K170370 · Syneron Medical, Ltd. · OHV · May 15, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170370
Device Name
ULTRASHAPE POWER SYSTEM
Applicant
Syneron Medical, Ltd.
Product Code
OHV · General, Plastic Surgery
Decision Date
May 15, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UltraShape Power System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) for lipolysis (breakdown of fat) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference and fat reduction in the flanks and thighs.
Device Story
System uses focused ultrasound energy to disrupt subcutaneous adipose tissue (SAT) for non-invasive lipolysis; aesthetic body contouring. Components include control unit and ultrasonic transducer; transducer converts electrical signals into mechanical (acoustical) energy. Used in clinical settings by trained operators. Real-time optical and acoustic feedback provided via tracking and guidance system. Energy beam targets SAT at specific focal depth/diameter/length; selective disruption of adipose tissue without damaging neighboring blood vessels, nerves, or muscle. Output is mechanical energy; healthcare provider uses system to achieve aesthetic reduction in abdominal circumference and fat in flanks/thighs. Benefits patient through non-invasive fat reduction with minimal discomfort and no required downtime.
Clinical Evidence
Supported by two prospective, single-arm, self-controlled, multicenter studies. Flank study (n=48) and thigh study (n=47) evaluated fat reduction via blinded photographic review and ultrasound/caliper measurements. Primary endpoint (80% correct identification of pre/post treatment) met in both studies (80% and 81% respectively). Secondary endpoints showed statistically significant fat thickness and circumference reduction compared to control. Safety profile demonstrated no adverse events; only mild, transient erythema/edema reported.
Technological Characteristics
Focused ultrasound stimulator; piezoelectric ceramic transducer. System console includes computer and U-Sculpt Power transducer. Parameters: frequency, burst duration, node duration, focal depth, diameter, and length consistent with predicate. Electrical safety per IEC 60601-1, IEC 60601-2-37; EMC per IEC 60601-1-2. Biocompatible patient-contacting components.
Indications for Use
Indicated for non-invasive reduction in abdominal circumference and fat reduction in the flanks and thighs in patients with subcutaneous adipose tissue of at least 1.5 cm and BMI between 22 and 30.
Regulatory Classification
Identification
A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W O66-G609 Silver Spring, MD 20993-0002
May 15, 2017
Syneron Medical Ltd. % Ms. Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K170370
Trade/Device Name: UltraShape Power System Regulation Number: 21 CFR 878.4590 Regulation Name: Focused Ultrasound Stimulator System for Aesthetic Use Regulatory Class: Class II Product Code: OHV Dated: April 12, 2017 Received: April 12, 2017
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known)
K170370
Device Name
UltraShape Power System
Indications for Use (Describe)
The UltraShape Power System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) for lipolysis (breakdown of fat) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference and fat reduction in the flanks and thighs.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="display:inline-block; margin-right: 5px;"> <svg height="15px" width="15px"> <rect height="15px" style="fill:lightgray;stroke-width:1;stroke:rgb(0,0,0)" width="15px"></rect> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="2" x2="13" y1="2" y2="13"></line> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="13" x2="2" y1="2" y2="13"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-right: 5px;"> <svg height="15px" width="15px"> <rect height="15px" style="fill:lightgray;stroke-width:1;stroke:rgb(0,0,0)" width="15px"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
| Department of Health and Human Services |
|----------------------------------------------------------------|
| Food and Drug Administration |
| Office of Chief Information Officer |
| Paperwork Reduction Act (PRA) Staff |
| <a href="mailto:PRAStaff@fda.hhs.gov">PRAStaff@fda.hhs.gov</a> |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (1/14) | Page 1 of 1 FDA | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-----------------|-------------------------------------------|
|----------------------|-----------------|-------------------------------------------|
{3}------------------------------------------------
### 510(k) SUMMARY
### Syneron Medical Ltd.'s UltraShape Power System
### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Ruthie Amir, MD, Global Vice President of Clinical, Requlatory and Education Syneron Medical Ltd. P.O.B. 550 Industrial Zone, Tavor Building Yokneam Illit, 20692 Israel Phone: 972-73-244-2200 Facsimile: 972-73-244-2202
Date Prepared: February 6, 2017
## Name of Device
Syneron UltraShape Power System
## Common or Usual Name
Focused Ultrasound Stimulator System for Aesthetic Use
## Classification
21 CFR 878.4590, Class II, product code OHV
### Predicate Devices
Syneron UltraShape System (K162163, K161952) (Primary Predicate)
Syneron UltraShape Power System (K160896)
# Intended Use / Indications for Use
The UltraShape Power System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) for lipolysis (breakdown of fat) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference and fat reduction in the flanks and thighs.
# Device Description
The UltraShape Power System is comprised of multiple components, including the control unit and an ultrasonic transducer. The UltraShape Power System selectively targets subcutaneous adipose tissue via focused ultrasound for the purpose of non-invasive aesthetic body contouring. The transducer is an electro-mechanical device that converts an electrical signal into mechanical (acoustical) energy. The operating parameters of the UltraShape Power System achieve selective disruption of adipose tissue without damaging neighboring tissues such as blood vessels, nerves, or muscle.
# Technological Characteristics
The UltraShape Power System has similar technological characteristics as its predicate. Both devices are comprised of the system console, including the computer and a small (U-Sculpt Power)
{4}------------------------------------------------
transducer. The transducer delivers the focused ultrasound energy beam to the targeted treatment area, and real-time optical and acoustic feedback (optional) on the treatment is provided via the tracking and guidance system. With both the UltraShape Power System and its predicate, the transducer's functionality is based on the piezoelectric effect implemented with the ceramic element.
In addition, the subject UltraShape Power System has the same treatment parameters as the previously cleared UltraShape Power device, including the same frequency, burst duration, node duration, and user interface. The power intensity levels were previously cleared for the UltraShape Power device, and the focal distribution of the energy beams delivered to the treatment area is consistent between the device and its predicate, including the focal depth, diameter, and length.
Therefore, the subject UltraShape Power System has very similar technological characteristics as its predicate.
### Performance Data
The following nonclinical performance testing was previously conducted to support the substantial equivalence of the UltraShape Power to its predicate devices, consistent with FDA's "Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use" (2011). In all instances, the UltraShape Power System functioned as intended.
- Beam profile testing demonstrated that the acoustic enerqy is delivered and concentrated in the desired tarqet location, at a focal depth similar to that of the predicate devices.
- Acoustic power testing demonstrated that the acoustic power of the transducer met specifications, is well defined, and presents low variability.
- . In vitro thermal evaluation was conducted and results were passing, demonstrating the absence of thermal coagulation or lesion in tissue.
- Software verification and validation was performed, and demonstrated that the software ● performs as intended.
- Electrical safety (IEC 60601-1, IEC 60601-2-37) and electromagnetic compatibility (IEC 60601-1-2) testing was conducted and results were passing.
- Biocompatibility of the patient-contacting components of the device was established.
- In vivo testing in an animal model was performed to evaluate the device treatment mode, ● and results supported the device safety and efficacy profile for the intended use.
In addition, clinical evaluation of the UltraShape System (cleared in K161952) in the intended population was performed in several separate prospective studies, including a single-arm, prospective, self-controlled study to evaluate the safety and effectiveness of the device for noninvasive lipolysis of the flanks. A total of 48 subjects were enrolled and treated in the study (of which 46 subjects completed the study) at 3 U.S. sites. The study included females (83%) and males (17%) across a range of ages, races, ethnicities, and skin types. The mean age was 45 years and the majority of the subjects were Caucasian. The baseline mean weight was 68±9 kg and mean BMI was 25.28±2.08 kg/m².
Each subject received 3 biweekly treatments on one randomized flank, while the other flank remained un-treated throughout the study. Subject follow up was conducted at 4 weeks, 8 weeks,
{5}------------------------------------------------
and 16 weeks after the last treatment. The study included visual assessment of the flanks by 2 blinded reviewers as well as fat thickness measurements of the post treatment flanks compared to baseline measurements.
Study results demonstrated that the UltraShape treatment at the flanks was accompanied with no or minimal discomfort, consistent with results observed for the prior UltraShape clearances. Throughout the study, no adverse events were reported, and the anticipated immediate responses after treatment (e.g., mild erythema, mild edema) were consistent with those observed with other similar devices, such as the previously cleared UltraShape systems (K133238, K160896) and Cynosure's SculpSure (K150230). The responses were mild and resolved completely within days without any intervention.
The primary endpoint was met, where 80% of treated versus control flanks and pre versus post treatment photographs were correctly identified by blinded reviewers. In addition, secondary effectiveness endpoints demonstrated that fat thickness reduction on the treated flank was statistically significantly greater compared to the control flank at each follow up as measured by ultrasound. Fat thickness reduction results following UltraShape treatment were greater compared to controls as measured by caliper (reaching statistical significance in absolute fat thickness reduction at 16 week follow up). At 4 week follow up, investigator satisfaction measured 70%, and subsequent results reflected lower satisfaction with increasing time from treatment (53% at 8 weeks follow up, 43% at 16 weeks follow up). In terms of subject satisfaction, at 4, 8, and 16 weeks following end of treatments. 43%, 47%, and 41% subjects were satisfied. The study design and results are further summarized in the table below.
| Study Design | Prospective, single-arm, self-controlled, multicenter clinical study |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sample size | 48 subjects at 3 sites were enrolled and treated in the study |
| Principal<br>Eligibility<br>Criteria | • Fat thickness of at least 1.5 cm in the treated area as measured by<br>calibrated caliper.<br>• BMI interval: $22 ≤ BMI ≤ 30$ (normal to overweight, but not obese).<br>• General good health confirmed by medical history and skin examination of<br>the treated area. |
| Follow up<br>intervals | 3 treatment visits and follow up visits at 4 weeks, 8 weeks, and 16 weeks. |
| Endpoints | Primary: Fat reduction in flanks was evaluated by visual assessment of<br>randomized photographs of before (baseline) versus after (16 weeks following<br>last treatment), as well as treated versus control flanks, by 2 blinded<br>reviewers. The primary endpoint was achieved when at least 80% of treated<br>versus control flanks and pre- versus post-treatment photographs were<br>correctly identified.<br><br>Secondary:<br>- Fat thickness reduction on the treated flank compared to the control flank,<br>measured by ultrasound.<br>- Fat thickness reduction on the treated flank compared to the control flank,<br>measured by caliper.<br>- Investigator satisfaction assessment.<br>- Subject satisfaction self-assessment.<br>- Subjects' comfort level assessed immediately after each treatment. |
| Effectiveness<br>Results | Primary: Met endpoint; blinded reviewers identified correctly (by agreement<br>between the blinded reviewers) both the pre-/post-treatment photographs as<br>well as the treated/un-treated flanks in 80% (36 of 45) subjects. |
{6}------------------------------------------------
| Study Design | Prospective, single-arm, self-controlled, multicenter clinical study |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Secondary:<br>- Fat thickness reduction results following UltraShape treatment were<br>statistically significantly greater compared to controls as measured by<br>ultrasound.<br>- Fat thickness reduction results following UltraShape treatment were greater<br>compared to controls as measured by caliper (reaching statistical significance<br>in absolute fat thickness reduction at 16 week follow up).<br>- Investigators were satisfied with results following UltraShape treatment for<br>70% of the subjects at 4 week follow up, 53% of the subjects at 8 week follow<br>up, and 43% of the subjects at 16 week follow up.<br>- At 4, 8, and 16 weeks following end of treatments, 43%, 47%, and 41% of<br>the subjects were satisfied or very satisfied.<br>- Subjects reported no to minimal pain for each of the three treatments. |
| Safety<br>Results | No adverse events were reported after 141 treatment sessions were<br>conducted.<br>The only immediate responses recorded were mild and resolved completely<br>within days without any intervention. |
In addition, clinical evaluation of the UltraShape System (cleared in K162163) in the intended population for the non-invasive lipolysis of the thighs was also performed in a prospective, singlearm, self-controlled study. A total of 47 subjects (all female) across a range of ages, ethnicities, and skin types were enrolled at 3 U.S. sites. The mean age was 46 years and the majority of the subjects were Caucasian. The mean weight at baseline was 70±10 kg (range 51-103 kg). Each subject received 3 biweekly treatments on one randomized thigh, while the other thigh remained un-treated throughout the study to serve as a control. Subject follow up was conducted at 4 weeks, 8 weeks, and 16 weeks after the last treatment.
The study met its primary effectiveness endpoint. where blinded reviewers who were shown randomized pairs of thigh photographs correctly identified both pre- versus post-treatment and treated versus untreated thighs in 81% of cases. Further, even in the Intent-to-Treat population, with worst case imputation where all missing data were treated as failures, 64% of the subjects were successes. UltraShape treated thighs showed statistically significantly greater circumference reduction compared to the control thighs at each follow-up visit. Fat thickness reduction results following UltraShape treatment were statistically significantly greater compared to controls as measured by ultrasound. Investigators were satisfied with results following UltraShape treatment for 77% of the subjects at 4 week follow up, 81% of the subjects at 8 week follow up, and 68% of the subjects at 16 week follow up. At 4, 8 and 16 weeks following end of treatments, 62%, 61% and 68% of the subjects were satisfied or very satisfied. Finally, treatment with the UltraShape System in the thigh area demonstrated a strong positive safety profile, with no adverse events reported after 128 treatment sessions and only mild anticipated treatment responses (e.g., erythema) that resolved without intervention. This is consistent with the results observed for prior UltraShape clearances, including the predicate device.
{7}------------------------------------------------
The study design and results are further summarized in the table below.
| Study Design | Prospective, single-arm, self-controlled, multicenter clinical study |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sample size | 47 subjects at 3 sites were enrolled in the study |
| Principal<br>Eligibility<br>Criteria | Fat thickness of at least 1.5 cm in the treated area as measured by calibrated caliper. BMI interval: 22 at least 1.5 cm in the treated area as measured). General good health confirmed by medical history and skin examination of the treated area. |
| Follow up<br>Intervals<br>Endpoints | 3 treatment visits and follow up visits at 4 weeks, 8 weeks, and 16 weeks.<br>Primary: Circumference and fat reduction in thighs was based on visual assessment of randomized photographs of before (baseline) versus after (16 weeks following last treatment), as well as treated versus untreated thighs by 2 blinded reviewers. The primary endpoint was achieved when at least 80% of treated versus control thighs and pre- versus post-treatment photographs were correctly identified.<br>Secondary:<br>- Circumference reduction on treated thigh compared to control thigh.<br>- Fat thickness reduction on the treated thigh compared to the control thigh, measured by ultrasound.<br>- Fat thickness reduction on the treated thigh compared to the control thigh, measured by caliper.<br>- Investigator satisfaction assessment.<br>- Subject satisfaction self-assessment.<br>- Subjects' comfort level assessed immediately after each treatment. |
| Effectiveness<br>Results | Primary: Met endpoint; blinded reviewers identified correctly (by agreement between the blinded reviewers) both the pre-/post-treatment photographs as well as the treated/un-treated thighs in 81% (30 of 37) subjects.<br>Secondary:<br>- UltraShape treated thighs showed statistically significantly greater circumference reduction compared to the control thighs at each follow up visit.<br>- Fat thickness reduction results following UltraShape treatment were statistically significantly greater compared to controls as measured by ultrasound.<br>- Fat thickness reduction results following UltraShape treatment were greater compared to controls as measured by caliper, although the difference was not statistically significant at all visits.<br>- Investigators were satisfied with results following UltraShape treatment for 77% of the subjects at 4 week follow up, 81% of the subjects at 8 week follow up, and 68% of the subjects at 16 week follow up.<br>- At 4, 8 and 16 weeks following end of treatments, 62%, 61% and 68% of the subjects were satisfied or very satisfied.<br>- Subjects reported no to minimal pain on average for each of the three treatments. |
| Safety<br>Results | No adverse events were reported after 128 treatment sessions were conducted.<br>The only immediate responses recorded were mild and resolved completely within days without any intervention. |
{8}------------------------------------------------
Therefore, clinical evaluation of the UltraShape design demonstrated its favorable performance and safety profile for lipolysis of the flanks and thighs. Results thus further support substantial equivalence of the device as compared to the predicate.
### Substantial Equivalence
The UltraShape Power has the same intended use and very similar indications for use, technological characteristics, and principles of operation as its predicate devices. The added indications for use in the thighs and flanks are supported by the clinical data. The technological differences between the UltraShape Power and its prior clearance mainly consist of minor improvements to the device to facilitate use and customer preference. Nonclinical and clinical studies of the device have demonstrated that the minor differences do not raise new types of safety or effectiveness questions. Thus, the UltraShape Power is substantially equivalent to the predicate devices.
### Conclusion
Syneron's UltraShape Power System is a Focused Ultrasound Stimulator System for Aesthetic Use Class II device that has been evaluated in nonclinical testing in accordance with FDA's Special Controls Guidance Document. Testing demonstrated that the device performs as intended. The UltraShape Power device is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.