K161776 · Lenstec, Inc. · MSS · Nov 29, 2016 · Ophthalmic
Device Facts
Record ID
K161776
Device Name
Lenstec LC Injection system
Applicant
Lenstec, Inc.
Product Code
MSS · Ophthalmic
Decision Date
Nov 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4300
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Lenstec's LC Injection System is intended for use in the implantation of the Softec I, Softec HDO, Softec HDM and any approved IOL in which labeling specifies use of this injector (s) to insert the capsular bag following extracapsular extraction.
Device Story
System facilitates intraocular lens (IOL) implantation during cataract surgery. Components include reusable, autoclavable syringe-based injectors and lens loaders, plus single-use, sterile disposable cartridges with silicone cushions. Injectors feature either smooth or threaded barrels; threaded versions allow plunger advancement via twisting motion. Cartridges utilize LubriMATRIX™ surface treatment for lubricity and are constructed from medical-grade polypropylene. Device is used by ophthalmologists in clinical settings. Surgeon loads IOL into cartridge, then uses injector to deliver lens into capsular bag. System provides controlled mechanical delivery of IOL, reducing manual handling and potentially minimizing incision size requirements. Changes from predicate include addition of LubriMATRIX™ coating, material substitution (polypropylene 7823M), and introduction of threaded injector barrels.
Clinical Evidence
Bench testing only. Biocompatibility testing performed on all contact materials with satisfactory results. Performance testing conducted with Softec HD, Softec HDO, and Softec HDM models. Cartridges passed USP <789> Particulate Matter in Ophthalmic Solutions (Light Obscuration Tests). No clinical data required.
Technological Characteristics
System comprises reusable stainless steel/plastic injectors, reusable lens loaders, and single-use polypropylene cartridges with silicone cushions. Injectors utilize manual mechanical actuation (push or twist-threaded). Cartridges feature LubriMATRIX™ surface treatment. Materials: Medical-grade polypropylene (Total 7823M). Sterilization: Cartridges provided sterile. Connectivity: None.
Indications for Use
Indicated for implantation of Softec I, Softec HDO, Softec HDM, and other approved IOLs into the capsular bag following extracapsular extraction.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
Predicate Devices
Lenstec Injection System for Intraocular Lens delivery (K122848)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them that resemble water or fabric.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 29, 2016
Lenstec Inc. Mr. Jimmy Chacko Vice President, Regulatory Affairs 1765 Commerce Ave. N. St. Petersburg, Florida 33716
Re: K161776
Trade/Device Name: Lenstec LC Injection System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: MSS Dated: November 2, 2016 Received: November 3, 2016
Dear Mr. Chacko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Lenstec LC Injection System
#### Indications for Use (Describe)
Lensted's LC Injection System is intended for use in the implantation of the Softec I, Softec HDO, Softec HDM and any approved IOL in which labeling specifies use of this injector (s) to insert the capsular bag following extracapsular extraction.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
### Section 6: 510(k) Summary
Date Summary Prepared: June 17th, 2016
# Lenstec Inc 510(k) Summary for Premarket Notification Submission for Lenstec LC Injection System
### Labeling:
Federal (United States) Law restricts this device to sale by or on the order of a physician
- 1. Applicant Information:
- a. Name: Lenstec Inc.
- b. Address: 1765 Commerce Avenue North St. Petersburg, FL 33716 Telephone Number: (727) 571-2272 Fax Number: (727) 571-1792
- c. Contact Person: Jimmy Chacko, Vice President, Regulatory Affairs Email: JChacko@Lenstec.com Additional Contact Person: Sheila Baker, Clinical Research Manager Email: SBaker@Lenstec.com
- 2. Name of Device
- a. Trade name: Lenstec LC Injection System
- Common name: Intraocular lens injection system b.
- Regulation name: Intraocular lens guide ﻥ
- d. Regulatory class: Class I, reserve
- e. Product code: MSS
- Regulation number: 21 CFR 886.4300 f.
- 3. Substantially Equivalent legally-marketed device:
- a. Lenstec Injection System for Intraocular Lens delivery (K122848)
{4}------------------------------------------------
# 4. Device Description
The proposed system consists of the following components:
| CART SERIES CARTRIDGE CHART | | | | |
|------------------------------------|------------|--------------------------------|----------------------|--------------------------------|
| Cartridge with<br>silicone cushion | IOL | Injector | Tip Diameter<br>(mm) | Lenstec IOL Power<br>range (D) |
| CART 20S | SOFTEC HD | I-9011S<br>I-9012<br>I-9012 FS | 1.6 | 5.0 - 26.0 |
| | SOFTEC I | | | |
| | SOFTEC HDO | | | SOFTEC HDO<br>5.0 - 22.0 |
| CART 45S | SOFTEC HD | I-9011S<br>I-9012<br>I-9012 FS | 1.6 | 5.0-26.0 |
| | SOFTEC I | | | |
| | SOFTEC HDO | | | |
| CART M | SOFTEC HDM | I-9011S<br>I-9012<br>I-9012 FS | 1.7 | 5.0-36.0 |
Three types of injectors are provided: they are syringe based and reusable and autoclavable. One type of lens loader is provided, which is reusable and autoclavable. The cartridges/silicone cushion are single-use and are provided sterile.
- 5. Use:
Lenstec's LC Injection System is intended for use in the implantation of the Softec HD (P090022), Softec I (P090022/S001), Softec HDO (P090022/S011), Softec HDM (P090022/S026) and any approved IOL in which the labeling specifies use of this injector to insert the lens into the capsular bag following extracapsular extraction. The predicate device was comprised of the nearly identical components.
Lenstec seeks to have cleared the addition of a surface treatment technology, LubriMATRIX™, to Cart 45S. Cart 20S and Cart M cartridges as well as a change of material due to the vendor discontinuing the current material. We also request the addition of two new lens injectors, which have barrels that are now threaded instead of smooth, allowing for a twisting motion to advance the plunger into the cartridge.
The only changes in this iniection system are
- a) the modified indication for inclusion for use of the Softec HD, Softec I, both through K122848. Softec HDO and HDM through this current application.
{5}------------------------------------------------
- b) additional new cartridges; Cart 45S, Cart 20S, Cart M
- c) The addition of surface treatment technology (LubriMATRIX™) on all "Cart" models (Cart 45S, Cart 20S, Cart M)
- d) A slight material change from the Atofina Polypropylene 7823MZ to the Total 7823M Polypropylene. The only reason for the change is that the vendor has discontinued the 7823MZ. The only difference is that the 7823M does not contain an antistat.
- 6. Indications for use:
Lenstec's LC Injection System is intended for use in the implantation of the Softec HD, Softec I, Softec HDO, Softec HDM and any approved IOL in which the labeling specifies use of this injector to insert the lens into the capsular bag following extracapsular extraction.
- 7. Technological characteristics:
The system has three major components: a reusable injector, a reusable lens loader and one of three disposable cartridges (Cart 45S, Cart 20S, Cart M) with a silicone cushion (SIC-01-02). Discussed here are the characteristics of the changes only. The lens loader and silicone cushion have not had any changes and are still identical to those cleared under K122848.
- a. The proposed injectors (1-9012 and I-9012 FS) both have threaded barrels instead of the smooth barrel ("syringe like") injection of the I-9011S. Both the I-9012 and I-9012 FS advance the IOL through the cartridge once the user twists the injector knob until the IOL is delivered into the eye. The I-9011S injector advances the IOL through the cartridge once the user pushes the injector knob until the IOL is delivered into the eye.
- b. The surface treated cartridges (Cart 45S, Cart 20S, Cart M) have LubriMATRIXTM applied for lubricity.
- The proposed cartridge is manufactured of medical grade polypropylene c. and is for single use as in the cleared K122848. however, a material change is necessary due to the vendor discontinuing the polypropylene we currently purchase (7823MZ). The proposed material is now 7823M without the antistat that the 7823MZ had.
- 8. Performance data:
- a. Non clinical tests
All contact materials have been tested for biocompatibility, with satisfactory results. Also, the system was tested with each of the following Lenstec intraocular lens models with satisfactory results: Softec HD, Softec HDO and Softec HDM. Due to the similarities of the Softec HD and Softec 1 IOL design, the Softec 1 will not undergo performance testing of its own. Based on the overall design shape of the Softec HD to the Softec 1, equivalence is expected and we can therefore include the Softec 1 as part of the indications for use as in the previously cleared K122848.
{6}------------------------------------------------
The Cart M was successfully tested with the Softec HDM and can also be included as part of the indications for use statement. Furthermore, the cartridges tested passed the USP <789> Particulate Matter in Ophthalmic Solutions- Light Obscuration Tests.
- 9. Clinical tests: Not required
- 10. Conclusions:
The Lenstec LC Injection System is substantially equivalent to the legally marketed predicate device, and the included testing validates the expanded indication for use.
17 June 16
Timmy Chacko Vice President, Regulatory Affairs Lenstec Inc.
Date
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.