K132741 · Kci USA, Inc. · OMP · Nov 15, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132741
Device Name
V.A.C. VIA NEGATIVE PRESSURE WOUND THERAPY SYSTEM
Applicant
Kci USA, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Nov 15, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The V.A.C. ® Via™ Therapy System is an integrated wound management system for use in acute, extended and home care settings. When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudates and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure, or venous insufficiency), flaps and grafts. When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy
Device Story
V.A.C. Via is a portable negative pressure wound therapy (NPWT) system. It applies controlled negative pressure to wound sites to promote healing by removing exudates, reducing edema, and stimulating granulation tissue. The system consists of a therapy unit, dressing components (Granufoam, interface pad, tubing, adhesive drape), and a canister. The modified device utilizes a single brushed DC-powered diaphragm pump controlled by software, replacing the dual piezoelectric pump design of the predicate. Used in acute, extended, and home care settings; operated by clinicians or patients. The system maintains a closed environment over the wound or incision. Healthcare providers monitor the wound status to guide therapy. Benefits include accelerated wound bed preparation and management of surgical incision drainage.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included performance verification of negative pressure delivery, software validation, electrical safety, electromagnetic compatibility, and battery life testing to ensure the modified pump design meets design specifications.
Technological Characteristics
NPWT system featuring a single brushed DC-powered diaphragm pump. System includes Granufoam dressing, interface pad, tubing, and adhesive drape. Dimensions: 3.5" W x 6.4" L x 2.8" D. Software-controlled pressure delivery. Battery-powered. Sterilization method not specified.
Indications for Use
Indicated for patients in acute, extended, or home care settings requiring wound management. Applicable to open wounds (chronic, acute, traumatic, subacute, dehisced, partial-thickness burns, ulcers, flaps, grafts) and closed surgical incisions with drainage.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
V.A.C. ® Via™ Negative Pressure Wound Therapy System (K093526)
V.A.C. ® Via™ Negative Pressure Wound Therapy System (K120033)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for KCI. The logo consists of a series of curved lines on the left, followed by the letters "KCI" in bold, sans-serif font. There is a period after the "I".
NOW 15 2013
## 510(k) SUMMARY
# V.A.C. ® V.A.C. ® Via™ | V.A.C. ® Via™
Negative Pressure Wound Therapy System
| Submitter Information [21 CFR 807.929(a)(1)] | |
|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | KCI USA, Inc. (Kinetic Concepts, Inc.) |
| Address | 6203 Farinon Drive<br>San Antonio, TX 78249 |
| Phone number | 210-515-4059 |
| Fax number | 210-255-6727 |
| Establishment<br>Registration Number | 1625774 |
| Name of contact person | Melanie Avila |
| Date prepared | 08/23/2013 |
| Name of the device [21 CFR 807.92(a)(2)] | |
| Trade or proprietary<br>name | V.A.C. ® Via™ Negative Pressure Wound Therapy System |
| Common or usual name | Negative Pressure Wound Therapy System |
| Classification name | Negative Pressure Wound Therapy Powered Suction Pump |
| Regulation | 878.4780 |
| Product Code(s) | OMP |
| Legally marketed<br>device(s) to which<br>equivalence is claimed<br>[21 CFR 807.92(a)(3)] | V.A.C. ® Via™ Negative Pressure Wound Therapy System, originally cleared<br>under 510(k) K093526 and K120033 |
| Device description<br>[21 CFR 807.92(a)(4)] | Negative pressure wound therapy system |
| Indications for use<br>[21 CFR 807.92(a)(5)] | The V.A.C. ® Via™ Therapy System is an integrated wound management<br>system for use in acute, extended and home care settings. When used on<br>open wounds, it is intended to create an environment that promotes wound<br>healing by secondary or tertiary (delayed primary) intention by preparing the<br>wound bed for closure, reducing edema, promoting granulation tissue<br>formation and perfusion, and by removing exudates and infectious material.<br>Open wound types include: chronic, acute, traumatic, subacute and dehisced<br>wounds, partial-thickness burns, ulcers (such as diabetic, pressure, or venous<br>insufficiency), flaps and grafts.<br>When used on closed surgical incisions, it is intended to manage the<br>environment of surgical incisions that continue to drain following sutured or<br>stapled closure by maintaining a closed environment and removing exudates<br>via the application of negative pressure wound therapy |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for KCI. The logo consists of a graphic on the left and the letters "KCI" on the right. The graphic on the left is a series of curved lines that are stacked on top of each other. The letters "KCI" are in a bold, sans-serif font.
## 510(k) SUMMARY
## V.A.C. ® Via™
## Negative Pressure Wound Therapy System
| Summary of the technological characteristics of the device compared to the predicate device<br>[21 CFR 807.92(a)(6)] | | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|--------|-------|---------------------------------------------------------------------------------------------|--------|-------|
| Feature | V.A.C.® Via™ NPWT System<br>(Modified Device) | | | V.A.C.® Via™ NPWT System<br>(Predicate ) | | |
| Therapy unit | Same as predicate | | | Software controlled pump for delivery of<br>negative pressure wound therapy | | |
| • Operating Principle | Same as predicate | | | Software controlled pump for delivery of<br>negative pressure wound therapy | | |
| • Pump Type | One brushed, DC powered, single<br>diaphragm pump | | | Two piezoelectric disc pumps | | |
| • Software | Software code drives the<br>operation of one conventional<br>diaphragm pump | | | Software code drives the operation of two<br>piezoelectric pumps | | |
| • Housing Dimensions<br>(inches) | Width | Length | Depth | Width | Length | Depth |
| | 3.5 | 6.4 | 2.8 | 3.5 | 6.4 | 2.4 |
| Therapy unit options | Same as predicate | | | Selectable negative pressure at either<br>-125mmHg or -75mmHg | | |
| Dressing system<br>components | Same as predicate | | | System consists of Granufoam ™<br>Dressing, interface pad and tubing, and<br>adhesive drape | | |
| Performance Data [21 CFR 807.92(b)] | | | | | | |
| Summary of non-clinical tests conducted for determination of substantial equivalence<br>[21 CFR 807.92(b)(1)] | | | | | | |
| The V.A.C.® Via™ NPWT System was evaluated to ensure conformance to the design specifications. | | | | | | |
| The following bench tests were conducted: | | | | | | |
| • Performance testing to confirm therapy unit delivers negative pressure therapy | | | | | | |
| • Software verification and validation testing to confirm that therapy unit performs as designed | | | | | | |
| • Testing to confirm device meets all applicable electrical safety and electromagnetic compatibility standards. | | | | | | |
| • Battery life testing to demonstrate that the battery will supply adequate power to provide the appropriate negative pressure to the wound. | | | | | | |
Information [21 CFR 807.92(b)(2)] Not applicable
Conclusions drawn [21 CFR 807.92(b)(3)]
Testing indicates that the V.A.C. ® Via™ NPWT System is substantially equivalent in terms of both
indications for use and fundamental scientific technology to the predicate p
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
· November 15, 2013
KCI USA, Incorporated Ms. Melanie Avila Regulatory Affairs Project Manager 6203 Farinon Drive San Antonio, Texas 78249
Re: K132741
Trade/Device Name: V.A.C. Via" Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: October 14, 2013 Received: October 17, 2013
Dear Ms. Avila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Ms. Melanie Avila
... ... ... ..
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
for Mark N. Melkerson
Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
#### 510(k) Number (if known) K 132741
#### Device Name
V.A.C. Via Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
The V.A.C. @ Via™ Therapy System is an integrated wound management system for use in acute, excended and home care settings. When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation lissue formation and perfusion, and by removing exudates and infectious material. Open wound types include: chronic, acute, traumatic, subacute and debissed wounds, partial-thickness burns, ulcers (such as diabetic, pressure, or venous insufficiency), flaps and grafts.
When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative wound therepy.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Rediological Health (CDRH) (Signature)
## Jivoung Dang -S
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.