VERIFY ASSERT STEAM Process Challenge Device for Gravity Cycles
Applicant
STERIS Corporation
Product Code
FRC · General Hospital
Decision Date
Apr 28, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The VERIFY™ ASSERT™ STEAM Process Challenge Device for Gravity Cycles is used for qualification, routine microbial monitoring, and cycle monitoring of steam sterilizers. The validated steam sterilization cycles include: - 250°F (121°C) 30-minute gravity - 270°F (132°C) 15-minute gravity The VERIFY Assert Self-Contained Indicator within the Process challenge device must be used with the VERIFY Incubator for Assert Self Contained Biological Indicators. When used in conjunction with the VERIFY Incubator for Assert Self Contained Biological Indicators, the VERIFY Assert Self-Contained Indicator within the Process challenge device provides a fluorescent result within 40 minutes.
Device Story
Device is a Process Challenge Device (PCD) for steam sterilization monitoring; contains a self-contained biological indicator (SCBI) and a chemical integrator sealed in a plastic tray with a foil cover. Tray features a molded channel creating a tortuous path to simulate air removal and steam penetration challenges equivalent to an ANSI/AAMI ST79 16-towel test pack. Used in clinical settings to monitor gravity steam sterilizers. The SCBI utilizes Geobacillus stearothermophilus spores; an enzyme produced by the organism reacts with a fluorogenic substrate in nutrient media to produce fluorescence. The device is placed in the sterilizer; after the cycle, the SCBI is removed and placed in a compatible incubator/reader. The reader detects fluorescence within 40 minutes, indicating sterilization efficacy. Provides healthcare providers with rapid confirmation of sterilization cycle success, ensuring patient safety by verifying that instruments are properly processed.
Clinical Evidence
Bench testing only. Simulated use testing compared the PCD performance against the AAMI reference pack and the predicate device. Results confirmed the PCD provides an equivalent or greater challenge to the sterilization process than the AAMI standardized test pack. Biological and chemical indicators within the pack were validated to perform as intended, with the chemical integrator not reaching its endpoint before the biological indicator is inactivated.
Technological Characteristics
Plastic tray with molded channel for steam/air resistance; contains SCBI and chemical integrator. SCBI uses Geobacillus stearothermophilus (>90% similarity to ATCC 7953) with 1.0-4.0 x 10^6 spores/SCBI. Mechanism: enzymatic reaction with fluorogenic substrate. Connectivity: requires external incubator/reader (K163587). Complies with ISO 11138-3 and ISO 11138-1.
Indications for Use
Indicated for qualification, routine microbial monitoring, and cycle monitoring of steam sterilizers in 250°F (121°C) 30-minute gravity and 270°F (132°C) 15-minute gravity cycles.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M Attest Super Rapid 5 Steam-Plus Challenge Pack (K121593)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 28, 2017
STERIS Corporation Mr. Anthony Piotrkowski Senior Manager, Regulatory Affairs 5960 Heisley Rd. Mentor, Ohio 44060
Re: K170070
Trade/Device Name: VERIFY™ Assert™ STEAM Process Challenge Device for Gravity Cycles Regulation Number: 21 CFR 880.2800 (b) Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: March 28, 2017 Received: March 29, 2017
Dear Mr. Anthony Piotrkowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K170070
Device Name
VERIFY Assert STEAM Process Challenge Device for Gravity Cycles
### Indications for Use (Describe)
The VERIFY™ ASSERT™ STEAM Process Challenge Device for Gravity Cycles is used for qualification, routine microbial monitoring, and cycle monitoring of steam sterilizers.
The validated steam sterilization cycles include:
- · 250°F (121°C) 30-minute gravity
- · 270°F (132°C) 15-minute gravity
The VERIFY Assert Self-Contained Indicator within the Process challenge device must be used with the VERIFY Incubator for Assert Self Contained Biological Indicators. When used in conjunction with the VERIFY Incubator for Assert Self Contained Biological Indicators, the VERIFY Assert Self-Contained Indicator within the Process challenge device provides a fluorescent result within 40 minutes.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the STERIS logo. The word "STERIS" is in large, bold, black letters at the top of the image, with a small registered trademark symbol to the right of the word. Below the word is a graphic of several horizontal blue lines that are slightly curved, resembling waves.
# 510(k) Summary For VERIFY™ Assert™ STEAM Process Challenge Device for Gravity Cycles
### Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
### Manufacturing Facility
STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896
Contact: Tony Piotrkowski
Telephone: (440) 392-7437 Fax No: e-mail: tony_piotrkowski.@steris.com
April 28, 2017 Submission Date:
Premarket Notification Number: K170070
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
### K170070/S001 STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION VERIFY" Assert " STEAM Process Challenge Device for Gravity Cycles
#### Device Name 1.
| Trade Name: | VERIFYTM AssertTM STEAM Process Challenge Device<br>for Gravity Cycles |
|------------------------|------------------------------------------------------------------------|
| Common/usual Name: | Biological Indicator Pack (PCD) |
| Device Classification: | Class II |
| Classification Name: | Indicator, Biological Sterilization Process<br>(21 CFR 880.2800, FRC) |
#### 2. Predicate Device
3M Attest Super Rapid 5 Steam-Plus Challenge Pack K121593
#### 3. Description of Device
The VERIFY"" Assert" |STEAM Process Challenge Device for Gravity Cycles (Gravity PCD), contains a VERIFY Assert Self-Contained Biological Indicator (SCBI) and a steam chemical integrator, sealed within a plastic tray with foil cover. The tray has a molded channel which serves as the tortuous pathway for air removal/steam penetration during steam sterilization. The PCD is designed and validated to be equivalent to the standard 16-towel test pack described in ANSI/AAMI ST79.
#### 4. Intended Use/ Indications for Use
The VERIFY™ ASSERT™ STEAM Process Challenge Device for Gravity Cycles is used for qualification, routine microbial monitoring, and cycle monitoring of steam sterilizers.
The validated steam sterilization cycles include:
- · 250°F (121°C) 30-minute gravity
- · 270°F (132°C) 15-minute gravity
The VERIFY Assert Self-Contained Indicator within the Process challenge device must be used with the VERIFY Incubator for Assert Self Contained Biological Indicators. When used in conjunction with the VERIFY Incubator for Assert Self Contained Biological Indicators, the VERIFY Assert Self-Contained Indicator within the Process challenge device provides a fluorescent result within 40 minutes.
{5}------------------------------------------------
# K170070/S001 STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION VERIFY™ Assert™ STEAM Process Challenge Device for Gravity Cycles
#### Summary of Technical Characteristics ട.
| Feature | Assert Gravity PCD | Attest 41482V | Comparison |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (proposed) | (K121593) Predicate | |
| Intended<br>Use | The VERIFY™<br>Assert™ STEAM<br>Process Challenge<br>Device (PCD) is used<br>for qualification, routine<br>microbial monitoring,<br>and load monitoring of<br>steam sterilizers.<br>270F (132°C), 15-<br>minute gravity;<br>250F (121°C), 30-<br>minute gravity<br>When used in<br>conjunction with the<br>VERIFY Incubator for<br>Assert Self Contained<br>Biological Indicators,<br>the VERIFY Assert<br>Self-Contained<br>Indicator within the<br>Process challenge<br>device provides a<br>fluorescent result within<br>40 minutes | Use 3M Attest Super<br>Rapid 5 Steam-Plus<br>Challenge Pack in<br>conjunction with the 3M<br>Attest Auto-reader 490<br>to qualify or monitor<br>dynamic air removal<br>(prevacuum) steam<br>sterilization cycles of 4<br>minutes at 270 F (132<br>C) and 3 minutes at 275<br>F (135 C).<br>The 3M Attest Super<br>Rapid Readout<br>Biological Indicator<br>contained within the<br>challenge pack provides<br>a final fluorescent result<br>in 1 hour. An optional<br>visual pH color change<br>result is observed in 48<br>hours. | Both are intended for<br>monitoring steam<br>sterilization cycles.<br>Testing to demonstrate<br>performance in the<br>claimed cycles has been<br>completed. |
| General<br>Design | Indicators are sealed in<br>plastic tray with channel<br>to limit steam<br>penetration/air removal. | Layers of medical index<br>cards, some of which<br>are die-cut to contain<br>indicators, overwrapped<br>and secured with a<br>label. | Both the devices are<br>contained in a steam<br>barrier. Testing to<br>demonstrate equivalence<br>has been completed. |
| Biological<br>Indicator | Verify Assert Self<br>Contained Biological<br>Indicator (40-minute<br>fluorescent result) | Attest 1492V Biological<br>Indicator (1-hour<br>fluorescent result) | Both contain 510(k)-<br>cleared Biological<br>Indicators with rapid<br>fluorescent results. |
| Chemical<br>Integrator | Non-commercial<br>Chemical Integrator | SteriGage Chemical<br>Integrator | Both contain chemical<br>integrators |
| Means to<br>distinguish<br>processed<br>PCD from<br>unprocessed | Proposed device's<br>internal integrator is<br>visible through the<br>pack. | External process<br>indicator. Yellow to<br>brown color change<br>when exposed to steam | Both have a means to<br>distinguish processed<br>PCD from unprocessed |
| Feature | Assert SCBI<br>(K162701) | Attest 1492V<br>(K121484) | Comparison |
| Indicator<br>organism | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus | Same criteria |
| Mechanism<br>of action | An enzyme, which is<br>produced by the organism,<br>reacts with a fluorogenic<br>substrate within the<br>defined nutrient media to<br>produce a fluorescent<br>moiety | An enzyme, which is<br>produced by the organism,<br>reacts with a fluorogenic<br>substrate within the<br>defined nutrient media to<br>produce a fluorescent<br>moiety | Same mechanism |
| Accessories | Automated incubator /<br>reader (K1635870)<br>• 40 minute incubation<br>time<br>• Holds temperature<br>range throughout<br>incubation time. | Automated incubator /<br>reader<br>• 60 minute incubation<br>time<br>• Holds temperature<br>range throughout<br>incubation time. | RIT testing performed with<br>the proposed<br>incubator/reader cleared<br>uner K163587. Inubator<br>readers meet criterion of<br>FDA BI 510(k) guidance. |
| Viable spore<br>population | 1.0 - 4.0 x 106 spore/SCBI | ≥ 1.0 x 106 spore/SCBI | Both proposed and predicate<br>meet criteria of ISO 11138-3<br>and FDA guidance |
| Resistance | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s | D132 ≥ 10 s<br>D135 ≥ 8 s | |
| Culture<br>Conditions | 55- 59 °C, media included<br>in SCBI, 40 minute<br>incubation time. | 55- 59 °C, media included<br>in SCBI, 60 minute<br>incubation time. | RIT Testing and ISO 11138<br>media testing verifies<br>performance |
| Primary<br>Packaging | Direct inoculum on plastic<br>vial, cap with recovery<br>media. | Inoculated paper in plastic<br>vial with cap and glass<br>ampoule with recovery<br>media in capped vial. | Similar configuration.<br>Component testing per ISO<br>11138-1 Annex B<br>demonstrates packaging is<br>compatible with indicator<br>and sterilization process |
### A comparison of technical characteristics of the PCD are in Table 3-1. Table 3-1 Summary of PCD Physical Description and Technological Properties
{6}------------------------------------------------
## K170070/S001 STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION VERIFY" Assert™ STEAM Process Challenge Device for Gravity Cycles
A comparison of technical characteristics BI components of the PCD are in Table 3-2.
### Table 5-2 Summary of SCBI Components of the PCD Physical Description and Technological Properties
Any differences between the predicate and proposed device as noted above are addressed through testing summarized below or submitted as part of the cleared 510(k): K162701 for the BI component and K163587 for the incubator.
#### 6. Summary of Nonclinical Tests
Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 3-3 below. Simulated use testing was done side-by-side with the predicate device and all other testing demonstrates the PCD meets the criteria of the FDA guidances for CI and BI as they pertain to PCD.
{7}------------------------------------------------
## K170070/S001 STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION VERIFY" Assert™ |STEAM Process Challenge Device for Gravity Cycles
| Test | Acceptance Criteria | Conclusion |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Simulated Use | Performance of the BI in the PCD is equivalent to the<br>performance of the BI in the AAMI reference pack in<br>their respective sterilization processes Performance of the chemical integrator in the PCD is<br>equivalent to the performance of the chemical integrator<br>in AAMI reference pack in their respective processes PCD provides an equivalent or greater challenge than<br>the AAMI standardized test pack | PASS |
| BI in pack vs BI<br>outside of pack | PCD provides a greater challenge to the process than<br>the BI itself. | PASS |
| CI in pack vs CI<br>outside of pack | PCD provides a greater challenge to the process than<br>the integrator by itself. | PASS |
| Chemical Integrator | Chemical integrator does not reach endpoint before BI<br>is inactivated. | PASS |
### Table 3-3. Summary of Non-clinical Testing vs Predicate
"PASS" indicates that acceptance criteria were met and that the proposed pack performed at least as effectively as the predicate device.
#### 7. Conclusion
The VERIFY Assert |STEAM Process Challenge Device for Gravity Cycles has met the established performance criteria. The results of the studies demonstrate that the biological indicator performs as intended, and based on the nonclinical tests performed, the subject device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate device, Class II (21 CFR 880.2800, Product code FRC).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.