K160896 · Syneron Medical, Ltd. · OHV · Jul 8, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160896
Device Name
ULTRASHAPE POWER SYSTEM
Applicant
Syneron Medical, Ltd.
Product Code
OHV · General, Plastic Surgery
Decision Date
Jul 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UltraShape Power System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference.
Device Story
UltraShape Power System uses focused ultrasound to disrupt subcutaneous adipose tissue (SAT) for non-invasive aesthetic body contouring. System comprises control unit, computer, and electromechanical ultrasonic transducers (piezoelectric ceramic core). Transducers convert electrical signals into mechanical acoustic energy to selectively disrupt adipose tissue without damaging neighboring blood vessels, nerves, or muscle. Operator uses real-time video tracking and guidance system to navigate treatment; system captures and processes treatment data. Device features include increased acoustic intensity (up to 660 W/cm²), additional treatment modes, and HD digital camera. Used in clinical settings by trained operators. Output provides aesthetic body contouring; clinical benefit includes reduction in abdominal circumference and fat thickness. Safety features include integrated temperature sensors.
Clinical Evidence
Prospective single-arm study (n=43) evaluated safety and effectiveness. Primary endpoint: reduction in abdominal circumference. Results showed statistically significant reduction (2.55 cm) from 2 to 12 weeks post-treatment. Fat thickness reduction reached 3.63 mm at 12 weeks. Secondary safety study (n=21) confirmed positive safety profile with only one minor adverse event. No serious adverse events reported in either study. 100% of subjects reported comfort during treatment.
Technological Characteristics
Focused ultrasound stimulator; piezoelectric ceramic acoustic core. Components: system control unit, computer, ultrasonic transducers. Features: real-time video tracking/guidance, HD digital camera, integrated temperature sensors. Connectivity: standalone system. Standards: IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2. Sterilization: not specified. Software: performs treatment guidance, power/temperature monitoring, and data processing.
Indications for Use
Indicated for non-invasive reduction in abdominal circumference in patients seeking aesthetic body contouring.
Regulatory Classification
Identification
A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized depiction of a human figure, represented by three overlapping profiles, with a wing-like shape extending from the head. The profiles are facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 8, 2016
Syneron Medical Ltd % Ms. Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K160896
Trade/Device Name: Ultrashape Power System Regulation Number: 21 CFR 878.4590 Regulation Name: Focused Ultrasound Stimulator System for Aesthetic Use Regulatory Class: Class II Product Code: OHV Dated: June 6, 2016 Received: June 6, 2016
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160896
Device Name UltraShape Power System
Indications for Use (Describe)
The UltraShape Power System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <div><span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
#### Syneron Medical Ltd.'s UltraShape Power System
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Syneron Medical Ltd. P.O.B. 550 Industrial Zone Tavor Building Yokneam Illit, 20692 Israel
Phone: 972-73-244-2200 Facsimile: 972-73-244-2202
Company Contact Person: Ruthie Amir, MD, Global Vice President of Clinical, Regulatory and Education
Submission Contact Person: Janice Hogan, Hogan Lovells US LLP
Date Prepared: July 7, 2016
### Name of Device
Syneron UltraShape Power System
#### Common or Usual Name
Focused Ultrasound Stimulator System for Aesthetic Use
#### Classification
21 CFR 878.4590, Class II, product code OHV
# Predicate Devices
Syneron Medical Ltd.'s UltraShape (K141708) Syneron Medical Ltd.'s Contour I V3.1 (K133238)
#### Intended Use / Indications for Use
The UltraShape Power System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference.
#### Device Description
The UltraShape Power System is comprised of multiple components, including the control unit and two ultrasonic transducers. The UltraShape Power System selectively targets subcutaneous adipose tissue via focused ultrasound for the purpose of non-invasive aesthetic body contouring. The transducers are electro-mechanical devices that convert an electrical signal into mechanical (acoustical) energy. The operating parameters of the
{4}------------------------------------------------
UltraShape Power System achieve selective disruption of adipose tissue without damaging neighboring tissues such as blood vessels, nerves, or muscle.
The primary purpose of this submission is to add an increased acoustic intensity level to the device. Additional minor changes since the prior clearances were also included as part of this submission.
# Technological Characteristics
The UltraShape Power System has very similar technological characteristics compared to its predicates. The device continues to be a focused ultrasound device intended for the same indications as the predicates. The proposed and predicate devices are all prescription use only devices and all require the same number of treatments per treatment session. The targeted anatomic area for the UltraShape Power System is identical to the predicate devices. Each of the device systems is comprised of the system control unit, including the computer, and ultrasonic transducer(s). The transducer is the patient-contacting component for both the proposed and predicates devices. The transducers for the UltraShape Power System and the predicates are electromechanical devices that convert the electrical signal into mechanical (acoustic) energy using the same piezoelectric ceramic acoustics core. The UltraShape Power System and the predicates all have a real-time video tracking and guidance system that guides the operator through treatment, and captures and processes the treatment information. The screen size for the real-time display is the same for the proposed and predicate devices. The UltraShape Power System and its predicates all include the same temperature sensors as a safety feature.
In addition, the UltraShape Power System has the same frequency, burst duration, node duration, and user interface as the predicates. The UltraShape Power and its predicates deliver ultrasound energy to the treatment area in a similar manner with similar focal distribution. The primary difference between the UltraShape Power and its predicates is the additional option for treatment at an increased acoustic intensity using the small transducer; the subject UltraShape small transducer includes acoustic intensity level of 660 W/cm', compared to the predicate small transducer's acoustic intensity level of 550 W/cm².
Further, the UltraShape Power System also includes minor modifications to the pulser and power distribution unit compared to the predicates. The Pulser component of the device was modified to be smaller with a new design, and this change did not require any changes in the treatment parameters. An additional treatment mode option was also added. Minor improvements were also made to the UltraShape Power System compared to the predicates, including improved accuracy for output power and temperature measurement and an improved video camera (HD digital instead of analog). None of these changes significantly impact device performance or raise new types of safety and effectiveness questions.
The technological differences between the UltraShape Power and its predicates have been evaluated in clinical testing, and results demonstrated that the device performs in a substantially equivalent manner with no adverse events. Therefore, the changes do not raise any new types of safety or effectiveness questions.
{5}------------------------------------------------
# Performance Data
The following nonclinical performance testing was conducted to support the substantial equivalence of the UltraShape Power System to the predicate device, consistent with FDA's "Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use" (2011). In all instances, the UltraShape Power System functioned as intended.
- Beam profile testing demonstrated that the acoustic energy is delivered and concentrated in the desired target location, at a focal depth similar to that of the predicate devices.
- . Acoustic power testing demonstrated that the acoustic power of the transducers met specifications, is well defined, and presents low variability.
- . In vitro thermal evaluation was conducted and results were passing, demonstrating the absence of thermal coagulation or lesion in tissue.
- . Software verification and validation was performed, and demonstrated that the software performs as intended.
- . Electrical safety (IEC 60601-1, IEC 60601-2-37) and electromagnetic compatibility (IEC 60601-1-2) testing was conducted and results were passing.
- . Biocompatibility of the patient-contacting components of the device was established.
- In vivo testing in an animal model was performed to evaluate the device treatment mode, and results supported the device safety and efficacy profile for the intended use.
In addition, clinical evaluation of the device in the intended population was performed in a prospective single arm study. The study evaluated the safety and effectiveness of the UltraShape Power in 43 enrolled subjects. Eligible subjects received treatments with continued follow up to 12 weeks after treatment. Mean age was 48 years and most of the subjects were females (95%). All subjects were Caucasian. The study primary efficacy endpoint was achieved with statistically significant reduction in abdominal circumference, progressing from 2 weeks following the first treatment to 12 weeks post treatment (2.55 cm reduction). The circumference reduction was consistent at all measurements (midline, 2 cm above midline and 2 cm below midline). Responder analysis results also demonstrated that the majority of subjects achieved circumference reduction with device treatment. The study data also demonstrated statistically significant fat thickness reduction compared to baseline, reaching 3.63 mm reduction at 12 weeks post treatment (mean change adjusted for baseline 3.77±1.11 mm). Subjects were asked to rate their overall satisfaction with treatment on a 5point scale (-2=very disappointed to +2=very satisfied). By 2 weeks after treatment, approximately two-thirds of subjects were satisfied (somewhat satisfied or very satisfied), which continued at 4 weeks follow up; subsequent results reflected slightly lower satisfaction with increasing time from treatment. At 2 week follow-up, 81% of subjects stated that they would recommend the procedure to a friend or family, again with slightly decreasing rates with increasing time from treatment. Approximately half of the subjects were satisfied with the time required for the treatment results at 2 weeks, slightly decreasing over subsequent
{6}------------------------------------------------
follow-up visits, and approximately one-third of subjects felt slimmer after treatment (30%), with similar results at subsequent follow-up visits. Regarding the comfort of the treatment, 100% of subjects felt it was comfortable at all time points, with the exception of a single subject at 8 and 10 weeks (96% of subjects felt comfortable). Overall, subjects reported that they experienced improvement following treatment with the UltraShape Power. Treatment with the UltraShape Power demonstrated a positive safety profile, with no adverse events.
Additional safety data from a prospective clinical study of 21 patients treated with added treatment mode) further confirm the safety profile of the device. Subjects underwent 3 treatment sessions each 2 weeks apart: follow-up visits occurred at 2, 4, 8 and 12 weeks after the last treatment visit. The study results further supported the safety profile of device treatment with only one minor event from which the patient completely recovered without the need for any intervention. Subjects reported minimal discomfort and overall satisfaction with the treatments.
Therefore, performance data demonstrated the safety and effectiveness profile of the UltraShape Power for its intended use.
# Substantial Equivalence
The UltraShape Power has the same indications for use and similar technological characteristics and principles of operation as its predicate devices. The technological differences between the UltraShape Power and previously cleared UltraShape devices, including the increased acoustic intensity, additional treatment mode, and software/hardware changes, do not raise new types of safety or effectiveness questions. Nonclinical and clinical studies have been conducted to evaluate the UltraShape Power, and results confirm that the device performs as intended and in a similar manner compared to the predicates. Thus, the UltraShape Power is substantially equivalent to the predicate devices.
### Conclusion
Syneron's UltraShape Power System is a Focused Ultrasound Stimulator System for Aesthetic Use Class II device that has been evaluated in nonclinical testing in accordance with FDA's Special Controls Guidance Document. Testing demonstrated that the device performs as intended. The UltraShape Power device is substantially equivalent to its predicate devices (K141708, K133238).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.