K162610 · Nexa Medical , Ltd. · OMP · Jun 8, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162610
Device Name
NEXA Negative Pressure Wound Therapy System
Applicant
Nexa Medical , Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
Jun 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NEXA NPWT system is intended for patients who may benefit from the application of negative pressure to the wound to promote wound healing through the removal of excess exudate, infectious material and tissue debris. It is intended for use in long-term care and acute settings only when prescribed by a Health Care Professional. Appropriate wound types include: - Chronic wounds - Pressure ulcers - Diabetic foot ulcers - Venous leg ulcers
Device Story
NEXA NPWT is a portable negative pressure wound therapy system for acute and long-term care settings. It consists of a battery-powered peristaltic pump, a flexible fluid container, and a sterile dressing kit. The system applies negative pressure to a sealed wound site to remove exudate, infectious material, and tissue debris into the container. Pressure is regulated via mechanical valves, with settings of -125 mmHg and -75 mmHg selected by a healthcare professional. The device includes an integrated pressure relief valve in the tubing to prevent over-pressurization. The system provides alerts for fault conditions. Healthcare providers use the device to manage wound healing; the output (exudate removal) facilitates tissue repair and reduces wound debris. The device is designed for an 8-week service life.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including fluid removal performance across viscosities, pressure generation verification, battery life validation, and alert system functionality. Biocompatibility was evaluated per ISO 10993-1. Electrical safety and EMC testing were performed per IEC 60601-1 and related standards.
Technological Characteristics
DC-powered peristaltic pump; mechanical pressure control valves. Materials: polyurethane reticulated foam, transparent polyurethane film, phthalate-free PVC tubing. Dimensions: 4.1" x 4.5" x 2.5". Power: battery and mains. Connectivity: none. Sterilization: not specified. Software: firmware drives pump motor and fault alerts.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to promote healing via removal of exudate, infectious material, and tissue debris. Applicable for chronic wounds, pressure ulcers, diabetic foot ulcers, and venous leg ulcers. Intended for use in acute and long-term care settings under prescription.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 8, 2017
Nexa Medical Ltd. Keith Heaton Managing Director Suite 5-6, Willow View, Southfields, Church Lane Bournemouth, BH22 8TR GB
Re: K162610
Trade/Device Name: Nexa Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: May 4, 2017 Received: May 8, 2017
Dear Keith Heaton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K162610
Device Name NEXA NPWT
Indications for Use (Describe)
The NEXA NPWT system is intended for patients who may benefit from the application of negative pressure to the wound to promote wound healing through the removal of excess exudate, infectious material and tissue debris.
It is intended for use in long-term care and acute settings only when prescribed by a Health Care Professional.
Appropriate wound types include:
- Chronic wounds
- · Pressure ulcers
- · Diabetic foot ulcers
- · Venous leg ulcers
| Type of Use (Select one or both, as applicable) | <table><tr><td> Prescription Use (Part 21 CFR 801 Subpart D) </td><td> Over-The-Counter Use (21 CFR 801 Subpart C) </td></tr></table> | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary (in accordance with 21 CFR 807.87(h) and 21 CFR 807.92)
#### Submitter Information [21 CFR 807.929(a)(1)]
#### Submitter's name and address:
Nexa Medical Limited Suite 5-6 Willow View, Southfields, Church Lane, Bournemouth, BH22 8TR UK
FDA Establishment Registration No. To be applied for after marketing clearance is given.
#### Submitter's telephone number and fax number:
Tel: 011 44 1202 238738 Fax: Nexa Medical Ltd. does not have a fax number
#### Contact person:
Mr. Keith Heaton, Managing Director
#### Date this 510(k) summary prepared:
8th June 2017
#### Name of the device [21 CFR 807.92(a)(2)1
Trade/proprietary name of the device: NEXA NPWT System
#### Classification name and number of the device:
FDA Class - II FDA Product Code: OMP FDA Classification Name - Powered suction pump FDA Regulation Number - 21CFR 878.4780
{4}------------------------------------------------
#### Predicate Device [21 CFR 807.92(a)(3)]
#### Legally marketed predicate devices to which substantial equivalence is claimed:
| Device Manufacturer | KCI USA, Inc. (Kinetic Concepts, Inc). San<br>Antonio, TX 78249, US |
|----------------------------------------------------|---------------------------------------------------------------------|
| Device Name | V.A.C.® Via™ NPWT System |
| FDA 510(k) No. | K132741 |
| Clearance given by FDA<br>to market this device on | November 15th 2013 |
| FDA Device Classification | Class II |
| FDA Product Code | OMP |
| FDA Classification Name | Powered suction pump |
| FDA Regulation Number | 21CFR 878.4780 |
#### Device Description [21 CFR 807.92(a)(4)]
The NEXA NPWT System is an integrated negative pressure wound management system for use in acute and extended care settings. The system applies a negative pressure to a sealed wound dressing and promotes wound healing through the removal of exudates into a disposable fluid container.
The system consists of the following key components:
- NEXA Device: A portable device that contains a pump and a 1. rechargeable Battery and is supplied with a Power Supply and a Carry Case.
- 2. NEXA Fluid Container Pack: A single use polymeric flexible fluid container that stores the exudate removed from the wound.
- 3. NEXA Dressing: Sterile single use wound dressing components that are in contact with the wound tissue and a means of sealing to the periwound area. Interconnect tubing includes an integrated negative pressure relief valve that limits the maximum pressure able to be applied to the wound and are pre-set and not adjustable by the user.
{5}------------------------------------------------
#### Device Intended Use [21 CFR 807.92(a)(5)]
The NEXA NPWT system is intended for patients who may benefit from the application of negative pressure to the wound to promote wound healing through the removal of excess exudate, infectious material and tissue debris.
It is intended for use in long-term care and acute settings only when prescribed by a Health Care Professional.
Appropriate wound types include:
- Chronic wounds .
- Pressure ulcers •
- Diabetic foot ulcers
- Venous leg ulcers
{6}------------------------------------------------
#
| Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)] | | |
|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Candidate Device<br>NEXA NPWT System | Predicate Device<br>V.A.C.® Via™ NPWT System |
| Operating Principle | DC powered pump removes fluid from the wound and generates a negative pressure at the wound site which is controlled by mechanical valves | DC powered pump removes fluid from the wound and generates a negative pressure at the wound site which is controlled electronically |
| Pump<br>Software | DC powered, peristaltic pump<br>Software code drives the pump motor and alerts user of a fault condition | DC powered, Diaphragm pump<br>Software drives the pump motor, alerts user of a fault condition and controls negative pressure |
| Pressure control | Mechanical pressure control valves | Electronic control system |
| Pressure range | -125 mmHg and -75 mmHg<br>Selected by prescriber | -125mmHg and -75mmHg<br>User selectable |
| Power | Battery and mains connected power supply | Battery and mains connected power supply |
| Housing Dimensions | Width 4.1 inches<br>Length 4.5 inches<br>Depth 2.5 inches | Width 3.5 inches<br>Length 8.4 inches<br>Depth 2.8 inches |
| Wound fluid container | Flexible fluid container 125ml capacity | Rigid polymeric canister 250ml capacity |
| Expected service life | 8 weeks | 7.5 days |
| Dressing System<br>components | Foam: Charcoal coloured polyurethane reticulated foam.<br>Drape: Transparent polyurethane film with adhesive backing.<br>Interconnect tubing: Clear PVC single lumen tubing Phthalate free | Foam: Charcoal coloured polyurethane reticulated foam.<br>Drape: Transparent polyurethane film with adhesive backing.<br>Interconnect tubing: Clear PVC multi lumen tubing contains Phthalates |
{7}------------------------------------------------
#### Performance data [21 CFR 807.92(b)]
#### Summary of Non Clinical tests undertaken [21 CFR 807.92(b)(1)]
#### Bench Testing
In order to demonstrate the NEXA NPWT System is substantially equivalent to the Predicate Device V.A.C.® Via™ NPWT System, the following bench testing was undertaken:
- candidate device effectively removes wound fluid in a clinically relevant setting across a range of viscosities and positions
- candidate device generates the stated negative pressure at the wound ● site
- device alert and warning system works as intended
- duration of the battery life is as stated in the Instructions ●
- device performs as intended over its stated life cycle ●
- components with a stated shelf life i.e. dressing components perform ● as intended over the stated life
- comparative performance testing against the predicate
Bench performance testing was conducted using simulated wound exudates using wound models over a range of clinically relevant conditions.
#### Bio-compatibility testing
- Of the items included within the NEXA NPWT System, biocompatibility is . only applicable to the NPWT Wound Dressing Kits.
- . An expert review was completed by an independent test facility that was certified according to EN 45011, ISO 17025, GLP and GMP standards. The purpose of this review was to evaluate the biocompatibility test requirements of the NEXA Medical NPWT Wound Dressing Kits against ISO 10993-1 requirements and FDA guidelines taking into account existing test data, history of use and application of the device.
- Bio-compatibility testing was undertaken with the final materials subjected ● to production processes.
{8}------------------------------------------------
#### Product testing: Electrical Safety, Usability, EMC & Packaging
- IEC 60601-1 (ed. 3.1) General requirements for basic safety and . essential performance (3rd Edition)
- IEC 60601-1-11(ed.1) Requirements for medical electrical equipment ● and medical electrical systems used in the home healthcare environment
- IEC 60601-1-6(ed.3) General requirements for basic safety and . essential performance. Collateral standard. Usability
- . IEC 62366 (ed. 1) Medical Devices - Application of usability engineering to medical devices
- EN 60601-1-2:2007 General requirements for safety Collateral ● standard: Electromagnetic compatibility - requirements and tests
- ISTA 2 (2011), Vibration and shock for packaged products 150lb ● (68kg) or less
- BS EN ISO 3744, Acoustic testing ●
#### Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]
Not applicable
{9}------------------------------------------------
#### Conclusions drawn regarding substantial equivalence [21 CFR 807.92(b)(3)]
The results of the nonclinical tests performed demonstrate the performance of the NEXA NPWT is substantially equivalent to the performance of the above named predicate device.
To further establish substantial equivalence to the predicate device, NEXA Medical Limited evaluated the indications for use, materials, technology, and product specifications for the components of the product. As a result of this analysis along with performance testing, NEXA Medical has demonstrated the NEXA NPWT is substantially equivalent to the above named predicate device for its indications for use.
This concludes the 510(k) Summary.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.