K162352 · Chemence Medical, Inc. · MPN · Dec 22, 2016 · General, Plastic Surgery
Device Facts
Record ID
K162352
Device Name
derma+flex QS High Viscosity Tissue Adhesive
Applicant
Chemence Medical, Inc.
Product Code
MPN · General, Plastic Surgery
Decision Date
Dec 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
derma+flex® QS™ High Viscosity Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incision, including incisions from minimally invasive surgery, and simple, thorough cleansed, trauma-induced lacerations. derma+flex® QS™ High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures.
Device Story
Sterile, liquid topical skin adhesive; monomeric 2-octyl cyanoacrylate formulation with D&C Violet #2 colorant. Supplied in single-use aluminum collapsible tube with dauber or nozzle applicator tips. Applied topically by clinicians to skin edges of surgical or trauma wounds; syrup-like viscosity prevents migration; polymerizes within minutes to hold wound edges closed. Intended to reduce risk of unintended adhesive placement. Benefits include effective wound closure for surgical and trauma applications.
Clinical Evidence
Bench testing only. Performance evaluated via wound closure strength (ASTM F2458-05), tensile strength (ASTM F2258-05), T-peel strength (ASTM F2256-05), lap shear strength (ASTM F2255-05), viscosity, polymerization time, purity, and water content.
Indicated for topical closure of easily approximated skin edges of surgical incisions (including minimally invasive surgery) and simple, cleansed, trauma-induced lacerations. May be used with deep dermal sutures.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2016
Chemence Medical Inc. Kenneth Broadley, Ph.D. Executive Vice President 200 Technology Drive Alpharetta, GA 30005 US
Re: K162352
Trade/Device Name: derma+flex QS High Viscosity Tissue Adhesive Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue adhesive Regulatory Class: Class II Product Code: MPN Dated: November 21, 2016 Received: November 22, 2016
Dear Dr. Broadley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR: Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162352
Device Name derma+flex® QSTM
#### Indications for Use (Describe)
derma+flex® QS ™ High Viscosity Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations.
derma+flex® QS ™ High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures.
Type of Use (Select one or both, as applicable)
| <div>☑ Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|-----------------------------------------------------------|
| <div>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Chemence Medical. The logo consists of a red and blue diamond shape on top, with the word "CHEMENCE" in red and the word "MEDICAL" in blue underneath. The diamond shape is divided into two triangles, with the top triangle being red and the bottom triangle being blue. There is a white plus sign in the red triangle.
# 510(k) Summary (In accordance with 21 CFR 807.92)
# Chemence Medical, Inc. Derma+flex® QS™ High Viscosity Tissue Adhesive
### 1. Submitter
| Submitted by: | Chemence Medical, Inc.<br>200 Technology Drive<br>Alpharetta GA 3005-3926<br>Phone: 844-633-4583<br>Fax: 678-820-3320 |
|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dr. Kenneth N. Broadley<br>Executive Vice President |
| Date of Summary: | December 20th, 2016 |
| 2. Device | |
| Device Trade Name:<br>Common or Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code: | derma+flex® QS™ High Viscosity Tissue Adhesive<br>Topical Skin Adhesive<br>Tissue Adhesive (21 CFR 878.4010)<br>Class II<br>MPN |
| 3. Predicate Devices | |
| Proprietary Name:<br>510(k) Clearance: | derma+flex® QS™<br>K101276 |
| 4. Device Description | |
derma+flex® QS™ High Viscosity Tissue Adhesive is a sterile, liquid topical skin adhesive containing a monomeric (2-octyl cyanoacrylate) formulation and the colorant D&C Violet #2. It is provided in a single use aluminum collapsible tube packaged in a RXM 48gaPET-200LDPE
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Chemence Medical. The logo consists of a red and blue diamond shape with a stylized "C" inside. The word "CHEMENCE" is written in red, and the word "MEDICAL" is written in blue below the diamond shape. The logo is simple and modern, and the colors are eye-catching.
Film/1095B uncoated Tyvek pouch also containing 2 Indothene HD Grade HD50MA 180 applicator tips. The dauber applicator is comprised of a self-puncturing cap and a foam surface, which allows spreading of the adhesive with uniformity.
The nozzle applicator is also a self-puncturing cap with an elongation that enables detailed application of the adhesive. As applied to the skin, the liquid is syrup-like in viscosity and polymerizes within minutes. The increased viscosity of derma+flex® QS™ is intended to reduce the risk of unintended placement of the adhesive during application due to migration of the liquid adhesive from the wound site.
## 5. Intended Use:
derma+flex® QS™ High Viscosity Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incision, including incisions from minimally invasive surgery, and simple, thorough cleansed, trauma-induced lacerations. derma+flex® QS™ High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures.
# 6. Comparison of Technological Characteristics with Predicate
derma+flex® QS™ High Viscosity Tissue Adhesive is practically identical to the predicate with regards to Indication for use, formulation, technology, target population, intended application, mechanism of action and performance at achieving their intended use. Both devices present that same main chemical characteristics (2-Octyl Cyanoacrylate) and the same formulating materials. The minor differences between derma+flex® QS™ High Viscosity Tissue Adhesive and the predicate center around the sterilization processes. Like the predicate, derma+flex® QS™ is sterilized in two steps during the production process involving dry heat and ethylene oxide to a Sterility Assurance Level of 106. While the sterilizing agents are the same the conditions have been changed. These differences do not raise new questions of safety and effectiveness.
### 7. Shelf-Life.
Shelf-life studies were performed to show that the minor differences in sterilization conditions did not change the performance of the device over time. Testing included:
Wound Closure Strength (according to ASTM F2458-05)
Tensile Strength (according to ASTM F2258-05) T-Peel Strength (according to ASTM F2256 -05) Lap Shear Strength (according to ASTM F2255-05)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Chemence Medical. The logo consists of a red and blue diamond shape above the words "CHEMENCE" and "MEDICAL". The word "CHEMENCE" is in red, and the word "MEDICAL" is in blue. The diamond shape is split in half, with the left half being red and the right half being blue. There is a white "C" in the center of the diamond shape.
Viscosity Polymerization Time Purity Water Content
### 8. Conclusion
Based on the studies carried out, the minor changes in the sterilization process between derma+flex® QS™ and the predicate device have not altered its performance indicating derma+flex® QS™ High Viscosity Tissue Adhesive is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.