K140582 · Shenzhen Jumper Medical Equipment Co., Ltd. · DQA · Aug 28, 2014 · Cardiovascular
Device Facts
Record ID
K140582
Device Name
FINGERTIP PULSE OXIMETER
Applicant
Shenzhen Jumper Medical Equipment Co., Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Aug 28, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The JPD-500A Fingertip Pulse Oximeter is non-invasive device intended for spotchecking of functional oxygen saturation of arterial hemoglobin (Spo2) and pulse rate. The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).
Device Story
Portable fingertip pulse oximeter; uses photoelectric oxyhemoglobin inspection technology; emits two light wavelengths (660nm red, 905nm infrared) through finger nail tip; photosensitive element detects transmitted light; microprocessor calculates SpO2 and pulse rate based on Lambert-Beer Law and capacity pulse scanning; output displayed on device screen; used by clinicians or patients for spot-checking; aids in assessing patient oxygenation status; battery-powered.
Clinical Evidence
Controlled desaturation study conducted on 10 healthy, non-smoking, light-to-dark-skinned subjects. SpO2 accuracy compared against CO-oximeter blood samples over 70-100% range. Accuracy calculated using root-mean-squared (Arms) per ISO 80601-2-61; results showed error within specified limits. Bland-Altman analysis confirmed agreement within 95% limits. No adverse effects reported.
Technological Characteristics
Materials: Silica gel (patient contact). Sensing: Photoelectric (660nm/905nm). Power: 2x AAA batteries. Form factor: Portable fingertip clip. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-61, ISO 10993-1/5/10. Software: IEC 62304 compliant.
Indications for Use
Indicated for adult and pediatric patients requiring spot-check monitoring of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate in home and hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, with the head tilted upwards.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 28, 2014
Shenzen Jumper Medical Equipment Co., Ltd. C/O Field Fu, Consultant Shenzen Joyantech Consulting Co., Ltd 4th Floor, Jinhui Building Nanhai Blvd. Nanshan District, Shenzen Guangdong, China 518000
Re: K140582
Trade/Device Name: JPD-500A Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: July 21, 2014 Received: July 31, 2014
Dear Mr. Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Erin | Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Chapter 04 Indications for Use
510(k) Number (if known): K140582
Device Name: Fingertip Pulse Oximeter Model: JPD-500A
Indications for Use:
The JPD-500A Fingertip Pulse Oximeter is non-invasive device intended for spotchecking of functional oxygen saturation of arterial hemoglobin (Spo2) and pulse rate. The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).
Prescription Use _____________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Shenzhen Jumper Medical Equipment Co., Ltd. Sponsor: Fingertip Pulse Oximeter, model: JPD-500A Subject Device:
## Chapter 05 510(k) Summary
510(K) Summary as required by 21 CFR 807.92(c).
#### 1.0 Information of Submitter and Correspondent
#### Submitter's information:
Shenzhen Jumper Medical Equipment Co., Ltd.
Address: 5th Floor, Building No.34, Baoyuan Industrial Zone, Xixiang Street, Baoan District, Shenzhen 518102, P.R. China.
Phone: +86-755-2669 2192
Fax: +86-755-2685 2025
Contact Person: Yi Hui
E-mail: info@jumper-medical.com
#### Submission correspondent's information:
Shenzhen Joyantech Consulting Co., Ltd. Address: 4th Floor, Jinhui Building, Nanhai BLVD, Nanshan District, Shenzhen, Guangdong, China. Contact person: Mr. Field.Fu E-mail: cefda13485@163.com
#### 2.0 Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|--------------------------|
| Trade Name: | Fingertip Pulse Oximeter |
| Model: | JPD-500A |
| Classification name: | Oximeter |
| Review Panel: | Anesthesiology |
| Product Code: | DQA |
| Device Class: | II |
| Regulation Number: | 870.2700 |
#### 3.0 Predicate Device Information
| Sponsor: | Contec Medical Systems Co., Ltd. |
|----------------|----------------------------------|
| Device: | CMS-50D Finger Pulse Oximeter |
| 510(K) Number: | K082641 |
{4}------------------------------------------------
Shenzhen Jumper Medical Equipment Co.. Ltd. Sponsor: Subject Device: Fingertip Pulse Oximeter, model: JPD-500A
#### 4.0 Device Description
The JPD-500A fingertip pulse oximeter is intended for spot-checking of functional pulse oxygen saturation (SpO2) and pulse rage (PR) of single adult and pediatric patient in home and hospital.
The fingertip pulse oximeter features small size, low power consumption, convenient operation and portability. Power consumption of the product is low and two AAA batteries can be operated continuously for 24 hours. It is only necessary for a patient to put one of his/her fingers into the fingertip clips for measurement, very easy to use it.
Principle of the fingertip pulse oximeter as follows:
A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones.
Operation principle of the instrument:
Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm glow and 905nm near infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. Relevant data is shown on the Oximeter's display through electronic circuits and a microprocessor.
#### 5.0 Intended Use
The JPD-500A Fingertip Pulse Oximeter is intended for measuring the functional oxygen saturation and pulse rate through patient's finger. It is applicable for spot-checking SPO2 and pulse rate of adult and pediatric patients in homes and clinics.
#### 6.0 Indications for Use
The JPD-500A Fingertip Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (Spo2) and pulse rate. The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).
## 7.0 Summary comparing technological characteristics with predicate device
{5}------------------------------------------------
| Technological Characteristics | Comparison result |
|-------------------------------|-------------------|
| Design principle | Same |
| Appearance | Similar |
| Patients contact materials | Same |
| Performance | Similar |
| Biocompatibility | Same |
| Mechanical safety | Same |
| Energy source | Same |
| Electrical safety | Same |
| Standards met | Same |
| EMC | Same |
| Function | Similar |
Shenzhen Jumper Medical Equipment Co., Ltd. Sponsor: Subject Device: Fingertip Pulse Oximeter, model: JPD-500A
#### 8.0 Performance Summary
JPD-500A Fingertip Pulse Oximeter conforms to the following standards:
- IEC 60601-1, Medical electrical equipment Part 1: General requirements for basic safety and essential performance, 2005, CORR.1:2006 + CORR.2: 2007;
- IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements for safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests;
- IEC 60601-1-11:2010 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment;
- ISO 80601-2-61:2011 Medical electrical equipment Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter;
- . IEC 62304:2006 Medical device software – Software life cycle processes;
- ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process;
- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for In Vitro cytotoxicity;
- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
{6}------------------------------------------------
Shenzhen Jumper Medical Equipment Co., Ltd. Sponsor: Fingertip Pulse Oximeter, model: JPD-500A Subject Device:
#### 9.0 Clinical testing
Clinical hypoxia accuracy testing (controlled desaturation study) was conducted during induced hypoxia studies on 10 healthy, nonsmoking, light-to-dlark-skinned subjects in an independent research laboratory. The measured arterial hemoglobin saturation value (SpO2) of the proposed device was compared with arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter. The accuracy of the device is in comparison with the CO-oximeter samples measured over the SpO2 range of 70-100%. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, the result showed that the error is far less than the scope specified in the STANDARD ; and the Agreement between Methods of Measurement with Multiple Observations per each subject was analyzed with the Bland and Altman statistics, the analysis demonstrated that the vast majority of data is within ±95% limit of agreement, the data points beyond or below this scope were regarded as outliers. By analyzing, these few outliers are occasional, which does not raise safety and performance concerns regarding the accuracy of the device.
15 subjects were enrolled for the clinical study, who are healthy, non-smoking, competent adults, between twenty-seven and forty-four (27-44) years of age, they were provided EC (Ethics Committee)-approved informed consent as documented on an informed consent form. 5 patients were excluded during test based on the exclusion criteria and the criteria for determining stability of the SaO2 at the pulse oximeter probe site.
In addition, there were no reported adverse effects during these investigations.
#### 10.0 Comparison to predicate device and conclusion
The subject device JPD-500A Fingertip Pulse Oximeter is substantially equivalent to CMS-50D Fingertip Pulse Oximeter whose 510(k) number is K082641.
| Characteristics | Subject device<br>JPD-500A | Predicate device<br>(K082641) | Judgment |
|-------------------------------|-------------------------------------------------------------------------------------------------------|--------------------------------------------------|----------|
| Device name | Fingertip Pulse Oximeter | Fingertip Pulse<br>Oximeter | Same |
| Model | JPD-500A | CMS-50D | -- |
| K Number | Pending | K082641 | -- |
| Characteristics | Subject device<br>JPD-500A | Predicate device<br>(K082641) | Judgment |
| Manufacturer | Shenzhen Jumper Medical | Contec Medical | |
| | Equipment Co., Ltd. | Systems Co., Ltd. | -- |
| Intended patient | Adult | | Same |
| population | | | |
| Intended | Fingertip | | Same |
| application site | | | |
| | The JPD-500A Fingertip | This Fingertip | |
| | Pulse Oximeter is intended | Oximeter is intended | |
| | for measuring the functional | for measuring the pulse | |
| | oxygen saturation and pulse | rate and functional | |
| | rate through patient's finger. | oxygen saturation | |
| | It is applicable for spot- | (SpO2) through | |
| Intended use | checking SPO2 and pulse | patient's finger. It is | Same |
| | rate of audit and pediatric | applicable for spot- | |
| | patients in homes and | checking SpO2 and | |
| | clinics. | pulse rate of adult and | |
| | | pediatric patients in | |
| | | | |
| | A mathematical formula is established making use of | homes and clinics. | |
| | | | |
| | Lambert Beer Law according to Spectrum Absorption<br>Characteristics of Reductive hemoglobin(RHb) and | | Same |
| | Oxyhemoglobin (HbO2) in glow and near-infrared | | |
| | zones. Operation principle of the instrument: | | |
| | Photoelectric Oxyhemoglobin Inspection Technology | | |
| | is adopted in accordance with Capacity Pulse | | |
| Design principle | Scanning and Recording Technology, so that two | | |
| | beams of different wavelength of lightscan be focused | | |
| | onto a human nail tip through a clamping finger-type | | |
| | sensor. A measured signal obtained by a | | |
| | photosensitive element, will be shown on the | | |
| | Oximeter's display through process in electronic | | |
| | circuits and microprocessor. | | |
| Prescription or OTC | Prescription | | Same |
| Contact material | Silica gel | | Same |
| Characteristics | Subject device<br>JPD-500A | Predicate device<br>(K082641) | Judgment |
| SpO2 measuring<br>range | 0-100% | 35%-99% | Similar |
| SpO2 accuracy | 70%-100%: $\pm$ 2%<br>0%-69%: no definition | 70%-100%: $\pm$ 2%<br>35%-69%: no definition | Similar |
| Pulse rate<br>measuring range | 25-250bpm | 30-240bpm | Similar |
| Pulse rate accuracy | $\pm$ 2bpm | $\pm$ 2bpm or $\pm$ 2% (whichever<br>is greater) | Similar |
| Operation<br>temperature | 5°C-40°C | 5°C-40°C | Same |
| Operation humidity | 15%-80% | 30-80%RH | Similar |
| Power source | 2 × AAA batteries | Same | Same |
| Other function | Low battery voltage alarm;<br>Automatically power off;<br>Alarm limits setting function | | Same |
{7}------------------------------------------------
#### Shenzhen Jumper Medical Equipment Co., Ltd. Sponsor: Subject Device: Fingertip Pulse Oximeter, model: JPD-500A
{8}------------------------------------------------
| Sponsor: Shenzhen Jumper Medical Equipment Co., Ltd. |
|-------------------------------------------------------------|
| Subject Device: Fingertip Pulse Oximeter, model: JPD-500A |
#### 11.0 Conclusions
As a comparison result for Clause 7.0, JPD-500A is same as its predicate device in the intended use, the design principle, the material, the performance and the applicable standards. Only their appearance, the SpO2 measuring range and pulse rate range are a little bit different.
However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness to the new device. The proposed device is Substantially Equivalent (SE) to the predicate device which is legally marketed in US. Therefore, the subject device is determined as safe and effective as predicate device.
### 12.0 Summary prepared date
August 28, 2014
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.