The ELECTROtorque TLC 4893 is intended to convert pneumatic output from a dental treatment center to electrical energy to drive the INTRAmatic KL 702 motor for operation of electricallydriven dental handpieces. These devices are designed for use by a trained professional in the field of general dentistry.
Device Story
The ELECTROtorque TLC 4893 is a software-driven dental control unit that converts pneumatic output from a dental treatment center into electrical energy to power the INTRAmatic KL 702 motor. The system includes a base unit, motor hose, electrical motor, transformer, and power cord. It is operated by dental professionals in a clinical setting. The device features the 'SAFEdrive' software function, which monitors power consumption to prevent handpiece overheating and reduce the risk of patient burns. The user interacts with the system via a control panel to navigate software menus. The device controls motor start/stop, speed, and performance, and monitors power consumption to ensure safe operation. By preventing overheating, the device enhances patient safety during dental procedures.
Clinical Evidence
No clinical data was conducted. Substantial equivalence is supported by bench testing, including temperature and energy studies to validate the SAFEdrive function, software validation, electromagnetic compatibility testing, electrical safety testing, and biocompatibility/sterilization studies for the INTRAmatic KL 702 motor.
Technological Characteristics
System consists of a base unit, motor hose, electrical motor, transformer, and power cord. Motor conforms to ISO 11498 type 2; handpiece connection conforms to ISO 3964. Operates via pneumatic-to-electrical energy conversion. Software-driven control unit with SAFEdrive monitoring feature. Connectivity is integrated into dental treatment centers via 4-hole tubing.
Indications for Use
Indicated for use by trained dental professionals in general dentistry to convert pneumatic output from a dental treatment center into electrical energy to power the INTRAmatic KL 702 motor for operating electrically-driven dental handpieces.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
# Section III - 510(k) Summary of Safety and Effectiveness
Image /page/0/Picture/2 description: The image contains a sequence of alphanumeric characters, specifically "K103027". The characters are written in a clear, handwritten style, with each digit and letter distinctly formed. The sequence appears to be a code or identifier, possibly a serial number or a reference code.
Image /page/0/Picture/3 description: The image shows the logo for "sybron dental specialties". The text is in a bold, sans-serif font. A curved line is above the text, possibly representing a smile or dental arch.
Submitter:
Sybron Dental Specialties, Inc. 1717 West Collins Drive Orange, CA 92656 Claudia Ortiz - Contact Person Phone number: 714.516.7981 Fax number: 714.516.7488
Date Summary Prepared: May 31, 2011
### Device Name:
- Trade Name ELECTROtorque TLC 4893 with INTRAmatic KL 702 .
- Common Name Dental Handpieces and Accessories .
- Classification Name Dental Handpieces and Accessories, per 21 CFR § 872.4200 .
Devices for Which Substantial Equivalence is Claimed:
- . Bien Air, Optima MX (K042759)
- Sirona Dental Systems GmbH, Sirotorque L (K031584) .
- . Kaltenbach & Voigt GmbH, COMFORTtronic 4894 (K080677)
#### Device Description:
The ELECTROtorque TLC 4893 dental control unit is a stand-alone system for operating electrically-driven KaVo handpieces. External power supply provides electric power to the unit. The 4-hole tubing connected to the unit supplies chip /cooling air, water and pressure signal. The speed of the electric handpiece is controlled by air pressure. The control unit is positioned close a treatment unit at the location preferred by the dentist. The ELECTROtorque TLC 4893 system consists of a base unit with a motor hose, an electrical motor, a transformer, and a power cord. The ELECTROtorque TLC 4893 dental control unit is a software-driven device. The software controls the following features: (1) Motor control, (2) Motor start / stop behavior, (3) Motor speed, (4) Motor performance, (5) Measuring power consumption, (6) Monitoring power consumption, and (7) SAFEdrive. The new SAFEdrive software feature monitors the power consumption of electrical hand pieces to reduce the probability or severity of overheating, thus minimizing the risk of burns to the patient. The user can navigate through the software menu via the control panel (see Figure 2.0).
{1}------------------------------------------------
The INTRAmatic KL 702 motor is an electrical low-voltage motor for dental purposes according to ISO 11498 type 2. The motor connects onto the KaVo specific tubing of the dental treatment unit ELECTROtorque. The speed of the INTRAmatic KL 702 is controlled by air pressure of the dental treatment center. The converted pneumatic output signal (electrical energy) from a dental treatment center drives the motor to operate an electrically-driven dental handpiece. Electricallydriven dental handpiece, which conform to ISO 3964, can be attached on the motor. The INTRAmatic KL 702 is intended only for dental treatment by a dental professional.
#### Intended Use of the Device:
The ELECTROtorque TLC 4893 is intended to convert pneumatic output from a dental treatment center to electrical energy to drive the INTRAmatic KL 702 motor for operation of electricallydriven dental handpieces. These devices are designed for use by a trained professional in the field of general dentistry.
#### Substantial Equivalence:
The ELECTROtorque TLC 4893 dental control unit is substantially equivalent to other legally marketed devices in the United States. The ELECTROtorque TLC 4893 functions in a manner similar to and is intended for the same use as the COMFORTtronic 4894 marketed by Kaltenbach & Voigt, the Optima MX marketed by Bien Air, and to the Sirotorque L marketed by Sirona Dental Systems.
The ELECTROtorque TLC 4893 is similar to all three predicate devices in that it is a softwaredriven dental control unit consisting of a base unit with a motor hose, an electrical motor, a transformer, and a power cord. As the other three predicate devices, it is integrated in a dental treatment center and it uses the same water system and the same power supply. The software in the ELECTROtorque TLC 4893 device as well as the software of the three predicate devices control the following features: (1) Motor control, (2) Motor start / stop behavior, (3) Motor speed, (4) Motor performance, (5) Measuring power consumption, and (6) Monitoring power consumption. The ELECTROtorque TLC 4893 differs from all three predicate devices in that the ELECTROrorque TLC 4893 software has an added software function called SAFEdrive. This SAFEdrive function monitors the power consumption to reduce the probability or severity of overheating of electrical handpieces, thus minimizing the risk of burns to the patient.
The INTRAmatic KL 702 motor is substantially equivalent to other legally marketed devices in the United States. The INTRAmatic KL 702 functions in a manner similar to and is intended for the same use as the COMFORTdrive 200XDA marketed by Kaltenbach & Voigt, the Mikromotor MX marketed by Bien Air, and to the motor BL ISO marketed by Sirona Dental Systems. The INTRAmatic KL 702 is similar to all three predicate devices in that it is an electrical low-voltage motor for dental purposes according to ISO 11498 type 2. The INTRAmatic KL 702 differs only from the COMFORTdrive 200 XDA in that it can also rotated counterclockwise and it can be used with any electrically-driven dental handpiece that has a handpiece connection that conforms to ISO 3964.
#### Non-Clinical Test Data:
Temperature and energy studies have been conducted to determine the parameters for the new SAFEdrive software function. The ELECTROtorque TLC 4893 software has been successfully validated to confirm the performance of the device. The software validation included testing of
{2}------------------------------------------------
the new SAFEdrive software function to demonstrate that with this added function, the ELECTROtorque TLC 4893 is capable of detective electrical handpieces to reduce the probability or severity of overheating, thus minimizing the risk of burns to the patient. Testing of electromagnetic compatibility and electrical safety has been conducted in accordance with applicable recognized consensus standards. Also sterilization and biocompatibility studies were done to determine the safety and effectiveness of the INTRAmatic KL 702.
## Clinical Test Data:
Clinical testing has not been conducted on these products.
### Conclusion:
Based upon similar technological / performance characteristics as compared to the predicate devices, and successful validation of the ELECTROtorque TLC 4893 software, the clinical performance of the ELECTROtorque TLC 4893 and INTRAmatic KL 702 is deemed to be substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three curved lines that resemble an abstract representation of a human figure or a flowing ribbon. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Kaltenbach & Voight GmbH C/O Ms. Claudia Ortiz Compliance Director, Regulatory Affairs & Quality Assurance Sybron Dental Specialties, Incorporated 1717 West Collins Orange, California 92687
AUG 1 9 2011
Re: K103027
Trade/Device Name: ELECTROtorque TLC 4893 with INTRAmatic KL 702 Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EBW, EKX Dated: July 22, 2011 Received: July 25, 2011
Dear Ms. Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 - Ms. Ortiz
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that PDA \$ Issuance of a substantials with other requirements
mean that FDA has made a determination that your her hederal agencies mean that FDA has made a decemination had your administered by other Federal agencies.
of the Act or any Federal statutes and regulations and realigning to: registration of the Act or any Federal statues and reginations and limited to: registration
You must comply with all the Act's requirements, including reporting You must comply with an the Act 3 requirement of an 801); medical device reporting and listing (21 CFR Fall 807), fabeling (21 CFR 803); good manufacturing
(reporting of medical device-related adverse events) (21 CFR Part 87 (reporting of medical device-related ad reros oversions (QS) regulation (21 CFR Part 820);
practice requirements as set forth in the quality systems (Sections 531-542 of practice requirements as set form in the quality systems (2) 3 feetions 531-542 of the Act); 21 CFR 1000-1050. J
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), and if you desire specific advice for your device of control of Co
please go to http://www.fda.gov/AboutFDA/Centers/CDRH/CDRH/s) please go to http://www.ida.gov/Abouts British in the lth 's (CDRH's) Office of
/ucm l 1 5809.htm for the Center for Devices and Ration by reference to Compliance. Also, please note the regulation entitled, "Misbranding by reference to Compliance. Also, please note the regarations regarding the reporting of
premarket notification" (21CFR Part 807.97). For questions regarding of premarket nothication (21 CFR R rail 800 (21 CFR Part 803), please go to adverse events under the MDR regulation (21 Or NaProblem/default.htm for the CDRH's
http://www.fda.gov/MedicalDevices/Safety/Reportallers/2017/11/2019 http://www.ida.gov/Medical/Dometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may obtain other general miorniation your your Assistance at its toll-free Division of Sman Manataeta.com ,
number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Section I - Indications for Use
510(k) Number (if known): K10 3027
Device Name: ELECTROtorque TLC 4893 with INTRAmatic KL 702
Indications for Use:
The ELECTROtorque TLC 4893 is intended to convert pneumatic output from a dental treatment center to electrical energy to drive the INTRAmatic KL 702 motor for operation of electricallydriven dental handpieces. These devices are designed for use by a trained professional in the field of general dentistry.
Suse Runoes
(Division Sign-Off)
Tivision of Anesthesiology, General Hospital nfection Control, Dental Devices
:10(k) Number: K103027
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.