K162802 · Covidien, LLC · GEI · Jan 24, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162802
Device Name
Barrx Anorectal RFA Wand
Applicant
Covidien, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Barry™ Anorectal RFA Wand is indicated for use in the coagulation of bleeding sites in the anal canal and rectum including, but not limited to, arteriovenous malformations, angiodysplasia, radiation proctitis (RP), and anal intraepithelial neoplasia (AIN).
Device Story
The Barrx Anorectal RFA Wand is a sterile, single-use, bipolar device designed to deliver radiofrequency (RF) energy to tissue within the anal canal and rectum. The device features a cap with an electrode, an EEPROM chip, a proximal shaft, and a rigid distal shaft with a handle. It is used exclusively with the Barrx Flex RFA Energy Generator. Upon connection, the generator reads the device's EEPROM to identify the device and establish operational parameters. The device is operated by clinicians to perform tissue coagulation for conditions such as AIN, arteriovenous malformations, and radiation proctitis. By delivering controlled RF energy, the device enables the ablation of soft tissue, providing a therapeutic benefit for patients with these anorectal conditions.
Clinical Evidence
No new bench or performance testing was conducted as the device is technologically identical to the predicates. The submission references three clinical studies regarding the use of the device for Anal Intraepithelial Neoplasia (AIN) to support the expanded indication.
Technological Characteristics
Sterile, single-use, bipolar RF ablation device. Components include a cap with electrode, EEPROM for generator communication, proximal shaft, and rigid distal shaft with handle. Operates via RF energy delivered by the Barrx Flex RFA Energy Generator. No changes to technological characteristics compared to predicate.
Indications for Use
Indicated for coagulation of bleeding and nonbleeding sites in the anal canal and rectum, including arteriovenous malformations, angiomata, angiodysplasia, radiation proctitis, and anal intraepithelial neoplasia (AIN).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing right, layered on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 24, 2017
Covidien LLC Rachel Silva Senior Regulatory Affairs Specialist 15 Hampshire Street Mansfield. MA 02048
Re: K162802 Trade/Device Name: Barrx Anorectal RFA Wand Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 9, 2016 Received: December 12, 2016
Dear Rachel Silva,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162802
Device Name Barrx™ Anorectal RFA Wand
#### Indications for Use (Describe)
The Barry™ Anorectal RFA Wand is indicated for use in the coagulation of bleeding sites in the anal canal and rectum including, but not limited to, arteriovenous malformations, angiodysplasia, radiation proctitis (RP), and anal intraepithelial neoplasia (AIN).
Type of Use (Select one or both, as applicable):
| <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------|
| <span> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5 510(k) SUMMARY
#### Submitter's Name and Address:
Covidien Ilc 15 Hampshire Street Mansfield, MA 02048
## Contact Person:
Rachel Silva Senior Regulatory Affairs Specialist Phone: (408) 328-7359 Fax: (408) 328-7359
Date Prepared: October 3, 2016
#### Name of Device:
Proprietary Name: Barrx™ Anorectal RFA Wand Common/Usual Name: Electrosurgical Coagulation Catheter Classification Panel: General & Plastic Surgery Device Regulation: 21 CFR 878.4400, Class II Product Code: GEI
#### Establishment Registration Number, Owner/Operator Number:
Establishment Registration Number: 3004904811 Owner/Operator Number: 1282497
## Predicate Devices:
K150251 Barrx™ Anorectal RFA Wand by Covidien K112454 Barrx™ 60 RFA Focal Catheter by Covidien
#### Reason for 510(k) Submission
The purpose of this submission is to include a specific indication for use, which is Anal Intraepithelial Neoplasia (AIN), and include clinical information in the labeling resulting from three AIN related clinical studies. The intent of this 510(k) submission aligns with information stated in the FDA Guidance document titled "Guidance for Industry: General/Specific Intended Use" dated November 4, 1998 as AIN is a specific type of tissue which has been treated by the clinical community under the generic indications for use of coagulating bleeding and nonbleeding sites within the gastrointestinal tract. Also, AIN is found within a narrow target population that falls within the broader population for the products intended use.
#### Device Description:
The Barrx™ Anorectal RFA Wand is a sterile, single-use, bipolar device used to deliver radiofrequency (RF) energy to treatment tissue in the anal canal and rectum. The Barrx™ Anorectal RFA Wand consists of a cap with electrode, EEPROM, a proximal shaft and a rigid distal shaft with handle. The device is used exclusively in connection with the Barrx™ Flex RFA
{4}------------------------------------------------
Energy Generator (K092487). Once the device is connected, the EEPROM is read by the generator, recognizes the device, and determines the operation parameters.
### Indications for Use:
The Barrx™ Anorectal RFA Wand is indicated for use in the coagulation of bleeding and nonbleeding sites in the anal canal and rectum including, but not limited to, arteriovenous malformations, angiomata, angiodysplasia, radiation proctitis (RP), and anal intraepithelial neoplasia (AIN).
The Indication for Use statement for the subject device, Barrx™ Anorectal RFA Wand is not identical to the predicate device, however the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both, the subject and the predicate devices have intended use for the ablation of soft tissue using RF energy.
#### Technological Characteristics of the Device Compared to Predicate Device
The Barrx™ Anorectal RFA Wand has identical technological characteristics to the predicate device cleared under K150251 as well as the secondary predicate device cleared under K112454. There have been minor changes to the Barrx™ Anorectal RFA Wand that did not require a submission and were documented in a letter to file.
#### Performance Data
The Barrx™ Anorectal RFA Wand subject device is technologically identical to the predicate device cleared under K150251 and reference predicate cleared under K112454. Accordingly, no performance testing was conducted.
#### Conclusion
Clinical studies referenced within this submission demonstrate the Barrx™ Anorectal RFA Wand may be used for anal intraepithelial neoplasia and is substantially equivalent to the primary predicate and reference predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.