K142142 · Sun Nuclear Corporation · IYE · Dec 5, 2014 · Radiology
Device Facts
Record ID
K142142
Device Name
QUALITY REPORTS
Applicant
Sun Nuclear Corporation
Product Code
IYE · Radiology
Decision Date
Dec 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Model 1216 Quality Reports is intended for quality assessment of radiotherapy treatment plans and the radiotherapy treatment planning process.
Device Story
Model 1216 Quality Reports is a software application for radiotherapy treatment plan quality assessment. It ingests DICOM data (images, RT Plan, RT Structure Set, RT Dose) from Treatment Planning Systems (TPS) or manual user input. Clinicians (physicists, dosimetrists, oncologists) use the software to audit Dose Volume Histograms (DVH), measure plan quality against institutional standards, track performance metrics over time, and generate reports for Electronic Medical Records (EMR). The device facilitates quantitative review of estimated doses and non-dosimetric plan parameters. By providing standardized metrics and benchmarking, the software supports continuous quality improvement in the radiotherapy planning process, helping clinicians ensure treatment plan consistency and safety.
Clinical Evidence
No clinical trials performed. Evidence limited to bench testing using simulated clinical workflows. Verification tests and regression testing were conducted against pass/fail criteria to ensure system design requirements and risk mitigations were met.
Technological Characteristics
Software-based oncology information system running on standard Windows PCs (Windows 7, 8, Server 2008/2012). Requires high-performance hardware (multi-core CPU, 8+ GB RAM) to process voluminous DICOM RT objects. Operates as a standalone application for data management, analysis, and reporting. No specific algorithm architecture (e.g., ML/DL) stated; functions via standardized metric calculation and data auditing.
Indications for Use
Indicated for quality assessment of radiotherapy treatment plans and the radiotherapy treatment planning process in clinical settings by trained clinicians including medical physicists, medical dosimetrists, and radiation oncologists.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 5, 2014
Sun Nuclear Corporation % Mr. James Luker Regulatory Affairs Consultant 2640 Nobility Avenue MELBOURNE FL 32934
Re: K142142
Trade/Device Name: Ouality Reports, Model 1216 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: November 5, 2014 Received: November 10, 2014
Dear Mr. Luker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142142
Device Name Model 1216 Quality Reports
Indications for Use (Describe)
Model 1216 Quality Reports is intended for quality assessment of radiotherapy treatment plans and the radiotherapy treatment planning process.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Sun Nuclear Corporation. The logo features a blue square with a yellow sun inside on the left. To the right of the square, the words "SUN NUCLEAR" are written in blue, with the word "corporation" written in a smaller font below.
# Section 5 - 510(k) Summary
Provided in accordance with 21 CFR 807.92 (c)
#### 1 General Provisions
Date Prepared:
August 2, 2014
Submitted by:
Sun Nuclear Corporation 3275 Suntree Blvd. Melbourne, FL 32940 Ph: 321-259-6862 Fax: 321-259-7979 Web: www.sunnuclear.com
Contact Person:
Jonathan Lee jonathanlee@sunnuclear.com
Ph: 321-259-6862 extension 423
Classification Name:
21 CFR § 892.5050 IYE Medical charged-particle radiation therapy system Class II
Common Name:
Oncology Information system
Proprietary Names:
Model 1216 Quality Reports
Establishment Registration Number:
1038814
Classification:
Regulation Number: 21 CFR 892.5050 Name: Medical charged-particle radiation therapy system, dosimetric quality control system Product code: IYE
Predicate Device(s):
| Model Name: | MOSAIQ Release 2.50 |
|---------------|-----------------------------|
| Common Name: | Oncology Information System |
| 510(k) # | K123230 |
| Manufacturer: | IMPAC Medical Systems, Inc. |
| Submitted: | November 6, 2012 |
{4}------------------------------------------------
Model Name: MOSAIQ Release 2.40 Oncology Information System Common Name: 510(k) # K120067 Manufacturer: IMPAC Medical Systems, Inc. Submitted: January 9, 2012
#### 2 Description and Use:
Model 1216 Quality Reports is a multi-functional information and data management software application intended to be used by trained clinicians who are familiar with radiation therapy, such as medical physicists, medical dosimetrists, and radiation oncologists.
Model 1216 Quality Reports receives DICOM data from Treatment Planning Systems (TPS) which are capable of providing such data. The data may include images, radiation therapy (RT) objects such as RT Plan, RT Structure Set and RT Dose. The user may also capture data through manual input.
The application allows the user to:
- Create reports and charts which may be sent to the institution's Electronic . Medical Record (EMR) system.
- Measure quality with standardized metrics that are based on the institution's ● chosen standards.
- Capture information from the clinical team during pre-treatment planning. ●
- Establish benchmark and progress data to demonstrate continuous improvement.
- Audit the Dose Volume Histogram (DVH) from the TPS.
- . Facilitate quantitative review of the estimated dose via a single DVH metric.
- Customize and save a library of non-dosimetric plan parameters and compare ● them against user defined constraints.
- . Track plan quality scores over time to assess performance and provide the basis for measuring continual improvement.
#### 3 Intended Use Statement:
Model 1216 Quality Reports is intended for quality assessment of radiotherapy treatment plans and the radiotherapy treatment planning process.
#### ব Technological Characteristics
As with the predicate devices, Model 1216 Quality Reports software runs on a standard PC running a Windows Operating System (OS). The complexity of DICOM data and the sheer data volume in images and DICOM RT objects (notably RT Structure Sets and RT Doses) demands high performance computer processing and memory. The software holds all vital data in memory during each analysis and report generation session to ensure high performance to optimize the user experience, and thus there are special requirements for the machine(s) on which the software is installed.
{5}------------------------------------------------
## System Recommendations
- . Operating Systems Supported: Windows 7 (32- or 64-bit), Windows 8 (32- or 64bit), and Windows Server 2008, 2008 R2, and 2012
- I CPU: 2.4+ GHz
- l CPU: Multi-core processor (4+ cores, 8+ threads)
- l Hard drive space: Software components fully installed require only ~20 MB, but storage requirements for voluminous patient data and archives are much larger but will vary clinic-to-clinic. A minimum of 900 GB hard drive is suggested with larger drives for DICIOM archives.
- Memory (RAM): 2+ GB x number of cores (e.g. 8+ GB for 4 cores)
- . Display Resolution: 1920 x 1080 screen resolution, 24- or 32-bit color depth.
#### 5 Performance Data and Comparison with Predicate
As with the predicate device, no clinical trials were performed for Model 1216 Quality Reports. Testing was limited to Bench testing using simulated clinical workflows.
Verification tests were written and executed to ensure that the system was working as designed. Pass/fail criteria were used to verify requirements and to ensure that risk mitigations functioned as intended. Regression tests were performed.
#### 6 Summary
Model 1216 Quality Reports is deemed substantially equivalent to the predicate devices due to the similarities in function, technology, and performance. The intended use, performance testing, safety and effectiveness reviews demonstrate that Quality Reports is as safe, as effective, and performs as well as the predicate device. The minor technological differences between Model 1216 Quality Reports and the predicate devices do not raise new types of safety or effectiveness questions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.