K152332 · Omm Imports Inc Dba Zero Gravity · OHS · Feb 25, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152332
Device Name
Perfectio LED infrared device
Applicant
Omm Imports Inc Dba Zero Gravity
Product Code
OHS · General, Plastic Surgery
Decision Date
Feb 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Perfectio LED infrared device is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.
Device Story
Perfectio LED infrared device is a battery-operated, handheld LED device for home-use treatment of periorbital wrinkles. Device emits light energy at red (633 ±5nm) and infrared (830 ±5nm) wavelengths using 25 LEDs distributed over 17cm². Device includes a handpiece, charging base, and AC adapter. User operates device by applying light energy to treatment zones for 15-20 minutes per session. No complex algorithms or software-based analysis; device provides uniform light distribution. Benefits include non-invasive treatment of periorbital lines and wrinkles.
Clinical Evidence
No clinical data included. Evidence consists of bench testing for electrical safety, electromagnetic compatibility, photobiological safety, and biocompatibility. A usability/label comprehension study with 35 participants demonstrated 100% success in device setup, operation, and cleaning, and 95% success in questionnaire accuracy.
Technological Characteristics
Handheld LED device; 25 LEDs; wavelengths 633 ±5nm and 830 ±5nm; constant waveform. Materials: ABS (body), Stainless steel 304 (head). Power: 100~240V AC adapter, 2x3.7V 900 mAh Lithium battery. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471. Biocompatibility: ISO 10993-1, -5, -10. No software-based algorithm.
Indications for Use
Indicated for individuals with periorbital wrinkles for use in dermatology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are black and are connected by a flowing line at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 25, 2016
Omm Imports Inc Dba Zero Gravity % Ms. Susan Anthoney-DeWet Aegis Regulatory, Inc. 2424 Dempster Drive Coralville, IA 52241
Re: K152332
Trade/Device Name: Perfectio LED Infrared Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHS Dated: January 27, 2015 Received: January 27, 2015
Dear Ms. Anthoney-DeWet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Post-market Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152332
Device Name Perfectio LED infrared Device
Indications for Use (Describe)
Perfectio LED infrared device is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5 510(k) Summary
[As required by 21 CFR 807.92]
#### 1. Submission Information:
| 510(k) Number: | K152332 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | Aug 10, 2015 |
| Type of 510(k) Submission: | Traditional |
| Basis for 510(k) Submission: | New device |
| Submitter/Manufacturer: | OMM IMPORTS INC DBA ZERO GRAVITY;<br>1945 S OCEAN DR APT 509 HALLANDLE BEACH BLVD,FLORIDA<br>33009, US |
| Contact: | Doris Dong (Consultant)<br>Shanghai CV Technology Co., Ltd.<br>Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, 201600 China<br>E-mail: doris_d@126.com<br>Tel: 86 21-31261348 |
#### 2. Device Description:
| Proprietary Name: | Perfectio LED infrared device |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Light Emitting Diode (LED) device |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Product Code: | OHS |
| Device Class: | II |
| Regulation Number: | 21 CFR 878.4810 |
| Review Panel: | General & Plastic Surgery |
| Indications for use: | Perfectio LED infrared device is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles. |
| Device Description: | Perfectio LED infrared device is a battery operated device that uses low power light spectrum at red and infrared LED, at wavelength of 633 ±5nm, 830 ±5nm emitting optical power in a uniform distribution with no hot spots. The device is composed of a handpiece for delivery of light energy, base unit for charging and storage when not in use, and A.C. charging adapter. It is a hand held light emitting diode (LED) device for the treatment of periorbital wrinkles designed for home-use. |
#### 3. Predicate Device Identification
510(k) Number: K110301 Clearing date: August 19, 2011 Product Name: Silk'n FX Manufacturer: Home Skinovations Ltd.
{4}------------------------------------------------
#### OMM IMPORTS INC DBA ZERO GRAVITY; 1945 S OCEAN DR APT 509 HALLANDLE BEACH BLVD,FLORIDA 33009
## 4. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device conforms with the following standards:
* AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.
* IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests. * IEC 60601-1-11:2010, Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
* IEC 62471:2006, Photobiological Safety of Lamps and Lamp Systems
The patient contact materials in Perfectio LED infrared device are the body housing material of ABS and the head housing material of Stainless steel 304. Both the two materials were tested and found to meet the biocompatibility standards of:
* ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
* ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity, and * ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and
delayed-type hypersensitivity
A Usability/Label Comprehension Study was conducted with 35 participants.
The results of the study showed that (100%) were able to:
-Correctly self-select as being an appropriate user of the device.
-Correctly demonstrate how to set up the device, perform the Light Sensitivity Test, operate the device (apply Tx), and clean the device.
And that (95%) of participants were able to correctly answer each question for the Questionnaire portion of the Study.
## 5. Substantial Equivalent Based on Assessment of Clinical Performance Data:
Clinical data was not included in this submission.
| | New Device | Predicate Device |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K152332 | K110301 |
| Product Code: | OHS | OHS |
| Proprietary Name: | Perfectio LED infrared device | Silk'n FX |
| Manufacturer: | OMM IMPORTS INC DBA ZERO<br>GRAVITY | Home Skinovations Ltd. |
| Indications for use: | Perfectio LED infrared device is an over<br>the counter device indicated to emit energy<br>in the red and IR region of the spectrum for<br>use in dermatology for the treatment of | Silkn FX is an over the counter device<br>indicated to emit energy in the red and IR<br>region of the spectrum for use in<br>dermatology for the treatment of |
### 6. Substantially Equivalent Comparison Conclusion
{5}------------------------------------------------
#### OMM IMPORTS INC DBA ZERO GRAVITY; 1945 S OCEAN DR APT 509 HALLANDLE BEACH BLVD,FLORIDA 33009
| | periorbital wrinkles. | periorbital wrinkles. |
|--------------------------|------------------------------------------|------------------------------------------|
| Handheld | Yes | Yes |
| Materials | ABS and stainless steel | ABS and stainless steel |
| Wavelengths | 633 ±5nm, 830 ±5nm | 633 ±5nm, 830 ±5nm |
| Light source | Light emitting diode(LED) | Light emitting diode(LED) |
| Waveform | Constant | Constant |
| Energy Source | 25 LEDs over 17cm² | 24 LEDs over 12 cm² |
| Power supply | Adaptor:100~240V AC 50/60Hz 0.2A | Adaptor:100~240V AC 50/60Hz 0.4A |
| | Lithium battery: 2x3.7V, 900 mAh | Lithium battery: 2x3.7V, 750±50mAh |
| Initial treatment course | For the first month (4 weeks), treatment | For the first month (4 weeks), treatment |
| | should be performed 3 times a week for | should be performed 3 times a week for |
| | 15-20 minutes each time.(5-7 minutes on | 15-20 minutes each time.(5-7 minutes on |
| | each treatment zone). | each treatment zone). |
| Maintenance regime | Once a week for 15-20 minutes | Once a week for 15-20 minutes |
| Target Population | Individuals with periorbital lines and | Individuals with periorbital lines and |
| | wrinkles | wrinkles |
| Location for use | OTC | OTC |
| Standard meet | IEC60601-1 | IEC60601-1 |
| | IEC60601-1-2 | IEC60601-1-2 |
| | IEC60601-1-11 | IEC62471 |
| | IEC62471 | |
### The Conclusions:
Taking into consideration the table for substantial equivalence, an analysis of safety, indications, intended uses, performance, and technological properties, the proposed device raises no new issues of safety or effectiveness and has been found to be substantially equivalent to the predicate device.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.