K030092 · Advanced Medical Optics, Inc. · LPN · Apr 4, 2003 · Ophthalmic
Device Facts
Record ID
K030092
Device Name
COMPLETE BRAND MULTI-PURPOSE SOLUTION
Applicant
Advanced Medical Optics, Inc.
Product Code
LPN · Ophthalmic
Decision Date
Apr 4, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
COMPLETE® brand Multi-Purpose Solution is indicated for the care of soft (hydrophilic) contact lenses. Use this product, as recommended by your eye care practitioner, to: Chemically (NOT HEAT) Disinfect . Clean ● Rinse . Store . Remove Protein . Condition .
Device Story
COMPLETE® brand Multi-Purpose Solution is a sterile, isotonic, buffered chemical solution for soft (hydrophilic) contact lens care. Formulation includes hydroxypropyl methylcellulose (HPMC) as a lubricant, polyhexamethylene biguanide (0.0001%) as a preservative, Poloxamer 237 as a surfactant, edetate disodium as a chelating agent, phosphate buffer, sodium chloride, potassium chloride, and purified water. Device functions by chemical disinfection and surface conditioning; HPMC adsorbs onto the lens surface during soaking to enhance moisture retention and wettability. Used by patients at home for daily lens maintenance. Output is a treated, cleaned, and disinfected contact lens. Benefits include improved lens comfort and moisture retention.
Clinical Evidence
Clinical evidence includes three subjective studies using questionnaires to rate comfort and acceptability of HPMC-containing formulations versus original COMPLETE® and competitor regimens. Results indicate comfort is substantially equivalent to other marketed solutions. An additional clinical study provided objective data demonstrating equivalence between 'no-rub' and 'rub' regimens.
Technological Characteristics
Sterile, isotonic, buffered aqueous solution. Ingredients: hydroxypropyl methylcellulose (lubricant), polyhexamethylene biguanide 0.0001% (preservative), Poloxamer 237 (surfactant), edetate disodium (chelating agent), phosphate buffer, sodium chloride, potassium chloride, purified water. No chlorhexidine or thimerosal. Packaged in plastic bottles with controlled dropper tips. Chemical disinfection principle; no heat required.
Indications for Use
Indicated for the care of soft (hydrophilic) contact lenses. Used for chemical disinfection, cleaning, rinsing, storage, protein removal, and conditioning as recommended by an eye care practitioner.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
Other contact lens care multi-purpose solutions (unspecified)
Reference Devices
B&L's ReNu® MultiPlus®
Alcon's OPTI-FREE® Express
Submission Summary (Full Text)
{0}------------------------------------------------
K030092
Image /page/0/Picture/1 description: The image shows the logo for AMO, which stands for Advanced Medical Optics. The logo features a stylized letter "A" with a curved line above the letters "AMO". Below the letters is the text "ADVANCED MEDICAL OPTICS" in a smaller font. The logo is simple and professional, and it is likely used to represent the company's brand.
APR 0 4 2003
# 510(k) SUMMARY COMPLETE® brand Multi-Purpose Solution
This summary uses the format provided in 21 CFR 807.92:
- Paul J. Nowacki Submitter: (a)(1) Manager Regulatory Affairs Advanced Medical Optics, Inc. 1700 E. St. Andrew Place P.O. Box 25162 Santa Ana, CA 92799-5162 Phone: (714) 247-8601 Fax: (714) 247-8677 Summary Prepared: December 16, 2002 (a)(2) Device Trade Name: COMPLETE® brand Multi-Purpose Solution
- Soft (Hydrophilic) Contact Lens Solution Device Common Name:
Device Classification Names: Accessories to Contact Lens Solution (86LPN)
- (a)(3) Identification of Predicate Device: COMPLETE® brand Multi-Purpose Solution is substantially equivalent to other contact lens care multi-purpose solutions.
- (a)(4) Device Description: COMPLETE® brand MULTI-PURPOSE Solution is a sterile, isotonic, buffered, solution containing hydroxypropyl methylcellulose as a lubricant, preserved with polyhexamethylene biguanide 0.0001%, a phosphate buffer, Poloxamer 237 as a surfactant, edetate disodium as a chelating agent, sodium chloride, potassium chloride, and purified water. COMPLETE® brand MULTI-PURPOSE Solution contains no chlorhexidine or thimerosal.
The product is a clear, colorless solution packaged in plastic bottles with controlled dropper tips.
. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.. ..
{1}------------------------------------------------
- Intended Use (Indications for Use): COMPLETE® brand Multi-Purpose (a)(5) Solution is indicated for the care of soft (hydrophilic) contact lenses. Use this product, as recommended by your eye care practitioner, to:
- Chemically (NOT HEAT) Disinfect .
- Clean ●
- Rinse .
- Store
- Remove Protein .
- Condition .
- Comparison of Technological Characteristics: No changes have been (a)(6) made to the product formulation subject to this application.
#### Discussion of Nonclinical: (b)(1)
In Vitro Studies were conducted to compare the physical and chemical properties of COMPLETE® Solution with other multi-purpose solutions in order to evaluate the effects of adding hydroxypropyl methylcellulose (HPMC). They were as follows:
- Compare viscosity of COMPLETE® Solution with HPMC to COMPLETE® . without HPMC, and to other multipurpose solutions.
- Investigate the uptake of HPMC onto the lens surface during the soak . period, and its subsequent release, over time, in a saline solution.
- . Compare the effect of three COMPLETE® Solution formulations to B&L's ReNu® MultiPlus® and Alcon's OPTI-FREE® Express on surface fluid retention.
- Examine the effect of various multipurpose solutions on receding contact . angle as a measure of wetting.
The studies show uptake of HPMC onto the lens surface during the soak period, and its subsequent release, over time, in saline. This laver of adsorbed HPMC appears to enhance the lens' ability to retain moisture and remain more wettable.
{2}------------------------------------------------
#### Discussion of Clinical Data: (b)(2)
Study subjects in three clinical studies received questionnaires asking them to rate the comfort and acceptability of two COMPLETE® formulations with hydroxypropyl methylcellulose (HPMC) demulcent/lubricant, vs. original COMPLETE® without HPMC and other Advanced Medical Optics and competitor regimens. This subjective data shows that COMPLETE® Solution, used with a rub regimen, has comfort substantially equivalent to other marketed contact lens care multipurpose solutions.
Objective data from an additional clinical study demonstrate the substantial equivalence of COMPLETE® Solution used in a "no-rub" regimen, with COMPLETE® Solution used in a "rub" regimen.
- Conclusions Drawn from Data Supporting Equivalence Determination: (b)(3) In view of the in vitro and clinical studies supporting this application, we believe the safety, efficacy and comfort of COMPLETE® brand Multi-Purpose Solution is substantially equivalent to other marketed contact lens care multipurpose solutions.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image contains the words "Public Health Service" in a simple, sans-serif font. The text is arranged on a single line, with each word clearly legible. The overall impression is clean and straightforward, suggesting an official or institutional context.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Advanced Medical Optics c/o Paul J. Nowacki Manager, Regulatory Affairs 1700 E. St. Andrew Place P.O. Box 25162 Santa Ana, CA 97299-5162
Re: K030092 Trade/Device Name: COMPLETE® brand Multi-Purpose Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN Dated: January 10, 2003 Received: January 10, 2003
Dear Mr. Nowacki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
APR 0 4 2003
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
A. Palgi Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) NUMBER: (IF KNOWN):
### K030092
DEVICE NAME:
## COMPLETE® brand Multi-Purpose Solution
INDICATIONS FOR USE:
COMPLETE® brand Multi-Purpose Solution is indicated for the care of soft (hydrophilic) contact lenses. Use this product, as recommended by your eye care practitioner, to:
- Chemically (NOT HEAT) Disinfect .
- Clean ●
- Rinse .
- Store ●
- Remove Protein ●
- Condition .
Mirafratt
on Sign-C lean of Ophthalmic Ear, ୍ୟ and Throat Devises
Number K030092
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109 OR
| Over-The-Counter-Use | X |
|--------------------------|---|
| (Optional Format 1-2-96) | |
js
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.