Bausch + Lomb NNR06 Multi-Purpose Solution is indicated for use in the daily cleaning, conditioning, removing protein deposits, rinsing, chemical (not heat) disinfection, and storage of soft (hydrophilic) contact lenses, including silicone hydrogel contact lenses, as recommended by your eye care practitioner.
Device Story
NNR06 Multi-Purpose Solution is a sterile, isotonic, aqueous chemical disinfecting solution for soft contact lenses. It utilizes a triple disinfection system (polyaminopropyl biguanide, polyquaternium, and alexidine) combined with dual surfactants (poloxamine and poloxamer 181) and a buffering system (edetate disodium, sodium citrate, diglycine). The solution is used by patients at home to clean, condition, disinfect, and store lenses. The user applies the solution to the lens, rubs, and rinses before storage. The device maintains lens hygiene, removes protein deposits, and ensures lens wettability, thereby supporting comfortable lens wear and preventing ocular complications. It is packaged in a plastic bottle with a tamper-evident seal.
Clinical Evidence
Three-month, active-control, parallel-group, masked randomized clinical study comparing NNR06 to Complete Multi-Purpose Solution Easy Rub Formula in habitual soft contact lens wearers. Primary endpoints included slit lamp findings, visual acuity, and adverse events. Results showed no adverse events in either group, similar visual acuities, and effective lens wettability and deposit removal. Bench testing included biocompatibility (ISO 10993), microbiological efficacy (ISO 14729, ISO 14730, ISO 18259, ISO 19045), and lens compatibility (ISO 11981, ISO 11986).
Technological Characteristics
Sterile, isotonic, aqueous chemical solution. Triple disinfection system: polyaminopropyl biguanide (0.00005%), polyquaternium (0.00015%), alexidine (0.0002%). Dual surfactants: poloxamine, poloxamer 181. Buffering system: edetate disodium, sodium citrate, diglycine. Additional components: boric acid, sodium borate, sodium chloride, purified water. Packaged in plastic bottles. Tested per ISO 10993, ISO 14729, ISO 14730, ISO 18259, ISO 19045, ISO 11981, ISO 11986, and ISO 11980.
Indications for Use
Indicated for daily cleaning, conditioning, protein removal, rinsing, chemical disinfection, and storage of soft (hydrophilic) contact lenses, including silicone hydrogel lenses, for habitual contact lens wearers.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
COMPLETE Multi-Purpose Solution Easy Rub Formula (K030092)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, which is a symbol often associated with healthcare.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2016
Bausch & Lomb Incorporated Glenn Davies, O.D. Senior Director Regulatory Affairs 1400 North Goodman Street Rochester, NY 14609
Re: K161819
Trade/Device Name: NNR06 Multi-purpose Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN Dated: October 19, 2016 Received: October 20, 2016
Dear Dr. Davies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Glenn Davies, O. D.
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K161819
Device Name NNR06 Multi-Purpose Solution
#### Indications for Use (Describe)
Bausch + Lomb® NNR06 Multi-Purpose Solution is indicated for use in the daily conditioning, cleaning, removal of protein deposits, rinsing, chemical (not heat) disinfection, and storage of soft (hydrophilic) contact lenses, including silicone hydrogel contact lenses, as recommended by your eye care practitioner.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K161819
#### 510(k) SUMMARY
#### 1. Submitter Information
Bausch & Lomb Incorporated 1400 North Goodman Street Rochester, NY 14609
Contact Person Glenn A. Davies, O.D. Senior Director Requlatory Affairs
# Telephone Number
(585) 338-8215
Summary Prepared: November 2016
## 2. Device Name
| Common Name: | Soft (hydrophilic) Contact Lens Care Solution |
|------------------------|-----------------------------------------------|
| Trade Name: | NNR06 Multi-Purpose Solution |
| Classification: | Accessories, Soft Lens Products |
| Device classification: | Class II (21 CFR §886.5928) |
| Product Code: | LPN |
## 3. Predicate Device
COMPLETE Multi-Purpose Solution Easy Rub Formula (K030092)
#### 4. Description of the Device
Bausch + Lomb NNR06 Multi-Purpose Solution is a sterile, isotonic solution containing a triple disinfection system (polyaminopropyl biguanide 0.00005%, polyquaternium 0.00015% and alexidine 0.0002%), dual surfactants, poloxamine, poloxamer 181, and a novel buffering system consisting of edetate disodium, sodium citrate, and diglycine. The formulation also contains boric acid, sodium borate, sodium chloride and purified water. The sterile solution is packaged in a plastic bottle with a tamper evident seal and labeled with a lot number and expiration date.
{4}------------------------------------------------
## 5. Intended Use
Bausch + Lomb NNR06 Multi-Purpose Solution is indicated for use in the daily cleaning, conditioning, removing protein deposits, rinsing, chemical (not heat) disinfection, and storage of soft (hydrophilic) contact lenses, including silicone hydrogel contact lenses, as recommended by your eye care practitioner.
## 6. Description of Safety and Substantial Equivalence
A series of preclinical and clinical testing was performed to demonstrate the safety and effectiveness of Bausch + Lomb NNR06 Multi-Purpose Solution as described in Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products, May 1, 1997 A brief summary of the test results is provided below:
#### Biocompatibility
Cytotoxicity, ocular irritation, oral toxicity, sensitization and in vivio ocular biocompatibility studies were completed for Bausch + Lomb NNR06 Multi-Purpose Solution. The test results demonstrated the biocompatibility of NNR06 Multi-Purpose Solution.
Testing was conducted according to Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products, May 1, 1997 and ISO Standard 10993-1 Biological evaluation of medical devices and related biocompatibility standards (ISO 9394. ISO 10993-5. ISO 10993-10. ISO 10993-11).
#### Microbiology
The sponsor conducted a series of studies according to EN ISO 14729:2001 Microbiological requirements and test methods for products and regimens for hygienic management of contact lenses, including Amendment 1 (2010), and EN ISO 14730:2014 Antimicrobial preservative efficacy testing and guidance on determining discard date, ISO18259 Method to assess contact lens care products with contact lens care products with contact lenses in a lens case, and ISO 19045 Method for evaluating Acanthamoeba encystment by contact lens care products and demonstrate Bausch + Lomb NNR06 Multi-Purpose Solution exceeds the criteria for disinfection and preservative efficacy.
In addition, ISO 17665-1 Sterilization of Health Care Products - Moist Heat and ISO 11137 Sterilization of Health Care Products - Radiation were used to assess the sterilization method for production.
#### Lens Compatibility
The results of lens compatibility studies demonstrate Bausch + Lomb NNR06 Multi-Purpose Solution is compatible with soft contact lenses including silicone hydrogel contact lenses.
Testing was conducted according to Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products, May 1, 1997 and ISO Standard 11981
{5}------------------------------------------------
Determination of physical compatibility of contact lens care products with contact lenses, ISO 11986 Determination of preservative uptake and release.
#### Cleaning Efficacy
The cleaning efficacy of the solution was evaluated through the determination of the Critical Micelle Concentration. The surfactant concentrations are well above the CMC for the individual surfactants. In vitro laboratory studies demonstrated the cleaning properties of NNR06 Multi-Purpose Solution with contact lenses artificially deposited with protein.
#### Clinical Study
Bausch + Lomb conducted a controlled clinical study with soft (hydrophilic) contact lenses, including silicone hydrogel lenses, comparing the safety and effectiveness of Bausch + Lomb NNR06 Multi-Purpose Solution to Complete Multi-Purpose Solution Easy Rub Formula. The study was a three-month, active control, parallel group, masked randomized study of habitual contact lens wearers.
Testing was conducted according to Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products, May 1, 1997 and ISO Standard 11980 Guidance for clinical investigations.
#### Safety
The primary endpoint of interest for slit lamp findings was achieved. There were no reported adverse events in either treatment group during the study. Visual acuities were similar between eyes of both groups.
Graded slit lamp findings; the incidence and type of slit lamp findings and symptoms requiring medical treatment; the comparison of visual acuities; and the absence of adverse events indicate that the NNR06 Solution is safe for use with soft contact lenses, including silicone hydrogel lenses.
#### Effectiveness
Comfort-related symptoms/complaints, Investigator-assessed lens wettability, and Investigator-assessed lens deposits indicate that the NNR06 Multi-Purpose Solution is effective for use with soft contact lenses including silicone hydrogel lenses.
The results of the study demonstrated similar safety and effectiveness profiles and support a substantial equivalence determination.
## 7. Substantial Equivalence
Bausch + Lomb NNR06 Multi-Purpose Solution has the same technological characteristics, sterile, isotonic, aqueous chemical disinfecting solution, as the predicate device. Bausch + Lomb NNR06 Multi-Purpose Solution does not raise new questions of safety and effectiveness.
{6}------------------------------------------------
The cumulative results of laboratory, in vivo, and clinical testing sponsored by Bausch + Lomb demonstrate that the safety, effectiveness and performance of Bausch + Lomb NNR06 Multi-Purpose Solution are substantially equivalent to AMO Complete Multi-Purpose Solution Easy Rub Formula for soft contact lenses, including silicone hydrogel contact lenses.
The similarities and differences between Bausch + Lomb NNR06 Multi-Purpose Solution and the predicate device, AMO Complete Multi-Purpose Solution Easy Rub Formula, are described in the following table.
| Feature | Bausch + Lomb<br>NNR06 Multi-Purpose<br>Solution | AMO<br>Complete Multi-<br>Purpose Solution<br>Easy Rub Formula |
|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Indicated for Soft Lenses | Yes | Yes |
| Indicated for Silicone<br>Hydrogel Lenses | Yes | No |
| Indicated for use in<br>conditioning, cleaning,<br>removal of protein rinsing,<br>disinfection and storage of<br>contact lenses | Yes | Yes<br>Except Conditioning |
| Rub Regimen | 3 Drops<br>20 Seconds | 3 Or More Drops<br>20 Seconds |
| Optional No-Rub Regimen | No | No |
| Rinse Required for<br>Cleaning | Thoroughly Rinse<br>5 Seconds Each Side | Rinse<br>5 Seconds Each Side |
| Minimum Disinfection Time | 4 hours | 6 hours |
| Triple Disinfection System | Yes | No |
| Preservatives /<br>Disinfectants | Polyaminopropyl Biguanide<br>0.5ppm<br>Polyquaternium<br>1.5ppm<br>Alexidine<br>2.0ppm | Polyhexamethylene Biguanide<br>1ppm |
# SUBSTANTIAL EQUIVALENCE SUMMARY TABLE
{7}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE SUMMARY TABLE (CONTINUED)
| Feature | Bausch + Lomb<br>NNR06 Multi-Purpose<br>Solution | AMO<br>Complete Multi-<br>Purpose Solution<br>Easy Rub Formula |
|-----------------------|--------------------------------------------------|----------------------------------------------------------------|
| Dual Surfactants | Yes | No |
| Surfactants | Poloxamer 181<br>Poloxamine | Poloxamer 237 |
| Lens Storage Period | 30 Days | 30 Days |
| Discard After Opening | 90 Days | No<br>Use Before Expiration Date |
| Primary Container | Clear Plastic Bottle<br>Multiple Sizes | White Plastic Bottle<br>Multiple Sizes |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.