HALOFLEX ENERGY GENERATOR, MODELS 1190A-115A AND 1190A-230A
K092487 · Barrx Medical, Incorporated · GEI · Nov 10, 2009 · General, Plastic Surgery
Device Facts
Record ID
K092487
Device Name
HALOFLEX ENERGY GENERATOR, MODELS 1190A-115A AND 1190A-230A
Applicant
Barrx Medical, Incorporated
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 10, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HALOFLEX Energy Generator is indicated for use for the coagulation of soft tissue. The HALOFLEX Energy Generator intended use is for coagulation of soft tissue, and bleeding on non bleeding sites in the gastrointestinal tract. The HALORLEX Energy Generator is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including, but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.
Device Story
HALOFLEX Energy Generator is an electrosurgical coagulation system; provides RF energy for soft tissue coagulation in GI tract. Modification of HALO360/HALO90 generators; integrates multiple generator functions onto single hardware platform. Input: electrical power; Output: RF energy for coagulation. Used in clinical settings by physicians. Design changes include CPU board with increased memory, consolidated pressure monitoring/RF MUX boards, and updated software for specific performance characteristics. Device maintains same fundamental scientific technology and clinical parameters as predicates; safety mechanisms remain unchanged. Benefits include reliable soft tissue coagulation and treatment of various GI lesions.
Clinical Evidence
Bench testing only. Performance testing conducted in compliance with international and domestic requirements demonstrated the device is at least as safe and effective as the predicate HALO360 and HALO90 generators.
Technological Characteristics
Electrosurgical coagulation system. Features: CPU board with increased memory, consolidated Pressure Monitoring and RF MUX boards, integrated hardware platform for HALO360/HALO90 functionality. Energy source: RF. Software: updated codes for specific performance characteristics. Sterilization: not specified.
Indications for Use
Indicated for coagulation of soft tissue and bleeding/non-bleeding GI tract sites (e.g., esophagus). Includes treatment for Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
HALO360 Energy Generator model 1100C-115B (K051168)
HALO® Energy Generator model 1100C-115C (K060169, K062441)
{0}------------------------------------------------
K092487)
NOV 1 0 2009
#### 510(k) SUMMARY
## BÂRRX Medical Inc.'s HALORLEX Energy Generator
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
BARRX Medical Inc. 540 Oakmead Parkway Sunnyvale, CA 94085
| Phone: | 408-328-7302 |
|------------|--------------|
| Facsimile: | 408-328-7395 |
Viorica (Vivi) Filimon Contact Person:
Date Prepared: August 12, 2009
#### Name of Device and Name/Address of Sponsor
HALOFLEX Energy Generator
BÂRRX Medical Inc. 540 Oakmead Parkway Sunnyvale, CA 94085
#### Common or Usual Name:
Electrosurgical Coagulation System
#### Classification Name
878.4400 Electrosurgical, cutting & coagulation & accessories, product code GEI
#### Predicate Devices
HALO360 Energy Generator model 1100C-115B
HALO® Energy Generator model 1100C-115C
HALO90 Energy Generator model 90-9000
RF 3000 Radiofrequency Generator
BÂRRX Medical Inc.
BARRX Medical Inc.
BARRX Medical Inc.
Radio Therapeutics Corporation (Boston Scientific Inc)
{1}------------------------------------------------
## Intended Use / Indications for Use
The HALOFLEX Energy Generator is indicated for use for the coagulation of soft tissue
The HALOFLEX Energy Generator intended use is for coagulation of soft tissue, and bleeding on non bleeding sites in the gastrointestinal tract.
The HALORLEX Energy Generator is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including, but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.
## Technological Characteristics
The HALORLEX Energy Generator is a modification to the HALO360 Energy Generator and HALO® Energy Generators that have already been cleared by the Food and Drug Administration ("FDA" or the "Agency") for use in coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus (K051168 and K060169, K062441 respectively). The Generator has the same intended use and fundamental scientific technology as the cleared HALO360 and HALO90 Energy Generators. In addition, the intended use was expanded to include "coagulation of soft tissue", for this indication for use the predicate device is the RF 3000 Radiofrequency Generator (K000241).
The design changes between the HALO360 and HALO® Energy Generators and the HALOFLEX Energy Generator do not affect the clinical parameters employed in the energy delivery or the safety mechanisms that control those parameters.
The HALOFLEX Energy Generator has a limited number of design changes compared to the predicate devices. These changes are minor and do not raise no questions or effectiveness. Further, bench testing demonstrated comparable performance and safety. These modifications were as follows:
- Add to the indications for use the "coagulation of soft tissue", to reflect the . general nature of the device.
- Modification of the CPU board, for increased memory .
- Consolidation of the Pressure Monitoring Board and RF MUX Board in the Hand Piece Interface Board
{2}------------------------------------------------
- Integration on the HALO360 and HALO90 Energy Generators on the same . hardware platform as HALO360 Energy Generator
- . Update the software codes to reflect the specific performance characteristics for each Energy Generator.
- Optimization of components for reliability, and/or obsolescence
## Performance Data
The HALOFLEX Energy Generator is technologically equivalent and clinically identical to the cleared HALO360 and HALO90 Energy Generators.
Performance testing was conducted in compliance with the applicable international and domestic requirements and the certifications (for harmonized standards) and the reports for performance testing were provided to FDA to support this submission. That data demonstrated that, when used in accordance with the instructions, the HALOFLEX is at least as safe and effective as the predicate devices HALO360 and HALO90 Energy Generators.
### Substantial Equivalence
The HALOFLEX Energy Generator is as safe and effective as the HALO360 Energy Generator (K051168) and HALO90 Energy Generator (K060169 and K062441), RF 3000 Radiofrequency Generator (K000241) the predicate devices. The HALOFLEX Energy Generator has the same intended uses and indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the HALORLES Energy Generator and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the HALOFLEX Energy Generator is as safe and effective as HALO360 Energy Generator (K051168) and HALO90 Energy Generator (K060169 and K062441), RF 3000 Radiofrequency Generator the predicate devices. Thus, the HALOFLEX Energy Generator is substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure with outstretched arms.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
BARRX Medical, Inc. % Ms. Viorica Filimon Vice President of Quality/Regulatory Affairs 540 Oakmead Parkway Sunnyvale, California 94085
NOV 1 0 2009
Re: K092487
Trade/Device Name: HALOFLEX Energy Generator Model 1190A-115A Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 12, 2009 Received: August 13, 2009
Dear Ms. Filimon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Ms. Viorica Filimon
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Sincerely yours,
Mark N Milliken
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name:
Indications for Use:
The HALOFLEX Energy Generator is indicated for use for the coagulation of soft tissue.
The HALOFLEX Energy Generator is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including, but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Die Hotor Lesions, and Angiodysplasia.
AND/OR
Prescription Use _ X Use (Part 21 C.F.R. 801 Subpart D)
Over-The-Counter
(21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
FOR M. MELKERSON
Division of Surgical, Orthopedic, and Restorative Devices
Page **_ of _**
510(k) Number K01248
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.