PASCAL Synthesis Twinstar Ophthalmic Scanning Laser System
K170409 · Topcon Medical Laser Systems, Inc. (Tmls) · GEX · Sep 12, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170409
Device Name
PASCAL Synthesis Twinstar Ophthalmic Scanning Laser System
Applicant
Topcon Medical Laser Systems, Inc. (Tmls)
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 12, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PASCAL Synthesis TwinStar Ophthalmic Scanning Laser System is intended for use to perform single-spot photocoagulation in the posterior segment (retina, choroid) and in the anterior segment (iris, trabecular meshwork) as well as pattern-scanning photocoagulation in the non-macular retina of the eye. Single-spot delivery may be performed using a slit lamp biomicroscope or an indirect ophthalmoscope. Pattern delivery may be performed using a slit lamp biomicroscope.
Device Story
Ophthalmic scanning laser system for photocoagulation; delivers laser energy to retina, choroid, iris, or trabecular meshwork. System utilizes 577 nm (yellow) and 638 nm (red) laser wavelengths. Physician operates system via slit lamp biomicroscope or indirect ophthalmoscope. Provides single-spot or selectable pattern-scanning delivery to enhance efficiency and reduce patient discomfort. Output is laser energy; healthcare provider uses visual feedback through microscope to guide treatment. Benefits include precise retinal/anterior segment photocoagulation for eye disease management.
Clinical Evidence
No clinical data. Evidence consists of bench performance testing and an in vivo animal study (rabbit retina) comparing the 638 nm red laser to the reference predicate. Study evaluated lesion morphology via ophthalmoscopy, fundus photos, and SD-OCT; results showed <20% difference in lesion width, supporting equivalence.
Technological Characteristics
Laser surgical instrument; Class II. Wavelengths: 577 nm (OPSL) and 638 nm (Diode). Top hat intensity profile. Spot diameter: 50-400 μm (577 nm), 60/200 μm (638 nm). Scan field: 4 mm x 4 mm. Aiming laser: 670 nm diode. Compliance: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1, IEC 60601-2-22, IEC 60825-1.
Indications for Use
Indicated for patients requiring photocoagulation in the posterior segment (retina, choroid) or anterior segment (iris, trabecular meshwork) of the eye. Includes single-spot and pattern-scanning delivery for non-macular retina.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
PASCAL® Synthesis™ Ophthalmic Scanning Laser System (K123542)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 12, 2017
Topcon Medical Laser Systems, Inc. % Pamela Buckman. MSN Regulatory Consultant 2800 Pleasant Hill Rd., Suite 175 Pleasant Hill, CA 94523
Re: K170409
Trade/Device Name: PASCAL® Synthesis™ TwinStar™ Ophthalmic Scanning Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, HQF Dated: August 4, 2017 Received: August 7, 2017
Dear Pamela Buckman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170409
Device Name
PASCAL® Synthesis™ TwinStar™ Ophthalmic Scanning Laser System
Indications for Use (Describe)
PASCAL Synthesis TwinStar Ophthalmic Scanning Laser System is intended for use to perform single-spot photocoagulation in the posterior segment (retina, choroid) and in the anterior segment (iris, trabecular meshwork) as well as pattern-scanning photocoagulation in the non-macular retina of the eye. Single-spot delivery may be performed using a slit lamp biomicroscope or an indirect ophthalmoscope. Pattern delivery may be performed using a slit lamp biomicroscope.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY
#### K170409
#### 1. General Information
Submission
- Sponsor: Topcon Medical Laser Systems, Inc. 606 Enterprise Court Livermore CA- 94550 USA T 925 245 8300 F 925 245 7723 Contact Person: Sweta Srivastava
Submission Correspondent:
Pamela M. Buckman, MSN 2800 Pleasant Hill Rd., Suite 175 Pleasant Hill, CA 94523 T 925 980 7007 F 925 705 7381
Date Prepared: September 1, 2017
#### 2. Device Identification
| Device Name: | PASCAL® Synthesis™ TwinStar™ Ophthalmic Scanning<br>Laser System |
|-------------------------------|--------------------------------------------------------------------------------------------------------|
| Classification Name: | Laser Surgical Instrument for use in General and Plastic<br>Surgery and Dermatology; Laser, Ophthalmic |
| Classification<br>Regulation: | 21 CFR 878.4810<br>21 CFR 886.4390 |
| Classification Panel: | Ophthalmology/General and Plastic Surgery |
| Product Code: | GEX, HQF |
| Device Class: | Class II |
#### 3. Predicate Devices
Primary Predicate Device
PASCAL® Synthesis™Ophthalmic Scanning Laser System (K123542)
Reference Predicate Device
NOVUS Multiwavelength Omni (K932468)
{4}------------------------------------------------
#### 4. Device Description
PASCAL® Synthesis™ TwinStar™ is an ophthalmic scanning laser system. The system can perform single shot photocoagulation as is performed conventionally. At the physician's discretion, the system provides for a single laser spot treatment or the option of using a selectable pattern of delivering multiple laser spots simultaneously. Pattern scanning allows the physician to enhanced efficiency and reduced patient discomfort.
The PASCAL Synthesis™ Laser System (parent product to the PASCAL Synthesis Twinstar™) included 532 nm/577 nm wavelengths and was cleared via K123542. The subject device for this submission is a line extension named PASCAL Synthesis Twinstar and will include 577 nm/638 nm laser wavelengths.
### 5. Intended Use
PASCAL Synthesis TwinStar Ophthalmic Scanning Laser System is intended for use to perform single-spot photocoagulation in the posterior segment (retina, choroid) and in the anterior segment (iris, trabecular meshwork) as well as pattern-scanning photocoagulation in the non-macular retina of the eye. Singlespot delivery may be performed using a slit lamp biomicroscope or an indirect ophthalmoscope. Pattern delivery may be performed using a slit lamp biomicroscope.
## 6. Comparison of Technological Characteristics
The technological characteristics of PASCAL® Synthesis™ TwinStar™ Ophthalmic Scanning Laser System are substantially equivalent to those of the Primary predicate and the Reference predicate devices.
The Pascal Synthesis Twinstar that is the subject of k170409 is a modification of Pascal Synthesis (primary predicate) with an added option of red treatment wavelength. Pascal Synthesis Twinstar is a combination wavelength device that offers two treatment wavelengths, i.e., 577nm and 638 nm.
Using red laser for the treatment of eye disease, the Pascal Synthesis Twinstar uses 638 nm red laser to perform single-spot photocoagulation. The Novus Multiwavelength (reference predicate) has been cleared for the use of similar red laser wavelength (647 nm) for the treatment of eye disease.
The following Table compares the technological characteristics of the Pascal Synthesis Twinstar to the Primary Predicate as well as the Reference Predicate.
{5}------------------------------------------------
| ITEM | NOVUS Multi-<br>Wavelength<br>(K932468)<br>Reference Predicate | PASCAL® Synthesis™<br>(K123542)<br>Primary Predicate | PASCAL Synthesis<br>TwinStar<br>(K170409) |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type | Treatment Laser:<br>Coherent super<br>INNOVA Ceramic<br>CoolDisk, Aiming<br>Laser: InGaAIP Diode | OPSL- Optically<br>Pumped Semiconductor<br>Laser; Diode Pumped;<br>Solid State | OPSL - Optically<br>Pumped Semiconductor<br>Laser; Diode Pumped;<br>Solid State (577)<br>Diode Laser (638) |
| Output<br>Wavelength3 | 521 and 531 nm<br>(Green)<br>647 nm (Red)<br>568 nm (Yellow)<br>521, 531 and 568 nm<br>(Yellow-Green) | 532 nm (Green)<br>577 nm (Yellow) | 577 nm (Yellow)<br>638 nm (Red) |
| Power Output4 | Green: 50 - 900 mW<br>Red: 50 – 1000 mW<br>Yellow: 75 - 600 mW<br>Yellow-Green: 75 -<br>1500 mW | 0, 30 mW - 2000 mW | 0, 30 mW - 2000 mW<br>(577)<br>0 to 600mW (638) |
| Duty Cycle | 100% | 100% | 100% |
| Exposure Time | 10 ms - 1000 ms, and<br>continuous | 2ms -1000 ms | 2ms -1000 ms |
| Repetition Rate<br>(Available with<br>single-spot<br>pattern only) | Available settings<br>range from 1 Hz to 9<br>Hz depending on rate<br>and exposure time<br>settings | Off, 1.0 Hz, 1.5 Hz, 2.0<br>Hz, 3.0 Hz, 4.0 Hz, 5.0<br>Hz, 6.0 Hz, 7.0 Hz, 8.0<br>Hz | Off, 1.0 Hz, 1.5 Hz, 2.0<br>Hz, 3.0 Hz, 4.0 Hz, 5.0<br>Hz, 6.0 Hz, 7.0 Hz, 8.0<br>Hz |
| Pulse Counter | 0 - 9999, reset<br>available | 0-99,999, reset available | 0-99,999, reset available |
| CDRH<br>Classification | Class IV | Class IV | Class IV |
| European<br>Classification | Class 4 | Class 4 | Class 4 |
| Aiming Laser<br>Type | InGaAIP<br>Diode<br>(Direct Diode) | Direct Diode | Direct Diode |
| Wavelength | 670 nm | 635 nm | 670 nm |
| Power Output | Adjustable to <1mW | Adjustable to <1mW | Adjustable to <1mW |
| CDRH<br>Classification | Class IIIa | Class II | Class II |
| MDD<br>Classification | Class 3a | Class 2 | Class 2 |
| ITEM | NOVUS Multi-Wavelength<br>(K932468)<br>Reference Predicate | PASCAL® Synthesis™<br>(K123542)<br>Primary Predicate | PASCAL Synthesis<br>TwinStar<br>(K170409) |
| Level of<br>Concern | Unknown | Major | Major |
| Spot Diameter | 50 μm – 500 μm<br>delivered to the focal<br>plane of the slit lamp<br>in the air (with<br>LaserLink Z Slit Lamp<br>Adapter) There is a<br>laser diameter<br>magnification factor for<br>the contact lens<br>utilized. | 50 μm – 400 μm<br>delivered to the focal<br>plane of the slit lamp in<br>the air. There is a laser<br>diameter magnification<br>factor for the contact<br>lens utilized. | 50 μm – 400 μm (577);<br>60, 200um (638)<br>delivered to the focal<br>plane of the slit lamp in<br>the air. There is a laser<br>diameter magnification<br>factor for the contact<br>lens utilized. |
| Intensity Profile | Top Hat | Top Hat | Top Hat |
| Spot Spacing | Not applicable (non-<br>scanning) | Adjustable, with a<br>minimum center-to-<br>center spacing of 1 spot<br>diameter. | Adjustable, with a<br>minimum center-to-<br>center spacing of 1 spot<br>diameter. |
| Scan Field | Not applicable (non-<br>scanning) | 4 mm X 4 MM (in air) | 4 mm X 4 mm (in air) |
| Scan Rate | Not applicable (non-<br>scanning) | ≤ 100 Hz | ≤ 100 Hz |
| Electrical Requirements: | | | |
| Voltage | 200-240 VAC or<br>380 to 415 VAC | 100-230 VAC ± 10% | 100-230 VAC ± 10% |
| Frequency | 50/60 Hz, single<br>phase or three-<br>phase | 50,60 Hz, single-phase | 50,60 Hz, single-phase |
| Current | 20, 35, or 60 A | < 10 Amperes | < 10 Amperes |
{6}------------------------------------------------
#### 7. Substantial Equivalence
PASCAL® Synthesis™ TwinStar™ Ophthalmic Scanning Laser System shares the same indications for use, device operation, overall technical and functional capabilities and therefore is substantially equivalent to the predicate devices. Performance and animal testing confirmed that the PASCAL® Synthesis™ TwinStar™ Ophthalmic Scanning Laser System is substantially equivalent to the identified predicates.
{7}------------------------------------------------
#### Safety and Effectiveness information 8.
The review of the indications for use and technical characteristics as well as animal study results and performance testing demonstrate that the PASCAL® Synthesis™ Twinstar™ Ophthalmic Scanning Laser System is substantially equivalent to the predicate devices. No new safety and effectiveness questions are applicable.
## 9. Non Clinical Testing
### Performance Testing - Bench
Performance testing was conducted in order to demonstrate compliance with recognized consensus standards:
• AAMI/ANSI ES60601-1:2005/(R) 2012 + A1:2012, C1:2009(R) 2012 and A2:2010/(R) 2012 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance
· IEC 60601-1-2:2007: Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests
• IEC 60601-1-6:2010+A1:2013 Medical electrical equipment-Part 1-6: General requirements for safety-Collateral Standard Usability
· IEC 62366-1 Medical Devices-Part 1: Application of usability engineering to medical devices • IEC 60601-2-22: 2007 (Third Edition) + A1:2012 Medical Electrical Equipment Part 2: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
· IEC 60825-1 2nd Edition Part 1: Safety of laser products-Part 1: Eguipment classification and requirements Additionally, hardware and software validation activities were performed to ensure the device performed as intended and software documentation appropriate for the Major level of concern was provided.
### Performance Testing - Animal
An animal study was performed for the novel features of the subject device relative to the PASCAL® Synthesis: the 638 nm red treatment laser. In the study grids of mild to moderate grade retinal lesions were delivered to rabbit retinae in vivo with both the subject device and the reference predicate (the Novus Omni compared with the red treatment laser).
{8}------------------------------------------------
In the case of the red laser study, both 60um and 200um spot-sizes were used. Ophthalmoscopic evaluation, fundus photos, and SD-OCT measurements were taken in vivo immediately after treatment, prior to euthanasia. No difference in lesion morphology between predicate and subject device was observed funduscopically or in OCT for equivalent lesion ophthalmoscopic grade. In an analysis of OCT width measurements, the difference in lesion width for predicate and subject device was <20% in all statistically significant single-eye comparative cases. The similarity in lesion character and size under matching delivery conditions supports substantial equivalence between predicate and subject devices for the red treatment laser.
### 10. Clinical Performance
There was no clinical testing required to support this medical device as the indications for use and technology are equivalent to those of the predicate devices.
### 11. Statement of Substantial Equivalence
A device is substantially equivalent when the subject device has the same intended use and the same technological characteristics as the previously cleared predicate device(s). The addition of the red laser to the PASCAL® Synthesis™ TwinStar™ Ophthalmic Scanning Laser System does not alter either of these characteristics and does not raise any new questions of safety or efficacy as compared to the predicates.
## 12. Conclusion
The PASCAL® Synthesis™ TwinStar™ Ophthalmic Scanning Laser System was found to be substantially equivalent to the predicate devices. It shares the same indications for use, similar design features, and functional features with, and thus is substantially equivalent to, the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.