K163402 · Nakanishi, Inc. · EBW · Aug 31, 2017 · Dental
Device Facts
Record ID
K163402
Device Name
NLZ Motor System
Applicant
Nakanishi, Inc.
Product Code
EBW · Dental
Decision Date
Aug 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
The NLZ Motor System is intended for use by dental professionals in the performance of dental restoration, prophylaxis and endodontic procedures. The NLZ Endo is intended for use by dental professionals in the performance of dental endodontic procedures.
Device Story
NLZ Motor System is an electric micromotor system for dental units; adds brushless electric micromotor with LED light to existing dental units. System comprises main unit, control unit, motor, and optional NLZ Endo contra-angle handpiece. Operated by dental professionals via foot pedal to regulate speed/rotation. Features include General Application Mode, Rotary Endo Mode, and Reciprocating Endo Mode (for WaveOne files). Safety features include Contra-SAFE (load-based motor stop), Contra-Check (manual friction/maintenance status check), and Contra-Restriction (speed control based on handpiece status). System provides feedback on handpiece status (OK, OIL, NG). Benefits include reduced risk of patient burns from overheated handpieces and precise torque/speed control for endodontic procedures.
Clinical Evidence
No clinical or animal testing performed. Evidence consists of bench testing, biocompatibility testing (cytotoxicity, sensitization, irritation per ISO 10993), electrical safety testing (AAMI/ANSI/ES60601-1), and electromagnetic compatibility testing (IEC 60601-1-2). Performance of overheating risk reduction features was verified via in-house bench testing using excessive motor loads.
Technological Characteristics
Electric micromotor system; brushless motor; LED light (>10,000 lx). Handpiece: stainless steel, 6:1 reduction, ISO 3964 coupling, ISO 14457 compliant. Connectivity: integrates with existing dental units. Sterilization: steam sterilization. Software-based safety functions: Contra-SAFE, Contra-Check, Contra-Restriction. Power: electric.
Indications for Use
Indicated for dental professionals performing dental restoration, prophylaxis, and endodontic procedures. No specific patient population age or gender restrictions provided.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
ELECTROtorque TLC 4893 with INTRAmatic KL 702 (K103027)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 31, 2017
Nakanishi Inc. % Izumi Maruo Senior Consultant MIC International Corp. 4-1-17 Hongo Bunkyo-ku, Tokyo 1130033 JAPAN
Re: K163402
Trade/Device Name: NLZ Motor System Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I Product Code: EBW Dated: July 21, 2017 Received: August 1, 2017
Dear Izumi Maruo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Andrew I. Steen -S
for Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163402
Device Name NLZ Motor System
Indications for Use (Describe)
The NLZ Motor System is intended for use by dental professionals in the performance of dental restoration, prophylaxis and endodontic procedures.
The NLZ Endo is intended for use by dental professionals in the performance of dental endodontic procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image features the logo of NSK, a company specializing in bearings and motion control technology. The logo consists of the letters 'NSK' in a bold, sans-serif typeface. The letters are white and are set against a solid blue background, creating a strong contrast that makes the logo easily recognizable.
# K163402
# 510(k) Summary
- a. Owner/Company name, address NAKANISHI INC. 700 Shimohinata, Kanuma-shi, Tochigi, 322-8666, Japan
# b. Contact
Kimihiko Satoh General Manager, Regulatory Affairs Department NAKANISHI INC. 700 Shimohinata, Kanuma-shi, Tochigi, 322-8666, Japan
| Phone: | +81-289-64-4325 |
|--------|-----------------------------|
| Fax: | +81-289-62-6665 |
| Email: | k-satoh@nsk-nakanishi.co.jp |
# c. Application Correspondent
Izumi Maruo Senior Consultant MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan
| Phone: | +81-3-3818-8577 |
|--------|------------------|
| Fax: | +81-3-3818-8573 |
| Email: | maruo@mici.co.jp |
# d. Date prepared
July 21, 2017
# e. Name of device
Trade Name: Classification Name: Regulation Name: Regulatory Class: Product Code:
NLZ Motor System Controller, Foot, Handpiece and Cord Dental handpiece and accessories Class I EBW
Image /page/3/Picture/16 description: The image shows a hand holding a blue object that resembles a pen or stylus. The hand is white and appears to be wearing a glove. The blue object is angled across the hand, and the words "Powerful Partners" are written in a small font below the hand. The logo is simple and modern, and the text suggests a business or professional partnership.
{4}------------------------------------------------
# f. Predicate devices and reference device
| Predicate device | |
|------------------|-----------------------------------|
| 510(k): | K090937 |
| Trade name: | E-STATIS 40 Electric Motor System |
| Company name: | SciCAN |
| Product code: | EBW |
The E-STATIS 40 Electric Motor System (K090937) does not include a handpiece. Therefore, the predicate device for the contra-angle handpiece, NLZ Endo is shown below.
| Predicate device for the NLZ Endo | |
|-----------------------------------|----------------------------------|
| 510(k): | K150750 |
| Trade name: | Contra-Angle Handpiece :Endo 6:1 |
| Company name: | Sirona Dental Systems GmbH |
| Product code: | EGS |
This submission uses following reference device for the reciprocating mode of the NLZ Motor System. The E-STATIS 40 Electric Motor System (K090937) is not equipped with that mode.
| Reference Device | |
|------------------|-----------------------------------------------|
| 510(k): | K103653 |
| Trade name: | e3TM Torque Control Motor |
| Company name: | DENTSPLY International Inc. |
| Product code: | EBW |
| 510(k): | K103027 |
| Trade name: | ELECTROtorque TLC 4893 with INTRAmatic KL 702 |
| Company name: | Kaltenbach & Voigt GmbH |
| Product code: | EBW, EKX |
# g. Description of the device
NLZ Motor Systems are available in two models;
NLZ Motor Set, which consists of the Main Unit, the Control Unit and the Motor.
NLZ E Motor Set, which consists of the Main Unit, the Control Unit, the Motor, and the Contra Angle Handpiece: NLZ Endo. The NLZ Endo is available only for NLZ E Motor Set and is not avialable for the NLZ Motor Set. The NLZ Endo is not equipped with any cooling system because it is a speed-reducing/low-speed handpiece.
The Control Unit drives an electric micromotor, and turns on off or regulates the speed of an electric motor by its foot pedal of the dental unit. This product can be connected to dental
{5}------------------------------------------------
units currently in use to add on a brushless electric micromotor with an LED light.
# h. Indications for Use
The NLZ Motor System is intended for use by dental professionals in the performance of dental restoration, prophylaxis and endodontic procedures.
The NLZ Endo is intended for use by dental professionals in the performance of dental endodontic procedures.
### Statement of substantial equivalence i.
The Indications for Use of the NLZ Motor System is identical to the E-STATIS 40 Electric Motor System (K090937).
The NLZ Motor System and the E-STATIS 40 Electric Motor System (K090937) have the similar functions and characteristics regarding the range of rotating speed, rotation direction, and auto reverse/auto stop.
The NLZ Motor System is equipped with similar reciprocating mode to the reference device.
Following table is comparison between the subject, the predicate and the reference devices.
{6}------------------------------------------------
| | Subject Device | Predicate Device | Reference Device | Reference Device |
|---------------------------|----------------------------------------------------------------|-------------------------------------------------|-------------------------------------------------|----------------------------------------------------------------------------------------|
| | NLZ Motor System | E-STATIS 40 Electric<br>Motor System (K090937) | ELECTROtorque TLC<br>4893 (K103027) | e3TM Torque Control Motor<br>(K103653) |
| Indications<br>for<br>Use | The NLZ Motor System is | The E-STATIS type 40 | The ELECTROtorgue TLC | The e3 Torque Control Motor |
| | intended for use by dental | dental system is intended | 4893 is intended to convert | is a medical device designed |
| | professionals<br>in<br>the | for use by dental | pneumatic output from a | for use by dentists for use |
| | performance<br>of<br>dental | professionals in the | dental treatment center to | with dental root canal |
| | restoration,<br>prophylaxis | performance of dental | electrical energy to drive | instruments in continuous |
| | and<br>endodontic | restoration, prophylaxis and | the INTRAmatic KL 702 | rotation with torque control or |
| | procedures. | endodontic procedures. | motor for operation of | in reciprocating movement. |
| | | | electrically driven | |
| | [Indications for Use for | | dental handpieces. These | |
| | NLZ Endo] | | devices are designed for | |
| | The NLZ Endo is intended | | use by a trained | |
| | for use<br>by dental | | professional in the field of | |
| | professionals<br>in the | | general dentistry. | |
| | performance<br>of | | | |
| | endodontic procedures. | | | |
| Drive | Electronic-micromotor | Electronic-micromotor | Electronic-micromotor | Electronic-micromotor |
| Components | Power supply, Motor<br>controller, Electric<br>micromotor, and | Power supply, Motor<br>controller, and Electric | Power supply, Motor<br>controller, and Electric | Power supply, Motor<br>controller, Foot pedal, Electric<br>micromotor and Contra-angle |
| | Contra-angle handpiece | micromotor | micromotor | handpiece (K972436) |
| | (NLZ Endo, NLZ E Motor | | | |
| | Set ONLY) | | | |
| Light | LED | Light bulb | LED | No light |
| - Light intensity | >10,000[lx] | Unknown | Unknown | NA |
Table 1. Comparison table for outline of the motor system
{7}------------------------------------------------
| Range of rotation speed (rpm) | 100 - 40,000 rpm | 100 - 40,000 rpm | 100 - 40,000 rpm | 250 - 1,000 rpm |
|--------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|---------------------------------------------|----------------------------------------------------------|
| Rotating direction | Forward and Reverse | Forward and Reverse | Forward and Reverse | Forward and Reverse |
| Standard operation mode | Yes | Yes | Yes | No |
| -Overheating risk reduction features | Yes* | No | Yes* | NA |
| Reciprocating mode | Yes (NLZ E Motor Set ONLY) | No | No | Yes |
| -Intended file | WaveOne | NA | NA | Wave One |
| -6:1 gear ratio handpiece designed for rotary and reciprocating modes is provided with the Motor | Yes | NA | NA | Yes |
| - Preset parameters for the reciprocating mode | Yes (User cannot alter the parameters) | NA | NA | Yes (User cannot alter the parameters) |
| Rotary Endo Mode | Yes (NLZ E Motor Set ONLY) | Yes | Yes | Yes |
| - Auto reverse/Auto Stop function | Yes | Yes | Yes | Yes |
| Motor<br>overheating<br>prevention<br>function | Yes | Unknown | Yes | Unknown |
| Maximum air/water pressure | | | | |
| -Coolant air<br>supply | 0.40 MPa | 0.40 MPa | 0.5 MPa | No coolant circuit |
| -Spray water<br>supply | 0.25 MPa | 0.25 MPa | 0.25 MPa | No spray circuit |
| -Spray air<br>supply | 0.25 MPa | 0.20 MPa | 0.20 MPa | No spray circuit |
| Sterilization | Sterilized by user<br>(steam sterilization) | Sterilized by user<br>(steam sterilization) | Sterilized by user<br>(steam sterilization) | The contra-angle handpiece<br>only is sterilized by user |
| Available<br>Handpiece type | E-type (ISO3964) | E-type (ISO3964) | ISO3964 conformed<br>attachment | E-type (ISO3964) |
{8}------------------------------------------------
* For detail comparison regarding overheating risk reduction features, please see Table 2.
{9}------------------------------------------------
Based on Table 1, the NLZ E Motor Set provides "Reciprocating Endo Mode" and "Rotary Endo Mode" in addition to standard operation mode (in the NLZ Motor System, named "General Application Mode"). The Reciprocating Endo Mode of the NLZ E Motor Set is designed for use with WaveOne file designed for unequal counterclockwise (150°) and clockwise (30°) rotation. In order to perform reciprocating endo treatment safely, we provide the NLZ Endo handpiece which was verified for use with the WaveOne and presetting parameters of this treatment for the NLZ Endo with the WaveOne on the NLZ E Motor Set. User cannot alter those presetting parameters. The reciprocating mode of the reference device, e3 M Torque Control Motor (K103653) has identical features to the Reciprocating Endo Mode, which means;
Both modes are designed for use with WaveOne file 1.
2. Both modes provide automatic unequal counterclockwise (150°) and clockwise (30°) rotation
3. Both modes provide a 6:1 handpiece which was verified for use with the WaveOne and presetting parameters for using the providing handpiece combined with the WaveOne file. 4. User cannot alter the presetting parameters
The Rotary Endo Mode of the NLZ E Motor Set provides the continuous one direction rotation for use with low-speed handpieces and three modes when the set torque reached. The values of rotation speed and torque can be adjusted. The features for "Endodontic Mode" of the E-STATIS 40 Electric Motor System (K090937) are identical to those for this mode.
Regarding standard operation mode, the provided rotation direction and range of rotation speed of the subject device are identical to those of the E-STATIS 40 Electric Motor System (K090937). Although the NLZ Motor System is equipped with the overheating risk reduction functions in order to reduce patient burns caused by overheated handpiece, the predicate device E-STATIS 40 Electric Motor System (K090937) is not equipped with such function. Therefore, such functions of the subject device are compared with a reference device, ELECTROtorque TLC 4893 (K103027).
{10}------------------------------------------------
| | Table 2. Comparison of overheating risk reduction functions | | |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Overheating<br>risk<br>reduction<br>function | NLZ Motor System (K163402/S001) | ELECTROtorque<br>4893 (K103027) | TLC |
| Detection of<br>defective<br>handpieces | Contra-SAFE | SAFEdrive | |
| - Detection<br>of<br>overheating | Indirectly by measuring load for the<br>handpiece and the duration of time | Indirectly by continuous<br>monitoring of the idling<br>properties of the<br>handpiece during its use. | |
| - Features | This function automatically stops motor<br>rotation or controls the speed when higher<br>and continued load of motor current are<br>detected. | If the protective function<br>is triggered, the<br>SAFEdrive initially<br>reduces the motor speed<br>and then stops the motor<br>altogether if the excessive<br>load persists | |
| Judge and<br>display of<br>the<br>handpiece<br>status | Contra-Check | The device does not have<br>this function | |
| - Features | 1) A manual feature for indirect detection<br>of excessive friction by measuring current<br>consumption of the Motor<br>2) Providing the status of handpiece<br>regarding potential overheating as follows;<br>● OK: The check result is acceptable<br>● OIL: Needs a maintenance such as<br>lubricating with oil<br>● NG: There is a serious<br>malfunction<br>● NG blinking: Damaged handpiece | NA | |
| Motor speed<br>control<br>function after<br>Contra-<br>Check | Contra-Restriction | NA | |
| -Features | This function restricts motor rotation when<br>handpiece is used for judge other than OK<br>through estimating temperature increase by<br>the Control Unit. The motor is slowed<br>down and stopped after speed control. | NA | |
Table 2. Comparison of overheating risk reduction functions
{11}------------------------------------------------
| ● If NG blinking is displayed, the | |
|------------------------------------|--|
| | |
| motor is stopped and cannot be | |
| activated. | |
| If NG is displayed, the motor is | |
| stopped approximately 10 seconds | |
| after continuous use. | |
| If Oil is displayed, the motor is | |
| stopped approximately 20 seconds | |
| after continuous use. | |
In addition, the subject device and the ELECTROtorque TLC 4893 (K103027) are equipped with the Motor overheat prevention function. The motor overheat prevention function indirectly detects overheat of the Motor itself in order to protect the Motor from failure of the coils. The purpose of design for this function is different from handpiece overheating risk reduction features described above.
This function of the subject device estimates temperature increase of the Motor by the motor current and rotation speed. The ELECTROtorque TLC 4893 (K103027) detects overload.
Regarding the handpiece, the comparison table is shown below;
| | NLZ Endo | Contra-Angle<br>Handpiece :Endo 6:1 (K150750 ) |
|------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Design features | Contra-Angle, Push-Button<br>Chuck, Motor coupling, Without<br>light and cooling system | Contra-Angle, Push-Button<br>Chuck, Motor coupling, Without<br>light and cooling system |
| Gear ratio | 6:1 Reduction | 6:1 Reduction |
| Maximum motor/drive<br>speed | 6,000 rpm | 40,000 rpm |
| Materials | Stainless Steel | Stainless Steel |
| Motor coupling | Compliant with ISO 3964 | Compliant with ISO 3964 |
| Source of power | Electronic-micromotor<br>NLZ E Motor Set Only | Electronic-micromotor |
| Dimension | Head Diameter: 8.6mm<br>Head Height: 11.3mm<br>Length: 81.9mm | Head Diameter: 8.7mm<br>Head Height: 11.4mm<br>Length: 92.3mm |
| Weight | 66 g | 70 g |
| Performance | Compliant with ISO 14457 | Compliant with ISO 14457 |
| -Eccentricity | 0.018 mm | Unknown |
| -Force for bur extraction | >45N | Unknown |
| Sterilization | Sterilized by user<br>(steam sterilization) | Sterilized by user<br>(steam sterilization) |
| Lubricant | PANA SPRAY Plus<br>(510(k) number is K131014) | Sirona T1 spray<br>(510(k)number is K113674) |
Table 3. Comparison table for the Contra-angle handpiece
{12}------------------------------------------------
| Bur | Contra-angle burs<br>ISO1797-1 Type1 | Contra-angle burs<br>ISO1797-1 Type1 |
|---------------------|--------------------------------------|--------------------------------------|
| -Maximum bur length | 46 mm | 25 mm |
Both the NLZ Endo and the Contra-Angle Handpiece :Endo 6:1 (K150750) are used for dental endodontic procedures. The main design features and gear ratio of the NLZ Endo are similar to those of the Contra-Angle Handpiece : Endo 6:1 (K150750). Although there are some differences between the NLZ Endo and the Contra-Angle Handpiece :Endo 6:1 (K150750), the performance characteristics of both devices are compliant with ISO 14457.
For the NLZ Motor System, performance testing including biocompatibility testing, Electric safety and EMC testing, bench testing was performed in order to demonstrate substantial equivalence to the predicates. Also, reprocessing validation was performed because both the Motor and the NLZ Endo are required cleaning and sterilization after use by user. Cleaning, sterilization, and drying time recommended in the instruction were validated using AAMI TIR 12:2010, AAMI TIR 30:2011, and FDA Reprocessing guidance. The NLZ Motor System met all requirements of the standards. Because the NLZ Motor System is equipped with the overheating risk reduction functions with which the E-STATIS 40 Electric Motor System (K090937) is not equipped, those functions are compared with the functions of the ELECTROtorque TLC 4893 (K103027). The Contra-SAFE function of the subject device has similar features to the SAFEdrive function of the ELECTROtorque TLC 4893 (K103027). The ELECTROtorque TLC 4893 (K103027) does not equipped the functions such as the Contra-Check and Contra-Restriction of the subject device. The performance testing for the Contra-Check and Contra-Restriction does not raise any new concern. Therefore, the NLZ Motor system demonstrated substantial equivalence to the predicates.
#### Biocompatibility j.
Following biocompatibility test was performed using the Motor and the NLZ Endo:
- Cytotoxicity in accordance with ANSI/AAMI/ISO 10993-5:2009
- Sensitization in accordance with ISO 10993-10:2010
- Irritation in accordance with ISO 10993-10:2010
In the biocompatibility testing reports, no biocompatibility concern was raised.
# k. Bench Testing
The bench tests were performed in order to verify conformity to ISO 14457:2012.
All samples were compliant with ISO 14457:2012.
Also, electromagnetic compatibility testing in accordance with IEC 60601-1-2:2014 and electrical safety testing in accordance with AAMI / ANSI/ES60601-1:2005/(R)2012 and A1:2012,, C1:2009/(R)2012 and A2:2010/(R)2012 were performed. Those test results indicate that the NLZ Motor System does not raise concern.
{13}------------------------------------------------
Performance testing for the overheating risk reductions was performed in accordance with in-house standard. This testing is verification test for software by putting excessive load to the Motor in order to confirm proper Contra-Check judge and motor speed control for Contra-Restriction and Contra-SAFE. The test results indicate that those safety functions do not raise any new concern. In addition, we analyzed temperature increase for returned handpieces due to overheating/burn event and confirmed that the overheating/burn event should be reduced by the overheating risk reduction features.
No animal or clinical testing was performed.
### l. Conclusion
The indications for use of the NLZ Motor System is identical to the E-STATIS 40 Electric Motor System (K090937). Although there are some differences in characteristics between the NLZ Motor System and the predicates, a number of performance testing indicated that the NLZ Motor System met the requirement of the standards. Based on above result, we conclude the NLZ Motor System is substantially equivalent to the predicates and does not raise any new safety and effectiveness concern.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.