K163296 · Siemens Medical Solutions USA, Inc. · JAK · Mar 21, 2017 · Radiology
Device Facts
Record ID
K163296
Device Name
SOMATOM go.Up, SOMATOM go.Now
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Mar 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Device Story
SOMATOM go.Now and go.Up are whole-body CT scanners using fan-beam geometry and rotating tube-detector systems. Inputs: X-ray transmission data. Processing: syngo CT VA10A software performs patient/data management, scan control, image reconstruction, and archiving. Outputs: DICOM-format CT images. Used in clinical settings by trained staff. Features include Stellar detector technology, 32kW generator, and optional wired/wireless remote scan control (Scan&GO) via mobile tablet. Provides interfaces for advanced post-processing and plugin tools. Clinicians use reconstructed images for diagnosis and treatment planning. Benefits include workflow enhancements, iterative reconstruction (SAFIRE, iMAR) for image quality, and low-dose screening capabilities.
Clinical Evidence
Bench testing only. Performance supported by integration, functional, and phantom testing. Conformance to standards including IEC 60601-1, 60601-2-44, 60601-1-2, and NEMA XR-29. Lung cancer screening indication supported by literature (National Lung Screening Trial, N Engl J Med 2011).
Indicated for generation and processing of cross-sectional patient images via computer reconstruction of x-ray data; intended for use by trained physicians as diagnostic aid; indicated for low-dose lung cancer screening in high-risk populations as defined by professional medical societies.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2017
Siemens Medical Solutions USA, Inc. % Ms. Kimberly Mangum Regulatory Affairs Specialist 40 Liberty Blvd., Mail Code 65-1A MALVERN PA 19355
Re: K163296
Trade/Device Name: Somatom Go.up, Somatom Go.now Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: February 21, 2017 Received: February 23, 2017
Dear Ms. Mangum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163296
Device Name Somatom Go.up, Somatom Go.now
Indications for Use (Describe)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a large, sans-serif font. The color of the text is a light teal. The text is the only element in the image, and it is centered.
# 510(K) SUMMARY FOR SOMATOM GO. PLATFORM SCANNERS Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Date Prepared: January 26, 2017
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
### I. Submitter
Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number 2240869
#### Location of Manufacturing Site (1)
Siemens Healthcare GmbH Siemensstr. 1 D-91301 Forchheim, Germany Establishment Registration Number 3004977335
#### Location of Manufacturing Site (2)
SIEMENS SHANGHAI, MEDICAL EQUIPMENT LTD 278 Zhou Zhu Rd Shanghai, CHINA, 201318 Establishment Registration Number: 3003202425
### Contact Person:
Kimberly Mangum Requlatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Phone: (610) 448-6477 Fax: (610) 640-4481 Email: kimberly.mangum@siemens.com
### II. Device Name and Classification
| Product Name: | SOMATOM go.Now |
|--------------------------------|---------------------------------------------|
| Propriety Trade Name: | SOMATOM go.Now |
| Classification Name: | Computed Tomography X-Ray System |
| Secondary Classification Name: | Picture Archiving and Communications System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Secondary CFR Section: | 21 CFR 892.2050 |
| Device Class: | Class II |
| Primary Product Code: | JAK |
| Secondary Product Code: | LLZ |
{4}------------------------------------------------
lll.
| Product Name: | SOMATOM go.Up |
|--------------------------------|-------------------------------------------------------------------------------------------|
| Propriety Trade Name: | SOMATOM go.Up |
| Classification Name: | Computed Tomography X-Ray System |
| Secondary Classification Name: | Picture Archiving and Communications System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Secondary CFR Section: | 21 CFR 892.2050 |
| Device Class: | Class II |
| Primary Product Code: | JAK |
| Secondary Product Code: | LLZ |
| Predicate Device | |
| Primary Predicate Device: | |
| Trade Name: | SOMATOM Perspective |
| 510(k) Number: | K142955 |
| Clearance Date: | November 24, 2015 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | All applicable recalls are considered and addressed as part of the design control process |
| Secondary Predicate Devices: | |
| Trade Name: | SOMATOM Drive |
| 510(k) Number: | K161196 |
| 510(k) Number: | K161196 |
|-----------------------|-------------------------------------------------------------------------------------------|
| Clearance Date: | August 24, 2016 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR § 892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | All applicable recalls are considered and addressed as part of the design control process |
# Trade Name: SOMATOM Scope
| 510(k) Number: | K142955 |
|-----------------------|----------------------------------------------------------------------------------------------|
| Clearance Date: | November 24, 2015 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR § 892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | All applicable recalls are considered and addressed as part of the design<br>control process |
# IV. Device Description
The Siemens SOMATOM go. Platform is comprised of 2 Computed Tomography (CT) Scanner Systems, SOMATOM go.Now and SOMATOM go.Up. These CT scanners feature one continuously rotating tube-detector system and function according to the fan beam principle. The system software is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation. The SOMATOM go.Now and SOMATOM go.Up scanners produce CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens and other vendors as an aid in diagnosis and treatment preparation. The computer system delivered with the CT scanner is able to run optional post processing applications.
{5}------------------------------------------------
The software version for the SOMATOM go.Now and SOMATOM go.Up scanners, syngo CT VA10A (SOMARIS/10 syngo CT VA10A), is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation. The software platform SOMARIS/10 syngo CT VA10A is designed to provide a plugin interface to integrate potential advanced post processing tasks, tools, or extendable functionalities.
The SOMATOM go.Now and SOMATOM go.Up will support the following modifications in comparison to the predicate devices:
- New Marketing Name: SOMATOM go.Now or SOMATOM go.Up ●
- New/Modified Hardware ●
- Gantry Mechanics (modified) ●
- Generator (modified) ●
- Patient Table (modified)
- . Stellar Detector Technology (modified)
- Remote Scan Control (wired/wireless) (new) ●
- Control Box (modified)
- X-Ray Tube (modified)
- . Software version SOMARIS/10 syngo CT VA10A
- Basic Post Processing Viewer (modified) ●
- Acquisition Application (modified) ●
- Image Reconstruction (modified)
- Interface for Advanced Post Processing Application (new) .
- Interface for Plugin (for future advanced visualization tools and extended functionalities) ● (new)
- . Interface to support an optional mobile workflow control application software (new)
- . Update 510(k) Information
A comparison of these modifications with respect to the predicate devices is provided the "Comparison of Technological Characteristics with the Predicate Device" section below. As with the primary predicate device, the SOMATOM go.Now will be available in a 16 row 32 slice configuration. The SOMATOM go.Up will be available in a 32 row 64 slice configuration. Both the SOMATOM qo.Now and the SOMATOM go.Up will include a UFC stellar-technology based detector system, with the SOMATOM go.Now supporting an 11.2 mm detector coverage in comparison to the SOMATOM go.Up which supports a 22.4 mm detector coverage. A summary of the differences between the SOMATOM go.Now and the SOMATOM go.Up is provided as Table 1 below:
| Key specifications | go.Now | go.Up |
|--------------------------|----------------------------------|---------------|
| Physical Rows | 16 | 32 |
| Slice width (mm) | 0.7 | 0.7 |
| z coverage (mm) | 11.2 | 22.4 |
| Rotation time (s) | 0.8, 1.0, 1.5 | 0.8, 1.0, 1.5 |
| FoV (cm) | 50 | 50 |
| Max scan range (cm) | 125<br>160 | 160 |
| Bore size (cm) | 70 | 70 |
| Gantry tilt | NO | YES |
| Table Vertical Movement | Standard: No<br>Optional: YES | YES |
| Table Load (kg) | Table Load: 160<br>Optional: 227 | 227 |
| Power (kW) | 32 | 32 |
| XTA X-ray Tube | Chronon | Chronon |
| Tube heat capacity (MHU) | 3.5 | 3.5 |
# Table 1: Subject Device Comparable Properties
{6}------------------------------------------------
| Key specifications | go.Now | go.Up |
|----------------------------------|-----------------|--------------|
| Recon speed (images/s) | 9 (optional 13) | 13 |
| CaScoring | N/A | optional |
| Detector width | 11.2 mm | 22.4 mm |
| Iterative Reconstruction Methods | SAFIRE, iMAR | SAFIRE, iMAR |
| Spiral Scan | yes | yes |
# V. Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
# VI. Comparison of Technological Characteristics with the Predicate Device
The SOMATOM go.Now and SOMATOM go.Up scanners provide the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the predicate devices. The software and hardware components of these scanners have been modified or improved in comparison to the predicate devices to support enhanced device functionality compared to the predicate devices. The hardware components of the subject device have been modified to include a Remote Scan Control (RSC) and a Control Device (Scan&GO) user interface (e.g. user interface via mobile tablet software application), modified gantry mechanics and patient tables, and a Stellar Technology detector.
Software version SOMARIS/10 syngo CT VA10A supports software features that are designed to enhance the user workflow such as, extended functionalities and GO technologies which provide interfaces to directly access optional post processing applications. The intended use and fundamental scientific technology for the SOMATOM go.Now and SOMATOM go.Up remains unchanged from the predicate devices.
At a high level, the subject and predicate devices are based on the following same technological elements:
- Scanner Principle- Whole body X-Ray Computed Tomography Scanner ●
- System Acquisition Continuously rotating tube detector system ●
- Iterative Reconstruction Support of various iterative reconstruction methods ●
- Workplaces - Support of workplaces that include reconstruction and image evaluation software
- Patient table ●
- Patient table foot switch for movement ●
- Tin filtration technology ●
- o Stellar detector technology
- . High Power 80 (High mA @low kV)
The following technological differences exist between the subject device and predicate devices:
- Software version SOMARIS/10 syngo CT VA10A ●
- Chronon X-ray Tube ●
- 32 kW maximum power Generator ●
- . Modified gantry that supports the integration of the ICS and IRS computers
- Support of additional cybersecurity features ●
- Support of workflow improvement features GO technologies ●
- Support of optional wired/wireless remote scan control module .
{7}------------------------------------------------
- . Support of interfaces to access optional post-processing applications
A tabular summary of the differences between the predicate and subject devices is provided as Table 2 below:
| Property | Subject<br>Device<br>SOMATOM<br>go.Up,<br>SOMATOM<br>go.Now | Primary<br>Predicate Device<br>SOMATOM<br>Perspective<br>(K142955) | Secondary<br>Predicate Device<br>SOMATOM Drive<br>(K161196) | Secondary<br>Predicate<br>Device<br>SOMATOM Scope<br>(K142955) | |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------------|
| Type of CT<br>Scanner | whole body X-<br>ray computed<br>tomography<br>scanner | whole body X-ray<br>computed<br>tomography<br>scanner | whole body X-ray<br>computed<br>tomography scanner | whole body X-ray<br>computed<br>tomography scanner | |
| Sytem<br>Hardware | continuously<br>rotating tube<br>detector system | continuously<br>rotating tube<br>detector system | two continuously<br>rotating tube detector<br>systems | continuously rotating<br>tube detector system | |
| | high voltage<br>generator with<br>max power<br>32kW | high voltage<br>generator with max<br>power 55kW | high voltage<br>generator with max<br>power 2x100kW | high voltage<br>generator with max<br>power 26kW | |
| X-Ray Tube | Chronon | DURA 688/422 | STRATON MX<br>Sigma | DURA 352/202 | |
| kV Steps | 80kV,<br>110kV,130kV | 80kV, 110kV,130kV | 70 kV, 80kV, 90kV,<br>100kV,<br>110kV, 120kV,<br>130kV,<br>140kV | 80kV, 110kV,130kV | |
| Selective<br>Photon Shield | Tin Filter<br>Technology | N/A | Tin Filter Technology | N/A | |
| HMI & Gantry<br>Display | realized as<br>wireless tablet<br>mobile medical<br>application<br>software,<br>remote scan<br>control<br>(wired/wireless) | realized as buttons<br>on both side of the<br>gantry;<br>touch display panel<br>on gantry | realized as touch<br>screen and buttons<br>on both sides of the<br>gantry | realized as buttons<br>on both side of the<br>gantry; one LCD<br>display on gantry<br>front | |
| Operating<br>System | Windows based<br>SOMARIS/10<br>syngo CT<br>VA10A | Windows based<br>SOMARIS/5 syngo<br>CT VC40 | Windows based<br>SOMARIS/7 syngo<br>CT VA62A | Windows based<br>SOMARIS/5 syngo<br>CT VC40 | |
| Software | Basic Post<br>Processing<br>Viewer<br>CT View&GO | Basic Post<br>Processing<br>provided by :<br>syngo Viewing<br>syngo Filming | Basic Post<br>Processing provided<br>by:<br>syngo Viewing<br>syngo Filming | Basic Post<br>Processing provided<br>by :<br>syngo Viewing<br>syngo Filming | |
| Property | Subject<br>Device<br>SOMATOM<br>go.Up,<br>SOMATOM<br>go.Now | Primary<br>Predicate Device<br>SOMATOM<br>Perspective<br>(K142955) | Secondary<br>Predicate Device<br>SOMATOM Drive<br>(K161196) | Secondary<br>Predicate<br>Device<br>SOMATOM Scope<br>(K142955) | |
| Interface for<br>Advanced Post<br>Processing<br>Application | Interface for<br>Advanced Post<br>Processing<br>Application | Optional post<br>processing and<br>visualization<br>applications<br>accessed via a<br>supported<br>workstation | Optional post<br>processing and<br>visualization<br>applications<br>accessed via a<br>supported<br>workstation | Optional post<br>processing and<br>visualization<br>applications<br>accessed via a<br>supported<br>workstation | |
| Interface for<br>Plugin (for<br>future advanced<br>visualization<br>tools and<br>extended<br>functionalities) | Interface for<br>Plugin (for<br>future advanced<br>visualization<br>tools and<br>extended<br>functionalities) | Optional post<br>processing and<br>visualization<br>applications<br>accessed via a<br>supported<br>workstation | Optional post<br>processing and<br>visualization<br>applications<br>accessed via a<br>supported<br>workstation | Optional post<br>processing and<br>visualization<br>applications<br>accessed via a<br>supported<br>workstation | |
| Interface to<br>support an<br>optional mobile<br>workflow control<br>application<br>software | Interface to<br>support an<br>optional mobile<br>workflow control<br>application<br>software | Support of optional<br>i-control<br>interventional<br>module that can be<br>operated as a wired<br>or Bluetooth<br>connection based<br>wireless table and<br>gantry control<br>module | Support of optional i-<br>control interventional<br>module that can be<br>operated as a wired<br>or Bluetooth<br>connection based<br>wireless table and<br>gantry control<br>module | N/A | |
| Image<br>Reconstruction | Image<br>Reconstruction | Image<br>Reconstruction | Image<br>Reconstruction | Image<br>Reconstruction | |
| IT Hardening | IT Hardening | IT features<br>supported that<br>protect against<br>cybersecurity<br>attacks | IT Hardening | IT features<br>supported that<br>protect against<br>cybersecurity attacks | |
| Data Exchange<br>with external<br>SW client<br>(Teamplay) -<br>allows to copy<br>scan protocols<br>from other<br>systems | Data Exchange<br>with external<br>SW client<br>(Teamplay) -<br>allows to copy<br>scan protocols<br>from other<br>systems | support of<br>availability of scan<br>protocols form<br>other systems | Data Exchange with<br>external SW client<br>(Teamplay) - allows<br>to copy scan<br>protocols from other<br>systems | support of availability<br>of scan protocols<br>form other systems | |
| Iterative<br>Reconstruction<br>Methods | Iterative<br>Reconstruction<br>Methods | SAFIRE<br>iMAR | SAFIRE<br>iMAR | ADMIRE<br>SAFIRE<br>iMAR | SAFIRE<br>iMAR |
Table 2: Comparison of Technological Characteristics
{8}------------------------------------------------
Any differences in technological characteristics do not raise different questions of safety and effectiveness. Siemens believes that the subject device is substantially equivalent to the predicate devices. Testing and validation is completed. Test results show that the subject devices, the SOMATOM go.Now and SOMATOM go.Up, are comparable to the predicate devices
{9}------------------------------------------------
in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the predicate devices.
# VII. Performance Data
#### Non Clinical Testing
Non-clinical test (integration and functional) including phantom tests were conducted for the SOMATOM go.Now and SOMATOM go.Up during product development. The modifications described in this Premarket Notification were supported with verification testing. Siemens claims conformance to the following performance standards: ISO 14791, NEMA XR-29, IEC 61223-2-6, IEC 61223-3-5, IEC 62304, NÈMA XR-25, and DICOM 3.1-3.20.
Electrical Safety and Electromagnetic Compatibility (EMC) testing were conducted on the SOMATOM go.Now and SOMATOM go.Up in accordance with the following standards: IEC 60601-1, 60601-2-44, and 60601-1-2. Completed Form FDA 3654 are provided within this submission.
### Verification and Validation
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claim of substantial equivalence.
Siemens Healthcare conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document "Content of Premarket Submissions for Management of Cybersecurity Medical Devices issues on October 2, 2014" is included within this submission. Additionally, Siemens conforms to the requirements for Radio Frequency Wireless Technology as defined in FDA guidance document "Radio Frequency Wireless Technology in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, issued on August 14, 2013" by adhering to the EMC and risk based verification and validation requirements in design, testing, and labeling of the wireless remote control components of the subject devices.
The Radio Frequency Wireless Technology of Remote Scan Control and Control Device (Scan&GO) which are options for SOMATOM go.Now and SOMATOM go.Up CT Systems complies to 47 CFR part 15 subpart c – Intentional Radiators. All Radio device labels will show an FCC ID code to show the compliance.
### Additional Supportive Data
The National Lung Screening Trial (NLST), sponsored by the National Cancer Institute, is used to support the additional lung cancer screening Indications for Use. The study was a randomized trial of screening with the use of low-dose CT compared to chest radiography to determine whether screening with low-dose CT could reduce mortality from lung cancer. The study start date was August, 2002 and the completion date was October, 2010. The interpretation task with CT for this study was to detect lung nodules of 4mm diameter or greater.
### Summary
The features described in this premarket notification are supported with verification and validation testing, dosimetry and imaging performance, and analysis of phantom images to assess device and feature performance during product development. The risk analysis was completed and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria. Verification and validation testing of the device was found acceptable to support the claim of substantial equivalence.
{10}------------------------------------------------
### General Safety and Effectiveness Concerns
The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a system related risk analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing according to the Risk Management process. In order to minimize electrical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
### VIII. Conclusions
The predicate devices were cleared based on the results of non-clinical testing including verification and validation, phantom tests, and supportive literature. The subject device is also tested using the same methods as used for the predicate devices. The non-clinical data supports the safety of the device and the hardware and software verification and validation demonstrates that the SOMATOM go.Now and SOMATOM go.Up should perform as intended in the specified use conditions. The data included in this submission demonstrates that the SOMATOM go.Now and SOMATOM go.Up scanners perform comparably to the predicate devices currently marketed for the same intended use. Since both devices were tested using the same methods, Siemens believes that the data generated from the SOMATOM go.Now and SOMATOM go.Up testing supports a finding of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.