K131549 · Icu Medical, Inc. · ONB · Jul 26, 2013 · General Hospital
Device Facts
Record ID
K131549
Device Name
CHEMOLOCK, CHEMOLOCK PORT
Applicant
Icu Medical, Inc.
Product Code
ONB · General Hospital
Decision Date
Jul 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The ChemoLock Closed System Drug Transfer Device prevents the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system. The ChemoLock is needlefree and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.
Device Story
ChemoLock is a needlefree, single-use Closed System Drug Transfer Device (CSTD) comprising a two-piece system (ChemoLock and Port). It facilitates the transfer of cytotoxic drugs between vials, bags, and syringes. The device uses passive, self-sealing elastomeric double-membrane mechanisms to maintain a closed system, preventing drug/vapor escape and environmental/bacterial ingress. It connects to standard ISO-594 luer systems. Operated by healthcare personnel in clinical environments, the device ensures a dry, leak-proof disconnection. By maintaining a closed fluid path, it protects clinicians from hazardous drug exposure and needlestick injuries, while protecting patients from microbial contamination.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including microbial ingress testing and chemical egress studies (cyclophosphamide compounding trials). Testing demonstrated effective prevention of bacterial ingress and surface contamination. Performance testing followed guidance for intravascular administration sets.
Technological Characteristics
Materials: Biocompatible thermoplastics and silicone. Principle: Passive, self-sealing elastomeric double-membrane mechanism for closed fluid transfer and pressure equalization. Connectivity: Standard ISO-594 luer compatibility with permanent locking features. Sterilization: E-beam. Form factor: Two-piece system (ChemoLock and Port).
Indications for Use
Indicated for healthcare personnel handling cytotoxic medications to prevent needlestick injuries and exposure to drugs or vapors, and to prevent environmental/bacterial contamination of the system. Applicable for use in clinical settings where cytotoxic drugs are prepared or administered.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
BD Phaseal Closed System Drug Transfer Device (K123213)
ChemoClave Cytotoxic Medication Preparation and Delivery System (K081361)
{0}------------------------------------------------
K131549
# JUL 2 6 2013
### Section 5 510(k) Summary
The following information is being submitted in accordance with the requirements of 21 CFR 807.92.
| Company Name: | ICU Medical, Inc.<br>4455 Atherton Drive |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Salt Lake City, UT 84123 |
| Contact Person: | Katherine Kenner<br>Regulatory Affairs Specialist<br>ICU Medical, Inc.<br>Phone: (801) 264 1702<br>Fax: (801) 264 1755 |
| Preparation Date: | May 28, 2013 |
| Device (Trade Name): | ChemoLock Closed System Drug Transfer Device |
| Common/Usual Name: | Closed System Drug Transfer Device |
| Classification Names: | 21 CFR 880.5440. Closed System Drug Transfer Device. Product Code:<br>800NB and 80FPA. |
| Predicate Device: | K123213 BD Phaseal Closed System Drug Transfer Device<br>K081361 ChemoClave Cytotoxic Medication Preparation and Delivery<br>System<br>K082806 Spinning Spiros |
| Device Description: | The ChemoLock is a needlefree, single-use, Closed System Drug Transfer<br>Device (CSTD). The ChemoLock has a mechanical means to prevent the<br>transfer of environmental contaminants into the system, and the escape of<br>drug or vapor concentrations outside the system. The system includes<br>closed vial and bag access devices, a closed syringe adapter and closed<br>patient administration sets. All components of the system include passive<br>self-sealing mechanisms which cannot be deactivated and remain<br>protective through disposal. |
| Intended Use: | The ChemoLock Closed System Drug Transfer Device prevents the<br>transfer of environmental contaminants, including bacterial and airborne<br>contaminants into the system, and the escape of drug or vapor<br>concentrations outside the system. The ChemoLock is needlefree and<br>cannot be deactivated, which will passively aid in preventing needlestick<br>injuries and the exposure to cytotoxic medications for healthcare<br>personnel. |
| Technology: | The ChemoLock Closed System Drug Transfer Device employs the same<br>fundamental scientific technology as its predicate devices. Please see<br>Table 1 for more information. |
| Feature | ChemoClave | Spinning Spiros | Phaseal | ChemoLock | Comparison |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| 510K<br>NUMBER | K081361 | K082806 | K123213 | K131549 | N/A |
| Equalizes<br>Vial Pressure<br>when<br>Accessed | Yes | Not Applicable | Yes | Yes | Identical to<br>K123213 and<br>K081361 |
| Feature | ChemoClave | Spinning Spiros | Phaseal | ChemoLock | Comparison |
| Prevents<br>Escape of<br>Drug or<br>Vapor<br>Concentratio | Yes | Yes | Yes | Yes | Identical |
| Closed Drug<br>Transfer<br>Mechanism | Elastomeric<br>single membrane<br>(ISO) | Elastomeric<br>double<br>membrane (ISO) | Elastomeric<br>double<br>membrane (non-<br>ISO) | Elastomeric<br>double<br>membrane (non-<br>ISO) | Identical to<br>K123213 |
| Activation<br>Mechanism | ISO Luer locking | ISO Luer Locking | Non-ISO push-<br>together<br>connection with<br>bayonet lock | Non-ISO push-<br>together<br>connection with<br>clip locks | Similar to<br>K123213 |
| Available<br>Accessories<br>and Useable<br>Components | Vial Access, Bag<br>Access, Syringe<br>Adapter and<br>Patient<br>Administration<br>Sets | Syringe Adapter<br>and Patient<br>Administration<br>Sets | Vial Access, Bag<br>Access, Syringe<br>Adapter, Injection<br>Port Adapter | Vial Access, Bag<br>Access, Syringe<br>Adapter and<br>Patient<br>Administration<br>Sets | Identical to<br>K081361 |
| NeedleFree | NeedleFree | NeedleFree | No, 18G needle<br>with safety sleeve<br>and/or steel spike | NeedleFree | Identical to<br>K081361 and<br>K082806 |
| Connection<br>to External<br>Equipment | Standard ISO luer<br>connection | Standard ISO Luer<br>connection with<br>permanent<br>locking feature<br>which prevents<br>removal and stays<br>protective<br>through disposal | Standard ISO Luer<br>Connections and<br>Vial Spikes | Standard ISO luer<br>connection with<br>permanent<br>locking feature<br>which prevents<br>removal and stays<br>protective<br>through disposal | Identical to<br>K082806 |
| Microbial<br>Barrier | Yes | Yes | Yes for Vial<br>Access<br>No for Syringe<br>Adapter | Yes | Identical to<br>K081361 and<br>K082806 |
| Sterilization | Ebeam | Ebeam | EO | Ebeam | Identical to<br>K081361 and<br>K082806 |
| Materials of<br>Construction | Biocompatible<br>thermoplastics<br>and silicone | Biocompatible<br>thermoplastics<br>and silicone | Biocompatible<br>thermoplastics,<br>silicone and<br>stainless steel | Biocompatible<br>thermoplastics<br>and silicone | Identical to<br>K081361 and<br>K082806 |
| Single Use | Yes | Yes | Yes | Yes | Identical |
#### Table 1: Summary Comparison of Technological Characteristics Table
{1}------------------------------------------------
·
・
•
.
{2}------------------------------------------------
Determination of Substantial Equivalence:
Non Clinical Performance Data Summary The demonstration of substantial equivalence is based on a comparison of features to the predicate devices and an assessment of non-clinical performance data. Information is included with this 510(k) submission that supports this determination.
The ChemoLock Closed System Drug Transfer Device is intended to be used with closed vial and bag access devices, closed syringe adapters, and closed patient administration sets. The two piece system, comprised of two components called the ChemoLock and the ChemoLock Port (or Port), is intended to allow the user to adapt an ISO-594 luer system to a custom adapted, Closed System Drug Transfer Device. The female luer end of the ChemoLock is compatible with all male 150 connections and the other side is only compatible with the Port. The male end of the Port is compatible with all female ISO connections while the other side is only compatible with the ChemoLock.
The ChemoLock and Port will prevent leakage of fluid when in the inactivated state. When the ChemoLock is connected to the Port, the fluid path is activated and becomes an onen two way conduit. When disconnected, the ChemoLock and Port will self-seal and offer a dry, leak proof disconnection.
Microbial Ingress testing as provided in this submission shows that the ChemoLock Closed System Drug Transfer Device has the ability to prevent bacterial ingress. Said testing shows that the device will prevent bacterial ingress completely if used in accordance with the directions for use. The prevention of bacterial ingress is specifically important for pediatrics and immune-compromised patients that would otherwise be at risk of infections.
The Evaluation of Closed-system Drug Transfer Devices as provided in this submission shows that the ChemoLock Closed System Drug Transfer Device is able to prevent the egress of chemicals both during preparation and transfer of the cytotoxic agent. Results of the study suggest that the ChemoLock system was effective in preventing detectible surface contamination during three separate trials of compounding activities with known amounts of cyclophosphamide. The prevention of environmental contamination is specifically important for the containment of cytotoxic pharmaceuticals which will passively prevent the exposure to such chemicals by healthcare workers.
The ChemoLock Closed System Drug Transfer Device has been tested in accordance with its performance specification which accommodates known functional requirements.
The ChemoLock Closed System Drug Transfer Device configurations are individually packaged and pre-sterilized in a peel type, Multivac pouch with sealed Tyvek lidding.. The device can be included as part of a closed vial or bag access device, closed syringe adapter, or closed patient administration set. This device may be included as part of a custom kit or set - as prescribed by a physician. ICU Medical performs analysis and design verification testing based on predetermined criteria, which is documented in the Performance Specification contained in this
{3}------------------------------------------------
submission. All testing meets these performance criteria as defined for the device.
ICU Medical has also performed testing that is recommended by the guidance document "Intravascular Administration Sets Premarket Notification Submissions [510(k)]" and successful testing is included as part of this submission in the Extended Use Protocol Results document.
Conclusion:
Performance testing included within this 510(k) demonstrates that the ChemoLock Closed System Drug Transfer Device is safe, effective and performs in an equivalent manner to the predicate devices and in accordance with its intended use.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 11, 2013
ICU Medical, Incorporated Ms. Katherine Kenner Regulatory Affairs Specialist 4455 Atherton Drive SALT LAKE CITY UT 84123
Re: K131549
Trade/Device Name: ChemoLock Closed System Drug Transfer Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: ONB, FPA Dated: May 28, 2013 Received: May 29, 2013
Dear Ms. Kenner:
This letter corrects our substantially equivalent letter of July 26, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Kenner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
**Mary S. Runner -S**
Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Section 4 Indications For Use
Page 1 of 1
510(k) Number (if known): K1315-49
ChamoLock Closed System Drug Transfar Device Device Name:
The ChemoLock Closed System Drug Transfer Device prevents the transfer of Indications for Use: environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system. The ChemoLock is needlefree and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use ✓<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use __ |
|--------------------------------------------|-------------------------------------------|--------------------------|
| | Richard C. Chapman<br>2013.07.25 09:44:14 | (Optional Format 1-2-96) |
Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| 510(k) Number: | K131549 |
|----------------|---------|
|----------------|---------|
1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.