicate Device
Device Name 510(k) number Manufacturer
VERIFY EO Valueline Chemical Indicator K991418 Steris Corporation
#### 5.6 Device Description
AN85/AN86 EO Indicators are adhesive-backed Class 1 process indicators for ethylene oxide sterilization that conform to AAMI/ANSI/ISO 11140-1.
AN85/AN86 EO Indicators contain a pH indicator. The pH indicator ink is printed in dots on the AN85 and in stripes on the AN86. The indicator ink is laminated between two layers of plastic material and changes color from yellow-green to blue by chemical reactions when exposed to ethylene oxide.
{5}------------------------------------------------
# 5.7 Indications for Use
The AN85/AN86 EO Indicators are single-use Class 1 process indicators used to distinguish between processed and unprocessed packaged medical devices, through a visible color change from yellow-green to blue. They are intended for the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc. Critical process parameters for the cycle are summarized in Table 5-1.
| | | | Table 5-1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer | |
|----------------|-------------|----------------------|--------------------------------------------------------------------------------------------------|---------------------|
| Ethylene Oxide | Temperature | Relative<br>Humiditv | Ethylene Oxide<br>Exposure Time | Total Cycle<br>Time |
| 17.6 g ± 5% | 50℃ ± 3℃ | 35-90% | 3 hours | 3.5 hours |
### 5.8 Technological Characteristics
AN85/AN86 EO Indicators contain a proprietary pH indicator which changes color by chemical reactions when exposed to ethylene oxide, allowing differentiation of ethylene oxide processed vs. unprocessed devices.
### 5.9 Performance Testing
The performance of AN85/AN86 EO Indicators was characterized in a Chemical Indicator Evaluator Resistometer (CIER) as well as in an EOGas 4 Ethylene Oxide Gas Sterilizer using the 3 hour cycles at 50 ℃. Performance testing is summarized in Table 5-2.
| Table 5-2. Summary of bench tests performed to demonstrate safety and effectiveness of |
|----------------------------------------------------------------------------------------|
| AN85/AN86 EO Indicators |
| Test | Description | Result |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Functionality | 1) Critical parameters include temperature, time, and gas concentration<br>under a relative humidity of 35-90%;<br>2) ISO 11140-1 Class 1 process indicator;<br>3) Indicate EO exposure in the CIER and in the EOGas 4 sterilizer. | Pass |
| Biocompatibility | Not direct or indirect patient-contacting devices;<br>Non-toxic ingredients;<br>Provides reasonable assurance for safety | Pass |
| Endpoint Color<br>Stability | Stable for at least 12 months at 20-25°C and 3 months at 50°C;<br>Stability demonstrates reasonable assurance for effectiveness | Pass |
| Shelf Life | Maintains performance specifications throughout the shelf life. For<br>storage, AN85/AN86 indicators should be kept away from EO or products<br>that affect pH such as acids or alkalis, and in an environment of 20-30°C<br>and 30-50% relative humidity. Under the conditions, AN85/AN86<br>indicators have a shelf life of 2 years from date of manufacture. | Pass |
{6}------------------------------------------------
The bench studies demonstrate that AN85/AN86 EO Indicators perform as intended to indicate that the devices have been exposed to ethylene oxide, and perform as safely and effectively as the legally marketed predicate device.
#### 5.10 Device Comparison
AN85/AN86 EO Indicators are substantially equivalent to the VERIFY EO Valueline Chemical Indicator (K991418). Both the subject and predicate indicators have the same intended use, design, and technical characteristics. The difference in endpoint color changes due to different pH indicators used does not raise different questions of safety and effectiveness. A comparison between the indicators is listed in Table 5-3.
| | VERIFY EO Valueline<br>Chemical Indicator | AN85/AN86 EO<br>Indicators | Comparison |
|--------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Intended Use | Process indicator to indicate<br>exposure to EO | Process indicator to indicate<br>exposure to EO | Same |
| | Insert labels | Adhesive-backed labels | AN85/AN86 may<br>be used to secure<br>wrapping materials |
| Sterilization<br>Method | Traditional EO sterilizer<br>100% and blended EO process | The EOGas 4 sterilizer<br>100% EO process 3 hr cycles at<br>50°C | Do not affect safety<br>or effectiveness |
| Design | Indicator changes color when<br>exposed to EO | Indicator changes color when<br>exposed to EO | Equivalent |
| Indicator<br>Agent | pH indicator | pH indicator | Equivalent |
| Device<br>Materials | Paper | Plastic films | AN85/AN86 :<br>plastic layer above<br>and below the ink<br>prevents direct<br>contact with the ink |
| Endpoint<br>Color Change | Orange to red color | Yellow-green to blue color | Equivalent |
| Technology | Chemical reactions with EO<br>changes the pH and the color<br>of the indicator ink | Chemical reactions with EO<br>changes the pH and the color<br>of the indicator ink | Equivalent |
| Performance | Correctly indicated EO<br>exposure; Class 1 process<br>indicator. | Correctly indicated EO<br>exposure; Class 1 process<br>indicator. | Equivalent |
# Table 5-3. Comparison between AN85/AN86 EO Indicators and the predicate device
# BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
Based on the intended use, technological characteristics, performance data, and nonclinical tests performed, the subject AN85/AN86 EO Indicators are substantially equivalent to, and are as safe and as effective as, the legally marketed predicate device, VERIFY EO Valueline Chemical Indicator, cleared under K991418.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.