K123507 · Kci USA, Inc. · OMP · Apr 4, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123507
Device Name
KCI NPWT GAUZE DRESSING
Applicant
Kci USA, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Apr 4, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The KCI NPWT Gauze Dressing is intended to be used with the following KCI Therapy Units (ActiV.A.C.®, InfoV.A.C.®, V.A.C. Simplicity™, V.A.C. ® Freedom, V.A.C. ATS® and V.A.C.Ulta™ Therapy Systems). The system is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and by removing exudate and infectious material. Wound types include chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts. The KCI NPWT Gauze Dressing is not intended for use with instillation therapy, intermittent therapy or over closed incisions.
Device Story
KCI NPWT Gauze Dressing is a convenience kit for negative pressure wound therapy (NPWT). Components include antimicrobial gauze (Kerlix AMD), SensaT.R.A.C. Pad, V.A.C. Drapes, and a measuring ruler. The dressing connects to KCI therapy units (ActiV.A.C., InfoV.A.C., V.A.C. Simplicity, V.A.C. Freedom, V.A.C. ATS, V.A.C.Ulta) to deliver negative pressure to the wound bed. The system removes exudate and infectious material while promoting granulation tissue formation and perfusion. Used in clinical settings by healthcare providers to prepare wounds for closure. The provider applies the dressing and monitors the wound; the therapy unit manages pressure delivery. Benefits include an environment conducive to healing for various wound types.
Clinical Evidence
Bench testing only. No clinical data required for substantial equivalence. Testing verified that the dressing, when used with KCI V.A.C. units, delivers negative pressure within specifications for 72 hours, effectively removes exudate, and does not trigger false blockage alarms.
Technological Characteristics
Convenience kit containing antimicrobial gauze, polyurethane film drape with acrylic adhesive, and SensaT.R.A.C. interface pad. Designed for single-use. Compatible with KCI V.A.C. powered suction pumps. Non-sterile packaging; sterile components.
Indications for Use
Indicated for patients with chronic, acute, traumatic, subacute, dehisced wounds, partial-thickness burns, ulcers (diabetic, pressure, venous), flaps, and grafts. Used to promote healing via secondary or tertiary intention, reduce edema, promote granulation/perfusion, and remove exudate/infectious material. Contraindicated for use with instillation therapy, intermittent therapy, or over closed incisions.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
Smith and Nephew's RENASYS™ Gauze NPWT Wound Dressing Kit with Softport (K110647)
Submission Summary (Full Text)
{0}------------------------------------------------
K123507
## 510(k) SUMMARY
APR 4 2013
## KCI NPWT Gauze Dressing
| Date prepared | March 4, 2013 | | |
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| 510(k) owner | | | |
| • Name | KCI USA, Inc. (Kinetic Concepts, Inc.) | | |
| • Address | 6203 Farinon Drive; San Antonio, Texas 78249 | | |
| • Phone number | 210 255-6345 | | |
| • Fax number | 210 255-6727 | | |
| • Name of<br>contact person | Anona Goebel | | |
| Name of the device | | | |
| Trade or<br>proprietary<br>name | KCI NPWT Gauze Dressing | | |
| Common or<br>usual name | Negative pressure wound therapy system (gauze dressing component) | | |
| Classification<br>name | Negative pressure wound therapy powered suction pump<br>(and components) | | |
| Legally<br>marketed<br>device(s) to<br>which<br>equivalence is<br>claimed | Smith and Nephew's RENASYS™ Gauze NPWT Wound Dressing Kit with<br>Softport<br>510(k) Number: K110647 | | |
| Device description | | | |
| • Device design | The KCI NPWT Gauze Dressing is a convenience kit that consists of the<br>following currently marketed products:<br>• One Kerlix AMD Roll, identical to predicate (antimicrobial gauze wound<br>dressing)<br>• One V.A.C. SensaT.R.A.C.™ Pad (connection to KCI NPWT source)<br>• Two V.A.C. Drapes<br>• One wound measuring ruler | | |
| | The currently marketed sterile products are kitted in a non-sterile package.<br>The kit is provided in one size and is intended to be single use. | | |
| • Indications for<br>Use | The KCI NPWT Gauze Dressing is intended to be used with the following<br>KCI Therapy Units (ActiV.A.C.®, InfoV.A.C.®, V.A.C. Simplicity™, V.A.C.®<br>Freedom, V.A.C. ATS® and V.A.C.Ulta™ Therapy Systems). The system is<br>intended to create an environment that promotes wound healing by<br>secondary or tertiary (delayed primary) intention by preparing the wound<br>bed for closure, reducing edema, promoting granulation tissue formation<br>and perfusion, and by removing exudate and infectious material. Wound<br>types include chronic, acute, traumatic, subacute and dehisced wounds,<br>partial-thickness burns, ulcers (such as diabetic, pressure or venous<br>insufficiency), flaps and grafts. | | |
| | The KCI NPWT Gauze Dressing is not intended for use with instillation | | |
| Summary of the<br>technological<br>characteristics<br>of the device<br>compared to<br>the predicate<br>device | Feature | KCI NPWT Gauze<br>Dressing | Predicate Renasys<br>Gauze NPWT<br>Dressing Kit |
| | Dressing | | |
| | | Material | Same as predicate |
| | | Configuration | Identical to the large size of<br>the predicate |
| | | Drape | Currently marketed V.A.C.<br>Drape (polyurethane film<br>with acrylic adhesive) |
| | | Interface pad and<br>tubing | Currently marketed<br>V.A.C. SensaT.R.A.C.TM<br>Pad |
| | | Wound measuring<br>ruler | Same as predicate |
| | | Accessories: Saline,<br>skin prep, non-<br>adherent layer | Not provided |
| | NPWT Therapy Units<br>Compatibility | Compatible with the<br>following KCI V.A.C.<br>Negative Pressure Wound<br>Therapy Units:<br>ActiV.A.C.<br>InfoV.A.C.<br>V.A.C. ATS<br>V.A.C. Freedom<br>V.A.C. Simplicity<br>V.A.C.Ulta | Compatible with S&N<br>therapy units:<br>RENASYS EZ<br>RENASYS GO |
| Summary of non-<br>clinical tests<br>conducted for<br>determination of<br>substantial<br>equivalence | The KCI NPWT Gauze Dressing was evaluated under a number of design<br>verification and validation tests to assure performance requirements for<br>delivery of negative pressure wound therapy were met. | | |
| | ● The KCI NPWT Gauze Dressing when used with the V.A.C. Negative<br>Pressure Wound Therapy Units delivers negative pressure within<br>specification to the top of the wound site for a period of 72 hours.<br>● The KCI NPWT Gauze Dressing when used with the V.A.C. Negative<br>Pressure Wound Therapy Units removes exudate from the wound for<br>a period of 72 hours.<br>● The KCI NPWT Gauze Dressing when used with the V.A.C. Negative<br>Pressure Wound Therapy Units do not contribute to false blockage<br>alarms. | | |
| Summary of<br>clinical tests<br>conducted for | None required for determining substantial equivalence | | |
{1}------------------------------------------------
## [ [ [ [ ] [ ] ] ] therapy, intermittent therapy or over closed incisions.
{2}------------------------------------------------
| determination of<br>substantial<br>equivalence | |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusions<br>drawn | Testing demonstrates that the KCI NPWT Gauze Dressing and the<br>RENASYS™ Gauze NPWT Wound Dressing Kits with Softport are<br>substantially equivalent in terms of both indications for use and delivery of<br>negative pressure wound therapy. |
and the comments of the comments of the comments of the comments of
:
:
.
·
:
and the comments of the comments of the comments of the comments of the comments of the comments of the contraction of the contribution of the contribution of the contributio
and the comments of the comments of the comments of the comments of
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The logo is black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
April 4, 2013
KCI USA, Inc. % Ms. Anona Goebel Senior Manager, Regulatory Affairs 6203 Farinon Drive San Antonio, Texas 78249
Re: K123507
Trade/Device Name: KCI NPWT Gauze Dressing Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP, FRO Dated: March 04, 2013 Received: March 07, 2013
Dear Ms. Goebel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Anona Goebel
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, FOR
# Peter Di Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
510(k) Number (if known): K123507 Device Name: KCI NPWT Gauze Dressing
Indications for Use:
The KCI NPWT Gauze Dressing is intended to be used with the following KCI Therapy Units (ActiV.A.C.®, InfoV.A.C.®, V.A.C. Simplicity™, V.A.C. ® Freedom, V.A.C. ATS® and V.A.C.Ulta™ Therapy Systems). The system is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and by removing exudate and infectious material. Wound types include chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
The KCI NPWT Gauze Dressing is not intended for use with instillation therapy, intermittent therapy or over closed incisions.
Prescription Use Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page of of a
(Posted November 13, 2003)
Jiyoung Dang -S
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K123507
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.