K133383 · Omron Healthcare, Inc. · DXN · Mar 4, 2014 · Cardiovascular
Device Facts
Record ID
K133383
Device Name
HEM-7320
Applicant
Omron Healthcare, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Mar 4, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
Device Story
Automatic non-invasive blood pressure system; battery or AC-powered. Device inflates cuff via integral pump; electronically controlled valve manages deflation. During deflation, device monitors cuff pressure and extracts pulse waveform data. Software analyzes waveform to determine systolic/diastolic pressure and pulse rate. Features irregular heartbeat detection and body movement detection. Stores up to 100 measurements for two users; displays average of last three values. Intended for home use by adult patients. Healthcare providers review stored data to monitor blood pressure trends, aiding in hypertension management. Benefits include convenient, self-administered blood pressure monitoring.
Clinical Evidence
Clinical accuracy study conducted in accordance with ANSI/AAMI/ISO 81060-2. Study included 85 adult patients (40 males, 45 females). Standard auscultation method used as reference. Measurements were taken alternatively with the device and auscultation in the same arm. Results confirmed clinical accuracy equivalent to predicate devices.
Technological Characteristics
Cuff oscillometric measurement principle. Semiconductor pressure sensor. Powered by 4 AA batteries or AC adapter. Patient-contacting cuff materials tested per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity). Complies with AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 80601-2-30, and IEC 60601-1-11. Features LCD display, active electronic control valve, and DC rolling pump.
Indications for Use
Indicated for adult patients for the measurement of blood pressure and pulse rate. Includes detection and warning of irregular heartbeats during measurement.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
510(k) Summary Page 1 of 8 1/30/2014
# MAR 0 4 2014
| Omron Healthcare, Inc.<br>1925 West Field Court<br>Lake Forest, IL 60045 USA | Tel - 847-247-5626<br>Fax- 847-680-6269 | MAR 0 4 201 |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------|
| Official Contact: | Renee Thornborough – Director QA/RA | |
| Proprietary or Trade Name: | Model HEM-7320 | |
| Common/Usual Name: | Noninvasive blood pressure measurement system. | |
| Classification Name/Code:<br>system. | DXN – Noninvasive blood pressure measurement<br>21CFR 870.1130<br>Class II | |
| Device: | Model HEM-7320 | |
| Predicate Devices: | Omron – HEM -780N3 (HEM-7080-Z)- K061822<br>Omron – BP742 (HEM-7200-Z) – K121932 | |
### Device Description:
The device is an automatic non-invasive blood pressure system. The device is battery powered and can also be powered from an IEC 60601-1 compliant AC adaptor. The device inflates a cuff with an integral pump, then deflates the cuff via an electronically controllable valve. During deflation the cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic pressure. The device is intended to be used with an Omron specified cuff to encompass arms ranging from 9 to 17 inches in circumference.
The device also detects the appearance of irregular heartbeats during measurement.
The device has provisions for selecting two users, measurements from these users are stored in memory. The memory stores up to 100 of the latest measurements. It can also display an average of the last three values.
Note that a number of documents in this submission include reference to the Cuff Wrapping Guide (or Cuff Wrapping Check). This feature has been removed from the device and is not part of this submission.
Intended User Home user
{1}------------------------------------------------
#### 510(k) Summary Page 2 of 8 1/30/2014
#### Patient Population
This device is intended for use on adults.
### Indications for Use:
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population.
The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
Environment of Use: Home
Contraindications: There are no known contraindications.
#### Predicate Device Comparison:
The HEM-7320was compared to the predicates HEM -780N3 (HEM-7080-Z) - K061822 and BP742 (HEM-7200-Z) - K121932 in the device comparison table below.
{2}------------------------------------------------
evice Compariso.
| Model Name:<br>510(k) Number | Predicate Device<br>HEM-780N3 (HEM-7080-Z)<br>K061822 | Predicate Device<br>BP742(HEM-7200-Z)<br>K121932 | New Device<br>HEM-7320<br>TBD |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | The device is a digital monitor intended<br>for use in measuring blood pressure and<br>pulse rate in adult patient population with<br>arm circumference ranging from 9 inches<br>to 17 inches (22 cm - 42 cm).<br>The device detects the appearance of<br>irregular heartbeats during measurement<br>and gives a warning signal with readings.<br>This blood pressure monitor compares<br>average blood pressure results to pre-<br>established AHA (American Heart<br>Association) hypertension guideline of<br>135/85 mmHg. The Omron 780N3 model<br>is not intended to be a diagnostic device.<br>Contact your physician if hypertensive<br>values are indicated. | The device is a digital monitor intended<br>for use in measuring blood pressure and<br>pulse rate in adult patient population.<br>The device detects the appearance of<br>irregular heartbeats during measurement<br>and gives a warning signal with readings. | The device is a digital monitor intended<br>for use in measuring blood pressure and<br>pulse rate in adult patient population.<br>The device detects the appearance of<br>irregular heartbeats during measurement<br>and gives a warning signal with readings.<br>Identical to K121932 |
| Environment of Use | Home | Home | Home Identical to both predicates |
| Patient Population | Adult | Adult | Adult Identical to both predicates |
| Specifications / Features | | | |
| Measurement method /<br>Principal of operation | HEM-7080-Z(HEM 780N3)<br>Cuff oscillometric method | BP742(HEM-7200-Z)<br>Cuff oscillometric method | HEM-7320<br>Cuff oscillometric method. Identical to<br>both predicates |
| Measurement range | Pressure: 0 to 294 mmHg<br>Pulse Rate: 40 to 180 beats/min. | Pressure: 0 to 299 mmHg<br>Pulse Rate: 40 to 180 beats/min. | Pressure: 0 to 299 mmHg<br>Pulse Rate: 40 to 180 beats/min.<br>Identical to K121932 |
| Pressure sensor | Electrostatic capacitive sensor | Semiconductor pressure sensor | Semiconductor pressure sensor. Identical<br>to K121932 |
PDF Page 21 of 158
Page 3
{3}------------------------------------------------
| , Summary 1/30/2014<br>ist d (1)0<br>l<br>5 |
|---------------------------------------------|
|---------------------------------------------|
| | HEM-7080-Z(HEM 780N3) | BP742(HEM-7200-Z) | HEM-7320 |
|----------------------------------------|----------------------------------------------------------|-----------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Applicable cuff<br>(Arm Circumference) | 22-42cm | 22-42cm | 22-42cm, identical to both predicates |
| Accuracy of pressure indicator | Within $\pm$ 3 mmHg or 2 % of reading | Within $\pm$ 3 mmHg or 2 % of reading | Within $\pm$ 3 mmHg or 2 % of reading<br>Identical to both predicates |
| Accuracy of pulse rate | Within $\pm$ 5% of reading | Within $\pm$ 5% of reading | Within $\pm$ 5% of reading<br>Identical to both predicates |
| Inflation method | Automatic by electric pump | Automatic by electric pump | Automatic by electric pump.<br>Identical to both predicates |
| Deflation method | Active electronic control valve | Automatic pressure release valve | Automatic pressure release valve Identical<br>to K121932 |
| Display | LCD digital display | LCD digital display | LCD digital display<br>Identical to both predicates |
| Power Source | 4"AA"batteries or AC adapter | 4"AA"batteries or AC adapter | 4"AA"batteries or AC adapter<br>Identical to both predicates |
| Operating conditions | 10 to 40 °C (50 to 104 °F)<br>30 to 85 %RH | 10 to 40 °C<br>15 to 90 %RH | 10 to 40 °C<br>15 to 90 %RH<br>Identical to K121932 |
| Storage conditions | -20 to 60 °C<br>10 to 95 %RH | -20 to 60 °C<br>10 to 95 %RH | -20 to 60 °C<br>10 to 95 %RH<br>Identical to K121932 |
| Dimensions (mm) | 131 (W) x 155 (D) ×84 (H) mm<br>without cuff and battery | 124(W)×161(D)×90(H) mm | 124 (W) x 161 (D) × 90 (H) mm, size is<br>not a factor in function of the device |
| Weight | Approx. 420g<br>(not including cuff and batteries) | Approx. 340g<br>(not including battery) | Approx. 380g (13 3/8 oz) (not including<br>battery), weight is not a factor in function<br>of the device. |
| Irregular Heart beat Feature | Yes | Yes | Yes, identical to both predicates |
| Body movement detection | Yes | Yes | Yes, identical to both predicates |
| Hypertension indicator | No | No | Yes, identical to K121932 |
| TruRead™ | Yes | Yes | Yes identical to K061822 |
Page 4
PDF Page 22 of 158
{4}------------------------------------------------
Technology / Features Comparisons
| Technology / Features Comparison | HEM-7080-Z(HEM 780N3) | BP742(HEM-7200-Z) | HEM-7320 |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Power supply | Regulates power voltage regardless of battery voltage. | Regulates power voltage regardless of battery voltage. | Regulates power voltage regardless of battery voltage. Identical to both predicates |
| Microprocessor | determines blood pressure and pulse rate<br>controls the pump, the valve, and the display<br>detects switch operations<br>stores measurement results<br>manages date and time | determines blood pressure and pulse rate<br>controls the pump, the valve, and the display<br>detects switch operations<br>stores measurement results<br>manages date and time | determines blood pressure and pulse rate<br>controls the pump, the valve, and the display<br>detects switch operations<br>stores measurement results<br>manages date and time<br>Identical to both predicates |
| Pressure sensor | Electrostatic capacitive sensor | Semiconductor pressure sensor | Semiconductor pressure sensor, identical to K121932 |
| Rapid exhaust valve | Active electronic control valve that performs cuff air bleeding and release | Automatic rapid air release valve. | Active electronic control valve that performs cuff air bleeding and release, identical to K061822 |
| Deflation Valve | | Automatic pressure release valve | |
| Inflation source | DC rolling pump | DC rolling pump | DC rolling pump Identical to both predicates |
| Display | LCD (Liquid Crystal Display) displays;<br>current cuff pressure<br>systolic blood pressure<br>diastolic blood pressure<br>pulse rate<br>error messages<br>measurement results in the memory | LCD (Liquid Crystal Display) displays;<br>current cuff pressure<br>systolic blood pressure<br>diastolic blood pressure<br>pulse rate<br>error messages<br>measurement results in the memory | LCD (Liquid Crystal Display) displays;<br>current cuff pressure<br>systolic blood pressure<br>diastolic blood pressure<br>pulse rate<br>error messages<br>measurement results in the memory<br>Identical to both predicates |
| Controls | START/STOP Button<br>Setting Button<br>Memory button<br>User ID Selections Button<br>Morning/Evening Average Buttons | START/STOP Button<br>Date/Time setting Button<br>Up/Down Button<br>User ID Selections Button | START/STOP Button<br>Date/Time setting Button<br>Up/Down Button<br>User ID Selections Button<br>Identical to K121932 |
| | | | And American Manager States of State Annual Property of Children |
| | HEM-7080-Z(HEM 780N3 | BP742(HEM-7200-Z) | HEM-7320 |
| Cuff | between 9-17 inches(22- 42 cm). The cuff has<br>The cuff is applied to the arm circumference<br>an elasticity board to hold the arm.<br>A PRODUCTION CONTROLLERS CONTRACT CONTROLLET<br>A A M A M A M A B A B A B A B A B A L A L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L L | Wide Range Cuff, Standard Adult Arm Cuff ComFit Cuff(HEM-FL31) 22-42cm, new cuff<br>Large Cuff | to reduce dependence on bladder placement<br>relative to the brachial artery |
| Materials | Patient contact materials of the cuff have<br>been cleared in the referenced 510(k) | Patient contact materials of the cuff have been<br>cleared in the referenced 510(k) | been tested in accordance with ISO 10993 and<br>Patient contact matcrials of the cuff have<br>FDA guidance |
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | |
# K133383 Page 5 of 8
PDF Page 23 of 158
Page 5
{5}------------------------------------------------
510(k) Summary
Page 6
{6}------------------------------------------------
#### 510(k) Summary Page 7 of 8 1/30/2014
# Differences Between Other Legally Marketed Predicate Devices:
The Omron HEM-7320 is viewed as substantially equivalent to the predicate devices because: The HEM-7320 uses the exact same technology and has identical indications for use. The differences that exist between the devices are insignificant in the terms of safety or effectiveness.
#### Indications -
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population.
The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
Discussion - These indications are identical to the predicate Omron BP742 (K121932).
Prescriptive - The HEM-7320 and predicates are all OTC.
Design and Technology - The HEM-7320 has equivalent design and features when compared to the predicates and has the identical technology to the predicate.
Performance and Specifications - The HEM-7320 has equivalent specifications of performance when compared to the predicate.
Compliance with standards - The predicate devices declare compliance with SP10, IEC 60601-1 and IEC 60601-1-2. The HEM-7320 complies with AAMI ANSI ES6060-1 (which replaced IEC 60601-1), IEC 60601-1-2 and ANSI/AAMI/ISO 81060-2 (which replaced SP10). The HEM-7320 also complies with IEC 80601-2-30 and IEC 60601-1-11 for home healthcare.
#### Materials -
The patient contacting materials of the cuffs has been tested in accordance with ISO 10993-1 and FDA Guidance. The tests included Cytotoxicity, Sensitization, and Intracutancous Reactivity.
#### Patient Population -
The HEM-7320 and predicates are indicated for adults
Environment of Use - Home, Identical to the predicates
#### Non-Clinical Testing Summary:
We have performed bench tests and found that the HEM-7320 met all requirements specifications and standards requirements and were found to be equivalent in comparison to the predicate. Testing includes the following:
- . Verification Testing to insure the device meets its specifications
{7}------------------------------------------------
#### 510(k) Summary Page 8 of 8 1/30/2014
- Testing of hazard mitigations .
- Testing for compliance to AAMI ANSI ES60601-1 .
- . Testing for compliance to IEC 60601-1-2
- Testing for compliance to IEC 80601-2-30 .
- Testing for compliance to IEC 60601-1-11 .
- Comparative Testing to the predicates .
#### Clinical Testing Summary:
Testing to insure clinical accuracy of the device in accordance with ANSI/AAMI/ISO 81060-2 as documented in Section 20.
Eighty five patients (40 males and 45 females) were recruited for the study.
Standard auscultation method was used as the reference blood pressure (BP) measuring in the left upper arm. BP measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in AAMI.
#### Substantial Equivalence Conclusion
Omron maintains that the HEM-7320 is substantially equivalent to the predicate device in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards
{8}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2014
Omron Healthcare, Inc. c/o Paul Dryden 24301 Woodsage Dr. Bonita Springs, FL 34134 US
Re: K133383
> Trade/Device Name: HEM-7320 Regulation Number: 21 CFR 870.11307 Regulation Name: Noninvasive Blood Pressure Regulatory Class: Class II Product Code: DXN Dated: January 31, 2014 Received: November 6, 2013
#### Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2 - Paul Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number its Internet address 796-7100 or at 638 2041 or (301) (800) http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/9/Picture/7 description: The image shows the phrase "Sincerely yours,". Below the phrase is a signature that is illegible. The signature is written in black ink. The background is white.
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K133383
Device Name Omron HEM-7320
Indications for Use (Describe)
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
> K133383 Page 1 of 1
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE CONSULTION CONSULTION FOR FOR FOR FOR FOR CONLY CONLY - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2 - 1 - 3 -
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/10/Picture/14 description: The image shows a signature and a date stamp. The date on the stamp reads "2014.03.04" and the time is "14:53:17-05'00'". The text "for Bram Zuckerman" is printed below the date stamp.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.