K161196 · Siemens Medical Solutions USA, Inc. · JAK · Aug 24, 2016 · Radiology
Device Facts
Record ID
K161196
Device Name
SOMATOM Drive
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Aug 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K161196 · Aug 24, 2016
SOMATOM Drive
Siemens Medical Solutions USA, Inc.
National Lung Screening Trial (NLST) clinical literature
The National Lung Screening Trial (NLST) is cited as clinical evidence to support the device's indication for use in low-dose lung cancer screening for high-risk populations.
Lung cancer screening; NLST; Clinical literature
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
National Lung Screening Trial (NLST); Randomized trial; Study Period: August 2002 - October 2010
High-risk populations for lung cancer
Chest radiography
Reduction in lung cancer mortality; detection of lung nodules of 4mm diameter or greater
Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of X-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations. * * As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature. for further information.
Device Story
SOMATOM Drive is a dual-source computed tomography (CT) system using fan beam geometry; features two continuously rotating STRATON MX Sigma X-ray tubes. System generates cross-sectional images via computer reconstruction of X-ray transmission data. Operated by trained physicians/technicians in clinical settings. Includes Dual Power mode (full detector width) and expanded kV steps (70-140 kV). Produces DICOM-format images for diagnostic aid. Software (syngo CT VA62A) manages patient data, scan control, and image reconstruction. Benefits include improved mA at lower kV levels and enhanced lung cancer screening capabilities. Healthcare providers use output images for clinical diagnosis and nodule detection.
Clinical Evidence
Bench testing only. Non-clinical verification and validation included dosimetry, image performance, and phantom testing. Software specifications met acceptance criteria. Clinical performance for Dual Source Dual Power mode was evaluated via clinical images. Lung cancer screening indication supported by literature, specifically the National Lung Screening Trial (NLST) (N Engl J Med 2011; 365:395-409).
Technological Characteristics
Dual-source CT; two STRATON MX Sigma X-ray tubes; high voltage generator (100kW). Features movable tin filters on both tubes. kV range: 70-140 kV. Software: Windows-based SOMARIS/7 VA62A. Connectivity: DICOM. Standards: IEC 60601-1, IEC 60601-2-44, XR-29, ISO 14971, IEC 62304, DICOM NEMA PS 3.1-3.20.
Indications for Use
Indicated for patients requiring cross-sectional imaging via computer reconstruction of X-ray transmission data; includes low-dose lung cancer screening for high-risk populations as defined by professional medical societies.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 24, 2016
Siemens Medical Solutions USA, Inc. Kimberly Mangum Regulatory Affairs Specialist 65 Valley Stream Parkway Malvern, Pennsylvania 19355
Re: K161196
Trade/Device Name: SOMATOM Drive Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: July 22, 2016 Received: July 25, 2016
Dear Kimberly Mangum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The color of the text is a light blue or teal. The letters are large and bold, filling most of the frame. The background is plain and white, providing a clean contrast to the text.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K161196
Device Name
SOMATOM Drive
Indications for Use (Describe)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of X-ray transmission data.
The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations. *
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature. for further information.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
{3}------------------------------------------------
# SIEMENS
#### 510(k) Summary: SOMATOM Drive
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355
Date Prepared: July 20, 2016
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### 1. General Information:
Importer / Distributor: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number: 2240869
#### Location of Manufacturing Site
Siemens Healthcare GmbH Siemensstrasse 1 D-91301 Forchheim. Germany Establishment Registration Number: 3004977335
#### Contact Person:
Kimberly Mangum Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Phone: (610) 448-6477 Fax: (610) 640-4481 Email: kimberly.mangum@siemens.com
#### 2. Device Name and Classification:
| Product Name: | SOMATOM Drive |
|-----------------------|----------------------------------|
| Trade Name: | SOMATOM Drive |
| Classification Name: | Computed Tomography X-Ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
#### 3. Legally Marketed Predicate Devices
SOMATOM Definition Flash Trade Name: 510(k) #: K143416 Clearance Date: April 16, 2015 Computed Tomography X-ray System Classification Name: Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Class II Device Class: Product Code: JAK
{4}------------------------------------------------
| Recall Information: | This predicate device has not been the subject of any design related recalls. |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open, SOMATON Emotion 6/16, SOMATOM Sensation 64/Sensation Cardiac , SOMATOM Perspective SOMATOM Scope/Scope Power |
| 510(k) #: | K142955 |
| Clearance Date: | November 24, 2015 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | This predicate device has not been the subject of any design related recalls |
| Reference Device Trade Name: | SOMATOM Force |
| 510(k) #: | K133589 |
| Clearance Date: | April 17, 2014 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | This reference device has not been the subject of any design related recalls |
#### 4. Device Description:
The Siemens SOMATOM Drive is a Computed Tomography X- ray System, which features two continuously rotating tube-detector systems and functions according to the fan beam principle. The SOMATOM Drive produces CT images in DICOM format, which can be used by post-processing applications commercially distributed by Siemens and other vendors. The computer system delivered with the CT scanner is able to run the post processing applications optionally. syngo CT VA62A (SOMARIS/7 VA62A) is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation available on the SOMATOM Drive CT system.
#### 5. Indication for Use:
This computed tomography system is intended to generate and process crosssectional images of patients by computer reconstruction of X-ray transmission data.
The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
{5}------------------------------------------------
# SIEMENS
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
## 6. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Devices:
The SOMATOM Drive provides the same technological characteristics in terms of materials, energy source and control mechanisms when compared to the predicate devices. The performance data of the X-ray tube and software features for the subject device SOMATOM Drive have been improved in comparison to the predicate device as described below:
## X-Ray Tube/kV Steps Improvements:
The SOMATOM Drive includes two STRATON MX Sigma X-ray tubes, which are a modification of the X-ray tube to the predicate device (SOMATOM Definition Flash). This modification enables higher mA at lower kV levels. 70 kV and 80 kV (High Power 70 and High Power 80) in comparison to predicate device (SOMATOM Definition Flash).
The STRATON MX Sigma also enables consistent 10 kV steps from 70 to 140 kV, which were introduced as part of the predicate device (SOMATOM Force) and adds 90 kV. 110 kV and 130 kV steps in comparison to the predicate device (SOMATOM Definition Flash).
#### Tube Collimator Improvement
The SOMATOM Drive includes the Tin Filter (selective photon shield) as cleared in the predicate device (SOMATOM Definition Flash) on the collimator of both X-ray tubes, whereas the predicate device (SOMATOM Definition Flash) only had this on one X-ray tube.
## Dual Power (full detector width)
The SOMATOM Drive allows the use of both tubes at the same kV with a routine pitch and full detector width, providing a doubling of the mA available in the examination (Dual Power Mode). The predicate device (SOMATOM Definition Flash) allows the use of both tubes at the same kV with a routine pitch only at half detector width.
#### Comparison of the technological characteristics
A tabular comparison of the technological characteristics between the subject device and predicate devices is provided as Table 1 below:
| Property | Subject Device:<br>SOMATOM Drive | Primary Predicate<br>Device: SOMATOM<br>Definition Flash<br>K143416 | Secondary Predicate<br>Device: SOMATOM<br>Force cleared as part of<br>K142955 and Reference<br>Device K133589 |
|----------------|---------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Generator | high voltage generator<br>with max power<br>100kW | high voltage generator<br>with max power 100kW | high voltage generator<br>with max power 120kW |
| User Interface | Touch based Human<br>Machine Interface | LCD Front gantry<br>display | LCD Front gantry display |
| Table 1: Subject Device Comparable Properties | |
|------------------------------------------------|--|
|------------------------------------------------|--|
{6}------------------------------------------------
| Property | Subject Device:<br>SOMATOM Drive | Primary Predicate<br>Device: SOMATOM<br>Definition Flash<br>K143416 | Secondary Predicate<br>Device: SOMATOM<br>Force cleared as part of<br>K142955 and Reference<br>Device K133589 |
|----------------------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Patient tables | standard patient table<br>(PHS4n)<br>OPTIONAL<br>Multipurpose Table<br>(MPT2n) | standard patient table<br>(PHS4)<br>OPTIONAL<br>Multipurpose Table<br>(MPT2) | standard patient table<br>(PHS5)<br>OPTIONAL Multipurpose<br>Table (MPT4) |
| x-ray tube | STRATON MX Sigma | STRATON MX P | Vectron |
| kV Steps | 70 kV, 80kV, 90kV,<br>100kV, 110kV, 120kV,<br>130kV, 140kV | 70kV, 80kV,<br>100kV,120kV, 140kV | 70kV, 80kV, 100kV,<br>120kV, 140kV |
| tube<br>collimator | including movable tin<br>filters for both tubes | including movable tin<br>filters only for tube B | including movable tin<br>filters for both tubes |
| Software | Windows based<br>SOMARIS/7 VA62A | Windows based<br>SOMARIS/7 VA48A | Windows based<br>SOMARIS/7 VA50 |
| Adjustable kV<br>Settings | support of additional<br>kV steps | Support of adjustable<br>kV settings | Support of adjustable kV<br>settings |
| Post-<br>processing<br>application<br>Calcium<br>Scoring | support for Calcium<br>Scoring with SN 100kV<br>tube power setting | Support of Calcium<br>Scoring Post-<br>Processing Application | Support of Calcium<br>Scoring Post-Processing<br>Application |
| iterative<br>reconstruction<br>methods | ADMIRE<br>iMAR | SAFIRE<br>ADMIRE<br>iMAR | SAFIRE<br>ADMIRE |
The intended use and fundamental scientific technology remain unchanged from the predicate devices. The Indication for Use and technological characteristics are similar for the subject and predicate devices. Any differences in technological characteristics between the subject device and predicate devices do not raise different questions of safety and effectiveness. As such, Siemens believes that the subject device is substantially equivalent to the predicate devices.
## 7. Nonclinical Testing
The features described in this premarket notification are supported with verification and validation testing. Non clinical tests, including dosimetry and image performance, were conducted for the SOMATOM Drive during product development. The risk analysis was completed and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria. Verification and validation testing of the device was found acceptable to support the claim of substantial equivalence.
Siemens claims conformance to the following standards: IEC 60601-1/A2: 2005; IEC 60601-2-44:2009 +A1:2012; XR-29: 2013; ISO/IEC 14971:2007; IEC 62304: Ed. 1.0 2006, IEC 61223-2-6: 2006, 61223-3-5: 2004, XR 25, ISO/IEC 10918-1: 1994, DICOM NEMA PS 3.1 - PS 3.20.
{7}------------------------------------------------
# SIEMENS
The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests (integration and functional). including phantom test were conducted during the SOMATOM Drive product development. The Risk analysis was completed, and risk control implemented, to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria.
#### Additional provided supporting data
The National Lung Screening Trial (NLST), sponsored by the National Cancer Institute, is used to support the additional lung cancer screening Indications for Use. The study was a randomized trial of screening with the use of low-dose CT compared to chest radiography to determine whether screening with low-dose CT could reduce mortality from lung cancer. The study start date was August. 2002 and the completion date was October. 2010. The interpretation task with CT for this study was to detect lung nodules of 4mm diameter or greater.
#### 8. Clinical Testing
Clinical images were evaluated to demonstrate performance for Dual Source Dual Power mode. Clinical experience phrased in peer reviewed articles support the performance and effectiveness of several features provided with SOMATOM Drive.
#### 9. General Safety and Effectiveness Concerns:
The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. In order to minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
## 10.Conclusion as to Substantial Equivalence:
The subject device non-clinical data similarly supports the safety of the software with verification and validation testing. Verification and validation testing demonstrates that the SOMATOM Drive performs as intended. The non-clinical test data demonstrates that the SOMATOM Drive performance is comparable to the predicate devices that are currently marketed for the same intended use.
For SOMATOM Drive, Siemens used the same testing with the same workflows as was used to clear the predicate devices. Since both devices were tested using the same methods. Siemens believes that the data generated from the SOMATOM Drive testing supports a finding of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.