The IOB Temperature Management system is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
Device Story
IOB Temperature Management System consists of warming unit and disposable blankets. Unit draws ambient air through 0.2-micron filter, warms air to selected temperature (32°C, 38°C, 43°C, or ambient), and delivers it via hose to blanket. Blankets feature distribution channels and perforations to disperse air over patient. Servo-controlled by thermistor at hose-blanket connection; control thermistor in unit adjusts heater power. Used in clinical settings to manage patient temperature. Healthcare providers select temperature settings on control panel. System maintains temperature despite ambient variations. Benefits include thermal regulation for patients undergoing surgery or requiring temperature therapy.
Clinical Evidence
No clinical data. Bench testing only, including temperature uniformity, accelerated stability (3-year shelf life), electrical safety/EMC, software validation, and biocompatibility (ISO 10993).
Technological Characteristics
Materials: non-woven polypropylene fabric coated with polyethylene. Sensing: thermistor-based servo-control. Energy: 110-120V/230V AC electric heater. Connectivity: standalone. Sterilization: non-sterile (select models sterile). Software: embedded control logic for temperature regulation and safety monitoring.
Indications for Use
Indicated for hypothermic or normothermic patients requiring induced hyperthermia or localized temperature increase. Applicable to adult and pediatric populations.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
Predicate Devices
Bair Hugger Temperature Management System (K021473)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 28, 2017
IOB Medical, Inc. % Joe Shia Regulatory Consultant LSI International, Inc. 504 East Diamond Ave., Suite F Gaithersburg, Maryland 20877
Re: K162679
Trade/Device Name: IOB Temperature Management System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: March 10, 2017 Received: March 20, 2017
Dear Joe Shia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando
Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162679
#### Device Name
IOB Temperature Management System
Indications for Use (Describe)
The IOB Temperature Management system is indicated for hypothermic patients or normothermic for whom induced hyperthermia or localized increase in temperature is clinically indicated.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
| 1. Date: | April 28, 2017 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Submitter: | IOB Medical Inc<br>504E Diamond Ave., Suite I<br>Gaithersburg, MD 20877 |
| 3. Contact person: | Joe Shia<br>LSI International Inc.<br>504 East Diamond Ave., Suite F<br>Gaithersburg, MD 20877<br>Telephone: 240-505-7880<br>Fax: 301-916-6213<br>Email: shiajl@yahoo.com |
- 4. Device Name: IOB Temperature Management System
- 5. Classification: Class:
Class II
| Product Code | CFR # | Product Name |
|--------------|----------|---------------------------|
| DWJ | 870.5900 | Thermal Regulating System |
- 6. Predicate Devices: K021473 Bair Hugger Temperature Management System
- 7. Intended Use
The IOB Temperature Management system is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
- 8. Device Description
The IOB Temperature Management System consists of the IOB warming unit and the IOB warming The IOB warming unit draws ambient- temperature air through a 0.2 micron air filter. The blankets. filtered air is warmed to a selected temperature. The warmed air enters the IOB Warming Blanket through the hose and is distributed through delivery channels. Perforations on the patient side of the air delivery channels in the warming blanket gently disperse the warmed air over and around the patient. The warming unit has three temperature settings of 32°C, 38°C, and 43°C. These temperature settings are servo-controlled by a thermistor placed at the end of hose where the hose connects to the blanket. The unit can also delivers ambient-temperature air. The temperature indicated on the control panel is the temperature of air being delivered to the blanket at the end of the hose. A control thermistor in the warming unit adjusts the power applied to the heater to maintain the selected temperature. This enables the system to maintain the selected temperature under variations in ambient temperature. Besides, the warming unit has high and low air flow for choose.
The IOB Warming Blankets in this submission are the following:
{4}------------------------------------------------
Torso IOB-001 Lower Body IOB-002 Upper Body IOB-003 Full Body IOB-004 Pediatric Underbody IOB-005 Adult Underbody IOB-006 Pediatric Full Body IOB-007 Full Body Surgical IOB-008 Large Pediatric Underbody IOB-009 Spinal Underbody IOB-010 Lithotomy Underbody IOB-011 Pediatric Lower Body IOB-012 Pediatric Long IOB-014 Cath Lab IOB-015 Surgical Access IOB-016 Chest Access IOB-017 Multi-Access IOB-018 Dual Port Torso IOB-019 Cardiac Access IOB-020 XL Upper Body IOB-021 Outpatient Care IOB-022 Cardiac IOB-023
These blankets are single-use and disposable. Each blanket consists of two lavers of non-woven polypropylene fabric coated with a layer of polyethylene. The lavers are bonded together to form a distribution network of air delivery channels. The warm air is distributed around the blanket through the delivery channels and exits the blanket through a specially designed series of perforations in the patient side of the blanket.
- 9. Substantial Equivalence Information
A summary comparison of features of the IOB Temperature Management system and the predicate devices is provided in following Tables.
| Parameters | Predicate devices K021473 | Proposed devices |
|------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Bair Hugger blankets (models<br>540,525,522,300,555,635,310,610,550,575,<br>585,537,530,560,570,305,315,542,645,523,<br>110,630) | IOB Warming Blankets<br>(IOB-001,IOB-002,IOB-003,IOB-004,IOB-005,<br>IOB-006,IOB-007,IOB-008,IOB-009,IOB-010<br>IOB-011, IOB-012,IOB-014,IOB-15,IOB-16<br>IOB-17, IOB-18, IOB-019,IOB-020,IOB-021,<br>IOB-022,IOB-023) |
| | | Table 1: Comparison between IOB Warming Blankets and predicate devices. | | |
|--|--|-------------------------------------------------------------------------|--|--|
| | | | | |
{5}------------------------------------------------
| INDICATIONS<br>FOR USE | The Bair Hugger temperature management<br>system is indicated for hyper- or hypothermic<br>patients or normothermic patients for whom<br>induced hyper- or hypothermia or localized<br>temperature therapy is clinically indicated. In<br>addition, the Bair Hugger temperature<br>management system can be used to provide<br>patient thermal comfort when conditions exist<br>that may cause patients to become too warm or<br>too cold. The Bair Hugger temperature<br>management system can be used with adult and<br>pediatric patients. | The IOB Temperature Management system is<br>indicated for hypothermic patients or<br>normothermic patients for whom induced<br>hyperthermia or localized increase in<br>temperature is clinically indicated. |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MATERIAL<br>DESIGN | Consists of two layers of non- woven<br>polypropylene fabric bonded to a fusion layer of<br>polyethylene.<br>The layers are bonded together to form a<br>distribution network of air delivery channels<br>The warm air is distributed around the patient's<br>body through the delivery channels and exits the<br>blanket through a specially designed series of<br>perforations in the patient side layer of the<br>blanket.<br>The distribution of air is designed to minimize<br>temperature differences of delivered air at<br>different blanket locations. | SAME |
| Shelf Life | 3 years | SAME |
| Sterility | Non-sterile except 645 and 630 | Non-sterile except IOB-14, IOB-16, IOB-020 and<br>IOB-023 |
| Blanket<br>Dimensions<br>(approximate) | 540 107×91cm<br>525 152×91cm<br>522 188×61cm<br>300 213×91cm<br>635 221×91cm<br>610 183×91cm<br>575 188×91cm<br>585 188×91cm<br>560 23×335cm<br>570 213×91cm<br>305 183×91cm<br>315 213×91cm<br>542 107×91cm<br>523 213×91cm<br>110 213×91cm<br>310 152×91cm<br>555 91×84cm<br>550 152×81cm<br>537 89×61cm<br>530 188×21cm<br>645 23x274cm<br>630 152x91cm | IOB-001 142×120cm<br>IOB-002 142×120cm<br>IOB-003 195×80cm<br>IOB-004 210×120cm<br>IOB-006 215×100cm<br>IOB-008 210×120cm<br>IOB-010 215×100cm<br>IOB-011 170×100cm<br>IOB-015 17x180cm<br>IOB-016 210×120cm<br>IOB-017 180×120cm<br>IOB-018 210×120cm<br>IOB-019 142×120cm<br>IOB-021 215×80cm<br>IOB-022 210×120cm<br>IOB-007 170×100cm<br>IOB-005 91×80cm<br>IOB-009 160×80cm<br>IOB-012 142×100cm<br>IOB-014 110×17cm<br>IOB-020 17x150cm<br>IOB-023 142x120cm |
| Air Velocity | 28-30cfm | Same |
| Air<br>Temperature | Hose End Temperature<br>43°C +/- 3.0°C<br>38°C +/- 3.0°C<br>32°C +/- 3.0°C<br>Ambient | Hose End Temperature<br>43°C +/- 3.0°C<br>38°C +/- 3.0°C<br>32°C +/- 3.0°C<br>Ambient |
| System<br>Power | 110-120VAC,60Hz,10A<br>220-240VAC,50Hz,4.5A | 110-120V, 60 Hz, 12A<br>230V, 50/60 Hz, 8A |
| Heater<br>Power | 850 Watts | 1000 W |
| Dimensions | 33X 25 X 28cm | 28 X 22X 22cm |
| Weight | 5.2kg | 4.5kg |
| EMI/EMC<br>Compliant | IEC60601-1, IEC 60601-1-2 | IEC60601-1, IEC 60601-1-2 |
| Forced air<br>Over<br>Temperature | Auto-shuts heater off at 47°C +/- 2°C | Auto-shuts heater off at 47°C +/- 2°C. |
| Hose with<br>Secure<br>Locking | Yes | Yes |
| Air Filter | Replaceable 0.2 micron | Replaceable 0.2 micron |
| Temperature<br>Display | Front panel LCD display | Front panel digital display |
{6}------------------------------------------------
{7}------------------------------------------------
10. Safety and Performance Characteristics
- Temperature uniformity tests show equivalence between the IOB warming blankets and the 1. predicate.
- 2. Accelerated stability tests show three years shelf-life of the IOB warming blankets.
- 3. Electrical safety/EMC testing, and software validation show that the device meets its design specifications, performs as intended.
- 4. Biocompatibility testing (cytotoxicity, irritation and sensitivity) according to ISO 10993 for a limited contact device was demonstrated to be suitable for the intended use of the product.
- 5. Clinical Studies Not applicable
## 11.Conclusion
Based on the information presented in this 510K premarket notification, the IOB Temperature Management System is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.