K120582 · Home Skinovations , Ltd. · ILY · Jul 6, 2012 · Physical Medicine
Device Facts
Record ID
K120582
Device Name
HEATLUX
Applicant
Home Skinovations , Ltd.
Product Code
ILY · Physical Medicine
Decision Date
Jul 6, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
HeatLux is over the counter hand held device intended to emit energy in the visible and near IR spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local and blood circulation, and temporary relaxation of muscles.
Device Story
HeatLux is a handheld, battery-operated device for OTC topical heating. It utilizes LED technology to emit light at 645 nm (±10 nm) to elevate tissue temperature. The device features a tip temperature stabilizer designed to maintain a surface temperature of 41°C (±2°C) with uniform power distribution (60 mW/cm²) to prevent hot spots. Operated by the patient, the device is applied to the skin for 3-5 minute treatment sessions. By providing localized heat, the device aims to increase blood circulation, relax muscles, and provide temporary relief from minor pain and stiffness associated with muscles, joints, or arthritis. The device is constructed with stainless steel (17-4PH) and rigid ABS plastic.
Clinical Evidence
Bench testing only. Testing verified the device's ability to reach and maintain an idle temperature of 41°C. Temperature sensors were attached to the metal surface to monitor temperature over time. Thermal camera imaging was used to verify skin temperature and uniformity across multiple body areas and skin types.
Technological Characteristics
Handheld, battery-operated LED device. Wavelength: 645 nm (±10 nm). Power density: 60 mW/cm². Target skin temperature: 41°C (±2°C). Materials: Stainless steel 17-4PH and rigid ABS. Treatment area: 7 cm². Complies with 21 CFR 1010 performance standards.
Indications for Use
Indicated for OTC use by individuals seeking temporary relief of minor muscular and joint pain, stiffness, minor arthritis pain, muscle spasm, and for temporary increase in local blood circulation and muscle relaxation.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
K120582
Page 1 of 3
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Home Skinovations Ltd. HeatLux
6 2012 JUL
This summary of safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
### Submitter's information
| Name: | Home Skinovations Ltd. |
|----------|-------------------------------------------------|
| Address: | Tavor building, POB 533, Yokneam 20692, Israel. |
| Contact: | Dr. Amir Waldman VP Regulatory Affairs |
# Device information
Trade/Proprietary name: HeatLux
Regulation Number: 21 CFR 890.5500
Regulation Name: Infrared lamp
Regulatory Class: Class II
Product Code: ILY
## Predicate devices
- . Quantum WARP10 Light Delivery system by Quantum Devices Inc. K032229.
- HVR Pain Relief Device by HVR LLC, K101716. .
- Tanda Restore, by Pharos Life Corporation, K090008. �
- LightStim, by LED Intellectual Properties LLC, K083580. ●
#### Intended use:
HeatLux is over the counter hand held device intended to emit energy in the visible and near IR spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local and blood circulation, and temporary relaxation of muscles.
{1}------------------------------------------------
K120582
## Device Description& technology comparison to predicate device:
HeatLux is a hand held device battery operated that uses low power light spectrum, LED, at wavelength of 645 ±10 nm, combine with tip temperature stabilizer at 41°C. The emitting optical power is in a uniform distribution with no hot spots.
Technology comparison of HeatLux with predicate devices:
| | HeatLux | Quantum<br>WARP10 | HVR | Tanda<br>Restore | LightStim |
|--------------------------------------------|---------------------------------------------|-------------------|-----------|------------------|-------------------------|
| Wavelengths (nm) | 645 | 650-950 | 650-950 | 870 | 630,<br>660,855,<br>940 |
| Power mW/cm2 | 60 | 50-80 | 50-80 | 60 | 50-80 |
| Treatment area<br>cm2 | 7 | 10 | 10 | 27 | 37 |
| Type of Energy<br>source | LED | LED | LED | LED | LED |
| Targeted skin<br>temperature °C | 41±2 | 39-45 | 39-45 | 39-43 | 39-43 |
| Recommended<br>treatment time<br>(minutes) | 3-5 | | | | |
| Patient contacting<br>material | Stainless<br>steels<br>17-4PH,<br>Rigid ABS | Rigid ABS | Rigid ABS | Rigid ABS | Rigid ABS |
#### Performance data:
The device complies with the following U.S. Food and Drug Administration performance standards: 21 CFR 1010.
#### Bench testing summary:
Bench testing preformed in order to verify the HeatLux performance to reach to the idle temperate of 416C and to maintain the temperature over time.
In the first test temperature sensor was attached to the metal surface of the device and monitors the temperature as function of time.
In the second setup a thermal camera was used in order to monitor the skin temperate and uniformity. The measurement was done on multiple body areas and skin types.
#### Substantial Equivalence:
The HeatLux is substantial equivalent to its predicate device. The data in this 510(k) submission demonstrate that the HeatLux device has compatible output as the predicate devices, and identical intended use. Therefore is substantial equivalent to its predicate devices.
{2}------------------------------------------------
K 120582
Page 3 of 3
Based upon an analysis of the overall performance characteristic for the device, Home Skinovations Ltd. believes that no significant differences exit. Therefore the HeatLux should raise no new issues of safety or effectiveness.
June 21, 2012
Date
Dr. Amir Waldman,
VP Regulatory Affairs
Home Skinovations Ltd.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's commitment to health, human services, and well-being. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Home Skinovations Limited % Dr. Amir Waldman Vice President, Regulatory Affairs Tavor Building POB 533 Yokneam Illit, Israel 20692
Re: K120582
Trade/Device Name: HeatLux Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: June 21, 2012 Received: June 26, 2012
Dear Dr. Waldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
6 2012 JUL
{4}------------------------------------------------
## Page 2 - Dr. Amir Waldman
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known)______________________________________________________________________________________________________________________________________________________
Device Name HeatLux
Indications For Use:
HeatLux is over the counter hand held device intended to emit energy in the visible and near IR spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local and blood circulation, and temporary relaxation of muscles.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k 120582
Over The Counter Use_X
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.