The Bair Hugger temperature management system is indicated for hyper- or hypothermic patients or normothermic patients for whom induced hyper- or hypothermia or localized temperature therapy is clinically indicated. In addition, the Bair Hugger temperature management system can be used to provide patient thermal comfort when conditions exist that may cause patients to become too warm or too cold. The Bair Hugger temperature management system can be used with adult and pediatric patients.
Device Story
System consists of forced-air temperature management unit (Series 200, 500, or 700), disposable Bair Hugger blanket, and Model 459 patient cooling set. Unit delivers air through flexible hose; cooling set uses ice-filled plastic bag to cool ambient air before delivery to blanket. Blanket placed around, over, or under patient; perforations disperse air. Used in clinical settings; operated by healthcare providers. Cooling set provides localized temperature therapy or thermal comfort by cooling air to 3.5°C–room temperature range. Benefits include non-invasive temperature regulation for patients at risk of thermal instability.
Clinical Evidence
Bench testing only. Performance testing demonstrated that room-temperature air flowing through the ice-filled Model 459 cooling set is cooled to <10°C for approximately 1 hour. Testing confirmed that delivered temperatures stabilize between 5°C and 10°C, with brief initial exposures to temperatures near 0°C deemed non-harmful.
Technological Characteristics
System includes forced-air unit, flexible hose, disposable blankets, and ice-filled plastic cooling set. Cooling principle: passive heat exchange via ice. Air temperature range: 3.5°C to ambient. Connectivity: standalone. Power: electrical (unit).
Indications for Use
Indicated for adult and pediatric patients who are hyperthermic, hypothermic, or normothermic requiring induced temperature therapy or thermal comfort management.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
Predicate Devices
Augustine Medical, Bair Hugger® Model 600 Hyper/hypothermia unit
Seabrook Medical Systems, Tropi-Cool Hyper/hypothermia unit
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| Submitted by | Augustine Medical, Inc. |
|--------------|-------------------------|
| | 10393 West 70th Street |
| | Eden Prairie, MN 55344 |
- Contact person: Scott Augustine, MD Chief Executive Officer
Telephone: 952-947-1200 Fax: 952-947-1300
#### Device name
Trade name:
| Bair Hugger® temperature management system (Bair |
|--------------------------------------------------------|
| Hugger Series 200, 500, and 700 forced-air temperature |
| management units used with Bair Hugger blankets) and |
| Model 459 patient cooling set. |
| Common/usual name: | Hyper/Hypothermia system |
|--------------------|--------------------------|
|--------------------|--------------------------|
#### Classification name: System, Thermal, Regulating DWJ
### Predicate Devices
- 1. Augustine Medical, Bair Hugger® Model 600 Hyper/hypothermia unit used with Bair Hugger blankets.
- 2. Seabrook Medical Systems, Tropi-Cool Hyper/hypothermia unit used with Temp-pad blankets.
# Device Description
The Bair Hugger temperature management system consists of:
- · a portable forced-air temperature management unit (200, 500, or 700 series),
- · a disposable Bair Hugger forced-air blanket (various models), and
- · the Model 459 patient cooling set (new disposable component of the system).
The temperature management unit delivers warmed or room-temperature air directly to a Bair Hugger blanket via a flexible hose, or it delivers room-temperature air to the Model 459 patient cooling set, which, when filled with common ice, cools the air before delivering it to a Bair Hugger blanket. Depending on the blanket model used, the blanket
{1}------------------------------------------------
## Device Description (continued)
is placed around, over, or underneath the patient. Small perforations in the patient-side of the blanket disperse the air over the patient.
The new disposable Model 459 patient cooling set consists of an ice receptacle (flexible plastic bag) with an attached flexible hose assembly. The patient cooling set also has a handle to allow hanging it on an IV pole or similar stand near the temperature management unit.
The ice receptacle has:
- · a hose port that accepts the hose of a Bair Hugger® temperature management unit,
- · a resealable opening for adding and retaining ice,
- · a valve for draining meltwater, and
- · an integral divider that directs the flow of room-temperature air from the temperature management unit through the ice, which cools the air to a temperature ranging between 3.5℃ and room temperature, as required.
#### Indications for use
The Bair Hugger temperature management system is indicated for hyper- or hypothermic patients or normothermic for whom induced hyper- or hypothermia or localized temperature therapy is clinically indicated. In addition, the Bair Hugger temperature management system can be used to provide patient thermal comfort when conditions exist that may cause patients to become too warm or too cold. The Bair Hugger temperature management system can be used with adult and pediatric patients.
### Description of Safety and Effectiveness
Performance testing of the Model 459 patient cooling set was conducted to demonstrate the effectiveness of the component when used with the other Bair Hugger temperature management system components. The performance testing indicates that room-temperature air flowing from a temperature management unit through the ice-filled patient cooling set would be cooled to less than 10°C for about 1 hour. Ice may be added to the patient cooling set to prolong cooling therapy. A temperature management unit may also deliver ambient temperature air directly to a blanket for cooling therapy.
Because the modification to the device does not change any of the safety features of the device, testing was done only to determine the safety of temperatures delivered by the Model 459 patient cooling set, the new disposable component of the device.
{2}------------------------------------------------
The delivered temperatures are safe for two reasons: first, the temperature of air delivered to a Bair Hugger blanket is limited by the temperature and duration of the ice added to the Model 459 patient cooling set. Most ice in clinical settings is produced and stored in an icemaker at 0°C, and performance testing indicates that 15 pounds of ice will only deliver temperatures under 10°C for about an hour before the temperatures begin to rise. Second, tests show that if users add ice that has been stored below 0°C to the patient cooling set, the temperature of the delivered air might initially drop to close to 0°C, but it will quickly rise and stabilize between 5°C and 10°C, the standard temperature range for the device. The few minutes that a patient may be exposed to temperatures below 5°C are not harmful to the patient.
#### Substantial equivalence
The modified Bair Hugger® temperature management system (including the new Model 459 patient cooling set) is substantially equivalent to the predicate devices in safety and effectiveness. The modified device has the same intended use and patient population as the predicate devices. The warming modes of the previously cleared device(s) are not affected by the modification so the comparison of the device to predicate devices focuses on the substantial equivalence of the cooling modes of the devices.
#### Substantial equivalence to the Bair Hugger Model 600 hyper/hypothermia unit
The modified Bair Hugger temperature management system and the Bair Hugger Model 600 hyper/hypothermia unit (first predicate device) use the same technology to deliver air to forced-air blankets. The only significant difference between the two devices is that. in the cooling mode, the modified Bair Hugger temperature management system uses an external, ice-filled component (Model 459 patient cooling set) to cool the air before delivering it to a Bair Hugger blanket, and the predicate device has a thermoelectric heat exchanger that cools air and delivers it directly to a Bair Hugger blanket. This difference has no negative effect on the safety and effectiveness of the modified device because the temperatures delivered to the patient are similar.
#### Substantial equivalence to the Tropi-Cool hyper/hypothermia unit
The only significant difference between the modified Bair Hugger temperature management system (modified device) and the Tropi-Cool hyper/hypothermia unit (second predicate device) is that, in the cooling mode, the modified device uses an external, ice-filled component (Model 459 patient cooling set) to cool air before delivering it to a Bair Hugger forced-air blanket. The predicate device has a thermoelectric heat exchanger that cools water and delivers it directly to a Temp-pad water pad. This difference has no negative effect on the safety and effectiveness of the modified device because the temperatures delivered to the patient are similar.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three wing-like shapes, positioned to the right of a circular text element. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" and is arranged around the circle's perimeter. The eagle design is simple yet recognizable, and the overall logo has a clean and official appearance.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ர்பட 9 2002
Augustine Medical, Inc. c/o Scott Augustine, MD Chief Executive Officer 10393 West 70th Street Eden Prairie, MN 55344
Re: K021473
> Trade Name: Bair Hugger® Temperature Manager System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal regulating system Regulatory Class: Class II (two) Product Code: DWJ Dated: May 7, 2002 Received: May 8, 2002
Dear Dr. Augustine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Scott Augustine, MD
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Ralston
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) number (Traditional): KO2 | 4773
Device name:
Augustine Medical, Inc. Bair Hugger® temperature management system
Indications for use:
The Bair Hugger temperature management system is indicated for hyper- or hypothermic patients or normothermic patients for whom induced hyper- or hypothermia or localized temperature therapy is clinically indicated. In addition, the Bair Hugger temperature management system can be used to provide patient the comfort when conditions exist that may cause patients to become too warm or too cold. The Bair Hugger temperatur management system can be used with adult and pectratr patients.
/00/HV000/
PLEASE DO NOT WRITE BELOW THIS LINE---- CONTINUE ON ANOTHER PAGE IF NEEDED.
Concurrence of CDRH, Office of Device Evaluation (ODE)
0(k) Numbe
Prescription Use X or Over the Counter Use **__**
(Per 21 CFR 801-109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.