K123542 · Topcon Medical Laser Systems, Inc. · HQF · May 15, 2013 · Ophthalmic
Device Facts
Record ID
K123542
Device Name
PASCAL SYNTHESIS (532NM/577NM) OPTHALMIC LASER
Applicant
Topcon Medical Laser Systems, Inc.
Product Code
HQF · Ophthalmic
Decision Date
May 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PASCAL® Synthesis Ophthalmic Scanning Laser System is intended for use in the treatment of ocular pathology in both the posterior and anterior segments. Intended for use in the posterior segment to perform retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including: (532nm) - proliferative and non-proliferative diabetic retinopathy - macular-edema - choroidal neovascularization associated with wet age-related macular degeneration - age-related macular degeneration - lattice degeneration - retinal tears and detachments (577nm) - proliferative and non-proliferative diabetic retinopathy - macular edema - choroidal neovascularization associated with wet age-related macular degeneration - age-related macular degeneration - lattice degeneration - retinal tears and detachments Intended for use in the treatment of ocular pathology in the anterior segment including: (532 nm and 577nm) - iridotomy - trabeculoplasty.
Device Story
PASCAL Synthesis is a stand-alone ophthalmic laser delivery console for use with slit lamps, LIOs, or probes. System delivers 532nm or 577nm laser energy to treat retinal/choroidal vascular/structural abnormalities and anterior segment pathologies. Operator (physician) controls system via LCD/touchscreen interface; laser activation via wired footswitch. Safety features include three-position key switch, emergency off switch, and door interlock. Device transforms electrical input into controlled laser output for photocoagulation. Clinical benefit includes precise, automated scanning delivery for retinal/anterior segment conditions. Output allows physician to perform targeted photocoagulation, iridotomy, or trabeculoplasty to manage ocular disease.
Clinical Evidence
Bench testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and software validation (EN 60601-1-4, 62304). Animal study (GLP-compliant) compared retinal lesion appearance between PASCAL Synthesis and PASCAL Streamline. Results showed comparable lesion morphology, diameter, and clinical grade, with no systematic trend in diameter differences. No clinical human data provided.
Technological Characteristics
OPSL (Optically Pumped Semiconductor Laser) technology; 532nm or 577nm wavelengths; 30-2000 mW power output; 2-1000 ms exposure time; 100% duty cycle. Includes 635nm aiming beam. Stand-alone console with LCD/touchscreen, slit lamp connector, and wired footswitch. Electrical: 100-230 VAC, 50/60 Hz. Class IV laser device.
Indications for Use
Indicated for patients requiring treatment of ocular pathology in posterior and anterior segments, including proliferative/non-proliferative diabetic retinopathy, macular edema, wet age-related macular degeneration, lattice degeneration, retinal tears/detachments, iridotomy, and trabeculoplasty.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
Predicate Devices
Optimedica Ophthalmic Scanning Laser System (K092621)
Iridex TxCell Adaptor/Photocoagulator System (K121475)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510 (K) SUMMARY - K123542
### General Information I.
Submitter:
Topcon Medical Laser Systems, Inc. 3130 Coronado Drive Santa Clara, CA 95054, USA
| Contact Person: | Pamela M. Buckman |
|-----------------|-----------------------|
| | Regulatory Consultant |
| | T 925 980 7007 |
| | F 925 705 7381 |
| | pmbuckman@gmail.com |
May 12, 2013 Summary Preparation Date:
## Names II.
| Device Name(s): | PASCAL® Synthesis™ Ophthalmic Scanning Laser<br>System |
|-------------------------|-----------------------------------------------------------------------------------------------------|
| Common Name: | Laser, Ophthalmic |
| Classification Name(s): | Laser Surgical Instrument for use in General and Plastic<br>Surgery and Dermatology; Laser, Powered |
### Predicate Devices III.
- Optimedica Ophthalmic Scanning Laser System (K092621) .
- . Topcon Ophthalmic Scanning Laser System (K111108)
- Iridex TxCell Adaptor/Photocoagulator System (K121475) .
### Product Description IV.
The PASCAL® Synthesis™ Ophthalmic Scanning Laser System is a stand-alone laser delivery console for use with a slit lamp. LIO or probe delivery. Controls and displays include an I.CD/Touchscreen user interface and a slit lamp connector (SLC). The system is equipped with a three-position key switch that enables laser and controls as well as an emergency off switch. An externally connected wired footswitch activator is provided with a shrouded housing, and contains a set of redundant switches. An external door interlock connection allows the laser to be terminated with an open circuit.
{1}------------------------------------------------
# V. Summary of Non Clinical Tests
Topcon utilized hardware testing (60601-1, -2-x; EN ISO14971-C; ASTM 4169-09&6653-01; ASTM D4169-09). electromagnetic testing (IEC 60601-1-2. 4. 6) software testing (EN 60601-1-41 62304 and animal testing to establish a basis for the determination of equivalence. The PASCAL® Synthesis performance characteristics were established by referencing the known performance characteristics of the predicate device. Specifications for the PASCAL® Synthesis system were established to assure that the predicate system and the Synthesis system performed in an equivalent manner. Performance specifications were set utilizing national and international standards for such devices with respect to output, indications for use, safety features and electromagnetic interference where it was established as suitable for the environment of use. All testing was conducted and established that the PASCAL® Synthesis system met or exceeded its design specifications and performed equally or better than the stated performance of the predicate.
The objective of the animal study was to compare the appearance of retinal laser lesions produced by the PASCAL Synthesis system with the predicate PASCAL Streamline system in an animal model. This prospective comparative study included delivery of laser lesions to two animals in accordance with Good Laboratory Practices (GLP) standards and the local Animal Institutional Review Board. Multiple lesions using varying spot sizes and intensities were created with each of the two lasers and assigned a clinical grade by the laser operator on a 5 point scale. When comparing lesions of the same clinical grades, the appearance of the lesions was similar. The diameter of the lesions was <20% in mean diameter. Within lesions of the same clinical grade, the variability was approximately 10%. Lesions were created using spot sizes of 100, 200 and 400 micron (50 micron was not part of the investigation). Powers ranged from 70 mW to 450 Endpoint Management algorithms were not used as part of this investigation. No mW. systematic trend in lesion diameter was observed. The Streamline device and the Synthesis device were comparable in a preclinical setting.
Conclusions drawn from these non-clinical tests demonstrate that the device is as safe, as effective and performs at least as safely and as effectively as the legally marketed device identified in this Summary.
#### VI. Indications for Use
The PASCAL® Synthesis Ophthalmic Scanning Laser System is intended for use in the treatment of ocular pathology in both the posterior and anterior segments. Intended for use in the posterior segment to perform retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
(532nm)
- proliferative and non-proliferative diabetic retinopathy .
- . macular-edema
- choroidal neovascularization associated with wet age-related macular degeneration .
- age-related macular degeneration t
- . lattice degeneration
- retinal tears and detachments .
{2}------------------------------------------------
# (577nm)
- proliferative and non-proliferative diabetic retinopathy ●
- . macular edema
- choroidal neovascularization associated with wet age-related macular degeneration �
- age-related macular degeneration .
- lattice degeneration .
- . retinal tears and detachments
Intended for use in the treatment of ocular pathology in the anterior segment including: (532 nm and 577nm)
- iridotomy .
- trabeculoplasty. .
## VII. Summary of Technological Characteristics .
The technological characteristics of the PASCAL® Synthesis Ophthalmic Scanning Laser System are substantially equivalent to those of the predicate.
| Technical<br>Characteristics | K092621 | K111108 | K121475/K071687 | K123542<br>PASCAL®<br>Synthesis |
|--------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Type of Laser | OPSL -<br>Optically<br>Pumped<br>Semiconductor<br>Laser; Diode<br>Pumped; Solid<br>State | OPSL -<br>Optically<br>Pumped<br>Semiconductor<br>Laser; Diode<br>Pumped; Solid<br>State | OPSL - Optically<br>Pumped<br>Semiconductor<br>Laser; Diode<br>Pumped; Solid State | OPSL -<br>Optically<br>Pumped<br>Semiconductor<br>Laser; Diode<br>Pumped; Solid<br>State |
| Output<br>Wavelength | 532 nm | 577 nm | 532 nm or 577 nm | 532 nm or 577<br>nm |
| Power output | 0, 100 mW -<br>2000 mW | 0, 100 mW -<br>2000 mW | 0-2000 mW | 0, 30 mW - 2000<br>mW |
| Duty cycle | 100% | 100% | Variable | 100% |
| Exposure time | 5 ms - 1000 ms | 10 ms - 1000 ms | 0.05 ms - 3000 ms | 2 ms - 1000 ms |
| Repetition Rate<br>(available with<br>single spot<br>pattern only) | Off, 1.0hz,<br>1.5hz, 2.0hz,<br>3.0hz, 4.0hz,<br>5.0hz, 6.0hz,<br>7.0hz, 8.0hz | Off, 1.0hz,<br>1.5hz, 2.0hz,<br>3.0hz, 4.0hz,<br>5.0hz, 6.0hz,<br>7.0hz, 8.0hz | Available only with<br>single spot pattern | Off, 1.0hz,<br>1.5hz, 2.0hz,<br>3.0hz, 4.0hz,<br>5.0hz, 6.0hz,<br>7.0hz, 8.0hz |
| Pulse counter | 0 - 99,999 with<br>Automatic reset | 0 - 99,999, reset<br>available | Not Reported | 0 - 99,999, reset<br>available |
| CDRH<br>Classification | Class IV | Class IV | Class IV | Class IV |
| European MDD<br>Classification | Class 4 | Class 4 | Class 4 | Class 4 |
| Type | Direct diode | Direct diode | Direct diode | Direct diode |
{3}------------------------------------------------
| Technical<br>Characteristics | K092621 | K111108 | K121475/K071687 | K123542<br>PASCAL®<br>Synthesis |
|------------------------------|----------------------------|----------------------------|----------------------|---------------------------------|
| Wavelength | 635 nm | 635 nm | 635 nm | 635 nm |
| Power output | Adjustable to <1<br>mW | Adjustable to <1<br>mW | Adjustable to <1 mW | Adjustable to <1<br>mW |
| CDRH laser<br>Classification | Class II | Class II | Class II | Class II |
| Electrical Requirements | | | | |
| Voltage | 100-240 VAC ±<br>10% | 100-230 VAC ±<br>10% | 100-240 VAC ±<br>10% | 100-230 VAC ±<br>10% |
| Frequency | 50/60 Hz, single-<br>phase | 50/60 Hz, single-<br>phase | 50/60 Hz | 50/60 Hz, single-<br>phase |
| Current | <5 Amperes | <10 Amperes | Not Reported | <10 Amperes |
# VIII. Rationale for Substantial Equivalence
The PASCAL® Synthesis Ophthalmic Scanning Laser System shares the same indications for use, device operation, overall technical and functional capabilities, and therefore is substantially equivalent to the predicate devices.
## IX. Safety and Effectiveness Information
The review of the indications for use and technical characteristics demonstrates that the PASCAL® Synthesis Ophthalmic Scanning Laser System is substantially equivalent to the predicate devices. No new safety or effectiveness questions are applicable.
## X. Conclusion
The PASCAL® Synthesis Ophthalmic Scanning Laser System was found to be substantially equivalent to the predicate devices. The PASCAL® Synthesis Ophthalmic Scanning Laser System shares the same indications for use, similar design features, and functional features with, and thus is substantially equivalent to, the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is in a sans-serif font.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-002
May 15, 2013
Topcon Medical Laser Systems, Inc. % Buckman Company, Inc. Ms. Pamela M. Buckman Regulatory Consultant 2800 Pleasant Hill Road, Suite 175 Pleasant Hill, California 94553
Re: K123542
Trade/Device Name: PASCAL® Synthesis" Ophthalmic Scanning Laser System . Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: Class II Product Code: HQF, GEX Dated: April 15, 2013 Received: April 22, 2013
Dear Ms. Buckman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Ms. Pamela M. Buckman
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely yours, For
Peter D Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
| 510(k) Number: K123542 | |
|------------------------|-----------------------------------------------------------|
| Device Name: | Topcon PASCAL® Synthesis Ophthalmic Scanning Laser System |
The PASCAL® Synthesis Ophthalmic Scanning Laser System is intended for use in the treatment of ocular pathology in both the posterior and anterior segments. Intended for use in the posterior segment to perform retinal photocoagulation, panretinal photocoagulation. focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including: .
- (532nm)
- proliferative and non-proliferative diabetic retinopathy
- macular edema t
- choroidal neovascularization associated with wet age-related macular degeneration .
- . age-related macular degeneration
- . lattice degeneration
- . retinal tears and detachments
(577nm)
- . proliferative and non-proliferative diabetic retinopathy
- . macular édema
- . choroidal neovascularization associated with wet age-related macular degeneration
- . age-related macular degeneration
- ◆ lattice degeneration
- � retinal tears and detachments
Intended for use in the treatment of ocular pathology in the anterior segment including: (532 nm and 577nm)
- iridotomy �
- trabeculoplasty .
X Prescription Use
(Per 21CFR 801 Subpart D)
OR
Over-The-Counter Use (21 CFR 801 Subpart C) ·
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
# Neil R Ogdens 2013.05.15 11:05:52 -04'00'
(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number K123542
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.