Coldplay Cryoballoon; Focal Ablation System, Full Ablation System; Swipe Ablation System
K152329 · C2 Therapeutics, Inc. · GEH · Apr 13, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152329
Device Name
Coldplay Cryoballoon; Focal Ablation System, Full Ablation System; Swipe Ablation System
Applicant
C2 Therapeutics, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Apr 13, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K152329 · Apr 13, 2016
Coldplay Cryoballoon; Focal Ablation System, Full Ablation System; Swipe Ablation System
C2 Therapeutics, Inc.
Published clinical literature
The sponsor utilized published clinical literature data to support the safety and effectiveness of the device for the ablation of Barrett's Esophagus with high grade dysplasia, supplementing the prospective clinical study data.
Barrett's Esophagus; Ablation; Literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published literature data; Literature review
Patients with Barrett's Esophagus
Not applicable for this study
Safety and effectiveness for ablation of Barrett's Esophagus with high grade dysplasia
Indications for Use
The Coldplay CryoBalloon™ Focal Ablation System, Coldplay CryoBalloon™ Full Ablation System, and Coldplay CryoBalloon™ Swipe Ablation System are intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett's Esophagus with high grade dysplasia.
Device Story
Cryosurgical system for endoscopic tissue ablation; utilizes nitrous oxide-cooled balloon. System components: sterile catheter, non-sterile handle, non-sterile cartridge. Compatible with endoscopes (min 3.7 mm working channel). Physician inserts catheter through endoscope; inflates balloon with nitrous oxide to contact tissue; visualizes treatment site through balloon; adjusts handle to select site. Nitrous oxide spray cools balloon to 360°, 90°, or 45° circumference; exhaust vented through handle. Used in clinical settings for Barrett's Esophagus treatment. Output is physical tissue ablation; enables physician to remove dysplastic tissue; provides therapeutic benefit by converting Barrett's epithelium to neosquamous epithelium.
Clinical Evidence
Prospective, multicenter clinical study of 37 patients with Barrett's Esophagus (56 focal ablations). Ablation durations: 6s (n=10), 8s (n=28), 10s (n=18). Primary endpoint: conversion of Barrett's epithelium to neosquamous epithelium. Results: 60% (6s), 82% (8s), and 100% (10s) conversion rates (P=0.04). No serious adverse events reported. Post-procedure pain reported in 27% (immediate) and 14% (2 days). Data supports safety and effectiveness for ablation of Barrett's Esophagus with high grade dysplasia.
Technological Characteristics
Cryosurgical unit; nitrous oxide-cooled compliant balloon. Components: catheter, handle, cartridge. Energy source: nitrous oxide gas. Connectivity: standalone. Software-controlled. No changes to materials or design from predicates.
Indications for Use
Indicated for patients with Barrett's Esophagus with high grade dysplasia requiring endoscopic cryosurgical ablation.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Predicate Devices
Coldplay CryoBalloon™ Ablation System (K131523, K151054)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. To the left of the caduceus is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2016
C2 Therapeutics, Inc. Ms. Theresa Brandner-Allen VP of Regulatory Affairs and Quality Assurance 303 Convention Way, Suite 1 Redwood City, California 94063
Re: K152329
Trade/Device Name: Coldplay Cryoballoon™ Focal Ablation System; Coldplay Cryoballoon™ Full Ablation System; Coldplay Cryoballoon™ Swipe Ablation System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryogenic unit and accessories Regulatory Class: Class II Product Code: GEH Dated: March 07, 2016 Received: March 08, 2016
Dear Ms. Brandner-Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152329
### Device Name
Coldplay CryoBalloon™ Focal Ablation System, Coldplay CryoBalloon™ Full Ablation System, Coldplay CryoBalloon™ Swipe Ablation System
### Indications for Use (Describe)
The Coldplay CryoBalloon™ Focal Ablation System, Coldplay CryoBalloon™ Full Ablation System, and Coldplay CryoBalloon™ Swipe Ablation System are intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett's Esophagus with high grade dysplasia.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K152329 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR §807.92.
#### SUBMITTER I.
| | Submitter Name: C2 Therapeutics, Inc. |
|-----------------|--------------------------------------------------------------------------|
| Address: | 303 Convention Way, Suite 1<br>Redwood City, CA 94063 |
| Phone Number: | 650-521-5921 |
| Fax Number: | 650-556-1145 |
| Contact Person: | Theresa Brandner-Allen<br>VP of Regulatory Affairs and Quality Assurance |
| Date Prepared: | April 12, 2016 |
#### II. DEVICE
| Name of Device: | Coldplay CryoBalloon™ Focal Ablation System |
|-----------------|---------------------------------------------|
| | Coldplay CryoBalloon™ Full Ablation System |
| | Coldplay CryoBalloon™ Swipe Ablation System |
Common Name: Cryosurgical Unit, Cryogenic Surgical Device
Classification Name: Cryosurgical Unit, Cryogenic Surgical Device 21 CFR§878.4350(a)(2)
Regulatory Class: Class II
Product Code: GEH
#### III. PREDICATE DEVICES
Coldplay CryoBalloon™ Ablation System: K131523, K151054 CSA Medical truFreeze: K143625 Barrx HALO: K083737
CSA Medical has reported two design-related recalls for the truFreeze device. Barrx Medical has reported three design-related recalls for the HALO device. The
{4}------------------------------------------------
Coldplay CryoBalloon™ Ablation System predicate devices have not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The subject device is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with commercially available endoscopes with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The subject device is a system comprised of three components including a Catheter (sterile), Handle (nonsterile), and Cartridge (non-sterile).
The subject device is used to ablate unwanted tissue by application of extreme cold to a 360° circumference, 90° circumference, or 45° circumference and is identical to the respective Coldplay CryoBalloon™ Ablation System predicates. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the pre-inflated balloon, and the treatment site is selected by adjusting the endoscope and Handle. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Handle.
### INDICATIONS FOR USE V.
The Coldplay CryoBalloon™ Focal Ablation System, Coldplay CryoBalloon™ Full Ablation System, and Coldplay CryoBalloon™ Swipe Ablation System are intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett's Esophagus with high grade dysplasia.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Cryoablation is the fundamental technological principle for both the subject Coldplay CryoBalloon™ Focal Ablation System. Coldplay CryoBalloon™ Full Ablation System, and Coldplay CryoBalloon™ Swipe Ablation System and the predicate Coldplay CryoBalloon™ Ablation Systems and the predicate CSA Medical truFreeze. Both the subject device and all predicate devices are used as endoscopic instrumentation for ablating unwanted tissue.
The only modification to the subject device is to update the labeling to modify the indication for use to list treatment of Barrett's Esophagus and to include the treatment parameters, similar to the indications for use and treatment parameter type updates made by the predicate devices and to reflect physician use. The subject device has similar technological characteristics to the legally marketed predicates. The subject device and predicate devices are based on the following same technological elements:
{5}------------------------------------------------
- . Inserted through an endoscope to access the treatment site
- . Apply extreme temperature (cryogen or heat) to ablate (freeze or cauterize) the unwanted tissue
- Use of a compliant balloon to position the treatment diffuser .
- User-controlled activation to release and apply the therapy ●
- . Software controlled
There have been no technological changes to the subject device, and there are no technological differences between the subject device and predicate devices. The subject device remains substantially equivalent to the predicate devices.
Per Section III, design-related recalls for the CSA Medical truFreeze predicate' 2 and Barrx Medical HALO predicate3,4,5 have been reported. The Coldplay CryoBalloon™ Ablation System predicate devices have not been subject to a designrelated recall.
#### VII. PERFORMANCE DATA
Because there have been no technological changes to the subject device, no additional bench testing, preclinical testing, sterilization, and biocompatibility testing was performed.
A clinical study was performed to support the modification of the indications for use to include ablation of Barrett's Esophagus with high grade dysplasia. The study conducted was prospective, multicenter clinical data for 37 patients with Barrett's Esophagus. There were a total of 56 focal ablations performed in 37 patients. For the 56 ablations performed, the duration of the ablation was 6 seconds (n=10), 8 seconds
<sup>1</sup> Recall number Z-1513-2015 was conducted due to active and passive venting during the procedure. Because the Coldplay CryoBalloon Ablation System fully contains the cryogen during the procedure, active or passive venting is not required, and this design malfunction or potential patient injury is not relevant to the Coldplay CryoBalloon Ablation System. This recall has been terminated.
<sup>4</sup> Recall number Z-0941-2015 was conducted due to an increase in complaint trending for a loss of audible sound associated with the timer. The software with audible sounder used in the Coldplay CryoBalloon Ablation System has been validated, and this design malfunction has not been observed with the Coldplay CryoBalloon Ablation System. This recall has been terminated.
<sup>3</sup> Recall number Z-2745-2011 was conducted due device pivot mechanism's inability to return the electrode cap to a flat neutral position. Because the Coldplay CryoBalloon Ablation System's diffuser is fully contained within the balloon during the procedure, this design malfunction is not relevant to the Coldplay CryoBalloon Ablation System. This recall has been terminated.
<sup>4</sup> Recall number Z-0192-2009 was conducted because some units may contain the wrong filter, which does not have the proper lock and may result in a leak. The Coldplay CryoBalloon System does not have a filter, and this design malfunction is not relevant to the Coldplay CryoBalloon Ablation System. This recall has been terminated. 3 Recall number Z-2379-2012 was conducted due to a potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on prior to use on the patient. The software used in the Coldplay CryoBalloon
Ablation System has been validated, and this design malfunction has not been observed with the Coldplay CryoBalloon Ablation System. This recall has been terminated.
{6}------------------------------------------------
(n=28), or 10 seconds (n=18). No patients required any further interventions. Pain in the treatment area immediately post procedure was reported in 10 patients (27%) with median score of 2.5 (IQR 2-3). None of these patients required additional pain medication. At 2 days post-procedure, 5 patients (14%) reported pain in the treatment area with a median score of 4 (IQR 3-6) and pain when swallowing with median score of 4 (IQR 2-5). In the days following the ablation procedure, 3 patients (8%) used additional pain medication. There were no serious adverse events reported. At the time of follow-up endoscopy, full conversion of Barrett's epithelium to neosquamous epithelium was observed significantly more frequently in 6 (60%) of the 6second ablations, 23 (82%) of the 8-second ablations, and 18 (100%) of the 10second ablations (P=0.04). Conversion was confirmed through biopsy. During the follow-up period, no adverse events occurred. The clinical study results indicate, when used in accordance with the Instructions for Use, the subject device is effective at demonstrating complete response of high grade dysplasia at follow-up. The subject device prospective clinical study data and other published literature data indicate it is at least as safe and effective as the legally marketed predicates for ablation of Barrett's Esophagus with high grade dysplasia. The clinical performance data support the indication for use modification and the substantial equivalence determination.
#### VIII. CONCLUSION
The subject Coldplay CryoBalloon™ Focal Ablation System, Coldplay CryoBalloon™ Full Ablation System, and Coldplay CryoBalloon™ Swipe Ablation System are substantially equivalent to predicate devices. Use of the Coldplay CryoBalloon™ Focal Ablation System, Coldplay CryoBalloon™ Full Ablation System, and Coldplay CryoBalloon™ Swipe Ablation System for treatment of Barrett's Esophagus for ablation of high grade dysplasia is supported by a clinical study and performance data derived from literature.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.