K162701 · STERIS Corporation · OWP · Mar 24, 2017 · General Hospital
Device Facts
Record ID
K162701
Device Name
VERIFY Assert Self-Contained Biological Indicator
Applicant
STERIS Corporation
Product Code
OWP · General Hospital
Decision Date
Mar 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2805
Device Class
Class 2
Indications for Use
The VERIFY Assert Self-Contained Biological Indicator (SCBI) is for routine monitoring, qualification testing and product testing of the following steam sterilization processes. Cycle Type Temperature Time Dynamic Air Removal 270°F (132°C) 4 minutes Dynamic Air Removal 275°F (135°C) 3 minutes Gravity 250°F (121°C) 30 minutes Gravity 270°F (132°C) 15 minutes When used in conjunction with the Reader for the VERIFY Assert Self-Contained Biological Indicator, the VERIFY Assert Self-Contained Indicator provides a fluorescent result within 40 minutes.
Device Story
Device monitors steam sterilization efficacy; utilizes Geobacillus stearothermophilus spores and nutrient media. Reporter enzyme produced by organism reacts with fluorogenic substrate to generate fluorescent signal. Used in conjunction with proprietary reader/incubator; reader detects fluorescence within 40 minutes to confirm spore viability. Healthcare providers use output to verify sterilization cycle success; failure to detect fluorescence indicates sterilization process effectiveness. Benefits include rapid confirmation of sterilization, ensuring patient safety by preventing use of non-sterile instruments.
Clinical Evidence
Bench testing only. Performance validated via RIT testing (>97% alignment with 7-day conventional incubation), viable spore population (1.6-1.9 x 10^6), resistance testing (D121, D132, D135 values), survival/kill time studies, carrier growth inhibition, and simulated use testing in sterilizers under pass/fail conditions.
Technological Characteristics
Biological indicator containing Geobacillus stearothermophilus spores and nutrient media with fluorogenic substrate. Plastic vial primary packaging. Automated incubator/reader system. Complies with ISO 11138-3 and ISO 11140-1:2005 (Class 1 process indicator).
Indications for Use
Indicated for routine monitoring, qualification testing, and product testing of steam sterilization processes (dynamic air removal and gravity cycles) in healthcare settings.
Regulatory Classification
Identification
A Recombinant-DNA plasmid biological indicator is intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, possibly black or a dark shade of blue. The graphic is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 24, 2017
STERIS Corporation Mr. Anthony Piotrkowski Senior Manager, Regulatory Affairs 5960 Heisley Rd. Mentor, Ohio 44060
Re: K162701
Trade/Device Name: VERIFY Assert Self-Contained Biological Indicator Regulation Number: 21 CFR 880.2805 Regulation Name: Recombinant Biological Indicator Regulatory Class: Class II Product Code: OWP Dated: February 17, 2017 Received: February 21, 2017
Dear Mr. Anthony Piotrkowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162701
### Device Name
VERIFY Assert Self-Contained Biological Indicator
Indications for Use (Describe)
The VERIFY Assert Self-Contained Biological Indicator (SCBI) is for routine monitoring, qualification testing and product testing of the following steam sterilization processes.
| Cycle Type | Temperature | Time |
|---------------------|---------------|------------|
| Dynamic Air Removal | 270°F (132°C) | 4 minutes |
| Dynamic Air Removal | 275°F (135°C) | 3 minutes |
| Gravity | 250°F (121°C) | 30 minutes |
| Gravity | 270°F (132°C) | 15 minutes |
When used in conjunction with the Reader for the VERIFY Assert Self-Contained Biological Indicator, the VERIFY Assert Self-Contained Indicator provides a fluorescent result within 40 minutes.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for STERIS. The word "STERIS" is written in a bold, serif font at the top of the image. Below the word, there is a graphic of several horizontal, wavy lines in blue. The lines are stacked on top of each other, creating a visual effect of water waves.
## 510(k) Summary For K162701 VERIFY Assert Self-Contained Biological Indicator
### Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
### Manufacturing Facility
STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896
Tony Piotrkowski Contact: Senior Manager, Regulatory Affairs
Telephone:(440) 392-7437 Fax No: (440) 357-9198 e-mail: tpiotrko@steris.com
Submission Date: March 15, 2017
Premarket Notification Number: K162701
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
#### 1. Device Name
| Trade Name: | VERIFY Assert Self-Contained Biological Indicator |
|------------------------|---------------------------------------------------------|
| Common/usual Name: | Biological Indicator (BI, SCBI) |
| Device Classification: | Class II |
| Classification Name: | Recombinant Biological Indicator (21 CFR 880.2805, OWP) |
#### 2. Predicate Device
VERIFY CRONOS Self-Contained Biological Indicator K102469/DEN110006
### Reference Devices
3M Attest Super Rapid Readout Biological Indicator, Model 1492 K121484 Attest Rapid Readout Biological Indicator, Model 1292 K090569 Attest Rapid Readout Biological Indicator, Model 1291 K926364
#### 3. Description_of Device
The product is intended to monitor the critical parameters of steam sterilization cycles described in the indications for use by producing an optical change (signal) that is detected by the STERIS proprietary reader, Incubator for VERIFY Assert Self-Contained Biolgoical Indicator in 40 minutes to confirm the viability of the biological indicator at the end of a steam sterilization process. The product consists of a biological organism known to be resistant to steam (Geobacillus stearothermophilus) and a defined nutrient media. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.
#### 4. Intended Use/ Indications for Use
The VERIFY Assert Self-Contained Biological Indicator (SCBI) is for routine monitoring, qualification testing and product testing of the following steam sterilization processes.
| Cycle Type | Temperature | Time |
|---------------------|---------------|------------|
| Dynamic Air Removal | 270°F (132°C) | 4 minutes |
| Dynamic Air Removal | 275°F (135°C) | 3 minutes |
| Gravity | 250°F (121°C) | 30 minutes |
| Gravity | 270°F (132°C) | 15 minutes |
{5}------------------------------------------------
### K162701/S001 STERIS Response to 11/23/16 Request for Additional Information VERIFY Assert Self-Contained Biological Indicator
When used in conjunction with the Reader for the VERIFY Assert Self-Contained Biological Indicator, the VERIFY Assert Self-Contained Indicator provides a fluorescent result within 40 minutes.
#### 5. Summary of Technical Characteristics
A comparison of technical characteristics are summarized in Table 5-1.
| Feature | Assert SCBI | VERIFY Cronos SCBI | |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (proposed -K162701) | (predicate-K102469) | Comparison |
| Intended<br>Use | The VERIFY Assert Self-<br>Contained Biological<br>Indicator (SCBI) is for<br>routine monitoring,<br>qualification testing and<br>product testing of the<br>following steam<br>sterilization processes:<br>270F, 4-minute dynamic<br>air removal; 275F, 3-<br>minute dynamic air<br>removal; 250 F, 30 minute<br>gravity; 270, 15-minutes<br>gravity.<br>When used in conjunction<br>with the Reader for the<br>VERIFY Assert Self-<br>Contained Incubator, the<br>VERIFY Assert Self-<br>Contained Indicator<br>provides a fluorescent<br>result within 40 minutes. | For use by health care<br>providers for qualification<br>and routine monitoring of<br>steam sterilization<br>processes. The VERIFY<br>Cronos SCBI is only<br>validated to be used in the<br>following steam<br>sterilization cycles:<br>132 C vacuum-assisted<br>steam with 4-minute<br>sterilization time; 135 C<br>vacuum-assisted steam with<br>3-minute sterilization time;<br>121 C gravity steam with<br>30-minute sterilization<br>time; 132 C gravity steam<br>with 15-minute<br>sterilization time; 135 C<br>gravity steam with 10-<br>minute sterilization time;<br>132 C Steam Flush Press<br>ure Pulse with 4- minute<br>sterilization time. | Both are intended for<br>monitoring steam<br>sterilization cycles.<br>Use statement for the<br>proposed device has been<br>aligned with AAMI ST79,<br>Table 6 recommendations<br>for use of BI.<br>The Assert SCBI is claiming<br>different cycles compared to<br>the predicate (no 10-minute<br>135 gravity adding 135<br>SFPP). Simulated use<br>testing in sterilizers with a<br>load under pass and fail<br>conditions demonstrate<br>performance of the Assert<br>SCBI in these additional<br>cycles. |
| Indicator<br>organism | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus | Same |
| Mechanism<br>of action | An enzyme, which is<br>produced by the indicator<br>organism, reacts with a<br>fluorogenic substrate<br>within the defined nutrient<br>media to produce a<br>fluorescent moiety | An enzyme, which is<br>produced by the plasmid,<br>reacts with a fluorogenic<br>substrate within the defined<br>nutrient media to produce a<br>fluorescent moiety | Same mechanism of action,<br>testing provided<br>demonstrates conformance<br>with ISO 11138 and the<br>FDA guidance on BI<br>submissions |
| Accessories | Automated incubator /<br>reader | Automated incubator /<br>reader | RIT testing performed with<br>the proposed<br>incubator/reader. |
| Viable spore<br>population | 1.0 - 4.0 x 106 spore/SCBI | 1.0 - 5.0 x 106 spore/SCBI | Proposed population range<br>within that of the predicate<br>and both meet FDA<br>guidance and ISO 11138<br>requirements |
| Feature | Assert SCBI<br>(proposed -K162701) | VERIFY Cronos SCBI<br>(predicate-K102469) | Comparis on |
| Resistance | $D_{121} ≥ 1.5$ min<br>$D_{132} ≥ 10$ s<br>$D_{135} ≥ 8$ s | 121°C - 1.5 to 3.0 min<br>132°C - 10 to 30 sec<br>134°C - 8 sec to 20 sec<br>135°C - 8 sec to 20 sec | Both proposed and predicate<br>meet criteria of ISO 11138-3<br>and FDA guidance |
| Culture<br>Conditions | 55- 59 °C, media included<br>in SCBI, 40 minute<br>incubation time. | 55- 59 °C, media included<br>in SCBI, 120 minute<br>incubation time. | RIT Testing and ISO 11138<br>media testing verifies<br>performance |
| Primary<br>Packaging | Direct inoculum on plastic<br>vial, cap with recovery<br>media. | Inoculated paper in plastic<br>vial with cap and glass<br>ampoule with recovery<br>media in capped vial. | Similar configuration.<br>Component testing per ISO<br>11138-1 Annex B<br>demons trates packaging is<br>compatible with indicator<br>and sterilization process. |
| Process<br>indicator | STERIS STEAM Π<br>(K112256) | STERIS STEAM Π<br>(K112256) | Same indicator on both<br>products |
### Table 5-1 Summary of SCBI Physical Description and Technological Properties
{6}------------------------------------------------
### K162701/S001 STERIS Response to 11/23/16 Request for Additional Information VERIFY Assert Self-Contained Biological Indicator
#### 6. Summary of Nonclinical Tests
Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-2 below.
| Test | Acceptance Criteria | Conclusion |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Reduced<br>Incubation Time<br>(RIT) Testing | Meets FDA's requirement of > 97%<br>alignment of the 40-minute results with the<br>conventional incubation time of 7 days | PASS |
| Viable spore<br>population | 1.0 - 4.0 x 10 spore/SCBI | 1.6 - 1.9 x 106 spore/SCBI |
| Resistance | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s | D121 ≥ 2.29 min<br>D132 ≥ 49 s<br>D135 ≥ 40 s |
| Survival Time | Meets the longer of FDA and ISO 11138-3<br>requirements | 121 C ≥ 9.81 min<br>132 C ≥ 3.51 min<br>135 C ≥ 2.83 min |
| Kill Time | Meets the shorter of FDA and ISO 11138-3<br>requirements | 121 C ≤ 25.57 min<br>132 C ≤ 9.21 min<br>135 C ≤ 7.73 min |
| Carrier growth<br>inhibition / media<br>growth promotion | Positive growth of less than 100 spores after<br>primary packaging and media are subject to<br>worst case steam exposure | PASS |
| Hold Time | Performance not affected if incubated<br>within 72 hours of exposure to steam<br>sterilization | PASS |
| Simulated Use | Demonstrate growth when exposed to<br>abbreviated cycle and all kill in a full cycle | Abbreviated cycle -<br>growth<br>Full cycle - no growth |
### Table 5-2. Summary of Non-clinical Testing
{7}------------------------------------------------
### K162701/S001 STERIS Response to 11/23/16 Request for Additional Information VERIFY Assert Self-Contained Biological Indicator
| Test | Acceptance Criteria | Conclusion |
|-------------------|-----------------------------------------------------------------------------|----------------------------------------|
| Process indicator | Meets requirements for a “Class 1” process<br>indicator of ISO 11140-1:2005 | PASS<br>Testing provided in<br>K112256 |
#### 7. Conclusion
The VERIFY Assert Self-Contained Biological Indicator has met the established performance criteria. The results of the studies demonstrate that the biological indicator performs as intended, and based on the nonclinical tests performed, the subject device is substantially equivalent and is as safe and as effective as the legally marketed predicate device Class II (21 CFR 880.2805, Product code OWP).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.