K131455 · Restore Medical Solutions, Inc. · KCT · Aug 30, 2013 · General Hospital
Device Facts
Record ID
K131455
Device Name
RESTORE MODULAR STERILIZATION TRAY SYSTEM
Applicant
Restore Medical Solutions, Inc.
Product Code
KCT · General Hospital
Decision Date
Aug 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Restore™ Modular Sterilization Tray System is intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses when used in conjunction with a FDA cleared sterilization wrap. The Modular Sterilization Tray System may be used with common surgical instruments such as hammers, rasps, drivers, chisels, ringed-handled instruments, rongeurs, pickups, including but not limited to short-lumened devices etc. The validated total weight for tray and contents was 20 pounds.
Device Story
Restore™ Modular Sterilization Tray System is a reusable aluminum tray system designed for organizing and protecting surgical instruments during steam sterilization, transport, and storage. The system includes base trays (10x20x4 or 10x10x4 inches), lids, insert tray assemblies, silicone dividers, and an adjustable racking/stringer device. It features an evenly distributed perforated hole pattern to facilitate steam penetration and air removal. Used in healthcare facilities by sterile processing staff; instruments are placed in the tray, which is then wrapped in FDA-cleared sterilization wrap or placed in a rigid container. The system is processed via high-vacuum steam sterilization (270°F, 4-minute exposure). The tray ensures organized instrument delivery to the OR, maintaining sterility until use. Benefits include improved instrument organization, protection during handling, and validated sterilization efficacy for complex instrument sets.
Clinical Evidence
Bench testing only. Cleaning and sterilization validation performed under GLP conditions per AAMI TIR30:2011 and AAMI TIR12:2010. Validated parameters include high-vacuum steam sterilization (270°F, 4-minute exposure, 20-minute dry time) for a 20-pound load. Biocompatibility confirmed per ISO 10993.
Indicated for use in healthcare facilities to organize, enclose, sterilize, transport, and store surgical instruments (e.g., hammers, rasps, drivers, chisels, ringed-handled instruments, rongeurs, pickups, short-lumened devices) weighing up to 20 lbs. Must be used with FDA-cleared sterilization wrap.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
PolyVac Surgical Instrument Delivery System (K012105)
PolyVac Surgical Instrument Delivery System (K040223)
Submission Summary (Full Text)
{0}------------------------------------------------
# KIJ1455
Restore™ Modular Sterilization Tray System Traditional 510(k) — K131455
| | 510(k) SUMMARY | | | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
| Submitter Name: | Restore Medical Solutions, Inc. | | | | | | |
| Submitter Address: | 100 Peabody PL STE 150<br>P.O. Box 3118<br>Memphis, TN 38103 | | | | | | |
| Contact Person: | Shawn M. Flynn, President | | | | | | |
| Phone Number: | 404-805-8305 | | | | | | |
| Fax: | 636-680-3644 | | | | | | |
| Date Prepared: | May 20, 2013 | | | | | | |
| Device Trade Name: | Restore™ Modular Sterilization Tray System | | | | | | |
| Classification Name: | Sterilization Wrap Containers, Cassettes, and Accessories | | | | | | |
| Classification Class and Regulation Number: | Class II<br>21 CFR 880.6850 | | | | | | |
| Product Code: | KCT | | | | | | |
| Predicate Device(s): | K012105, PolyVac Inc., PolyVac Surgical Instrument Delivery System<br>K040223, Symmetry Medical Inc., PolyVac Surgical Instrument Delivery System | | | | | | |
| Statement of Intended Use: | The Restore™ Modular Sterilization Tray System is intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses when used in conjunction with a FDA cleared sterilization wrap.<br><br>The Modular Sterilization Tray System may be used with common surgical instruments such as hammers, rasps, drivers, chisels, ringed-handled instruments, rongeurs, pickups, including but not limited to short-lumened devices etc. The validated total weight for tray and contents was 20 pounds. | | | | | | |
| Reusable Medical Device Challenge Conditions | | | | | | | |
| | Minimum Inner Diameter Maximum Length Number of lumens 1.8 mm 4 inches 2 | | | | | | |
| High Vacuum (pre-vacuum, three pulse, standard):<br>Temp: 270° F | | | | | | | |
| Exposure Time: 4 Minutes<br>Cycle Dry Time (wrapped): 20 Minutes (minimum)<br>Cool Time: Varies according to load contents | | | | | | | |
| NOTE: Cool drafts from air ducts or other air currents should be avoided during the cooling phase to avoid post-sterilization moisture caused by rapid cooling syndrome. | | | | | | | |
{1}------------------------------------------------
| Device Description: | The Restore ™ Modular Sterilization Tray System is designed using<br>materials that can be reused with steam sterilization methods. The<br>Restore ™ system has an even distribution of holes in relation to its<br>size for optimal steam penetration.<br><br>The basic system includes either a 10 x 20 x 4 inch or 10 x 10 x 4 inch<br>aluminum tray with lids and a variety of insert tray assemblies and lids.<br>The central function is an adjustable racking and multi-functioning<br>stringer device, which is designed for the organization and alignment<br>of surgical instruments for pre-sterilization cleaning decontamination,<br>and sterilization cycles of surgical devices in healthcare facilities.<br><br>For the sterilization process, the tray system may be either wrapped in<br>FDA-cleared sterilization wrap designed for this purpose or placed in a<br>rigid container. |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Testing: | Cleaning and Sterilization Validation testing was performed by under<br>GLP conditions by a contract test laboratory according to AAMI<br>TIR30:2011 and AAMI TIR12:2010 standards. The testing validated<br>the parameters for vacuum settings, temperature, exposure time, cycle<br>dry time wrapped and in a sterilization container, and cool time. |
| Comparison to the<br>Predicate Devices: | The Restore ™ Modular Sterilization Tray System was compared to the<br>predicate devices by review of intended use and technological<br>characteristics - product design, device characteristics and materials.<br>Please refer to the Comparison Table for Substantial Equivalence on<br>pages 3 and 4 for the comparison parameters. |
| Substantial<br>Equivalence<br>Discussion: | The Restore tray system was found to be substantially equivalent<br>based on the same intended use and the side-by-side comparison of<br>the technological characteristics.<br><br>The additional Stringer feature offered by the Restore ™ Modular<br>Sterilization Tray System does not impact the intended use in<br>comparison to the predicate devices.<br><br>Further the validation testing of the loaded Restore tray system<br>assures the device is safe and effective for its intended use. |
| Verifications<br>regarding this 510(k)<br>Summary: | The summary includes only information that is also covered in the<br>body of the 510(k). The summary does not contain any puffery or<br>unsubstantiated labeling claims. The summary does not contain any<br>raw data, i.e., contains only summary data. The summary does not<br>contain any trade secret or confidential commercial information. The<br>summary does not contain any patient identification information |
| Substantial<br>Equivalence<br>Conclusion: | It is concluded that the Restore ™ Modular Sterilization Tray System is<br>substantially equivalent, based on the nonclinical testing (discussed<br>above) that demonstrates that the device is as safe, as effective, and<br>performs as well as or better than the predicate device. |
{2}------------------------------------------------
·
| | Proposed Device | Predicates | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Product Name | Restore™ Modular Sterilization Tray System | PolyVac Surgical Instrument Delivery System<br>K040223, K012105 | |
| 510(k) Number | | | |
| Product Code,<br>Regulation #,<br>Name | KCT<br>21 CFR 880.6850<br>Sterilization Wrap Containers, Trays, Cassettes & Other Accessories | KCT<br>21 CFR 880.6850<br>Sterilization Wrap Containers, Trays, Cassettes & Other Accessories | |
| Manufacturer | Restore Medical Solutions, Inc. | Symmetry Medical, Inc. (Previously PolyVac, Inc for K012105) | |
| Indications for<br>Use Statement: | The Restore™ Modular Sterilization Tray System is intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses when used in conjunction with a FDA cleared sterilization wrap.<br><br>The Modular Sterilization Tray System may be used with common surgical instruments such as hammers, rasps, drivers, chisels, ringed-handled instruments, rongeurs, pickups, including but not limited to short-lumened devices etc. The validated total weight for tray and contents was 20 pounds. | The PolyVac's delivery systems consist of perforated trays with lids, which are intended to enclose and protect medical device instrumentation, and to facilitate the sterilization process by allowing steam penetration and air removal. When used in conjunction with an approved sterilization wrap, sterility of the enclosed medical devices is maintained until used.<br><br>PolyVac's delivery systems are to be sterilized in one of the following cycles:<br>Prevacuum Steam: 132°C - 4 minutes minimum<br>Gravity Stream: 132°C - 30 minutes minimum<br>Gravity Stream: 121°C - 55 minutes minimum | |
| | | | |
| | | Sterilization Method: Ethylene Oxide (for K040223) | |
| | Reusable Medical Device Challenge Conditions | | |
| | Minimum Inner Diameter | Maximum Length | Number of lumens |
| | 1.8 mm | 4 inches | 2 |
| | High Vacuum (pre-vacuum, three pulse, | | |
| | standard):<br>Temp: 270° F | | |
| | Exposure Time: 4 Minutes | | |
| | Cycle Dry Time (wrapped): 20 Minutes (minimum)<br>Cool Time: Varies according to load contents | | |
| | NOTE: Cool drafts from air ducts or other air currents should be avoided during the cooling phase to avoid post-sterilization moisture caused by rapid cooling syndrome. | | |
| Device<br>Description:<br>Design,<br>Components,<br>Configurations,<br>Dimensions | Base trays (10 x 20 x 4 & 10 x 10 x 4 inches) Modular insert trays and lids. Evenly distributed perforated hole pattern. Silicone Dividers In-situ Racking Device/Stringer | Base tray (17.3 x 7.25 x 4 inches) Modular insert trays and lids. Evenly distributed perforated hole pattern. Silicone mats | |
| Material<br>Composition | Trays & Lids: Aluminum, 300 Series<br>stainless steel, biomedical grade<br>silicone<br>Stringer: Ultem*HU1004 Resin | Trays & Lids: Aluminum, 300 Series<br>stainless steel, biomedical grade<br>silicone; or Radal® R Plastic | |
| Sterilant<br>penetration | Each tray and lid contains evenly<br>distributed hole pattern in relation to its<br>size. | Each trays and lids contains evenly<br>distributed hole pattern in relation to its<br>size. | |
| Pre-Vac Steam<br>Sterilization<br>Validation<br>Parameters | High Vacuum (pre-vacuum, three pulse,<br>standard):<br>Temp: 270°F<br>Exposure Time: 4 Minutes<br>Cycle Dry Time: 15 Minutes (minimum)<br>Cool Time: Varies according to load<br>contents | Prevacuum Steam: 132°C - 4 minutes<br>minimum<br>Gravity Stream: 132°C - 30 minutes<br>minimum<br>Gravity Stream: 121°C - 55 minutes<br>minimum | |
| Reusable | Yes | Yes | |
| Material<br>Compatibility<br>with sterilization<br>process | Yes | Yes | |
| ISO 10993<br>Biocompatible | Yes | Yes | |
# Comparison Table for Substantial Equivalence
(table continues to next page)
{3}------------------------------------------------
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### August 30, 2013
Restore Medical Solutions, Incorporated C/O Ms. Patsy J. Trisler President Consultant Trisler Consulting 5600 Wisconsin Ave. #509 CHEVY CHASE Maryland 20815
Re: K131455
Trade/Device Name: Restore™ Modular Sterilization Tray System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: July 24, 2013 Received: July 26, 2013
## Dear Ms. Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For Kwame Ulmer
# Lester W. Schultheis Jr
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use Form
510(k) Number (if known): K131455
#### Device Name: Restore™ Modular Sterilization Tray System
Indications for Use:
The Restore™ Modular Sterilization Tray System is intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses when used in conjunction with a FDA cleared sterilization wrap.
The Modular Sterilization Tray System may be used with common surgical instruments such as hammers, rasps, drivers, chisels, ringed-handled instruments, rongeurs, pickups, including but not limited to short-lumened devices etc. The validated total weight for tray and contents was 20 pounds.
| Reusable Medical Device Challenge Conditions | | |
|----------------------------------------------|-------------------|---------------------|
| Minimum Inner<br>Diameter | Maximum<br>Length | Number of<br>lumens |
| 1.8 mm | 4 inches | 2 |
High Vacuum (pre-vacuum, three pulse, standard): Temp: 270° F Exposure Time: 4 Minutes Cycle Dry Time (wrapped): 20 Minutes (minimum) Cool Time: Varies according to load contents
NOTE: Cool drafts from alr ducts or other air currents should be avoided during the cooling phase to avoid post-sterilization moisture caused by rapid cooling syndrome.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Sreekanth Gutala -S Sreekanth Gutala -5
2013.08.29 14:43:28 -04'00' - 14:00 -
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: ≤1314555
Section 4.0
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.