K110132 · Ge Medical Systems Information Technologies, Inc. · DQK · May 25, 2011 · Cardiovascular
Device Facts
Record ID
K110132
Device Name
MUSE CARDIOLOGY INFORMATION SYSTEM
Applicant
Ge Medical Systems Information Technologies, Inc.
Product Code
DQK · Cardiovascular
Decision Date
May 25, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The MUSE Cardiology Information System is intended to store, access and manage cardiovascular information on adult and pediatric patients. The information consists of measurements, text, and digitized waveforms. The MUSE Cardiology Information System provides the ability to review and edit electrocardiographic procedures on screen, through the use of reviewing, measuring, and editing tools including ECG serial comparison and interpretive 12-lead analysis. The MUSE Cardiology Information System is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or facility providing patient care. The MUSE Cardiology Information System is not intended for real time monitoring. The MUSE Cardiology Information System is not intended for pediatric serial comparison.
Device Story
MUSE Cardiology Information System is a network PC-based client/server system for centralized storage and management of cardiovascular data (Resting ECG, Stress, Holter, HiRes) from GE and non-GE diagnostic equipment. Operators (trained staff) use the system to review, measure, and edit digitized waveforms and reports on-screen. Key features include interpretive 12-lead analysis and ECG serial comparison for adults. The system generates management, ad-hoc database, and clinical patient reports. It is used in hospital or clinical settings to support healthcare practitioners in clinical decision-making by providing organized access to diagnostic cardiology information. It is not for real-time monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including risk analysis, requirements/design reviews, code inspection, module verification, system integration testing, performance verification, safety testing, and simulated use validation.
Technological Characteristics
Network PC-based client/server architecture. Manages diagnostic cardiology data (ECG, Stress, Holter, HiRes). Features include interpretive 12-lead analysis and serial comparison tools. Complies with voluntary standards for safety and performance.
Indications for Use
Indicated for storage, access, and management of cardiovascular data (measurements, text, digitized waveforms) for adult and pediatric patients. Used in hospitals or care facilities by trained operators under licensed practitioner supervision. Contraindicated for real-time patient monitoring and pediatric serial comparison.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
MUSE Cardiology Information System w/ VMWare (K083639)
Submission Summary (Full Text)
{0}------------------------------------------------
K110132 pg 1 of 2
MAY 2 5 2011
Image /page/0/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The letters are stylized and appear to be in a sans-serif font.
## GE Healthcare
510(k) Premarket Notification Submission
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
January 12, 2011 Date:
| Submitter: | GE Medical Systems Information Technologies<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Kristin Pabst<br>Regulatory Affairs Manager<br>GE Medical Systems Information Technologies<br>Phone: (414) 721-3104<br>Fax: (414) 721-3863 |
| Secondary Contact<br>Person: | Patricia Taige<br>Regulatory Affairs Leader<br>GE Medical Systems Information Technologies<br>Phone (414) 721-3222<br>FAX: (414) 721-3863 |
| Device: Trade Name: | MUSE Cardiology Information System |
| Common/Usual Name: | ECG Analysis Computer |
| Classification Names: | Programmable Diagnostic Computer (21 CFR 870.1425) |
| Product Code: | DQK |
| Predicate Device(s): | MUSE Cardiology Information System K#072502<br>MUSE Cardiology Information System w/ VMWare K#083639 |
| Device Description: | The MUSE Cardiology Information System is a network PC-<br>based system comprised of a client workstation/server<br>configuration that manages adult and pediatric diagnostic<br>cardiology data by providing centralized storage and ready<br>access to a wide range of data/reports (e.g. Resting ECG,<br>Stress, Holter, HiRes) from GE and non-GE diagnostic and<br>monitoring equipment. The device provides the ability to:<br>• Review and edit stored data consisting of measurements, text,<br>and digitized waveforms on screen, through the use of reviewing,<br>measuring, and editing tools including ECG serial comparison<br>and interpretive 12-lead analysis.<br>• Generate formatted management reports, ad-hoc database<br>search reports and clinical patient reports on selected stored<br>data. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and appear to be in a cursive font. The logo is black and white.
## GE Healthcare
### 510(k) Premarket Notification Submission
The MUSE Cardiology Information System is intended to store, Intended Use: access and manage cardiovascular information on adult and pediatric patients. The information consists of measurements, text, and digitized waveforms. The MUSE Cardiology Information System ' provides the ability to review and edit electrocardiographic procedures on screen, through the use of reviewing, measuring, and editing tools including ECG serial comparison and interpretive 12-lead analysis. The MUSE Cardiology_Information_System_is_intended_to_be_used_under_the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or facility providing patient care. The MUSE Cardiology Information System is not intended for real time monitoring. The MUSE Cardiology Information System is not intended for pediatric serial comparison.
The proposed MUSE Cardiology Information System employs Technology: the same functional scientific technology as the predicate devices MUSE Cardiovascular Information System (K072502) and MUSE Cardiovascular Information System w/ VMWare (K083639).
#### Determination of Summary of Non-Clinical Tests:
Substantial Equivalence:
The MUSE Cardiology Information System complies with voluntary standards as detailed in Section 9 and 17 of this premarket submission. The following quality assurance measures are applied to the development of the system:
- Risk Analysis .
- Requirements Reviews .
- . Design Reviews
- . Code Inspection
- Testing on unit level (Module verification) .
- . Integration testing (System verification)
- Performance testing (Verification) .
- Safety testing (Verification)
- Simulated use testing (Validation)
### Summary of Clinical Tests:
The subject of this premarket submission, MUSE Cardiology Information System, did not require clinical studies to support substantial equivalence.
Conclusion:
GE Healthcare considers the MUSE Cardiology Information System to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002
GE Medical Systems Information Technologies, Inc c/o Ms. Kristin Pabst Regulatory Affairs Manager 9900 West Innovation Drive Wauwatosa, WI 53226
MAY 2 5 2011
Re: K110132
> Trade/Device Name: Muse Cardiology Information System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: May 4, 2011 Received: May 6, 2011
Dear Ms. Pabst:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Kristin Pabst
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K110132 pg lofl
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The letters and the border are in black, contrasting with the white background.
# GE Healthcare
510(k) Premarket Notification Submission
510(k) Number (if known):
MUSE Cardiology Information System Device Name:
Indications for Use:
The MUSE Cardiology Information System is intended to store, access and manage cardiovascular information on adult and pediatric patients. The information consists of measurements, text, and digitized waveforms. The MUSE. Cardiology Information System provides the ability to review and edit electrocardiographic procedures on screen, through the use of reviewing, and editing tools including ECG serial comparison and interpretive 12-lead analysis. The MUSE Cardiology Information System is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or facility providing patient care. The MUSE Cardiology Information System is not intended for real-time patient monitoring. The MUSE Cardiology Information System is not intended for pediatric serial comparison.
Prescription Use_X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular-Device
510(k) Number K110132
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.