K162763 · Vevazz, LLC · ILY · Jan 5, 2017 · Physical Medicine
Device Facts
Record ID
K162763
Device Name
Vevazz LED Heat Lamp
Applicant
Vevazz, LLC
Product Code
ILY · Physical Medicine
Decision Date
Jan 5, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vevazz is a device intended to emit energy in the visible and near-IR spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm; the temporary increase in local blood circulation and temporary relaxation of muscles.
Device Story
Vevazz LED Heat Lamp is a prescription-use device providing topical heating via infrared energy. It consists of a base unit connecting up to 16 large and 4 small LED paddles. Each paddle emits 650 nm light at 50 mW/cm2. The device operates by elevating tissue temperature to 41°±2°C to provide temporary relief of minor muscle/joint pain, stiffness, arthritis, or spasms, and to increase local blood circulation. It is intended for use in clinical settings. Healthcare providers apply the paddles to the patient's skin; the device maintains a constant temperature during a default 17-minute treatment cycle. The heat application aims to relax muscles and alleviate pain, providing therapeutic benefit through thermal effects.
Clinical Evidence
Bench testing only. Performance testing verified the device reaches and maintains an idle temperature of 40°C (±2°C) for a minimum of 10 minutes. An IR camera monitored skin temperature and uniformity on the abdominal region across multiple skin types. The device complies with IEC 60601-1, IEC 60601-2-57, IEC 62471, and IEC 60601-1-2 standards.
Technological Characteristics
Base unit supports up to 16 large (28 LEDs) and 4 small (4 LEDs) paddles. LEDs emit 650 nm (+/- 25 nm) light at 50 mW/cm2. Patient contact material is rigid ABS. Power input 120/240 VAC; output 16 VDC @ 0.5 amps per port. Constant waveform. Complies with IEC 60601-1, IEC 60601-2-57, IEC 62471, and IEC 60601-1-2.
Indications for Use
Indicated for patients requiring topical heating for temporary relief of minor muscle and joint pain, arthritis, stiffness, or muscle spasms; temporary increase in local blood circulation; and temporary muscle relaxation.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 5, 2017
Vevazz, LLC Mr. Jamie Fettig, DC Owner 30520 Rancho California Rd, Suite 107-46 Temecula, California 92591
Re: K162763
Trade/Device Name: Vevazz Led Heat Lamp Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: ILY Dated: November 3, 2016 Received: November 8, 2016
Dear Dr. Fettig:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) 162763
Device Name Vevazz LED Heat Lamp
### Indications for Use (Describe)
The Vevazz is a device that emits energy in the infrared spectrum and is to be used to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain, arthritis, stiffness or muscle spasms; the temporary increasing of local blood circulation and/or the temporary relaxation of muscle.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span style="letter-spacing: normal;"> <span style="line-height: normal;"> <span style="orphans: auto;"> <span style="text-align: start;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: auto;"> <span style="word-spacing: 0px;"> <span style="-webkit-text-stroke-width: 0px;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510-(k) Summary
#### 1. Submitters Information
Name: Vevazz, LLC 30520 Rancho California Road Suite 107-46 Temecula, CA 92591 Tel: 773-665-4005 Contact: Jamie Fettig, Owner 773-620-9500 jamie@healthyhuman.org
#### 2. Date Prepared:
January 3, 2017
#### 3. Device Name and Classification
| Device | Vevazz LED Heat Lamp |
|------------------------|----------------------|
| Trade Name | Vevazz LED Heat Lamp |
| Regulation Description | Infrared lamp |
| Regulation Number | 21 CFR 890.5500 |
| Device Class | II |
| Review Panel | Physical Medicine |
| Product Code | ILY |
#### 4. Intended Use
The Vevazz is a device intended to emit energy in the visible and near-IR spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm; the temporary increase in local blood circulation and temporary relaxation of muscles.
{4}------------------------------------------------
#### 5. Device Description
The Vevazz device is a moderately complex device that consists of a base unit that plugs into 120 VAC or 240 VAC main power with Ground (earth) and a 2 amp 250 VAC Fuse. The base unit can support up to 16 large LED paddles and 4 small LED paddles. The connections on the base are female phono jacks that are low voltage having a maximum electrical output of 16 VDC @ .5 amps for a total of 8 Watts per port. The large paddles consist of 28 LEDs in a series/parallel electrical connection that has a cord with a male phono jack. Small paddles consist of 4 LEDs and have matching electrical characteristics as the large paddles. Each LED produces 50 mw/cm2 of light energy with a wavelength of 650 nm: +/- 25 pm uniformly over the treatment area without producing hot spots.
#### Predicate Devices 6.
- HVR Infrared Lamp by HVR, LLC K101716 ●
- Tanda Restore by Pharos Life Corporation K090008 .
- . HEATLUX by Home Skinnovations, LTD. - K120582
#### 7. Technical and Performance Substantial Equivalency
The submission and predicate devices are fully compliant with USFDA recognized standards for electrical safety and emit a substantially equivalent amount of power and energy. Differences where they occur are minor and do not represent any risk or hazard.
| Performance<br>Characteristic | Vevazz | HVR Lamp | Tanda Restore | HeatLux |
|--------------------------------|-------------------------|-------------------|---------------|-------------|
| Power | 50 mW/cm2 | 50-80 mW/cm2 | 60 mW/cm2 | 60 mW/cm2 |
| Wavelength | 650 nm +/-<br>25nm | 650 nm to 950 nm | 870 nm | 645 nm |
| Waveform | Constant | Constant | Constant | Constant |
| Energy Source | LED | LED | LED | LED |
| Treatment<br>Area | 7.5 cm2 | 1000 mm2 (10 cm2) | 27 cm2 | 7 cm2 |
| Treatment<br>Time | 17 minutes<br>(default) | 90 seconds | 3 minutes | 3-5 minutes |
| Target Skin<br>Temperature | 41°±2°C | 42°±3°C | 41°±2°C | 41°±2°C |
| Patient<br>Contact<br>Material | Rigid ABS | Rigid ABS | Rigid ABS | Rigid ABS |
# 8. Bench Testing Summary
The Vevazz is in compliance with the following standards:
- IEC 60601-1:2005 +C1:2006, +C2:2007, AM1:2012, Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-57:2011, Medical electrical equipment Part 2-57: Particular requirements for basic safety and essential performance of non-laser lightsource equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- IEC 62471 Ed. 1, International Standard for Photobiological Safety for Lamps and Lamp Systems - General Requirements
{5}------------------------------------------------
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests.
## 9. Clinical Testing Summary
The Vevazz heat performance was tested to verify it will reach an idle temperature of 40℃ and will maintain that temperature throughout the treatment for a minimum duration of 10 minutes ±2℃. An IR camera was used to monitor skin temperature and uniformity on the abdominal region and multiple skin types.
## 10. Conclusion
The Vevazz is substantially equivalent to its predicate devices. Based on the technological comparison between the Vevazz and the predicates plus the human test data, it can be concluded that the Vevazz is equivalent to the identified predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.