K163162 · Jiangxi Hongda Medical Equipment Group , Ltd. · MEG · Mar 20, 2017 · General Hospital
Device Facts
Record ID
K163162
Device Name
Sterile Hypodermic Safety Syringes for Single Use
Applicant
Jiangxi Hongda Medical Equipment Group , Ltd.
Product Code
MEG · General Hospital
Decision Date
Mar 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile Hypodermic Safety Syringes for Single Use is intended to be used for medical purpose to inject fluids into or withdraw fluids from the body. After injection, the anti-needlestick feature is manually activated to aid in the prevention of accidental needle stick injuries.
Device Story
Sterile hypodermic safety syringe for single use; manual operation; used to inject or withdraw fluids. Features integrated anti-needlestick mechanism activated manually post-injection. Available in various needle gauges (21G-23G) and lengths (1"-1.5"); Luer Lock connector. Intended for clinical use by healthcare professionals. Benefits include reduction of accidental needle stick injuries during fluid administration or withdrawal.
Clinical Evidence
No clinical study included. Evidence consists of bench testing (physical, mechanical, chemical) per ISO 9626, ISO 7886, ISO 594, and ISO 7864, plus biocompatibility testing per ISO 10993. A simulated clinical study was performed per FDA guidance to evaluate the safety mechanism, demonstrating the device met pre-established criteria.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, hub, cap), Polyisoprene (piston), Stainless Steel SUS304 (needle tube). Manual operation; Luer Lock connector. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 9626, ISO 7886, ISO 594, ISO 7864, ISO 10993, ASTM F88, ASTM F1140.
Indications for Use
Indicated for medical use to inject fluids into or withdraw fluids from the body, incorporating a manual anti-needlestick feature to prevent accidental injuries.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Retractable Auto-Disable Syringe for single use, with/without needle (K072739)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 20, 2017
JiangXi HongDa Medical Equipment Group Ltd. c/o Ms. Diana Hong Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 CHINA
Re: K163162
Trade/Device Name: Sterile Hypodermic Safety Syringes for Single Use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG, FMI Dated: February 14, 2017 Received: February 15, 2017
Dear Ms. Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely, Tina Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163162
Device Name
Sterile Hypodermic Safety Syringes for Single Use
Indications for Use (Describe)
The Sterile Hypodermic Safety Syringes for Single Use is intended to be used for medical purpose to inject fluids into or withdraw fluids from the body. After injection, the anti-needlestick feature is manually activated to aid in the prevention of accidental needle stick injuries.
Type of Use (Select one or both, as applicable)
| <span></span> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------|
| <span></span> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K163162
- 1. Date of Preparation: 03/15/2017
- 2. Sponsor Identification
## JiangXi HongDa Medical Equipment Group Ltd.
No. 39 South Shengli Road, Jinxian County, Nanchang City, Jiangxi Province, 331700, China
Establishment Registration Number: Not yet registered
Contact Person: Yang Xiaohua Position: Vice-general Manager Tel: +86-791-85628918 Fax: +86-791-85628881 Email: yxh2821006@126.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
# Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Sterile Hypodermic Safety Syringes for Single Use Common Name: Piston Syringe with Needle
Regulatory Information Classification Name: Syringe Antistick Classification: II Product Code: MEG Subsequent Product Code: FMI Regulation Number: 21 CFR 880.5860 Review Panel: General Hospital
### Indication for Use:
The Sterile Hypodermic Safety Syringes for Single Use is intended to be used for medical purpose to inject fluids into or withdraw fluids from the body. After injection, the anti-needlestick feature is manually activated to aid in the prevention of accidental needle stick injuries.
# Device Description
The proposed devices are for single use only, which are comprised of syringes and needles with various specifications. The proposed devices are available in a variety combination of needle sizes and syringe volumes. All specifications of proposed devices subject to the same design. The proposed devices are intended for manual use only.
#### 5. Identification of Predicate Device
510(k) Number: K072739 Product Name: Retractable Auto-Disable Syringe for single use, with/without needle
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test provided in this submission include:
Physical, Mechanical and Chemical Tests performed on the proposed device
| Materials | Clause 3 of ISO 9626:1991/AMD-1:200 |
|----------------|-------------------------------------|
| Surface finish | Clause 4 of ISO 9626:1991/AMD-1:200 |
| Cleanliness | Clause 5 of ISO 9626:1991/AMD-1:200 |
{5}------------------------------------------------
| Limits for acidity and alkalinity | Clause 6 of ISO 9626:1991/AMD-1:2001 |
|-----------------------------------|---------------------------------------|
| Size designation | Clause 7 of ISO 9626:1991/AMD-1:2001 |
| Dimensions | Clause 8 of ISO 9626:1991/AMD-1:2001 |
| Stiffness | Clause 9 of ISO 9626:1991/AMD-1:2001 |
| Resistance to breakage | Clause 10 of ISO 9626:1991/AMD-1:2001 |
| Resistance to corrosion | Clause 11 of ISO 9626:1991/AMD-1:2001 |
### Cleanliness
- Limits for acidity or alkalinity Limits for extractable metals Size designation Colour coding Needle hub Sheath Needle tube Needle point Performance
### Cleanliness
| Limits for acidity or alkalinity | Clause 6 of ISO 7886:1993 |
|----------------------------------|----------------------------|
| Limits for extractable metals | Clause 7 of ISO 7886:1993 |
| Lubricant | Clause 8 of ISO 7886:1993 |
| Tolerance on graduated capacity | Clause 9 of ISO 7886:1993 |
| Graduated scale | Clause 10 of ISO 7886:1993 |
| Barrel | Clause 11 of ISO 7886:1993 |
| Piston/plunger assembly | Clause 12 of ISO 7886:1993 |
| Nozzle | Clause 13 of ISO 7886:1993 |
| Performance | Clause 14 of ISO 7886:1993 |
Dimension Gauging Liquid leakage Air leakage Separation force Stress cracking
Gauging Clause 4.1 of ISO 594-2:1998 Leakage Clause 4.2 of ISO 594-2:1998 Separation force Clause 4.3 of ISO 594-2:1998 Unscrewing torque Clause 4.4 of ISO 594-2:1998
Clause 4 of ISO 7864:1993 Clause 5 of ISO 7864:1993 Clause 6 of ISO 7864:1993 Clause 7 of ISO 7864:1993
Clause 8 of ISO 7864:1993 Clause 9 of ISO 7864:1993 Clause 10 of ISO 7864:1993 Clause 11 of ISO 7864:1993 Clause 12 of ISO 7864:1993 Clause 13 of ISO 7864:1993
Clause 5 of ISO 7886:1993
ISO 7886:1993 ISO 7886:1993 ISO 7886:1993 f ISO 7886:1993 f ISO 7886:1993 f ISO 7886:1993 f ISO 7886:1993 f ISO 7886:1993 Clause 3 of ISO 594-1:1986 Clause 4.1 of ISO 594-1:1986 Clause 4.2 of ISO 594-1:1986
Clause 4.3 of ISO 594-1:1986 Clause 4.4 of ISO 594-1:1986 Clause 4.5 of ISO 594-1:1986
{6}------------------------------------------------
| Ease of assembly | Clause 4.5 of ISO 594-2:1998 |
|--------------------------|------------------------------|
| Resistance to overriding | Clause 4.6 of ISO 594-2:1998 |
| Stress cracking | Clause 4.7 of ISO 594-2:1998 |
Sterile Barrier Packaging Testing performed on the proposed device:
| Seal strength | ASTM F88/F88-09 |
|-------------------|----------------------|
| Internal pressure | ASTM F1140/F1140M-13 |
Sterilization and Shelf Life Testing performed on the proposed device:
| EO residue | ISO 10993-7:2008 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ECH residue | ISO 10993-7:2008 |
| Sterility test | ISO 11737-2: 2009 |
| Bacteria Endotoxin Limit | USP 36-NF 31<85> |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package and Sterility<br>Tests were performed on real time aging samples to<br>verify the claimed shelf life of the device |
### Biocompatibility Testing
The patient-contact materials of Sterile Hypodermic Safety Syringes for Single Use are identified and biocompatibility testing is performed according to ISO 10993 standards
# Simulated Clinical Study
A simulated clinical study was performed on proposed device according to FDA Guidance, Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 to evaluate the safety mechanism of the proposed device. The results demonstrated that the proposed device met the pre-established criteria.
### Safety Feature Test
The safety feature test was performed on both proposed device and predicate device to determine its safety feature. The results demonstrated that the proposed device did not show a significant difference from predicate device.
{7}------------------------------------------------
#### 7. Clinical Test Conclusion
No clinical study is included in this submission.
#### Substantially Equivalent (SE) Comparison 8.
# Table 1 Comparison of Technology Characteristics
| Item | Proposed Device | Predicate Device<br>K072739 | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|----------------|
| Product Code | MEG | Same | |
| Regulation Number | 21 CFR 880.5860 | Same | |
| Class | II | Same | |
| Indication for Use | The Sterile Hypodermic Safety Syringes for<br>Single Use is intended to be used for medical<br>purpose to inject fluids into or withdraw fluids<br>from the body. After injection, the<br>anti-needlestick feature is manually activated to<br>aid in the prevention of accidental needle stick<br>injuries. | Similar | |
| Configuration<br>and material | Barrel | Polypropylene (PP) | Same |
| | Plunger | Polypropylene (PP) | |
| | Piston | Polyisoprene | |
| | Needle hub | Polypropylene (PP) | |
| | Needle cap | Polypropylene (PP) | |
| | Needle tube | Stainless Steel, SUS304 | |
| Operation Mode | For manual use only | | Same |
| Syringe Volume | 5ml | | 3ml, 5ml, 10ml |
| Connector Type | Luer Lock | | Same |
| Needle Gauge and<br>Length | Gauge:21G, 22G, 23G<br>Length: 1", 1 1/4", 1 1/2" | | Similar |
| Biocompatibility | Cytotoxicity | No cytotoxicity | Same |
| | Irritation | No intracutaneous reactivity | |
| | Sensitization | No skin sensitization | |
| | Systemic Toxicity | No systemic toxicity | |
| | Hemolysis | No Hemolysis | |
| | Pyrogen | No Pyrogen | |
| Sterilization | EO Sterilization | | Same |
| SAL | $10^{-6}$ | | Same |
| Single Use | Yes | Same | |
| Label/Labeling | Complied with 21 CFR part 801 | Same | |
{8}------------------------------------------------
The Sterile Hypodermic Safety Syringe for Single Use is similar to the predicate device in device design, Indications for use, materials, sterilization, method of operation and technological characteristics. The differences are in needle lengths/gauges and barrel sizes. The predicate device submission K072739 also includes insulin syringes and non-safety svringes but the subject device is claiming substantial equivalence to the predicates safety syringe. Through performance testing comparison the subject device and predicate device have demonstrated substantial equivalence.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the bench performance testing, comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.