K072739 · Shandong Weigao Group Medical Polymer Co., Ltd. · MEG · Mar 21, 2008 · General Hospital
Device Facts
Record ID
K072739
Device Name
JIERUI SYRINGES AND NEEDLES
Applicant
Shandong Weigao Group Medical Polymer Co., Ltd.
Product Code
MEG · General Hospital
Decision Date
Mar 21, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile Hypodermic Syringe for Single Use With/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body. The Retractable Auto-Disable Syringe for single use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body. Its secondary intended use is to retract inside the safety barrel, contaminated needle and aid in the prevention of accidental needle stick injuries. The sterile Insulin Syringe for single use with fixed needle is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin. The Sterile Hypodermic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
Device Story
Manual syringes and hypodermic needles for fluid injection or aspiration. Includes standard piston syringes, retractable auto-disable safety syringes, and insulin syringes with fixed needles. Operated by clinicians or patients for insulin delivery. Retractable models feature a safety mechanism to retract the needle into the barrel post-injection to prevent needle stick injuries. Devices are single-use, sterile, and disposable. Output is the delivery or withdrawal of fluids. Benefits include accurate dosing and reduced risk of accidental injury via safety features.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1:2003(E), including cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, and haemo-compatibility. All results met standard requirements.
Technological Characteristics
Materials: Polypropylene (PP) for syringes, stainless steel for needles. Luer slip and Luer lock nozzle configurations. Sizes: 0.5ml to 100ml syringes; 16G to 29G needles. Sterilization: Not specified. Energy source: Manual. Connectivity: None.
Indications for Use
Indicated for medical professionals to inject or withdraw fluids from the body, including insulin administration for diabetic patients. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
K072739
l of 5
# 510(k) Summary
## MAR 2 1 2008
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 9807.92.
The Assigned 510(k) Number is
## 1. Applicant Device Information
Trade/Proprietary Name: Jierui Syringes and Needle
#### Classification Information:
- a. Sterile Hypodermic Syringe for single use, with/without needle
- (1) Classification Name: Syringe, Piston
- (2) Regulation Number: 880.5860
- (3) Product Code: FMF
- (4) Class: II
- (5) Review Panel: General Hospital
b. Retractable Auto-Disable Syringe for single use, with/without needle
- (1) Classification Name: Syringe, Antistick
- (2) Regulation Number: 880.5860
- (3) Product Code: MEG
- (4) Class: II
- (5) Review Panel: General Hospital
c. Sterile Insulin Syringe for single use, with fixed needle
(1) Classification Name: Syringe, Piston
- (2) Regulation Number: 880.5860
- (3) Product Code: FMF
- (4) Class: II
- (5) Review Panel: General Hospital
d. Sterile Hypodermic Needle for single use
- (1) Classification Name: Needle, Hypodermic, Single Lumen
- (2) Regulation Number: 880.5570
- (3) Product Code: FMI
- (4) Class: II
- (5) Review Panel: General Hospital
{1}------------------------------------------------
K07739
2 of 5
## 2. Submitter Information
#### Manufacturer Name:
ShanDong WeiGao Group Medical Polymer Products Co., LTD No.312, Shichang Road Weihai, Shandong, China, 264209
#### Contact Person of the Submission:
Ms. Diana. Hong
Mr. Eric. Chen
Suite 8D, Zhongxin Zhongshan Mansion,
No. 19, Lane 999, Zhongshan No.2 Road(S)
Shanghai, China 20030
Phone: +86-21-64264467 x 152
Fax: +86-21-64264468 x 809
- Email: Diana.hong@mid-link.net Eric.chen@mid-link.net
## 3. Predicate Device
#### a. K number: K070936
Trade Name: Welmed Hypodermic Syringe (various sizes) Common Name: Syringes, Hypodermic Classification Name: Piston Syringe Product Code: FMF
#### b. K number: K071630
Trade Name: TERUMO 31G ThinPro Insulin Syringe Classification Name: Piston syringe with fixed hypodermic single lumen needle Product Code: FMF
#### c. K number: K053519
Trade Name: Safety Syringe Common Name: Syringe Classification Name: Syringe, Antistick Product Code: MEG
#### d. K number: K070440
Trade Name: BD Hypoint Common Name: Hypodermic Needle Classification Name: Single Lumen Hypodermic Needle Product Code: FMI
{2}------------------------------------------------
Premarket Notification 510(k) Submission—510(k) Summar
Report No.: A20071005
# Device Descriptio 4.
| Device Name | Intended Use | Nozzel | Volume | Material | Remark |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|---------------------------------------------------------|--------------------|---------------------------|
| Sterile Hypodermic<br>Syringe for single use | The Sterile Hypodermic Syringe for Single Use<br>With/without needle is intended to be used for medical<br>purposes to inject fluid into or withdraw fluid from<br>body. | Luer Slip | 1,2,3,5,10,20,30,50,100<br>(ml) | PP | With or Without<br>Needle |
| Sterile Insulin<br>Syringe for single use | The sterile Insulin Syringe for single use with needle is<br>a device intended for medical purposes for the manual<br>aspiration of insulin, and for the injection of insulin into<br>parts of the body below the surface skin. | Fixed | 0.5,1<br>(ml) | PP | With Fixed Needle |
| Retractable<br>Auto-Disable Syringe<br>for single use | The Retractable Auto-Disable Syringe for single use<br>with/without needle is intended to be used for medical<br>purposes to inject fluid into or withdraw fluid from<br>body. Its secondary intended use is to retract inside the<br>safety barrel, contain the contaminated needle and aid in<br>the prevention of accidental needle stick injuries. | Luer Lock | 3,5,10<br>(ml) | PP | With or Without<br>Needle |
| Sterile Hypodermic<br>Needle for single use | The Sterile Hypodermic Needle for single use is<br>intended for use with syringes and injection devices for<br>general purpose fluid injection/aspiration | Luer Slip<br>Luer Lock | 16G,18G,19G,20G,<br>21G,22G,23G,24G,<br>25G,26G,27G,29G | Stainless<br>Steel | |
K072739
3 of 5
3
{3}------------------------------------------------
| Premarket Notification 510(k) Submission—510(k) Summary<br>Report No.: A20071005 | 4 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|
| <b>5. Substantially Equivalence Determination</b> | |
| Comparison Analysis:<br>The Applicant device has the same classification information, intended use, sterilization specifications, performance, biocompatibility, chemical<br>specifications and similar physical and mechanical specifications with the predicate device. The only difference between applicant device and predicate<br>device is some physical specifications variant which is too slight to influence the effectiveness and safety. | |
| Conclusion:<br>The applicant device is Substantially Equivalent (SE) to the predicate device in terms of Effectiveness and Safety. | |
| Shanghai Midlink Business Consulting Co.,Ltd | |
{4}------------------------------------------------
K072733
5 of 5
## 6. Effectiveness and Safety Considerations
#### Effectiveness:
All the variant models of the applicant device are evaluated regarding the performance.
#### Safety Considerations:
With accordance with the Table 1 Initial Evaluation Tests for Consideration and Table 2 Supplementary Evaluation Tests for Consideration in ISO 10993-1:2003(E), Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing, the necessary tests for Biocompatibility Testing includes: Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Systemic Toxicity (Acute), Haemo-compatibility.
Conclusion: The all conducted Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices". The compatibility of all the possible skin-contact component material in the finished product meets the requirement of Biocompatibility
The applicant device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the applicant device is determined as safe and effectiveness,
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### MAR 2 1 2008
ShanDong WeiGao Group Medical Polymer Products Company, Limited C/O Ms. Diana Hong General Manager Shanghai Mid-Link Business Consulting Company, Limited Suite 8D, No. 19, Lane 999 Zhongshan No. 2 Road (S) Shanghai 200030 CHINA
Re: K072739
Trade/Device Name: Sterile Hypodermic Syringe for Single Use With/Without Needle Retractable Auto-Disable Syringe for Single Use With/Without Needle Sterile Insulin Syringe for Single Use With Fixed Needle Sterile Hypodermic Needle for Single Use
Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF, MEG, FMI Dated: March 11, 2008 Received: March 11, 2008
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{6}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Senate Y. Michael Yeds.
Chia-Lin Pi, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{7}------------------------------------------------
510(k) Number: ___ K072739 ___________________________________________________________________________________________________________________________________________________
Device Name: Sterile Hypodermic Syringe for Single Use With/without need le
Indications for Use:
The Sterile Hypodermic Syringe for Single Use With/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
Prescription Use _ √ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anestheslology, General Hospital Infection Control, Dental Devices
Page
510(k) Number: _ K472739
{8}------------------------------------------------
510(k) Number:
Device Name: Retractable Auto-Disable Syringe for single use With/without needle
Indications for Use:
The Retractable Auto-Disable Syringe for single use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body. Its secondary intended use is to retract inside the safety barrel, contaminated needle and aid in the prevention of accidental needle stick injuries.
Prescription Use _ √ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cinta Veres
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 2_of 4
510(k) Number: __ Kyn 2039
2
{9}------------------------------------------------
510(k) Number: K072739
Device Name: Sterile Insulin Syringe for single use with fixed needle
Indications for Use:
The sterile Insulin Syringe for single use with fixed needle is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use .. (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Camen ha
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 3_ of __ 4
510(k) Number: Kyn2739 -----
:
{10}------------------------------------------------
510(k) Number: K072739
Device Name: Sterile Hypodermic Needle for single use
Indications for Use:
The Sterile Hypodernic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton vr. my
Page _4_ of __ 4
4
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: __ K472734
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.