K122848 · Lenstec, Inc. · MSS · Oct 18, 2013 · Ophthalmic
Device Facts
Record ID
K122848
Device Name
LENSTEC LC INJECTION SYSTEM
Applicant
Lenstec, Inc.
Product Code
MSS · Ophthalmic
Decision Date
Oct 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4300
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The Lenstec LC Injection System is intended for use in the implantation of the Softec HD, Softec I and Softec HD PS model Lenstec IOLs, as well as any IOL which has indicated for its use in its approved labeling, into the capsular bag following extracapsular extraction.
Device Story
The Lenstec LC Injection System is a manual, syringe-based surgical instrument used by ophthalmologists during cataract surgery. It facilitates the delivery of foldable intraocular lenses (IOLs) into the capsular bag. The system comprises a reusable titanium injector, a reusable plastic lens loader, and single-use disposable polypropylene cartridges with silicone cushions. The surgeon loads the IOL into the cartridge using the lens loader, then uses the injector to push the lens through the cartridge tip into the eye. The system offers different tip diameters (1.6mm and 2.4mm) and tip angulations (20° or 45°) to accommodate various IOL power ranges and surgical preferences. By enabling a smaller incision for lens insertion, the device aims to improve surgical outcomes and patient recovery. The device is intended for clinical use in an operating room setting.
Clinical Evidence
No clinical data. Substantial equivalence was established through non-clinical bench testing, including biocompatibility testing of all contact materials and performance testing of the system with Softec HD, Softec I, and Softec HD PS IOL models.
Technological Characteristics
System components: reusable titanium injector, reusable plastic lens loader, single-use medical grade polypropylene cartridges, and single-use medical grade silicone cushions. Manual operation. Autoclavable reusable components. Three cartridge configurations (LC 16, LC 1620, LC 2420) with 1.6mm or 2.4mm tip diameters and 20° or 45° tip angulations.
Indications for Use
Indicated for patients undergoing extracapsular cataract extraction requiring implantation of Softec HD, Softec I, or Softec HD PS intraocular lenses (IOLs), or other IOLs labeled for use with this system.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
Predicate Devices
Lenstec Injection System for STAAR Collamer Plate Intraocular Lens (K063802)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo that says "LENS-TECO" in a stylized, blocky font. The letters are outlined in black, with a white fill and a textured, slightly distressed appearance. The "TM" symbol is present in the upper right corner of the logo. The letters are stacked on top of each other, with "LENS" on the top line and "TECO" on the bottom line.
1765 Commerce Ave. N. · St. Petersburg, FL 33716 · U. Telephone: (727) 571-2272 · Fax: (727) 571-1792 · E-mail: lenstec@lenstec.com
### Section 5: 510(k) Summary
Date Summary Prepared: May 25, 2012
# Lenstec Inc 510(k) Summary for Premarket Notification Submission for Lenstec LC Injection System
## Labeling:
Federal (United States) Law restricts this device to sale by or on the order of a physician
- 1. Applicant Information:
- a. Name: Lenstec Inc.
- b. Address: 1765 Commerce Avenue North St. Petersburg, FL 33716
- ・ Telephone Number: (727) 571-2272 Fax Number: (727) 571-1792
- c. Contact Person: Jimmy Chacko, Vice President, Regulatory Affairs Email: JChacko@Lenstec.com
### 2. Name of Device
- Trade name: Lenstec LC Injection System a.
- b. Common name: Intraocular lens injection system
- Regulation name: Intraocular lens guide c.
- Regulatory class: Class I, reserve ರ್.
- e. Product code: MSS
- Regulation number: 21 CFR 886.4300 ﺕ
- 3. Substantially Equivalent legally-marketed device:
- a. Lenstec Injection System for STAAR Collamer Plate Intraocular Lens (K063802)
Section 5: pg 1 of 3 510(k) Number (if known):
13
OCT 1 8 2013
{1}------------------------------------------------
## 4. Device Description
The system consists of the following components:
| Cartridge with<br>silicone cushion | IOL | Injector | Tip Diameter<br>(mm) | Lenstec IOL Power<br>range (D) |
|------------------------------------|---------------------------------------|----------|----------------------|--------------------------------|
| LC 16 | Softec HD<br>Softec I<br>Softec HD PS | I-9011S | 1.6 | 5.0 - 26.0 |
| LC 1620 | Softec HD<br>Softec I<br>Softec HD PS | I-9011S | 1.6 | 5.0 - 26.0 |
| LC 2420 | Softec HD<br>Softec I<br>Softec HD PS | I-9011S | 2.4 | 5.0 - 36.0 |
One type of injector is provided: it is syringe based and is reusable and autoclavable. One type of lens loader is provided, which is reusable and autoclavable. The cartridges/silicone cushion is single-use and provided sterile.
- S. Use:
The Lenstec LC Injection System is intended for use in the implantation of the Softec HD, Softec I and Softec HD PS model Lenstec IOLs, as well as any IOL which has indicated for its use in its approved labeling, into the capsular bag following extracapsular extraction. The predicate device was comprised of the nearly identical components.
Lenstec has added a pair of cartridges to the LC Injection system (LC 1620 and LC 2420}. These two additional cartridges only differ from the originals in that their tips are angulated to 20°, as opposed to the LC 16, which has a 45° tip.
This submission also removes the LC24 model cartridge.
The only changes in this injection system are
- a) the modified indication for inclusion for use of the Lenstec model IOLs identified above, and removal of STAAR Collamer Plate IOL previously indicated for use with
- b) Change in the models of cartridge validated for use
- c) the replacement of the lens fork loader with the lens loader (this is discussed in greater detail elsewhere in this submission)
### 6. Indications for use:
The Lenstec LC Injection System is intended for use in the implantation of the Softec HD, Softec I and Softec HD PS model Lenstec IOLs, as well as any IOL which has indicated for its use in its approved labeling, into the capsular bag following extracapsular extraction.
> Section 5: pg 2 of 3 510(k) Number (if known):
{2}------------------------------------------------
7. Technological characteristics:
The system has three major components: a reusable injector, a reusable lens loader and one of three disposable cartridges (LC 16, LC 1620, LC 2420) with a silicone cushion (SIC-01-02)
a. the injector is manufactured of titanium and can be autoclaved
- b. the lens loader is manufactured from plastic and can be autoclaved
- the cartridge is manufactured of medical grade polypropylene and is for C. single use
- d. the silicone tip is manufactured of medical grade silicone and is for single use
- 8. Performance data:
- a. Non clinical tests
- All contact materials have been tested for biocompatibility, with satisfactory results. Also, the system was tested with each of the following Lenstec intraocular lens models with satisfactory results: Softee HD, Softec I and Softec HD PS.
- 9. Clinical tests: Not required
10. Conclusions:
The Lenstec LC Injection System is substantially equivalent to the legally marketed predicate device, and the included testing validates the expanded indication for use.
Jimmy Chacko Vice President, Regulatory Affairs Lenstec Inc.
25 MAY 12
Date
Section 5: pg 3 of 3 510(k) Number (if known):
14
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of a caduceus, a symbol often associated with healthcare. The caduceus is depicted with a staff entwined by two snakes and topped with wings.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
October 18, 2013
・・・
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Lenstec, Inc. % Mr. Hans-Gerd Evering . Lead Technical Reviewer British Standards Institute (BSI) Kitemark Court, Davy Avenue Knowlhill, Milton Keynes MK5 8PP, UK
Re: K122848
Trade/Device Name: Lenstec LC Injection System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I (reserve) Product Code: MSS Dated: October 2, 2013 Received: October 4, 2013
Dear Mr. Evering:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
## Page 2 - Mr. Hans-Gerd Evering
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Section 4: Indications for Use Statement
510(k) Number (if known): not yet known
Device Name: Lenstec LC Injection System
- Indications for Use: The Lenstec LC Injection System is intended for use in the implantation of the Softec HD, Softec I and Softec HD PS model Lenstec IOLs, as well as any IOL which has indicated for its use in its approved labeling, into the capsular bag following extracapsular extraction
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Picture/9 description: The image shows the text "Tieuv H. Nguyen" on the top line. The second line of text shows the date and time as "2013.10.17 15:40:20.04'00'". The text is surrounded by a decorative border.
Section 4: pg 1 of 1 510(k) Number (if known):
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.