K161894 · Surgtech, Inc. · HWC · Feb 21, 2017 · Orthopedic
Device Facts
Record ID
K161894
Device Name
SurgTech Trauma System
Applicant
Surgtech, Inc.
Product Code
HWC · Orthopedic
Decision Date
Feb 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SurgTech Trauma System plates and screws are intended to be used for internal bone fixation for bone fractures, fusions, osteotomies and non-unions appropriate for the size and design of the device. The large plate set components are indicated for the use including the femur and tibia. The small plate set is indicated for use in various bones including the clavicle, scapula, humerus, radius, ulna, pelvis, tibia and fibula. The mini plate set is indicated for use in various small bones including the metacsals, phalanges, calcaneus and those of the ankle. In addition, the SurgTech Trauma System cancellous, cannulated lag, headless compression and non-locking cortical screws are indicated for use in bone reconstruction, arthrodesis, joint fusion, fracture repair and fracture fixation, appropriate for the size of the device.
Device Story
SurgTech Trauma System comprises metallic bone plates and screws for internal fixation; used for bone fractures, fusions, osteotomies, and non-unions. Plates provided in mini, small, and large sets; screws provided in locking cortical, non-locking cortical, cancellous, and headless versions. Device used by orthopedic surgeons in clinical/surgical settings. Surgeon selects appropriate plate/screw size based on anatomical site; implants provide mechanical stabilization of bone segments to facilitate healing. System benefits patients by restoring skeletal alignment and promoting union in fractured or reconstructed bone.
Clinical Evidence
Bench testing only. Mechanical testing of worst-case plates included static and dynamic bending per ASTM F382. Mechanical testing of worst-case screws included torsion, insertion/removal, and pullout testing.
Technological Characteristics
Materials: Titanium alloys (Ti6Al4V per ASTM F136, Ti6Al7Nb per ASTM F1295) and stainless steel (ASTM F138). Design: Metallic bone plates and screws in mini, small, and large configurations. Mechanical performance validated via static/dynamic bending (ASTM F382) and screw torsion/pullout testing. System is non-powered, mechanical fixation hardware.
Indications for Use
Indicated for internal bone fixation in patients requiring bone fracture repair, fusion, osteotomy, or non-union treatment. Large plates: femur/tibia. Small plates: clavicle, scapula, humerus, radius, ulna, pelvis, tibia, fibula. Mini plates: metacarpals, metatarsals, phalanges, calcaneus, ankle. Screws: bone reconstruction, arthrodesis, joint fusion, fracture repair/fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 21, 2017
SurgTech Inc Karen E. Warden, Ph.D. President Backroads Consulting Inc. PO Box 566 Chesterland, Ohio 44026-0566
Re: K161894
Trade/Device Name: SurgTech Trauma System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, HRS Dated: January 23, 2017 Received: January 25, 2017
Dear Dr. Warden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Pg.1/2
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K161894
Device Name SurgTech Trauma System
Indications for Use (Describe)
The SurgTech Trauma System plates and screws are intended to be used for internal bone fixation for bone fractures, fusions, osteotomies and non-unions appropriate for the size and design of the device.
The large plate set components are indicated for the use including the femur and tibia. The small plate set is indicated for use in various bones including the clavicle, scapula, humerus, radius, ulna, pelvis, tibia and fibula. The mini plate set is indicated for use in various small bones including the metacsals, phalanges, calcaneus and those of the ankle.
In addition, the SurgTech Trauma System cancellous, cannulated lag, headless compression and non-locking cortical screws are indicated for use in bone reconstruction, arthrodesis, joint fusion, fracture repair and fracture fixation, appropriate for the size of the device.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
| Date: | 14 February 2017 |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | SurgTech Inc.<br>24600 Center Ridge Road, Suite 195<br>Westlake, OH 44145<br>Phone (216) 421-2613 |
| Sponsor Contact: | Xuegong Yu, General Manager |
| 510(k) Contact: | Karen E. Warden, PhD<br>BackRoads Consulting Inc.<br>PO Box 566<br>Chesterland, OH 44026<br>Office: 440.729.8457 |
| Trade Name: | SurgTech Trauma System |
| Common Name: | Bone plates & screws, Bone screws |
| Device Classification | Class II |
| Classification Name: | Smooth or threaded metallic bone fixation fastener & Single/multiple<br>component metallic bone fixation accessories |
| Regulation: | 888.3040, 888.3030 |
| Device Product Code: | HWC, HRS |
| Device Description: | The SurgTech Trauma System is a plate and screw fixation system. Plates<br>are offered in "mini," "small" and "large" set sizes in a variety of shapes<br>based upon the anatomical fixation required. Screws are also offered in<br>"mini" "small" and "large" sets and, in addition, in locking cortical, non-<br>locking cortical, cancellous and headless versions. Screws can also be<br>used for fixation without the plates. |
| Indications for Use: | The SurgTech Trauma System plates and screws are intended to be used<br>for internal bone fixation for bone fractures, fusions, osteotomies and non-<br>unions appropriate for the size and design of the device.<br>The large plate set components are indicated for the use in various long<br>bones including the femur and tibia. The small plate set is indicated for use<br>in various bones including the clavicle, scapula, humerus, radius,<br>ulna, pelvis, tibia and fibula. The mini plate set is indicated for use in various<br>small bones including the metacarpals, metatarsals, phalanges, calcaneus<br>and those of the ankle.<br>In addition, the SurgTech Trauma System cancellous, cannulated lag,<br>headless compression and non-locking cortical screws are indicated for use<br>in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair<br>and fracture fixation, appropriate for the size of the device. |
| Materials: | The SurqTech Trauma System implants are manufactured from titanium<br>alloys (Ti6Al4V per ASTM F136 and Ti6Al7Nb per ASTM F1295) and<br>stainless steel (ASTM F138). |
| Primary Predicate<br>(plate & screw system): | GPC Bone Plates and Bone Screws (GPC Medical K092493) |
| Additional Predicates<br>(plate & screw system): | The NCB Periprosthetic Femur Polyaxial Locking Plate System (Zimmer<br>Inc. K100111), I.T.S. Straight Plate (I.T.S. GmbH K060156), I.TS. Volar<br>Radius Plate (I.T.S. GmbH K033756), I.T.S. Olecranon Plate (I.T.S. GmbH<br>K052368), Synthes Variable Angle LCP Elbow System (Synthes, K120070),<br>I.T.S. Distal Humeral Plates (I.T.S. GmbH K080184), I.T.S. Straight Plate<br>(I.T.S. GmbH K060156), I.T.S. Claviculatplate (I.T.S. GmbH K050852),<br>Synthes (USA) Clavicle Hook Plate (Synthes, K031677), Synthes (USA)<br>LCP® Proximal Humerus Plates (Synthes K041860), the I.T.S. Pelvic<br>Reconstruction System (I.T.S. GmbH K063166), I.T.S. Fibula Plate (I.T.S.<br>GmbH K063672), DePuy ALPS (DePuy Orthopaedics K082300), Acumed<br>Hand Fracture System (Acumed LLC K141383), Synthes Locking<br>Calcaneal Plates (Synthes K991407), I.T.S. Calcaneus Repair System<br>(I.T.S. GmbH K051642), Synthes 2.4 mm / 2.7 mm Variable Angle (VA), -<br>LCP Forefoot /Midfoot System (Synthes K100776) and Medartis APTUS®<br>Foot System (Medartis K091479), Synthes Titanium Small Reconstruction<br>Plate (Synthes K915818), Synthes Large Fragment Dynamic Compression<br>Locking (DCL) System (Synthes K000682), Synthes Small Fragment<br>Dynamic Compression Locking (DCL) (System K000684), 1.5mm Mini<br>Fragment LCP System (Synthes K090047). |
| Primary Predicate<br>(standalone screws): | OsteoMed Headless Cannulated Screw System (OsteoMed LP K063298) |
| Additional Predicates<br>(standalone screws): | Treu Bone fixation Screws and Pins (Treu Instrumente GmbH K083912),<br>Synthes Cortical Screws (Synthes K112583), Synthes 3.0 mm Headless<br>Compression Screws (Synthes K050636), Synthes 4.5 mm and 6.5 mm<br>Headless Compression Screws (Synthes K080943), Vilex Bone Screw<br>(Vilex K014154), Synthes 7.0/7.3 mm Cannulated Screws (Synthes<br>K962011), SBi Foot and Ankle Cannulated Screws (Small Bone Innovations<br>K092754) |
| Performance Data: | Mechanical testing of the worst case SurgTech Trauma System plates<br>included static and dynamic bending performed according to ASTM F382.<br>Mechanical testing of the worst case SurgTech Trauma System screws<br>included torsion, insertion/removal and pullout performed according to<br> |
| Technological<br>Characteristics: | The SurgTech Trauma System implants possess technological<br>characteristics similar to one or more of the cited predicate devices. These<br>include: Mechanical performance (as described above), Basic design (bone plate & screw system), Use of implant grade materials (titanium, titanium alloys & stainless steel), and Dimensional sizes (comparable to those offered by the predicates). Minor dimensional differences between the subject and predicate devices,<br>such as longer or shorter device lengths, did not raise new questions of<br>safety and effectiveness. Therefore the fundamental scientific technology of<br>the SurgTech Trauma System is the same as previously cleared devices. |
| Conclusion: | The SurgTech Trauma System possesses the same intended use and<br>similar indications for use compared to the cited predicates. The SurgTech<br>Trauma System technological characteristics are similar to those of the<br>cited predicate devices. Therefore SurgTech Trauma System is<br>substantially equivalent for its intended use. |
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.