The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. The SimGrid is a technology that enhances the visibility of major clinical equipment by compensating for the decrease in contrast that is caused by scatter radiation of all kinds of anatomical regions. The Tube & Line Enhancement enhances the visibility of tube and line to support radiologists or ICU clinicians who may need to find tubes and line on chest radiographs by trained professionals, such as physician, radiologists, and technicians. The Pediatric Exposure Management support radiologists to select suitable patient's size or take steps to serve an optimal exposure conditions to the lowest possible levels especially for the pediatric patients.
Device Story
Mobile digital X-ray system; captures radiographic images of human anatomy. X-rays pass through patient; detector converts to electrical signals; signal processing device amplifies and converts to digital data; S-Station software saves as DICOM files for PACS. Operated by physicians, radiologists, or technicians in clinical settings. Features: SimGrid (software to compensate for scatter radiation contrast loss); Tube & Line Enhancement (TLE) (software to enhance visibility of tubes/lines on chest radiographs); Pediatric Exposure Management (software to optimize exposure conditions based on patient weight/size). TLE and SimGrid function as assistant tools for clinicians. Benefits include improved image contrast and clearer visualization of medical devices (tubes/lines) to support clinical decision-making in ICU or general radiology environments.
Clinical Evidence
Bench testing only. Evaluated using 30 phantom images covering eight body parts. Professional radiologist assessment confirmed equivalence to predicate devices. MTF and DQE measurements performed per IEC 62220-1. SimGrid and TLE evaluated using phantoms to demonstrate improved local contrast and tube/line visibility compared to original post-processing.
Technological Characteristics
Battery-powered mobile X-ray system; wheel-based; rotational collapsible column; extractable arm; rotational tube head. Detectors and image processing identical to predicates. Software: S-Station (operation), SimGrid (scatter compensation), TLE (enhancement), Pediatric Exposure Management. Standards: ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971, 21 CFR 1020.30/31.
Indications for Use
Indicated for generating radiographic images of human anatomy in patients of all sizes, including pediatric patients. Intended for use by qualified/trained doctors or technicians. Not for mammographic applications.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around the left side of the image. To the right of the text is a symbol consisting of three stylized human profiles facing right, connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 15, 2016
Samsung Electronics Co., Ltd. % Chulsin Kim Regulatory Affairs Manager 129, Samsung-ro, Yeongtong-gu Suwon-si, Gyeonggi-do 16677 REPUBLIC OF KOREA
Re: K162278
Trade/Device Name: GM85 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL, MQB Dated: October 24, 2016 Received: October 26, 2016
Dear Chulsin Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K162278
Device Name GM85
Indications for Use (Describe)
The GM85 Digital X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <div style="display:inline-block;"> <svg height="15px" width="15px"> <rect fill="white" height="15px" stroke="black" width="15px" x="0" y="0"></rect> <path d="M2,2 L13,13 M13,2 L2,13" stroke="black"></path> </svg> </div> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <div style="display:inline-block;"> <svg height="15px" width="15px"> <rect fill="white" height="15px" stroke="black" width="15px" x="0" y="0"></rect> </svg> </div> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Samsung logo. The logo is in blue and features the word "SAMSUNG" in white, uppercase letters. The logo is tilted slightly upwards.
510(k) Premarket Notification - Traditional
## Section 5: 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92
- 1. Date: August 10, 2016
## 2. Submitter
- Company Name: SAMSUNG ELECTRONICS Co., Ltd. A.
- Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, B. Republic of Korea
### 3. Primary Contact Person
- Name: KIM, CHULSIN A.
- B. Title: Regulatory Affairs Manager
- C. Phone Number: +82-31-200-7661
- FAX Number: +82-31-200-1199 D. E-Mail:chulsin.kim@samsung.com
## 4. Secondary Contact Person
- A. Name: Ninad Gujar
- B. Title: Director, Regulatory & Quality
- C. Phone Number: 978-564-8503
- D. FAX Number: 978-750-6677 E-Mail: ngujar@samsungneurologica.com
## 5. Proposed Device
- A. Trade Name: GM85
- B. Device Name: GM85
- C. Common Name: Digital Diagnostic Mobile X-ray System
- D. Classification Name: Mobile X-ray System
- E. Product Code: IZL & MQB
- Regulation: 21 CFR 892.1720 ட்
### 6. Predicate Devices
| | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 |
|----------------------|-------------------------|-------------------------|----------------------------------------|
| Device Name | GM60A-32S,<br>GM60A-40S | GC85A | ClearRead +Confirm |
| Classification Name | MobileX-ray system. | Stationary X-ray System | System, Image Processing, Radiological |
| Product Code | IZL | KPR | LLZ |
| Regulation | 21 CFR 892.1720 | 21 CFR 892.1680 | 21CFR 892.2050 |
| 510(K)# | K150097 | K160997 | K123526 |
| 510(K) Decision Date | February 19, 2015 | April 02, 2015 | December 27, 2012 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, and the letters are evenly spaced. The blue oval is tilted slightly to the left.
510(k) Premarket Notification - Traditional
#### 7. Device Description
The GM85 Digital Mobile X-ray imaging System is the equipment that captures images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device before being sent to the S-Station (Operation Software) and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images.
The SimGrid is an additional image processing software option which is able to compensate the contrast loss due to scatter radiations, primarily acquisitions without a physical anti-scatter grid of all kinds of anatomical regions.
The SimGrid is available as options to be exclusively installed in S-station, which is a Samsung Digital X-ray operation S/W, while this post-image processing software does not depend on how the image is acquired, or with what acquisition device.
The Tube & Line Enhancement (TLE) is that after generating a normal post processing image, pressing the TLE button will create a companion view. For making a better tube & line look in chest radiography, input raw image is processed by using stronger contrast and detail enhancement algorithm. This supports clearer visibility of tube & lines than normal post processing image.
TLE is available as options to be exclusively installed in S-station, which is a Samsung Digital X-ray operation S/W, while this post-image processing software does not depend on how the image is acquired, or with what acquisition device.
The Pediatric Exposure Management provides subdivided patient size and exposure conditions of chest and abdomen radiographs specially optimized for pediatric patients. It helps an optimal dose for radiographic procedures of pediatric patients by providing guidance of optimized exposure conditions according to patient's weight and procedures.
#### 8. Intended Use
The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
The SimGrid is a technology that enhances the visibility of major clinical equipment by compensating for the decrease in contrast that is caused by scatter radiation of all kinds of anatomical regions.
The Tube & Line Enhancement enhances the visibility of tube and line to support radiologists or ICU clinicians who may need to find tubes and line on chest radiographs by trained professionals, such as physician, radiologists, and technicians.
The Pediatric Exposure Management support radiologists to select suitable patient's size or take steps to serve an optimal exposure conditions to the lowest possible levels
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Samsung logo. The logo is in blue and the word "SAMSUNG" is written in white, bold letters. The logo is slightly tilted.
510(k) Premarket Notification - Traditional especially for the pediatric patients.
- 9. Summary of Technological characteristic of the proposed device compared with the predicate devices
The proposed device, GM85, has same detectors and image processing as predicate devices so it does not have significant changes in image characteristics compare to the predicate device #1 & #2, GM60A and GC85A.
SimGrid, optional post-image processing software, has same image processing technology as SimGrid of predicate device, SimGrid technology is possible to applying to all kinds of anatomical regions where scatter radiation cause the contrast decrease. Specific description is added to make it clear with the non-clinical data, because the description of the predicate device was not mentioned about the target anatomical regions specifically.
Tube & Line Enhancement (TLE) and ClearRead +Confirm (K123526) are the same kind of function in that both include software with the ability to enhance visibility of tube and line. These are also provided in addition to the application and used to assistant tool for radiologist.
Pediatric Exposure Management is subdivided patient size and exposure conditions especially for pediatric patients based on weight and protocols. It follows same methodologies to define preset of patient size compare to preset of standard patient size from predicate device but specially optimized for pediatric patients.
The gantry of GM85 is newly designed to enhance coverage of tube head and usability but it has same structures of battery powered system, wheel based movable body, builtin console, rotational collapsible column, extractable arm and rotational tube head as predicate device #1 GM60A so it does not have significant changes in acquiring radiographic images compare to predicate device #1 GM60A.
Comparisons of the following technological characteristics were assessed and the results demonstrate the substantial equivalence to the predicates.
A. Comparing with Predicate Device #1 and #2 The proposed device is shown as its parts are identical or equivalent with predicate device #1 or #2 while some differences are made as below, which does not show significant difference in safety and effectiveness.
| Specifica<br>tion | Predicate Device #2 | Proposed Device | Predicate Device #1 | Discussi<br>on |
|-------------------|----------------------------------|----------------------------------|----------------------------------|----------------|
| Device<br>Name | GC85A | GM85 | GM60A-32S, 40S | |
| Manufac<br>turer | SAMSUNG<br>ELECTRONICS co., ltd. | SAMSUNG<br>ELECTRONICS co., ltd. | SAMSUNG<br>ELECTRONICS co., ltd. | |
| 510(k)<br>Number | K160997 | - | K150097 | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the SAMSUNG logo. The logo is in a blue box with the word SAMSUNG in white letters. The logo is simple and recognizable.
510(k) Premarket Notification - Traditional
| Appearances | | | | Difference(1) |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Intended<br>Use | Image: GC85A digital X-ray imaging system | Image: GM85 Digital Mobile X-ray imaging System | Image: GM60A Digital Mobile X-ray Imaging System | Same |
| | The GC85A digital X-ray imaging system is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. | The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. | The GM60A Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. | |
| Manufacturer Contents | Predicate Device #2 | GM85 | Predicate Device #1 | Discussion<br>n | | |
|-----------------------|---------------------|--------------------------------------|--------------------------------------|----------------------------------|-----------------|---------------|
| (2)Tube assembly | | | | | | |
| Moving<br>Range | Horizontal | 1680~4180mm | 793~1355mm | 700~1250mm | Difference(2) | |
| | Vertical | 1030~3030mm | 550~2030mm | 550~2020mm | Difference(2) | |
| Tube | Target Angle | 12° | 14° | 12° 16° | Difference(3) | |
| Collimator | | SDR-<br>OGCL80U<br>212X300X1<br>79mm | SDR-<br>OGCL83U<br>212X306<br>X179mm | SDR-OGCL40U<br>212 X 306 X 179mm | RALCO R221 DHHS | Difference(4) |
| Manufacturer Contents | Predicate Device #2 | GM85 | Predicate Device #1 | Discussion |
|---------------------------------------|---------------------|---------|---------------------|---------------|
| (6) Software Feature | | | | |
| SimGrid | SimGrid | SimGrid | - | Difference(5) |
| Tube & Line<br>Enhancement(TLE) | - | TLE | - | Difference(6) |
| Pediatric Exposure<br>Management(PEM) | - | PEM | - | Difference(7) |
| S-Align | S-Align | S-Align | - | Difference(8) |
| No | Differences | Explanation |
|-----|-------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (1) | Appearance | The gantry of GM85 is newly designed to enhance coverage of tube head<br>and usability but it has same structures of battery powered system, wheel<br>based movable body, built-in console, rotational collapsible column,<br>extractable arm and rotational tube head as predicate device #1 GM60A.<br>So it does not have significant changes in acquiring radiographic images<br>compare to predicate device #1, GM60A. |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Samsung logo. The logo is in blue and white. The word "SAMSUNG" is written in white letters on a blue background.
510(k) Premarket Notification - Traditional
| (2) | Moving Range | Extended coverage of proposed device GM85 can be more helpful to set<br>position than predicate device #1, GM60A. |
|-----|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (3) | Tube Target Angle | Tube target angle is maximized while focal spot is maintained 0.6/1.2mm<br>for proposed device GM85. Larger target angle increases X-ray beam<br>coverage and enhances usability. Tube target angle 14° of proposed<br>device GM85 does not have significant changes in image characteristics<br>compare to tube target angle between 12° and 16° of predicate device #1<br>GM60A. |
| (4) | Collimator | Collimator (SDR-OGCL40U) of proposed device, GM85 has same design<br>compare to collimator (SDR-OGCL80U) of predicate device #2, GC85A but<br>collimator(SDR-OGCL40U) of proposed device, GM85 has some<br>mechanical changes to mount sub display unit on it from predicate device<br>#2 GC85A. So it does not have significant changes in performance<br>compare to predicate device #2 GC85A. |
| (5) | SimGrid | SimGrid, optional post-image processing software, has same image<br>processing technology as SimGrid of predicate device #2 GC85A. SimGrid<br>technology is possible to applying to all kinds of anatomical regions where<br>scatter radiation cause the contrast decrease. Specific description is added<br>to make it clear with the non-clinical data, because the description of the<br>predicate device #2 was not mentioned about the target anatomical regions<br>specifically. |
| (6) | Tube & Line<br>Enhancement (TLE) | Tube & Line Enhancement (TLE) is optional software to make an additional<br>image with enhanced tube and line besides original post processing image.<br>It focused on enhancement of tubes and lines regardless of patient size and<br>should be used only as an assistant tool. (Refer to the comparing with<br>predicate device #3) |
| (7) | Pediatric Exposure<br>Management (PEM) | Pediatric Exposure Management is subdivided patient size and exposure<br>conditions especially for pediatric patients based on weight and protocols. It<br>follows same methodologies to define preset of patient size compare to<br>preset of standard patient size from predicate device but specially<br>optimized for pediatric patients. |
| (8) | S-Align | S-Align of proposed device GM85 has same design but GM85 is operated<br>only in manual compare to predicate device GC85A and it doesn't<br>contribute adverse impacts to the device safety and performance. |
## B. Comparing with Predicate Device #3
| Specification | Predicate Device #3 | Proposed Device | Discussion |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Device Name | ClearRead +Confirm | GM85 | - |
| Manufacturer | Riverain Technologies, LLC | SAMSUNG ELECTRONICS co., ltd. | - |
| 510(k) Number | K123526 | None | - |
| S/W Feature<br>Name | ClearRead +Confirm 1.0 | Tube & Line Enhancement | - |
| Intended Use | ClearRead +Confirnm is intended to<br>generate an enhanced secondary digital<br>radiographic image of the chest to<br>facilitate confirmation of line/tubes. The<br>enhanced AP or PA image of the chest<br>provides improved visibility of lines and<br>tubes. The ClearRead +Confirm image<br>provides adjunctive information and is not | Tube & Line Enhancement (TLE)<br>software generates a companion image<br>which can provide clear visibility of tubes<br>and lines in addition to the original chest<br>radiograph. Since the image generated<br>by TLE is intended to be used to assist | Difference(6) |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, sans-serif font, set against a blue, oval-shaped background. The oval is tilted slightly, giving the logo a dynamic appearance.
510(k) Premarket Notification - Traditional
| a substitute for the original PA/AP image.<br>This device is intended to be used by<br>trained professionals, such as<br>physicians, radiologists, and technicians,<br>on patients with lines and tubes and is<br>not intended to be used on pediatric | provided with the original diagnostic<br>chest radiograph. TLE software is<br>intended to be used by qualified/ trained<br>doctors or technicians on patients with<br>tubes and lines. |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| No | Differences | Explanation |
|-----|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (6) | Intended Use | Since TLE software amplify specific frequencies suitable for tubes and lines<br>partially, it does not depend on the size of patients. Therefore TLE software<br>can be used for all patients including the pediatric patient. So it does not have<br>significant changes in performance compare to predicate device #3. |
### C. Safety, EMC and Performance Data
Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1,IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO14971, 21CFR1020.30 and 21CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. Wireless function was tested and verified followed by quidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfying the standards.
#### D. Non-clinical data
Non-clinical testing data was provided in conformance to the FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1.
The proposed device has same detectors so it shows no difference in non-clinical testing data such as MTF and DQE measurements from the predicate device #1 and #2.
SimGrid software, 510(k) cleared before with predicate device #2 was evaluated with various phantoms at various exposure conditions demonstrating that SimGrid processing gives the better local contrast than images acquired without a grid at expanded anatomical regions.
Tube & Line Enhancement(TLE) was evaluated with the phantom. After input image is acquired with chest phantom which is inserted tube line, the normal post image and TLE image are created. Then, the three images are compared in terms of the tube & line contrast. This evaluation proves that TLE processing gives the better tube and line visibility than original post processing.
### E. Clinical data
In clinical data, phantom image evaluations were performed in accordance with FDA guidance for the submission of 510(k)'s for Solid State X-ray Imaging Devices. They were evaluated by a professional radiologist and found to be equivalent to the predicate devices. Evaluated 30case of phantom image includes eight parts of the body.
### F. Conclusions
The non-clinical and clinical data demonstrate that the proposed device is as safe, as
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the text "SAMSUNG ELECTRONICS Co., Ltd." in a bold, sans-serif font. The text is arranged on a single line, with the company name in all caps. The words "Co." and "Ltd." are abbreviated and separated by a comma.
Image /page/9/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue oval shape.
510(k) Premarket Notification - Traditional
effective, and performs as well as the legally marketed predicate devices #1, #2 and #3.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.