ORTHO SOLUTIONS EXTREMITY FIXATION IMPLANTS FOR OSTEOSYSNTHESIS
Applicant
Ortho Solutions Limited
Product Code
HRS · Orthopedic
Decision Date
Feb 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Ortho Solutions Extremity Fixation Implants for Osteosymthesis System of bundled fixation device(s) is to draw two or more aligned small bone fragments together to facilitate healing. The Ortho Solutions Extremity Fixation Implants for Osteosynthesis System is used in adult patients. The Ortho Solutions Foot Plating System (FPS) is indicated for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the feet, ankles, and toes skeleton. The Ortho Solutions Memo Staple System is indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis. The Memo Staple is also indicated for use in fixation of soft tissue to bone such as anterior cruciate reconstruction. The Ortho Solutions Foot Osteotomy System (FOS) is indicated for alignment and stabilization of small bone fractures including: fixation of small bones, such as those in the foot and ankle for treatment of fractures, non-unions, or mal-unions; ligament reconstructions; osteochondritis dissecans; arthrodesis of the foot and ankle; small bone osteotomies, including first metatarsal head osteotomy and metatarsal osteotomies. The Ortho Solutions Staples System is indicated for fixation of bone fractures or bone reconstruction. The staples are also indicated for use in fixation of soft tissue to bone. The Ortho Solutions Compression Screws System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation of bones appropriate for the size of the device. The Ortho Solutions TwistFix™ Snap-Off Screws System is indicated for fixation of bone fractures or for bone reconstruction. Examples include: 1) Fixation of small bone fragments, 2) Weil Osteotomy, 3) Mono-cortical fixation, and 4) Osteotomies and fractures fixation in the foot and hand. The Ortho Solutions Extremity Fixation System (EFS) of a low profile head screw is indicated for use in fixation of small bone fractures, nonunions, arthrodesis, and osteotomies of the small bones in the hand or foot. The Ortho Solutions Extemity Fixation Implants for Osteosynthesis System is not intended for spinal use.
Device Story
System of metallic bone fixation implants including plates, screws, staples, and washers; used for small bone reconstruction, fracture fixation, and arthrodesis in hand/foot/ankle. Implants are surgically implanted by orthopedic surgeons in clinical settings. Principle of operation involves mechanical stabilization of bone fragments to facilitate healing; compression staples utilize superelastic nitinol properties or mechanical deformation via spreader instruments to provide bone compression. Devices are provided sterile for single-use. Output is physical stabilization of bone segments; clinical decision-making relies on surgeon assessment of fracture/reconstruction needs. Benefits include reduced tissue irritation via low-profile designs and improved bone alignment/healing.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics to legally marketed predicate devices.
Indicated for adult patients requiring stabilization, fixation, or reconstruction of small bones in the hand, foot, and ankle. Applications include fresh fractures, non-unions, mal-unions, osteotomies, joint fusion/arthrodesis, and soft tissue-to-bone fixation. Not for spinal use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K111678 ('15)
Ortho Solutions Limited - 510(k) Summary: .
# 510(k) Summary of Safety and Effectiveness
FEB - 7 2012
# SAFE MEDICAL DEVICES ACT OF 1990
510(k) Summary
| NAME OF FIRM: | Ortho Solutions Limited<br>West Station Buisness Park<br>Spital Road<br>Maldon<br>ESSEX, CM9 6FF<br>United Kingdom |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) FIRM CONTACT: | Al Lippincott<br>Engineering Consulting Services, Inc.<br>3150 E. 200th St.<br>Prior Lake, MN 55372<br>e-mail: allippincott@msn.com |
| DATE: | April 20, 2011 |
| TRADE NAME: | Ortho Solutions - Extremity Fixation Implants for Osteosynthesis |
| COMMON NAME: | Bone Fixation Plate, Bone Fixation Screw, Cannulated Bone Screw &<br>Washer, Memory Staple, Bone Screw, Bone Plate, Compression<br>Screw, Internal Fixation Device (non-spinal), Fixation Staple |
| CLASSIFICATION: | Smooth or threaded metallic bone fixation fastener<br>(per 21CFR888.3040) - Screw, Fixation, Bone |
| | Single/multiple component metallic bone fixation appliance and<br>accessories (per 21CFR888.3030) - Plate, Fixation, Bone;<br>Staple, Fixation, Bone; Washer, Bolt Nut |
| DEVICE PRODUCT CODE: | HRS |
| SUBSEQUENT PRODUCT CODE: | JDR, HWC |
| SUBSTANTIALLY<br>EQUIVALENT DEVICES | 'Foot Plating System' Ascension Orthopedics (K022325, K022324)<br>'Foot Plating System' Wright Medical Tech. (K061808, K100359)<br>'Memo Staple System' Memometal Technologies (K070031)<br>'Memo Staple System' DePuy Orthopedics (K060746, K964226)<br>'Memo Staple System' Biopro (K061798)<br>'Foot Osteotomy System' Biomet/EBI (K092670)<br>'Foot Osteotomy System' TriMed (K050681)<br>'Staples System' Wright Medical Tech. (K043059)<br>'Staples System' Small Bone Innovations (K071479)<br>'Compression Screws System' Wright Medical Tech. (K082320)<br>'TwistFix™ Snap-Off Screws System' Wright Medical Tech.<br>(K050819, K042583)<br>'Extremity Fixation System' Arthrex (K103705, K963118)<br>'Extremity Fixation System' Ascension Orthopedics (K060026)<br>'Extremity Fixation System' Acumed (K963118) |
Page 1 of 5
{1}------------------------------------------------
#### DEVICE DESCRIPTION:
The Ortho Solutions Extremity Fixation Implants for Osteosynthesis consists of Predicate Trauma Implant components commonly found with large companies with orthopedic markets in the United States. These 'small bone extremity trauma implant devices' consist of the following categories:
- 1. Foot Plating System (FPS)
- 2. Memo Staple System
- 3. Foot Osteotomy System (FOS)
- 4. Staples System
- 5. Compression Screws System
- 6. TwistFix™ Snap-Off Screws System
- 7. Extremity Fixation System (EFS)
A brief and concise description of each system is enclosed as follows:
1. Foot Plating System (FPS): The Ortho Solutions Foot Plating System (FPS) is designed to address a variety of indications in rearfoot, midfoot, and forefoot reconstuctive fixation surgery. The overall system is composed of a smaller 2.7mm and larger 3.5mm locking/nonlocking plating category of system(s). The 2.7mm system of three(3) plate designs consisting of a Universal Locking Plate 2.7 (ULP-2.7), a MetaTarso Phalangeal Plate (MPT), and a Open Wedge Locking Plate (OWL) - all utilized with either a locking or non-locking 2.7mm Screw in various lengths. The 3.5mm system of seven(7) plate designs consisting of a Universal Locking Plate 3.5 (ULP-3.5), a Lapidus Plate(LAP), a Arthrodesis Wedge Plate (AWP), a Rearfoot Reconstuction Plate (RRP), a Tarsal Fusion Plate (TFP), a Calcaneal Step Plate (CSP), and a Calcaneal Locking Plate (CLP) - all utilized with either a locking or non-locking 3.5mm Screw in various lengths. A 3.5mm Cannulated Screw System in various lengths with Washer is also included in the FPS System. Associated instrumentation such as trial plates, disposable drills & wires/guide wires, and ancillary instrumentation is available. All plates, screws, and washer are manufactured from high strength 6-4 Alloyed Titanium to ASTM F136. All plates, screws and washer are offered 'sterile' to the customer.
2. Memo Staple System: The Ortho Solutions Memo Staple System consists of four(4) staple sizes (at 10mm, 12mm, 20mm, and 25mm) manufactured from a Nickel-Titanium Nitinol material according to ASTM 2063 for foot reconstruction fixation/arthrodesis procedures. Only the room temperature superelastic properties of the nitinol material with the use of 'spreader/expander instrumention' is utilized in the surgical procedure to enact compression.
Page 2 of 5
Section XII
{2}------------------------------------------------
Associated instrumentation such as templates, spreaders, expanders, guides, disposable drills and ancillary instrumentation is available. All Memo Staples are offered 'sterile' to the customer.
3. Foot Osteotomy System (FOS): The Ortho Solutions Foot Osteotomy System(FOS) consists of a Nail Plate/Staple with Screws in various lengths and Cannulated/Non-cannulated Compression Screws in various lengths for performing metatarsal/phalangeal osteotomies and/or small bone fixation reconstruction. All components are manufactured from CP Titanium to ASTM F67 and high strength 6-4 Alloyed Titanium to ASTM F136. Associated instrumentation such as guide wires, disposable drills and ancillary instrumentation is available. All staple plate/screw and compression screws are offered 'sterile' to the customer.
4. Staples System: The Ortho Solutions Staples System is composed of a Varisation Staple (smooth), a Compression Varisation Staple (tapered and ribbed), a Compression Staple (in 11mm, 12mm, 13mm, 15mm, and 20mm and width sizes), and a Blount Staple (in 22mm and width sizes) that are barbed/ribbed, angled, and smooth staple components manufactured from high strength 316LVM surgical grade stainless steel material according to ASTM F138. These staple systems are utilized for fixation of bone fractures or bone reconstruction and for fixation of soft tissues. A 'spreader instrument' is used to deform the compression staple diamond shape for bone compression fixation. Associated instrumentation such as spreaders and expanders with disposable drills and ancillary instrumentation is available. All Staples are offered 'sterile' to the customer.
5. Compression Screws System: The Ortho Solutions Compression Screws System consists of Cannulated Meta Compression and Cannulated Compression Screws in various lengths at a diameter of 3.0mm and is used for bone reconstruction in small bone extremities. All compression screws are manufactured from high strength alloyed Titanium to ASTM F136. Associated instumentation such as guide wires and disposable drills and ancillary instrumentation is available. All compression screws are offered 'sterile' to the customer.
6. TwistFix™ Snap-Off Screws System: The Ortho Solutions TwistFix™ Snap-Off Screws System comes in a 2.0mm and 2.5mm diameter with various lengths. All TwistFix™ Snap-Off Screws are manufactured from high strength 6-4 Alloyed Titanium to ASTM F136 and CP Titanium to ASTM F67 and are utilized in metatarsal/ phalangeal small bone fixation/osteotomy procedures. Ancillary instrumentaion with disposable drills is available. All TwistFix™ Snap-Off screws are offered 'sterile' to the customer.
{3}------------------------------------------------
K11678(415)
#### Ortho Solutions Limited - 510(k) Summary: DEVICE DESCRIPTION CONTINUED:
7. Extremity Fixation System (EFS): The Ortho Solutions Extremity Fixation System (EFS) consists of various Cannulated and Non-cannulated Flat Head low profile Fixation Screws in diameter sizes of 2.0mm, 2.4mm, 2.7mm, and 3.5mm, in various lengths and are manufactured from high strength 6-4 Alloyed Titanium material to ASTM F136. These flat head screws allow for a low-profile head-on the cortex bone surface for reduction of tissue irritation. Associated instrumentation such as guide wires, disposable drills and ancillary instrumentation is available. All Flat Head Screws are offered 'sterile' to the customer.
INTENDED USE: The intended use of the Ortho Solutions Extremity Fixation Implants for Osteosymthesis System of bundled fixation device(s) is to draw two or more aligned small bone fragments together to facilitate healing. The Ortho Solutions Extremity Fixation Implants for Osteosynthesis System is used in adult patients.
> The Ortho Solutions Foot Plating System (FPS) is indicated for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the feet, ankles, and toes skeleton.
The Ortho Solutions Memo Staple System is indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis. The Memo Staple is also indicated for use in fixation of soft tissue to bone such as anterior cruciate reconstruction.
The Ortho Solutions Foot Osteotomy System (FOS) is indicated for alignment and stabilization of small bone fractures including: fixation of small bones, such as those in the foot and ankle for treatment of fractures, non-unions, or mal-unions; ligament reconstructions; osteochondritis dissecans; arthrodesis of the foot and ankle; small bone osteotomies, including first metatarsal head osteotomy and metatarsal osteotomies.
The Ortho Solutions Staples System is indicated for fixation of bone fractures or bone reconstruction. The staples are also indicated for use in fixation of soft tissue to bone.
The Ortho Solutions Compression Screws System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation of bones appropriate for the size of the device.
The Ortho Solutions TwistFix™ Snap-Off Screws System is indicated for fixation of bone fractures or for bone reconstruction. Examples include: 1) Fixation of small bone fragments, 2) Weil Osteotomy, 3) Mono-cortical fixation, and 4) Osteotomies and fractures fixation in the foot and hand.
Page 4 of 5 Section XII
{4}------------------------------------------------
Ortho Solutions Limited - 510(k) Summary: INTENDED USE CONTINUED:
> The Ortho Solutions Extremity Fixation System (EFS) of a low profile head screw is indicated for use in fixation of small bone fractures, nonunions, arthrodesis, and osteotomies of the small bones in the hand or foot.
#### The Ortho Solutions Extemity Fixation Implants for Osteosynthesis System is not intended for spinal use.
EQUIVALENCE: The Ortho Solutions Extremity Fixation Implants for Osteosynthesis is substantially equivalent to predicate systems from many orthopedic companies (as listed).
SUMMARY OF TECH-The Ortho Solutions Extremity Fixation Implants for Osteosynthesis NOLOGICAL CHARare identical in Material. Geometry Design/Markings, and Indications ACTERISTICS. to many predicate system marketed and sold in the U.S.
#### SUMMARY OF SAFETY AND EFFECTIVENESS:
The Ortho Solutions Extremity Fixation Implants for Osteosynthesis is shown to be safe and effective for use as 'sterile' and for single-use in a surgical setting.
Image /page/4/Picture/8 description: The image shows the text "Page 5 of 5" in a bold, slightly distressed font. The words are arranged on a single line, with "Page" on the left, followed by "5 of 5". The text appears to be part of a document or report, indicating that it is the fifth page of a five-page document.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains an image of an eagle with its wings spread. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are written around the edge of the circle. The eagle is facing to the right and has its head turned to the left.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ortho Solutions, Limited % Engineering Consulting Services, Incorporated Mr. Al Lippincott U.S. Agent and Official Correspondent to Ortho Solutions, Limited 3150 East 200th Street Prior Lake, Minnesota 55372
Re: K111678
Trade/Device Name: Ortho Solutions Extremity Fixation Implants for Osteosynthesis Regulation Number: 21 CFR 888.3030 Regulation Name: Single/ multiple component metallic bone fixation appliance and accessories Regulatory Class: Class II Product Code: HRS, HWC, JDR
Dated: January 18, 2012 Received: January 24, 2012
Dear Mr. Lippincott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
FEB - 7 2012
{6}------------------------------------------------
#### Page 2 - Mr. Al Lippincott
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
· Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the text "ORTHO (S) SOLUTIONS". The letter "S" is inside of a circle. The text is in a sans-serif font and is black.
West Station Business Park, Spital Road, Maldon, Essex CM9 6FF Tel: 0870 7777515 Fax: 0870 7777525 Email: sales@orthosol.com www.orthosolutions.com
# Indications for Use
## 510(k) NUMBER: K111678
#### DEVICE NAME: Ortho Solutions Extremity Fixation Implants for Osteosynthesis
# INDICATIONS FOR USE:
The intended use of the Ortho Solutions Extremity Fixation_ Implants for Osteosynthesis System of bundled fixation device(s) is to draw two or more aligned small bone fragments together to facilitate healing. The Ortho Solutions Extremity Fixation Implants for Osteosynthesis System are used in adult patients.
The Ortho Solutions Foot Plating System (FPS) is indicated for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of the feet, ankles, and toes skeleton.
The Ortho Solutions Memo Staple System is indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis. The Memo Staple is also indicated for use in fixation of soft tissue to bone such as anterior cruciate reconstruction.
The Ortho Solutions Foot Osteotomy System (FOS) is indicated for alignment and stabilization of small bone fractures including: fixation of small bones, such as those in the foot, ankle, and for treatment of fractures, non-unions, or mal-unions, ligament reconstructions; fraotares, more anthrodesis of the foot and ankle; small bone osteotomies, including first metatarsal head osteotomy and metatarsal osteotomies.
{8}------------------------------------------------
ORTHO SOLUTIONS
West Station Business Park, Spital Road, Maldon, Essex CM9 6FF Tel: 0870 7777515 Fax: 0870 7777525 Email: sales@orthosol.com www.orthosolutions.com
### Indications for Use Continued:
The Ortho Solutions Staples System are indicated for fixation of bone The Ortho Goldtions or bone reconstruction. The staples are also indicated for use in fixation of soft tissue to bone.
The Ortho Solutions Compression Screws System are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation of bones appropriate for the size of the device.
The Ortho Solutions TwistFix™ Snap-Off Screws System are indicated for fixation of bone fractures or for bone reconstruction. Examples include: 1) Fixation of small bone fragments, 2) Weil Osteotomy, 3) Mono-cortical fixation, and 4) Osteotomies and fractures fixation in the foot and hand.
The Ortho Solutions Extremity Fixation System (EFS) of a low profile head screw is indicated for use in fixation of small bone fractures, nonunions, arthrodesis, and osteotomies of the small bones in the hand or foot.
The Ortho Solutions Extremity Fixation Implants for Osteosynthesis System is not intended for spinal use.
Over-The-Counter-Use AND/OR x Prescription Use (21 CFR 801 Subpart C) (Per 21 CFR 801 Subpart D) (Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
H.H. for
---
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
**Page | 2 of 2**
KIll6781222
Registered Office: West Station Business Park, Spital Road, Maldon, Essex CM9 6FF Registered in England No: 4311933
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.