K162468 · Radformation · IYE · Dec 1, 2016 · Radiology
Device Facts
Record ID
K162468
Device Name
ClearCheck
Applicant
Radformation
Product Code
IYE · Radiology
Decision Date
Dec 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ClearCheck is intended for quality assessment of radiotherapy treatment plans.
Device Story
ClearCheck is a software plugin for Varian Eclipse Treatment Planning System (TPS) used by medical physicists, dosimetrists, and radiation oncologists to evaluate radiotherapy treatment plans. It ingests treatment plan data from Eclipse TPS; performs automated calculations based on user-defined dose constraints and plan check parameters; and presents results in a numerical format for clinical review. The device consists of a standalone Windows application for administrative settings and a dynamic link library (dll) plugin for plan evaluation. By providing a structured quality assessment, it assists clinicians in verifying plan quality before treatment delivery, potentially improving the accuracy and safety of radiation therapy for patients with malignant or benign diseases.
Clinical Evidence
No clinical trials were performed. Substantial equivalence is supported by bench-level verification and validation testing, including pass/fail criteria to ensure software functionality and hazard analysis.
Technological Characteristics
Pure software device; runs as a dynamic link library (dll) plugin on Varian Eclipse TPS and as a standalone Windows application. Compatible with Windows 7, 8, 10, and Windows Server 2008/2012. Requires 2.4+ GHz multi-core processor. Connectivity: interfaces with Eclipse TPS data. No hardware components.
Indications for Use
Indicated for quality assessment of radiotherapy treatment plans for patients with malignant or benign diseases requiring radiation therapy. Intended for use by medical physicists, medical dosimetrists, and radiation oncologists.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a caduceus, a traditional symbol of medicine, but with three figures instead of one.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2016
Radformation % Mr. Parul Chansoria Regulatory Consultant Elexes 6494 Tralee Village Drive DR DUBLIN CA 94568
Re: K162468 Trade/Device Name: ClearCheck Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: October 28, 2016 Received: October 28, 2016
Dear Mr. Chansoria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K162468
Device Name ClearCheck
Indications for Use (Describe)
ClearCheck is intended for quality assessment of radiotherapy treatment plans.
Type of Use (Select one or both, as applicable)
Radformation 510(k) Submission
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Section 5. 510(k) Summary
{5}------------------------------------------------
This 510(k) Summary has been created per the requirements of the Safe Medical Device Act (SMDA) of 1990, and the content is provided in conformance with 21 CFR Part 807.92.
# 5.1 Submitter's Information
| Table 1 : Submitter's Information | |
|-----------------------------------|------------------------------------------------------|
| Submitter's Name: | Kurt Sysock |
| Company: | Radformation, LLC |
| Address: | 429 East 52nd Street, Unit 4G,<br>New York, NY 10022 |
| Contact Person: | Parul Chansoria<br>Founder, Elexes |
| Phone: | 650-528-2445 |
| Fax: | ---------- |
| Email: | parul@elexes.com |
| Date of Summary<br>Preparation: | August 24th, 2016 |
# 5.2 Device Information
| Table 2: Device Information | |
|-----------------------------|-----------------------------------------------------------------------------------------|
| Trade Name: | ClearCheck |
| Common Name: | Oncology Information System |
| Classification<br>Name: | Class II |
| Classification | Medical charged-particle radiation therapy system, dosimetric<br>quality control system |
| Regulation Number: | 892.5050 |
| Product Code: | IYE |
| Classification<br>Panel: | Radiology |
# 5.3 Predicate Device Information
Model 1216 Quality Reports (K142142)
# 5.4 Device Description
The ClearCheck device (model RADCC) is a software intended to present treatment plans obtained from Eclipse Treatment Planning System (also referred to as Eclipse TPS) of Varian Medical Systems in a user friendly way (numerical form of data) for user approval of the treatment plan. ClearCheck runs as a dynamic link library (dll) plugin to Varian Eclipse.
{6}------------------------------------------------
It is designed to run on the Windows Operating System and generated reports can be viewed on Internet Explorer. ClearCheck performs calculations on the plan obtained from Eclipse TPS (Version 12 (K131891) and Version 13.5 (K141283)) which is a software used by trained medical professionals to design and simulate radiation therapy treatments for malignant or benign diseases.
#### ClearCheck has two components:
- A standalone Windows Operating System executable application that is used for 1. administrative operations to set specified default settings and user settings.
- 2. A plan evaluation application that is a dynamic link library (dll) file that is a plugin to the Varian Medical Systems Eclipse TPS. The plugin is designed to evaluate the quality of an Eclipse treatment plan. Plan quality is based on user specified Dose Constraints and Plan Check Parameters.
## 5.5 Indications for Use
ClearCheck is intended for quality assessment of radiotherapy treatment plans
## 5.6 Technological Characteristics
The ClearCheck (Subject Device) makes use of a Predicate Device, the Model 1216 Quality Reports (142142).
| Table 3 : Substantial Equivalence Table for ClearCheck | | | |
|--------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Parameters | Subject Device:<br>ClearCheck<br>Radformation | Predicate Device:<br>Model 1216 Quality Reports<br>Sun Nuclear Corporation<br>(K142142) | Equivalence |
| Indications for use | ClearCheck is intended for<br>quality assessment of<br>radiotherapy treatment plans | Model 1216 Quality Reports is<br>intended for quality assessment of<br>radiotherapy treatment plans and the<br>radiotherapy treatment planning<br>process. | Equivalent |
| Pure Software<br>Device | Yes | Yes | Equivalent |
| Intended users | Medical physicists, medical<br>dosimetrists, and radiation<br>oncologists | Medical physicists, medical<br>dosimetrists, and radiation<br>oncologists | Equivalent |
| OTC/Rx | Rx | Rx | Equivalent |
#### 5.6.1 ClearCheck vs. Model 1216 Quality Reports (K142142)
{7}------------------------------------------------
| Table 3 : Substantial Equivalence Table for ClearCheck | | | |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Parameters | Subject Device:<br>ClearCheck<br>Radformation | Predicate Device:<br>Model 1216 Quality Reports<br>Sun Nuclear Corporation<br>(K142142) | Equivalence |
| | ClearCheck software runs on a<br>standard PC running a Windows<br>Operating System (OS). The<br>software holds all vital data in<br>memory during each analysis and<br>report generation session. | Model 1216 Quality Reports software<br>runs on a standard PC running a<br>Windows Operating System (OS). The<br>software holds all vital data in memory<br>during each analysis and report<br>generation session. | Equivalent |
| Operating<br>System | Windows 7 ( 32- or 64-bit),<br>Windows 8 (32- or 64-bit),<br>Windows 10 (32- or 64-bit),<br>Windows Server 2008, 2008 RS,<br>and 2012. | Windows 7 (32- or 64-bit), Windows 8<br>(32- or 64- bit), and Windows Server<br>2008, 2008 R2, and 2012. | Equivalent |
| CPU | 2.4+ GHz and Multi-core<br>processors (2+ cores, 4+ threads) | 2.4+ GHz and Multi-core processors (2+<br>cores, 4+ threads) | Equivalent |
| Hard Drive<br>Space | Software components fully<br>installed require only ~3.5MB,<br>but storage requirement for<br>patient data are much larger and<br>vary from clinic-to-clinic. A<br>minimum of 100 GB hard drive<br>is suggested for larger patient<br>template sets. | Software components fully installed<br>require ~20 MB, but storage requirements<br>for patient data and archives are much<br>larger but will vary clinic-to-clinic. A<br>minimum of 900 GB hard drive is<br>suggested with larger drives for DICOM<br>archives. | Differ |
| Display<br>Resolution and<br>Color Depth | 1280 x 1024, 24- or 32- bit color<br>depth. | 1920 x 1080 screen resolution, 24- or 32-<br>bit color depth. | Differ |
#### 5.6.2 Similarities
- The intended use is same for the Subject and the Predicate Device. ●
- Both the Predicate and the Subject Devices are pure software devices meant for ● prescription use
- The Subject and the Predicate Devices are classified under the Product Code IYE ● and the common name "oncology information system"
- The intended user of the Subject Device is the same as the Predicate Device ●
- The Subject and Predicate Devices require similar processors and operating systems. However, ClearCheck is supported by an additional Operating System Windows 10 (32- Or 64-bit)
- . Both Subject and Predicate Devices run on a standard PC with a Windows Operating System (OS). The software devices hold all vital data in memory during each analysis and report generation session.
{8}------------------------------------------------
#### 5.6.3 Differences
- The install size of the Subject Device is only ~3.5MB whereas the install size of . the Predicate Device is 20 MB of storage requirement. ClearCheck recommends 100GB of storage space and Predicate Device recommends 900GB of storage space. This is because ClearCheck only saves constraint templates whereas Model 1216 Quality Reports stores DICOM datasets which are much larger when compared to constraint template files. Thus, ClearCheck requires much less storage. However, these differences don't raise any new questions for safety and effectiveness of the Subject Device w.r.t. the Predicate Device.
- The color depth for both the Predicate and Subject Devices are the same. The resolution required for the Subject Device (1280x1024: D-VHS, HD DVD, Bluray, HDV (miniDV), and the Predicate Device (1920x1080: HDV (miniDV), AVCHD, HD DVD, Blu-ray, HDCAM SR) are different. The Subject Device ClearCheck can work with the small size monitors without any impact on the image quality (1280 x 1024). Hence, this difference doesn't raise any new questions for safety and effectiveness of the Subject Device w.r.t. the Predicate Device.
# 5.7 Performance Data
As with the Predicate Device, no clinical trials were performed for ClearCheck. Verification tests were performed to ensure that the software works as intended and pass/fail criteria were used to verify requirements.
# 5.8 Conclusion
ClearCheck is deemed substantially equivalent to the Predicate Device. Model 1216 Quality Reports (K142142) due to the similarities in intended use and function. Verification and Validation testing and Hazard Analysis demonstrate that ClearCheck is as safe and effective as the Predicate Device. The minor technological differences between ClearCheck and the Predicate Device do not raise any questions on the safety and effectiveness of the Subject Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.