K170110 · Icu Medical, Inc. · ONB · Aug 10, 2018 · General Hospital
Device Facts
Record ID
K170110
Device Name
Diana Medication Transfer Set
Applicant
Icu Medical, Inc.
Product Code
ONB · General Hospital
Decision Date
Aug 10, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Diana ChemoLock Transfer Set is a sterile, single-use closed system transfer device used for drug preparation to transfer drug from a drug vial to an IV bag for intravenous drug administration. It prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system.
Device Story
Diana ChemoLock Transfer Set is a sterile, single-use ancillary device for the Diana Pharmacy Compounding System. It consists of a stopcock, tubing, ChemoLock connector, syringe, and cassette handle. The device facilitates the transfer of fluids between drug vials and IV bags for drug reconstitution or medication preparation. It functions as a closed system transfer device (CSTD) to prevent environmental contamination and the escape of hazardous drug vapors. Used in hospital pharmacies by trained personnel, the device is placed into the compounding system to automate fluid distribution. The healthcare provider uses the system to prepare medications safely, reducing exposure to hazardous drugs and maintaining sterility during the transfer process.
Clinical Evidence
Bench testing only. Testing included fluid flow, accuracy, chemical compatibility, seal integrity, component bond strength, packaging integrity (ASTM D4169, F1980, F2096, F88), biocompatibility (ISO 10993-1), microbial ingress, hazardous drug exposure evaluation, particulate testing (USP 788), and sterility validation (ISO 11137-1/2, AAMI ST72, USP 85).
Technological Characteristics
Multicomponent fluid transfer set including stopcock, tubing, ChemoLock, syringe, and cassette handle. Materials: Non-DEHP PVC tubing. Sterilization: E-beam (SAL 10^-6). Biocompatibility per ISO 10993-1. Closed system design prevents environmental contamination and vapor escape.
Indications for Use
Indicated for trained healthcare personnel in hospital pharmacy settings for the preparation and transfer of drugs from vials to IV bags for intravenous administration.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 10, 2018
ICU Medical, Inc. Jason Ma Regulatory Affairs Manager 600 N. Field Drive Lake Forest, Illinois 60045
Re: K170110
Trade/Device Name: Diana ChemoLock Transfer Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: July 10, 2018 Received: July 11, 2018
Dear Jason Ma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Alan M.
Stevens -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170110
Device Name
Diana ChemoLock Transfer Set
Indications for Use (Describe)
The Diana ChemoLock Transfer Set is a sterile, single-use closed system transfer device used for drug preparation to transfer drug from a drug vial to an IV bag for intravenous drug administration. It prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
A summary of 510(k) substantial equivalence information in accordance with the requirements of 21 CFR 807.92 for the Diana ChemoLock Transfer Set.
| I. SUBMITTER | |
|------------------|--------------------------------------------------------------------------------|
| Name | ICU Medical |
| Address | 600 N. Field Drive, Lake Forest, Illinois 60045, USA |
| Phone number | (224)-706-2411 |
| Fax number | (224)-706-2917 |
| Contact Person | Jason Ma, Regulatory Affairs Manager |
| Date prepared | 08/01/2018 |
| II. DEVICE | |
| Name of Device | Diana ChemoLock Transfer Set |
| Regulation Name | Intravascular Administration Sets, 21 CFR 880.5440 |
| Common Name | Closed Antineoplastic And Hazardous Drug Reconstitution And<br>Transfer System |
| Regulatory Class | II |
| Product Code: | ONB |
## III. PREDICATE DEVICE
Predicate - K050339 SETS GRI-FILL
Reference - K131549 ChemoLock (utilized for vapor containment characteristics/methods)
## IV. DEVICE DESCRIPTION
Diana ChemoLock Transfer Set is intended for use with the Diana Compounding System. It is a disposable, single use ancillary device for Pharmacy Compounding Devices (PCD).
The transfer set is comprised of multiple components bonded together to form a single device. These components include the following: stopcock, tubing, ChemoLock, syringe, and cassette handle. When placed in the Diana Compounding System, the Diana ChemoLock Transfer Set allows the transfer of fluids from one container to the other container for reconstitution of lyophilized drug or transfer of stock drug solution to prepare medications.
The purpose of this submission is to release a new closed system transfer device (CSTD) drug transfer set to be used with Diana Pharmacy Compounding System
## V. INDICATIONS FOR USE
The Diana ChemoLock Transfer Set is a sterile, single-use closed systemÁda • ^ / Aâ ^; aa used for drug preparation to transfer drug from a drug vial to an IV bag for intravenous drug administration. It prohibits the transfer of environmental contaminants into the system and the escape of druq or vapor concentrations outside the system.
{4}------------------------------------------------
| VI. COMPARISON OF INTENDED USE WITH THE PREDICATE DEVICE | | |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate (K050339) |
| Feature | Diana ChemoLock Transfer Set | SETS GRI-FILL |
| Intended Use | Same | Same |
| Indications for Use | The Diana ChemoLock Transfer Set is a sterile, single-use<br>closed system transfer device used for drug preparation to<br>transfer drug from a drug vial to an IV bag for intravenous<br>drug administration. It prohibits the transfer of<br>environmental contaminants into the system and the<br>escape of drug or vapor concentrations outside the<br>system. | SETS GRI-FILL 3.0 1 WAY and 2 WAY fluid transfer sets are<br>ancillary devices used in conjunction with the GRI-FILL 3.0<br>pharmacy compounder in the hospital pharmacy to provide a<br>fluid pathway through which one or two source substances are<br>delivered into a final IV container or syringe.<br><br>SETS GRI-FILL 3.0 MULTIPLE fluid transfer sets are ancillary<br>devices used as fluid pathways in conjunction with the GRI-FILL<br>3.0 pharmacy compounder and associated I WAY or 2 WAY<br>transfer sets through which the same substance from up to 6<br>source containers may be delivered into a final IV container.<br><br>This device should not be used with lipids.<br><br>This device is intended to be used by trained health-care<br>personnel. It is restricted to sale by or on the order of a<br>physician. |
| Classification | Class II | Class II |
| Product Code | ONB | LHI |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 880.5440 |
| Use | Single Use | Single Use |
| Prescription/OTC<br>Use | Prescription Use | Prescription Use |
{5}------------------------------------------------
| VI. COMPARISON OF INTENDED USE WITH THE PREDICATE DEVICE | | |
|---------------------------------------------------------------------------------------|------------------------------------|------------------------------------|
| | Subject Device | Predicate (K050339) |
| Feature | Diana ChemoLock Transfer Set | SETS GRI-FILL |
| Compatible Set to a particular compounding system | Diana Automated Compounding System | GRI-FILL 3.0 Compounding System |
| Intended for Direct Connection to Patient During Compounding | No | No |
| Use Environment for Compounding | Standard Hospital Pharmacy Setting | Standard Hospital Pharmacy Setting |
| Target Users | Trained health-care personnel | Trained health-care personnel |
| Closed system (fluid not in contact with any reusable part of the compounding device) | Yes | Yes |
Overall, Diana ChemoLock Transfer Set has the same intended use as the predicate. The differences in indications (e.g. inclusion of hazardous drug transfer and vapor containment) do not alter the intended use of the compared to the predicate, as both devices are intended for fluid transfer and preparation, and do not raise different questions of safety and effectiveness.
| VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | | |
|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate (K050339) |
| Feature | Diana ChemoLock Transfer Set | SETS GRI-FILL |
| System | Syringe Unit | Syringe Unit |
| VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | | |
| | Subject Device | Predicate (K050339) |
| Feature | Diana ChemoLock Transfer Set | SETS GRI-FILL |
| Components | Connector to Source Container(Drug Vial) | Connector to Source Container(Drug Vial) |
| | Connector to Receiving Container (IV Bag) | Connector to Receiving Container (IV Bag) |
| | Connection Tubing | Connection Tubing |
| | Integral stopcock (distribution purpose) | Integral distributor (distribution purpose) |
| | Cassette Handle | Not Applicable |
| Principles<br>of<br>Operation | Multicomponent device compatible with compounding<br>system to distribute fluid from sourcing container to<br>receiving container | Multicomponent device compatible with compounding system to<br>distribute fluid from sourcing container to receiving container |
| Fluid<br>Transfer<br>Mechanism | Integrated Syringe with connectors - Single Channel | Integrated Syringe with connector - Single or multiple Channel |
| Closed<br>system<br>(fluid<br>not<br>in<br>contact<br>with<br>any<br>reusable<br>part<br>of<br>the<br>compounding<br>device) | Yes | Yes |
| Connections<br>to<br>Source Containers | proprietary closed connector | Male luer connector or vented bag spike |
| Connections<br>to<br>Final Containers | proprietary closed connector | Male luer connector. Also supplied with female to female luer to<br>allow for female to male luer connection |
| Fluid<br>Contact<br>Material for tubing | PVC with Non-DEHP | PVC with DEHP Plasticizer |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic |
| VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | | |
| | Subject Device | Predicate (K050339) |
| Feature | Diana ChemoLock Transfer Set | SETS GRI-FILL |
| Biocompatibility | Per FDA Guidance and ISO 10993-1<br>Biological Effect<br>· Hemocompatibility<br>· Cytotoxicity<br>· Sensitization<br>· Intracutaneous Irritation<br>· Acute Systemic Toxicity<br>· Pyrogenicity | Information Not Available |
| Sterilization | Radiation (E-beam) | Ethylene Oxide |
| SAL | 10-6 | Information Not Available |
{6}------------------------------------------------
{7}------------------------------------------------
The technological differences between the Diana ChemoLock Transfer Set and the predicate are:
- Use in an integrated system including syringe, stopcock and ChemoLock system for the proposed to an integrated syringe • with luer or proprietary closed connector for the predicate
- The Diana ChemoLock Transfer Set uses Non-DEHP PVC tubing compared to DEHP PVC tubing in the predicate
- Sterilization method
These differences in technology do not raise different questions of safety or effectiveness.
{8}------------------------------------------------
## VIII. PERFORMANCE DATA
### Functional Testing
Functional testing was conducted to demonstrate the functionality of the Diana ChemoLock Transfer set that consists of ChemoLock, stopcock, tubing, syringe, and cassette handle. This testing follows the FDA quidance document of "Intravascular Administration Sets Premarket Notification Submissions [510(k)]".
Functional Testing included
- Fluid Flow ● .
- . Accuracy Testing
- Proid i Trow
Positive Pressure
Chemical Compatibility .
- Internal Seal Integrity .
- Stopcock Handle Torque ●
- Component Bond Strenath Testing ●
- Visual Inspection
### Packaging Integrity and Shelf Life
Package integrity and Shelf Life was conducted according to FDA recognized consensus standards of ASTM D4169 Standard Practice For Performance Testing Of Shipping Containers And Systems, ASTM F1980 Standard Guide For Accelerated Aging Of Sterile Barrier Systems For Medical Devices, ASTM F2096 Standard Test Method For Detecting Gross Leaks In Packaging By Internal Pressurization (Bubble Test), and ASTM F88 Standard Test Method For Seal Strength Of Flexible Barrier Materials.
#### Biocompatibility
The biocompatibility evaluation for Diana ChemoLock Transfer Set device was conducted in accordance with the FDA Guidance for Industry and FDA Staff - "Use of International Standard ISO 10993-1, `Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process", June 16, 2016; and ISO 10993-1 "Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process", as recognized by FDA. Testing included:
- Hemocompatibility ●
- . Cytotoxicity
- Sensitization ●
- . Intracutaneous Irritation
- Acute Systemic Toxicity .
- Material Mediated Pyrogencity .
## Microbial Ingress Testing
Microbial inqress testing was conducted to demonstrate that needless access site of Diana ChemoLock Transfer Set maintains physical integrity and mechanically prohibits the transfer of environmental contaminants into the system in use after access (the device is intended for singleuse). The tests were conducted consistent with the recommendations for microbial ingress testing provided in FDA guidance titled "Intravascular Administration Sets Premarket Notification Submissions [510(k)]" issued on July 11, 2008.
#### Emission, Dry Disconnection, and CSTD Hazardous Drug Exposure Evaluation
Emission, Dry Disconnection, and CSTD Hazardous Drug Exposure Evaluation were performed to validate no escape of drug from the subject device outside the closed system.
#### Particulates
Particulate contamination testing was performed by following USP <788> (method 1) to demonstrate particulate levels in the subject device meet USP 788 requirements.
{9}------------------------------------------------
### Sterility Testing
E-beam sterilization process validation was conducted according to ISO 11137-1 and ISO 11137-2, Sterilization of health care products – Radiation: Requirements for development, validation and routine control of a sterilization process for medical devices.
Bacterial endotoxin testing was conducted based on AAMI ST72 and USP <85>, and followed FDA Guidance for Industry - Pyrogen and Endotoxins Testing: Questions and Answers.
### IX. CONCLUSION
The Diana ChemoLock Transfer Set is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.