Invia Foam Dressing Kits with FitPad, Invia Foam Dressing Kits with FitPad, Invia Gauze Dressing Kits with FitPad, Invia FitPad, Invia Transparent Film
K170088 · Medela AG · OMP · Apr 6, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170088
Device Name
Invia Foam Dressing Kits with FitPad, Invia Foam Dressing Kits with FitPad, Invia Gauze Dressing Kits with FitPad, Invia FitPad, Invia Transparent Film
Applicant
Medela AG
Product Code
OMP · General, Plastic Surgery
Decision Date
Apr 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Invia Foam Dressing Kit with FitPad in conjunction with the Invia Liberty Negative Pressure Wound Therapy (NPWT) systems is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it creates an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing granulation tissue formation and perfusion, and by removing exudate and infectious material. The Invia Foam Dressing Kit with FitPad is appropriate for use on the following wounds: Acute or sub-acute wounds, Chronic wounds, Dehisced wounds, Pressure ulcers, Diabetic/neuropathic ulcers, Venous insufficiency ulcers, Traumatic wounds, Partial thickness burns, Flaps and grafts. The Invia Gauze Dressing Kit with FitPad in conjunction with the Invia Liberty Negative Pressure Wound Therapy (NPWT) systems is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it creates an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing granulation tissue formation and perfusion, and by removing exudate and infectious material. Invia Gauze Dressing Kit with FitPad is appropriate for use on the following wounds: Acute or sub-acute wounds, Chronic wounds, Dehisced wounds, Pressure ulcers, Diabetic/neuropathic ulcers, Venous insufficiency ulcers, Traumatic wounds, Partial thickness burns, Flaps and Grafts.
Device Story
Dressing kits (foam or gauze) with FitPad suction interface and transparent film; used with Invia Liberty/Motion NPWT systems. System applies negative pressure to wound bed; promotes healing by removing exudate/infectious material, reducing granulation tissue formation, and preparing wound bed for closure. Used in acute, extended, and home care settings; applied by clinicians using aseptic/sterile technique. Output is negative pressure environment; healthcare providers monitor wound progress and system function. Benefits include wound bed preparation and accelerated healing.
Clinical Evidence
Clinical study conducted per ISO 10993-10 (Annex C) to evaluate skin irritation of the Invia transparent film (adhesive drape). Results demonstrated the film is not a significant skin irritant. Performance testing confirmed ability to generate/maintain pressure with intended suction pumps.
Technological Characteristics
Materials: reticulated flexible polyether-based polyurethane hydrophobic foam, gauze, transparent film. Components provided sterile, single-use. Interface: FitPad suction interface with quick-connect tubing. Standards: ISO 14971 (risk), ISO 10993 (biocompatibility), ISO 11135 (sterilization), ASTM F1980 (aging), DIN EN ISO 11607-1 (packaging), EN 62366 (usability).
Indications for Use
Indicated for patients requiring negative pressure wound therapy for acute, sub-acute, chronic, dehisced, pressure, diabetic/neuropathic, venous insufficiency, traumatic wounds, partial thickness burns, and flaps/grafts to promote healing by secondary or tertiary intention.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the address for the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The address is likely for a specific department or office within the FDA.
April 6, 2017
Medela, AG % Julian Thorns Regulatory Affairs & Project Manager Prosystem AG Beim Strohhause 17 Hamburg, 20097 DE
Re: K170088
Trade/Device Name: Invia Foam Dressing Kits With Fitpad, Invia Gauze Dressing Kits With Fitpad, Invia Fitpad, Invia Transparent Film Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: February 3, 2017 Received: February 6, 2017
Dear Julian Thorns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K170088
Device Name Invia Foam Dressing Kit with FitPad
#### Indications for Use (Describe)
The Invia Foam Dressing Kit with FitPad in conjunction with the Invia Liberty Negative Pressure Wound Therapy (NPWT) systems is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it creates an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing granulation tissue formation and perfusion, and by removing exudate and infectious material.
The Invia Foam Dressing Kit with FitPad is appropriate for use on the following wounds:
- Acute or sub-acute wounds
- Chronic wounds
- Dehisced wounds
- Pressure ulcers
- Diabetic/neuropathic ulcers
- Venous insufficiency ulcers
- Traumatic wounds
- Partial thickness burns
- Flaps and grafts
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size:10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K170088
Device Name Invia Gauze Dressing Kit with FitPad
#### Indications for Use (Describe)
The Invia Gauze Dressing Kit with FitPad in conjunction with the Invia Liberty Negative Pressure Wound Therapy (NPWT) systems is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it creates an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing granulation tissue formation and perfusion, and by removing exudate and infectious material.
Invia Gauze Dressing Kit with FitPad is appropriate for use on the following wounds:
- Acute or sub-acute wounds
- Chronic wounds
- Dehisced wounds
- Pressure ulcers
- Diabetic/neuropathic ulcers
- Venous insufficiency ulcers
- Traumatic wounds
- Partial thickness burns
- Flaps and Grafts
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "medela" in bold, black font. To the right of the word is an orange logo. The logo is made up of three orange teardrop shapes that are arranged in a circular pattern.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
DATE: January 24, 2017
# SUBMITTER
Medela AG Lättichstrasse 4b 6341 Baar / Switzerland Phone: +41 (0)41 561 66 71 Fax: +41 (0)41 769 51 00
# OFFICIAL CORRESPONDENT
| | Mr. Markus Bütler |
|--------|-------------------------|
| | Medela AG |
| | Lättichstrasse 4b |
| | 6341 Baar / Switzerland |
| Phone: | +41 (0)41 561 66 71 |
| Fax: | +41 (0)41 769 51 00 |
# DEVICE IDENTIFICATION
| OVERALL TRADE NAME: | Invia Dressing Kits with FitPad |
|----------------------|-------------------------------------------------------------------------|
| TRADE NAME: | Invia Foam Dressing Kits with FitPad |
| MODEL NUMBER: | 087.6221, 087.6222, 087.6223, 087.6224, 087.6225,<br>087.6226, 087.6227 |
| TRADE NAME: | Invia Gauze Dressing Kits with FitPad |
| MODEL NUMBER: | 087.7170, 087.7171, 087.7172 |
| TRADE NAME: | Invia FitPad, Invia Transparent Film |
| MODEL NUMBER: | 087.0028, 087.0030 |
| COMMON/USUAL NAME: | Negative Pressure Wound Therapy System |
| CLASSIFICATION NAME: | Negative Pressure Wound Therapy Powered Suction Pump |
| PRODUCT CODE: | OMP |
| REGULATION NUMBER: | per 21 CFR 878.4780 |
| REVIEW PANEL: | General and Plastic Surgery |
| DEVICE CLASS: | II |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "medela" in a bold, sans-serif font. To the right of the word is a logo consisting of three orange teardrop shapes arranged in a triangular pattern. The logo is simple and abstract, and it complements the text.
# PREDICATE DEVICE(S)
INVIA FOAM DRESSING KITS WITH FITPAD
| DEVICE TRADE NAME: | Invia Foam Dressing Kit |
|--------------------|-------------------------|
|--------------------|-------------------------|
APPLICANT: Medela AG
510(K) NO.: K151261
2. PREDICATE DEVICE
| DEVICE TRADE NAME: | Avance® Foam Dressing Kits |
|--------------------|------------------------------|
| APPLICANT: | Molnlycke Health Care US LLC |
| 510(K) NO.: | K141847 |
# INVIA GAUZE DRESSING KITS WITH FITPAD
| 1. PRIMARY PREDICATE DEVICE | |
|-----------------------------|----------------------------------------|
| DEVICE TRADE NAME: | Invia Motion |
| APPLICANT: | Medela AG |
| 510(K) NO.: | K113678 |
| 2. PREDICATE DEVICE | |
| DEVICE TRADE NAME: | KCI NPWT Gauze Dressing |
| APPLICANT: | KCI USA, Inc. (Kinetic Concepts, Inc.) |
| 510(K) NO.: | K123507 |
{6}------------------------------------------------
# DEVICE DESCRIPTION
### INVIA FOAM DRESSING KITS WITH FITPAD
The Invia Foam Dressing Kit with FitPad consists of a foam, a suction interface (Invia FitPad) and transparent film (one or more pieces depending on the kit size).
The Invia Foam Dressing Kit with FitPad is available in four sizes.
The foam is manufactured using a reticulate flexible polyether and polyurethane hydrophobic material, with a high tensile strength (20 PSI*) and charcoal color. The FitPad suction interface features a quick connecting interface to the canister tubing for easy and secure connection. The transparent film is designed with self-explanatory markings to facilitate the ease of use.
### INVIA GAUZE DRESSING KITS WITH FITPAD
The Invia Foam Dressing Kit with FitPad consists of a gauze pad, one or two transparent films (depending on the kit size), a suction interface (Invia FitPad), and saline.
The Invia Gauze Dressing Kit with FitPad is available in two sizes.
### INTENDED USE
### INVIA® FOAM DRESSING KIT WITH FITPAD
The Invia Foam Dressing Kit with FitPad is intended for use in conjunction with the Invia Motion and Invia Liberty Negative Pressure Wound Therapy (NPWT) systems. The Invia Motion and Invia Liberty NPWT systems are intended for use in acute, extended and home care settings. Users are directed to the Invia Motion and Invia Liberty NPWT system labeling for additional safety information and instruction for use. To help ensure safe and effective use, the Invia Foam Dressing Kits with FitPad are to be used only with the approved therapy units.
The components of the Invia Foam Dressing Kit with FitPad are packaged sterile and are for single use only. Do not use if sterile package is damaged or opened prior to use.
To apply Invia Foam Dressing Kit with FitPad, use clean/aseptic or sterile techniques in accordance with local protocol.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the word "medela" in bold, black font. To the right of the word is an orange graphic that appears to be three teardrop shapes arranged in a triangular pattern. The teardrop shapes are slightly curved and have a gradient effect, with the color fading from dark orange to light orange.
### INVIA® GAUZE DRESSING KIT WITH FITPAD
Invia Gauze Dressing Kit with FitPad is intended for use in conjunction with the Invia Motion and Invia Liberty Negative Pressure Wound Therapy (NPWT) systems. The Invia Motion and Invia Liberty NPWT systems are intended for use in acute, extended and home care settings. Users are directed to the Invia Motion and Invia Liberty NPWT System labeling for additional safety information and instruction for use. To help ensure safe and effective use, the Invia Gauze Dressing Kits with FitPad are to be used only with the approved therapy units.
The components of the Invia gauze dressing kit with FitPad are packaged sterile and are for single use only. Do not use if sterile package is damaged or opened prior to use.
To apply Invia gauze dressing kit with FitPad, use clean/aseptic or sterile techniques in accordance with local protocol.
# DETERMINATION OF SUBSTANTIAL EQUIVALENCE
# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
The Invia Foam Dressing Kit and the predicate devices, Avance Foam Dressing Kits (K141847) and Invia Foam Dressing Kits (K151261), have the same indications for use, all use a reticulated flexible polyether based polyurethane hydrophobic foam dressing material, are provided sterile, offer kit sizes in similar dimensions, are provided with an external suction interface tubing, and are used for negative pressure wound therapy.
The Invia Gauze Dressing Kit and the predicate devices, Invia Motion (K113678) and KCI NPWT Gauze Dressing (K123507), have the same indications for use, all use the same gauze as dressing material, are provided single packed and sterile, offer kit sizes in similar dimensions, are provided with a tubing with suction interface, and are used for negative pressure wound therapy.
### SUMMARY OF NON-CLINICAL TESTS:
The Invia Dressing Kits with FitPad comply with voluntary standards for risk management, biocompatibility, and sterilization. The following data were provided in support of the substantial equivalence determination:
- . Risk Analysis developed in accordance with ISO 14971:2007.
- . Biocompatibility studies in accordance with: EN ISO 10993-1, EN ISO 10993-5, EN ISO 10993-12, EN ISO 10993-18
- . Sterilization Validation in accordance with ISO 11135:2014
- Analysis on Accelerated Aging according to ASTM F1980
- Packaging Validation in accordance with DIN EN ISO 11607-1
- Transport Validation
- Usability testing in accordance with EN 62366:2008
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Medela, a company that specializes in breastfeeding products and medical vacuum technology. The logo consists of the word "medela" in black, sans-serif font, followed by a stylized orange flower-like symbol. The symbol is composed of three teardrop-shaped petals arranged in a triangular formation.
- . Performance testing which demonstrates that Invia Dressing Kits have the ability to accurately generate and maintain pressure under usage conditions in combination with the intended suction pumps
SUMMARY OF CLINICAL TESTS:
A clinical study was conducted according to the ISO 10993-10 (Annex C) to demonstrate that the Invia transparent film (adhesive drape) of the Invia Dressing Kits with FitPad is not a significant skin irritant.
### CONCLUSION:
The differences between the Invia Dressing Kits with FitPad and its predicate devices do not introduce a new intended use and do not raise new questions of safety and effectiveness. Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended.
From the results of clinical and nonclinical testing described, Medela AG concludes that the Invia Dressing Kits with FitPad are substantially equivalent to the legally marketed predicate devices and do not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.